In vivo and in vitro results of an automated preloaded delivery system for IOL implantation in cataract surgery.

Leonardo Mastropasqua1, Lisa Toto1, Erminia D'Ugo2, Manuela Lanzini1, Peter A Mattei1, Gennaro Falconio1, Emanuele Doronzo1, Francesca Diomede3, Oriana Trubiani3.   

Abstract

PURPOSE: To compare the corneal tissue trauma after the use of an automated preloaded injector and a manual injector and assess scanning electron microscope (SEM) and atomic force microscope (AFM) features of both injector cartridges.
SETTING: Ophthalmology Clinic and Laboratory of Stem Cells and Regenerative Medicine University "G. d'Annunzio" of Chieti-Pescara, Chieti, Italy;
DESIGN: Prospective randomized clinical study
METHODS: Forty eyes of 40 patients for phacoemulsification were divided into two groups: implantation of intraocular lens was performed with AutonoMe automated delivery system (AutonoMe group: 20 eyes) and Monarch III injector system (Monarch group: 20 eyes). In vivo confocal microscopy (IVCM) and anterior segment optical coherence tomography (AS-OCT) were performed before surgery, at 1 h, 1 day and 1 month post-operatively. In addition, SEM and AFM were performed on cartridges of both injector systems after injection of the IOL.
RESULTS: A greater increase in central corneal thickness and corneal thickness at the incision site were observed in Monarch group versus AutonoMe group 1 h and 1 day post-operatively (p < 0.05). Endothelial cell count loss was significantly higher in Monarch group compared with AutonoMe group (p < 0.05) at 1 and 24 h. AS-OCT showed less endothelial misalignment at 30 days (p < 0.05), and IVCM showed less tunnel inflammation at all time points (p < 0.05) in AutonoMe group compared with Monarch group; roughness analysis at AFM of the AutonoMe cartridge was significantly lower compared to Monarch D cartridge (p < 0.05).
CONCLUSIONS: The AutonoMe injector provided less corneal tissue trauma compared with Monarch III injector. The AutonoMe cartridge showed lower roughness at AFM compared to the Monarch D cartridge.

Entities:  

Keywords:  Automated preloaded delivery system; Corneal tissue trauma; Scanning electron microscopy and atomic force microscopy features of injector cartridges

Mesh:

Year:  2019        PMID: 31451986     DOI: 10.1007/s10792-019-01154-0

Source DB:  PubMed          Journal:  Int Ophthalmol        ISSN: 0165-5701            Impact factor:   2.031


  19 in total

1.  Profile of clear corneal cataract incisions demonstrated by ocular coherence tomography.

Authors:  I Howard Fine; Richard S Hoffman; Mark Packer
Journal:  J Cataract Refract Surg       Date:  2007-01       Impact factor: 3.351

2.  Three year prospective, randomized evaluation of intraocular lens implantation through 3.2 and 5.5 mm incisions.

Authors:  T Oshika; K Nagahara; S Yaguchi; K Emi; H Takenaka; S Tsuboi; F Yoshitomi; T Nagamoto; D Kurosaka
Journal:  J Cataract Refract Surg       Date:  1998-04       Impact factor: 3.351

3.  Incision sizes before and after implantation of SN60WF intraocular lenses using the Monarch injector system with C and D cartridges.

Authors:  Thomas Kohnen; Oliver K Klaproth
Journal:  J Cataract Refract Surg       Date:  2008-10       Impact factor: 3.351

4.  Astigmatism after small incision cataract surgery. A prospective, randomized, multicenter comparison of 4- and 6.5-mm incisions.

Authors:  R F Steinert; S F Brint; S M White; I H Fine
Journal:  Ophthalmology       Date:  1991-04       Impact factor: 12.079

5.  Preloaded and non-preloaded intraocular lens delivery system and characteristics: human and porcine eyes trial.

Authors:  Byunghoon Chung; Hun Lee; Moonjung Choi; Kyoung Yul Seo; Eung Kweon Kim; Tae-Im Kim
Journal:  Int J Ophthalmol       Date:  2018-01-18       Impact factor: 1.779

6.  Astigmatism and visual recovery after 'large incision' extracapsular cataract surgery and 'small' incisions for phakoemulsification.

Authors:  L Zheng; J C Merriam; M Zaider
Journal:  Trans Am Ophthalmol Soc       Date:  1997

Review 7.  Advances in cataract surgery: preserving the corneal endothelium.

Authors:  Joanne W Ho; Natalie A Afshari
Journal:  Curr Opin Ophthalmol       Date:  2015-01       Impact factor: 3.761

8.  [Architectural analysis of clear corneal incisions using Visante OCT in acute postoperative endophthalmitis].

Authors:  R Belazzougui; S Dupont Monod; C Baudouin; A Labbé
Journal:  J Fr Ophtalmol       Date:  2010-01-06       Impact factor: 0.818

9.  Wound Integrity of 2.0 mm Transconjunctival Single-Plane Sclerocorneal Incision: A Comparison between a Motorized Injector with/without Pause Time and a Manual Injector.

Authors:  Hisanori Imai; Ayaka Fujii; Emiko Tani; Atsushi Azumi
Journal:  J Ophthalmol       Date:  2016-02-17       Impact factor: 1.909

Review 10.  Comparison of efficacy between coaxial microincision and standard-incision phacoemulsification in patients with age-related cataracts: a meta-analysis.

Authors:  Lijun Wang; Xiao Xiao; Lin Zhao; Yi Zhang; Jianming Wang; Aiyi Zhou; Jianchao Wang; Qian Wu
Journal:  BMC Ophthalmol       Date:  2017-12-29       Impact factor: 2.209

View more
  4 in total

1.  Enhanced VEGF/VEGF-R and RUNX2 Expression in Human Periodontal Ligament Stem Cells Cultured on Sandblasted/Etched Titanium Disk.

Authors:  Guya Diletta Marconi; Francesca Diomede; Jacopo Pizzicannella; Luigia Fonticoli; Ilaria Merciaro; Sante D Pierdomenico; Emanuela Mazzon; Adriano Piattelli; Oriana Trubiani
Journal:  Front Cell Dev Biol       Date:  2020-05-14

2.  Experimental Study on Delivery Performance of an Automated Preloaded Intraocular Lens Injector System for Corneal and Sclerocorneal Incisions.

Authors:  Tetsuro Oshika; Noriyuki Sasaki
Journal:  J Ophthalmol       Date:  2021-03-31       Impact factor: 1.909

3.  Refractive stability of a new single-piece hydrophobic acrylic intraocular lens and corneal wound repair after implantation using a new automated intraocular lens delivery system.

Authors:  Kazuno Negishi; Sachiko Masui; Hidemasa Torii; Yasuyo Nishi; Kazuo Tsubota
Journal:  PLoS One       Date:  2020-09-02       Impact factor: 3.240

4.  One-year multicenter evaluation of a new hydrophobic acrylic intraocular lens with hydroxyethyl methacrylate in an automated preloaded delivery system.

Authors:  Tetsuro Oshika; Noriyuki Sasaki
Journal:  J Cataract Refract Surg       Date:  2022-03-01       Impact factor: 3.528

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.