| Literature DB >> 31425790 |
Fabian Eisenstein1, Radostin Danev2, Martin Pilhofer3.
Abstract
The power of cryo-electron tomography (cryoET) lies in its capability to characterize macromolecules in their cellular context. Structure determination by cryoET, however, is time-consuming compared to single particle approaches. A recent study reported significant acceleration of data acquisition by a fast-incremental single-exposure (FISE) tilt series scheme. Here we improved the method and evaluated its efficiency and performance. We show that (1) FISE combined with the latest generation of direct electron detectors speeds up collection considerably, (2) previous generation (pre-2017) double-tilt axis Titan Krios holders are also suitable for FISE data acquisition, (3) x, y and z-specimen shifts can be compensated for, and (4) FISE tilt series data can generate averages of sub-nanometer resolution. These advances will allow for a widespread adoption of cryoET for high-throughput in situ studies and high-resolution structure determination across different biological research disciplines.Entities:
Keywords: Calibration; Cryo-EM; Cryo-electron tomography; Data acquisition; Electron cryotomography; K2; K3; Throughput; Tilt series
Mesh:
Year: 2019 PMID: 31425790 PMCID: PMC6839400 DOI: 10.1016/j.jsb.2019.08.006
Source DB: PubMed Journal: J Struct Biol ISSN: 1047-8477 Impact factor: 2.867
K3 camera eliminates long processing times. Time measurements of FISE −60° to +60° tilt series with 3° increments on Titan Krios [DT/K2] and Titan Krios [ST/K3].
| Instrument | Tilt scheme | Unblank time | Prep time | Set Rec time | Cam Rec time | Total time |
|---|---|---|---|---|---|---|
| [ST/K3] | unidirectional | 1 s | 98 s | 91 s | 95 s | 201 s |
| [DT/K2] | unidirectional | 1 s | 136 s | 117 s | 253 s | 399 s |
| [ST/K3] | dose-symmetric | 1 s | 101 s | 130 s | 134 s | 243 s |
| [DT/K2] | dose-symmetric | 1 s | 134 s | 185 s | 339 s | 484 s |
Time the sample is exposed per tilt.
Time taken to refine eucentricity and autofocus.
Exposure time set for Record in SerialEM. Variations are due to the slower tilt speed on Titan Krios [DT/K2]. In all cases, the default tilt speed of the stage was not altered.
Time from start of acquisition until regain of microscope control.
Total run time of the tilt series including resetting the microscope state.
Fig. 1Performance increase of ST holder compared to DT holder is marginal. Representative examples of specimen shift throughout a FISE dose-symmetric 3° increment tilt series recorded using a single-tilt (ST) axis holder (A/B) or a dual-tilt (DT) axis holder (C/D). The tilt axis in both cases is approximately parallel to the x-axis. Overall, specimen shifts are comparable between both holders and x-shift is negligible compared to y-shift. The DT holder shows larger systematic shift upon reversal of tilt direction (arrowhead).
Fig. 2Image shift compensation can be used to minimize specimen shifts. A/B: Specimen shift throughout a FISE dose-symmetric 3° increment tilt series recorded without application of compensating image shifts (A) and with application of compensating image shifts (B) recorded on a Titan Krios [ST/K3]. Image shifts effectively compensate specimen shifts, especially at large tilt angles. C/D: A single calibration tilt series can be used to compensate specimen shifts for multiple grid squares. The schematic (C) shows squares of a Quantifoil Cu 200 R2/2 grid and refers to the relative positions of tilt series that were collected with (colored) and without (grey) compensating image shifts (D). Shown are specimen shifts perpendicular to the tilt axis (y-shift) within full tilt series. E: Y-shifts of 47 tilt series collected in less than four hours on eleven different squares (Fig. S3) using a single calibration tilt series and a Titan Krios [ST/K3]. Blue crosses mark tilt angles higher than ±51°. The arrowhead indicates the worst tilt series that is also shown in Movie S3. Tilt series are grouped by squares. No micrographs had to be discarded due to the loss of target.
Fig. 3Z-shifts can be partially compensated by pre-calibrated defocus offsets. Estimated defocus of motion-corrected micrographs within individual FISE dose-symmetric tilt series (A/C/E) and per tilt angle (B/D/F). Red crosses mark tilt angles higher than 45° and blues crosses mark tilt angles lower than −45°. Data was collected on a Titan Krios [ST/K3]. A/B: 47 tilt series with a cycling target defocus between 2 and 5 µm and a step size of 0.5 µm. Defocus change is over 2 µm per tilt series (A) and shows approximately linear dependence on tilt angle (B). C/D: 36 tilt series with a target defocus of 5 µm and tilt angle-dependent defocus correction determined by CTF analysis of a calibration tilt series. Calibrations can compensate for defocus change in an angular range of ±45°. E/F: 9 tilt series with a cycling target defocus between 2.5 and 5 µm and tilt angle-dependent defocus correction determined by beam tilt defocus measurement at four tilts of a calibration tilt series. Grey tilt series in (E) were taken at the same stage position without applying the defocus correction. Calibrations can reduce defocus change within ~1 µm at an angular range of ±45°.
Fig. 4FISE tilt series acquisition can generate a sub-nanometer average. A: Subtomogram average of the E. coli 70S ribosome obtained from FISE tilt series. Blue: large subunit, yellow: small subunit. B: Gold standard FSC curve calculated using emClarity. C: Subtomogram average fitted with the atomic model of a 70S ribosome (PDB 5MDZ). Note the visibility of secondary structure features like rod-shaped alpha helices (yellow) and ribosomal RNA double helices (blue).