Literature DB >> 29074291

Standardisation and half-life of 89Zr.

E García-Toraño1, V Peyrés2, M Roteta2, M Mejuto2, A Sánchez-Cabezudo2, E Romero3.   

Abstract

The nuclide 89Zr is being tested for the labelling of compounds with long blood circulation times. It decays by beta plus emission (22.8%) and by electron capture (77.2%) to 89Y. Its half-life has been determined by following the decay rate with two measurement systems; an Ionisation Chamber and an HPGe detector. The combination of six results gives a value of T1/2 = 78.333 (38) h, slightly lower than the DDEP recommended value of 78.42 (13) h. This radionuclide has also been standardised by liquid scintillation counting, 4πγ counting and coincidence techniques.
Copyright © 2017 Elsevier Ltd. All rights reserved.

Keywords:  Half-life; Positron emitters; Standardisation; Zr-89

Year:  2017        PMID: 29074291     DOI: 10.1016/j.apradiso.2017.10.033

Source DB:  PubMed          Journal:  Appl Radiat Isot        ISSN: 0969-8043            Impact factor:   1.513


  1 in total

1.  Standardization of I-124 by three liquid scintillation-based methods.

Authors:  D E Bergeron; J T Cessna; R Fitzgerald; L Pibida; B E Zimmerman
Journal:  Appl Radiat Isot       Date:  2019-08-09       Impact factor: 1.513

  1 in total