The hidden cost of colonoscopy including cost of reprocessing and infection rate: the implications for disposable colonoscopes.

Sara Larsen1, Anthony Kalloo2, Susan Hutfless3,4,5.   

Abstract

Entities:  

Keywords:  colonoscopy; endoscopy; health economics

Mesh:

Year:  2019        PMID: 31413166      PMCID: PMC6984057          DOI: 10.1136/gutjnl-2019-319108

Source DB:  PubMed          Journal:  Gut        ISSN: 0017-5749            Impact factor:   23.059


× No keyword cloud information.

Message

Multiple studies have documented a high rate of contaminated colonoscopes after reprocessing. Contaminated reusable colonoscopes may increase the risk of device-related patient infections. As disposable colonoscopes enter the market, they may play a role in infection prevention and may be cost-effective at some facilities or in high-risk patients. Using a micro-costing approach, this study found that the cost per colonoscopy including purchase, maintenance and reprocessing ranges from US$188.64 at high volume centres (3000 annual procedures) to US$501.16 at low volume centres (1000 annual procedures). Accordingly, per-procedure capital costs range from US$87.48 to US$262.45; repair costs range from US$68.77 to US$206.32; cleaning supplies and labour costs US$32.39 and infections requiring hospitalisation cost US$20.12 to US$46.52. As disposable colonoscopes enter the market, low volume centres are most likely to achieve cost savings. Determining if post-procedural infection rates differ with reusable vs disposable colonoscopes is needed.

In more detail

Each year more than 15 million colonoscopy procedures are performed in the USA and the number is increasing.1 Colonoscopy is generally thought to be safe; however, patients are sometimes hospitalised afterwards, due to infections that may have been transmitted via contaminated colonoscopes (MAUDE Adverse Event Report).2 3 Colonoscopy-related infections and complications have been reported in multiple studies, although at lower rates compared with endoscopic retrograde cholangiopancreatography (ERCP).1 4 5 Guideline revisions in response to endoscope-associated infections require healthcare institutions to invest more resources in the high-level disinfection process or conversion to sterilisation (Ofstead et al).5 6 Despite colonoscopy being the highest volume GI procedure, the true cost and time associated with reusable colonoscopes are unknown. The purpose of this study was to explore real-world costs associated with reusable colonoscopes including capital costs and costs associated with reprocessing, personnel, maintenance, repair and postprocedural hospitalisation due to infection. All cost data related to the usage of reusable colonoscopes were obtained at a high-volume outpatient endoscopy referral centre (Johns Hopkins Hospital, Baltimore, Maryland, USA) by tracking 25 colonoscope reprocessing procedures over a 3-day period. Cost data were collected in US dollars ($) using a micro-costing approach as this method allows for precise assessment of economic costs.7 Cost per use of reusable colonoscopes were calculated fora range of annual procedures (1000, 2000 and 3000) performed with a fleet of 20 colonoscopes (US$35 000 per colonoscope) (Addendum) and extrapolated to different numbers of colonoscopes and procedure volumes (table 1). Automated endoscopic reprocessor (AER) cost calculations assumed two AERs to reprocess the colonoscopes available at the endoscopy unit (US$47 646.80 per AER (US$13863.01/year)) regardless of volume. Capital costs of the colonoscopes and their associated hardware and software were amortised over a 5-year period, and a discount rate of 3.5% was used to calculate the present value of capital expenditures. The AER and drying cabinets were amortised over an 8-year period. Average time spend on manual reprocessing was calculated for each reprocessing step. Costs related to initial and recurring training, education of personnel, time spent handling documentation for repair and retraining for compliance with latest reprocessing guidelines were not included. Average cost of repairs per AER is US$7831.25 per year. Each colonoscope was repaired 3.04 times per year on average at US$8609.94 per year (based on Johns Hopkins Service Contract Performance Reports, 2018).
Table 1

Estimation of the per-procedure costs of reusable colonoscopes by varying number of annual procedures and number of colonoscopes

Summary of costs1000 procedures (US$)2000 procedures (US$)3000 procedures (US$)Costs that will remain with disposable endoscope
Capital costs Yes. Purchase of disposable colonoscope per procedure only and monitor (lifespan approximately 5 years)
10 colonoscopes184.9392.4761.64
20 colonoscopes 262.45 131.22 87.48
35 colonoscopes378.73189.36126.24
50 colonoscopes495.00247.50165.00
Repair costs No
10 colonoscopes120.2260.1140.07
20 colonoscopes 206.32 103.16 68.77
35 colonoscopes335.47167.74111.82
50 colonoscopes464.62232.31154.87
Precleaning, leak testing, manual cleaning, visual inspection, high-level disinfection, storage (including personal protective equipment) 25.23 No
Personnel time during pre-cleaning through high-level disinfection 7.16 No
Infection-related treatment* Per-procedure cost at different infection rates: 1.6/10001 and 0.37/1004 20.12– 46.52 Not for infections caused by endoscopic cross-contamination
Total
10 colonoscopes357.66–384.07205.08–231.49154.23–180.63
20 colonoscopes 521.28–547.69 286.89313.30 208.76235.17
35 colonoscopes766.71–793.11409.61–436.01290.57–316.98
50 colonoscopes1012.12–1038.54532.32–558.73372.38–398.79

Costs marked with bold font are the base-case numbers (ie, based on 20 colonoscopes). The message of this study is based on the base-case results. Other estimates are used to increase transparency and to make data comparable to other facilities.

*Assumption: all infected patients are treated at a hospital.

Estimation of the per-procedure costs of reusable colonoscopes by varying number of annual procedures and number of colonoscopes Costs marked with bold font are the base-case numbers (ie, based on 20 colonoscopes). The message of this study is based on the base-case results. Other estimates are used to increase transparency and to make data comparable to other facilities. *Assumption: all infected patients are treated at a hospital. Infection rates were 3.7 and 1.6 per 1000 procedures.1 4 Costs of infection-related hospitalisation were collected from HCUPnet8 which samples the 2016 National Inpatient Sample and the International Classification of Diseases 10th revision codes A04 and A09. Cost of infection was based on the assumption that all infected patients are treated at a hospital. Cost per treatment is US$12 574.28. Costs of postendoscopic infection hospitalisation per procedure were calculated by multiplying the infection rate and cost per hospitalisation (eg (1.16/1000)×US$12 574.28). The cost associated with reusable colonoscopes ranges from US$188.64 to US$501.16 per procedure based on 20 colonoscopes (table 1 and online supplementary table 1 (addendum)). The per-procedure cost is highly dependent on the number of annual procedures and colonoscopes available at the facility (table 2). The time associated with manual reprocessing was approximately 19 min per procedure. Divided into three major cost categories, the capital costs per procedure range from US$81.21 to US$243.63, costs of repair ranged from US$68.77 to US$206.59, and the costs of cleaning including labour ranged from US$39.91 to US$50.11. In addition, the costs of hospitalisation due to infection following colonoscopy ranged between US$20.12 and US$46.52 per procedure.
Table 2

Cost of colonoscopy (in US$) including reprocessing, capital cost and repair cost by varying number of colonoscopes and annual number of procedures

Annual number of procedures
10001200140016001800200022002400260028003000320034003600
Number of colonoscopes available5255.73218.51191.92171.98156.47144.06133.91125.45118.29112.15106.84102.1898.0894.43
6272.10232.14203.61182.21165.56152.24141.35132.27124.58118.00112.29107.30102.8998.97
7288.46245.78215.29192.43174.65160.42148.78139.08130.88123.84117.74112.41107.70103.52
8304.82259.41226.98202.66183.7468.60156.22145.90137.17129.69123.20117.52112.52108.06
9321.18273.05238.67212.88192.83176.78163.66152.72143.46135.53128.65122.64117.33112.61
10337.54286.68250.36223.11201.92184.97171.09159.54149.76141.37134.11127.75122.14117.15
11353.90300.32262.04233.34211.01193.15178.53166.35156.05147.22139.56132.86126.95121.70
12370.27313.95273.73243.56220.10201.33185.97173.17162.34153.06145.01137.98131.76126.24
13386.63327.59285.42253.79229.19209.51193.41179.99168.63158.90150.47143.09136.58130.79
14402.99341.22297.10264.01238.28217.69200.84186.81174.93164.75155.92148.20141.39135.33
15419.35354.86308.79274.24247.37225.87208.28193.62181.22170.59161.38153.31146.20139.88
16435.71368.49320.48284.47256.46234.05215.72200.44187.51176.43166.83158.43151.01144.42
17452.07382.13332.16294.69265.55242.23223.16207.26193.81182.28172.28163.54155.83148.97
18468.44395.76343.85304.92274.64250.41230.59214.08200.10188.12177.74168.65160.64153.51
19484.80409.40355.54315.14283.73258.59238.03220.89206.39193.96183.19173.77165.45158.06
20501.16423.03367.23325.37292.82266.77245.47227.71212.69199.81188.65178.88170.26162.60
21517.52436.67378.91335.60301.91274.96252.90234.53218.98205.65194.10183.99175.07167.15
22533.88450.30390.60345.82311.00283.14260.34241.34225.27211.49199.55189.11179.89171.69
23550.25463.94402.29356.05320.09291.32267.78248.16231.56217.34205.01194.22184.70176.24
24566.61477.57413.97366.28329.18299.50275.22254.98237.86223.18210.46199.33189.51180.78
25582.97491.21425.66376.50338.27307.68282.65261.80244.15229.02215.92204.45194.32185.33
26599.33504.84437.35386.73347.36315.86290.09268.61250.44234.87221.37209.56199.14189.87
27615.69518.48449.03396.95356.45324.04297.53275.43256.74240.71226.82214.67203.95194.42
28632.05532.11460.72407.18365.54332.22304.96282.25263.03246.56232.28219.78208.76198.96
29648.42545.75472.41417.41374.63340.40312.40289.07269.32252.40237.73224.90213.57203.51
30664.78559.38484.10427.63383.72348.58319.84295.88275.62258.24243.19230.01218.39208.05
31681.14573.01495.78437.86392.81356.76327.28302.70281.91264.09248.64235.12223.20212.60
32697.50586.65507.47448.08401.90364.95334.71309.52288.20269.93254.09240.24228.01217.14
33713.86600.28519.16458.31410.99373.13342.15316.34294.49275.77259.55245.35232.82221.69
34730.23613.92530.84468.54420.08381.31349.59323.15300.79281.62265.00250.46237.63226.23
35746.59627.55542.53478.76429.17389.49357.02329.97307.08287.46270.45255.58242.45230.78

The blue area indicates that the per-procedure cost has exceeded US$200. All estimates are minimum estimates and do not include cost of treatment due to postendoscopic infection (per-procedure cost of device-related infection ranges between US$20.12 and US$46.52). Costs related to repair and reprocessing will inevitably increase at high-volume facilities with high-volume colonoscopes.

Cost of colonoscopy (in US$) including reprocessing, capital cost and repair cost by varying number of colonoscopes and annual number of procedures The blue area indicates that the per-procedure cost has exceeded US$200. All estimates are minimum estimates and do not include cost of treatment due to postendoscopic infection (per-procedure cost of device-related infection ranges between US$20.12 and US$46.52). Costs related to repair and reprocessing will inevitably increase at high-volume facilities with high-volume colonoscopes.

Comments

It costs between US$188.64 and US$501.16 per colonoscopy including purchase, maintenance and reprocessing. Per-procedure costs increase an additional US$20.12 to US$46.52 if the postprocedural infections are included. Previous colonoscopy cost estimates range from US$114.07 to US$280.71 to reprocess the colonoscope; they did not consider the capital costs to purchase the equipment nor pay for postprocedural infections (Ofstead et al)6 When we included only reprocessing and repair, our results are similar to the previous estimate at US$101.16 to US$238.71. The cost per colonoscopy is volume dependent based on the capital requirements of the colonoscope, its hardware and software and the AER, which was also found for duodenoscope and bronchoscope costs (Bang et al,9 Ofstead et al 10). Our cost estimates are minimum estimates. The true cost may be even higher once overhead costs, additional reprocessing and equipment costs are considered (eg, additional reprocessing after 7 days storage, cost of disposing single-use accessories, conducting internal audits, water and electricity, etc). At facilities with many colonoscopes available, the cost of additional reprocessing after 7 days storage might be substantial due to a low volume per colonoscope. Previous reprocessing estimates range from US$20 to US$150 (average: US$69)11 which is similar to our cost estimate of US$32.39. This study was not able to account for the cost of disposable colonoscopes because they have not entered the market yet. Disposal costs for colonoscopes will be a new cost, although it is anticipated to be similar to the cost of disposable bronchoscope disposal (US$0.06 per procedure).12 Additionally, the environmental impact is somewhat equal for disposable bronchoscopes and reusable bronchoscopes but is highly dependent on the different reprocessing standards.13 The per-procedure cost is dependent on differences in repair rate and costs, staffing costs and capital costs which may modify these estimates. Time spent on reprocessing in this study was relatively low (19 min compared with 76 min) (Ofstead et al 14) and may reflect the focus on colonoscopes exclusively in this study. The cost of reusable colonoscopes is highly dependent on the settings at each facility and especially the number of colonoscopes and annual volume of procedures. Disposable colonoscopes may decrease device-related infection transmission and may prove cost-effective for some facilities, particularly those with low-volume and high-infection risk patients. The true impact of disposable colonoscopes on infection and complication rates, cost-effectiveness and functionality for more difficult therapeutic procedures will remain unknown until they are widely available.
  7 in total

Review 1.  What's new in reprocessing endoscopes: Are we going to ensure "the needs of the patient come first" by shifting from disinfection to sterilization?

Authors:  William A Rutala; Hajime Kanamori; Emily E Sickbert-Bennett; David J Weber
Journal:  Am J Infect Control       Date:  2019-06       Impact factor: 2.918

2.  Rates of infection after colonoscopy and osophagogastroduodenoscopy in ambulatory surgery centres in the USA.

Authors:  Peiqi Wang; Tim Xu; Saowanee Ngamruengphong; Martin A Makary; Anthony Kalloo; Susan Hutfless
Journal:  Gut       Date:  2018-05-18       Impact factor: 23.059

3.  Risk of infection following colonoscopy and sigmoidoscopy in symptomatic patients.

Authors:  Jiun-Nong Lin; Chang-Bi Wang; Chih-Hui Yang; Chung-Hsu Lai; Hsi-Hsun Lin
Journal:  Endoscopy       Date:  2017-05-24       Impact factor: 10.093

4.  Cost Utility of Competing Strategies to Prevent Endoscopic Transmission of Carbapenem-Resistant Enterobacteriaceae.

Authors:  Christopher V Almario; Folasade P May; Nicholas J Shaheen; Rekha Murthy; Kapil Gupta; Laith H Jamil; Simon K Lo; Brennan M R Spiegel
Journal:  Am J Gastroenterol       Date:  2015-11-03       Impact factor: 10.864

5.  Cost analysis of single-use (Ambu® aScope™) and reusable bronchoscopes in the ICU.

Authors:  S Perbet; M Blanquet; C Mourgues; J Delmas; S Bertran; B Longère; V Boïko-Alaux; P Chennell; J-E Bazin; J-M Constantin
Journal:  Ann Intensive Care       Date:  2017-01-03       Impact factor: 6.925

6.  Concept of disposable duodenoscope: at what cost?

Authors:  Ji Young Bang; Bryce Sutton; Robert Hawes; Shyam Varadarajulu
Journal:  Gut       Date:  2019-02-12       Impact factor: 23.059

7.  Micro-costing studies in the health and medical literature: protocol for a systematic review.

Authors:  Xiao Xu; Holly K Grossetta Nardini; Jennifer Prah Ruger
Journal:  Syst Rev       Date:  2014-05-21
  7 in total
  10 in total

1.  Robotic Autonomy for Magnetic Endoscope Biopsy.

Authors:  James W Martin; Lavinia Barducci; Bruno Scaglioni; Joseph C Norton; Conchubhair Winters; Venkataraman Subramanian; Alberto Arezzo; Keith L Obstein; Pietro Valdastri
Journal:  IEEE Trans Med Robot Bionics       Date:  2022-06-29

Review 2.  Gastrointestinal Endoscopy-Associated Infections: Update on an Emerging Issue.

Authors:  Anasua Deb; Abhilash Perisetti; Hemant Goyal; Mark M Aloysius; Sonali Sachdeva; Dushyant Dahiya; Neil Sharma; Nirav Thosani
Journal:  Dig Dis Sci       Date:  2022-03-09       Impact factor: 3.487

3.  Active Stabilization of Interventional Tasks Utilizing a Magnetically Manipulated Endoscope.

Authors:  Lavinia Barducci; Bruno Scaglioni; James Martin; Keith L Obstein; Pietro Valdastri
Journal:  Front Robot AI       Date:  2022-04-14

Review 4.  Gastrointestinal endoscope contamination rates - elevators are not only to blame: a systematic review and meta-analysis.

Authors:  Hemant Goyal; Sara Larsen; Abhilash Perisetti; Nikolaj Birk Larsen; Lotte Klinten Ockert; Sven Adamsen; Benjamin Tharian; Nirav Thosani
Journal:  Endosc Int Open       Date:  2022-06-10

Review 5.  Guidelines for Robotic Flexible Endoscopy at the Time of COVID-19.

Authors:  Onaizah Onaizah; Zaneta Koszowska; Conchubhair Winters; Venkatamaran Subramanian; David Jayne; Alberto Arezzo; Keith L Obstein; Pietro Valdastri
Journal:  Front Robot AI       Date:  2021-02-25

6.  The global, regional, and national burden and quality of care index (QCI) of colorectal cancer; a global burden of disease systematic analysis 1990-2019.

Authors:  Seyed Aria Nejadghaderi; Shahin Roshani; Esmaeil Mohammadi; Moein Yoosefi; Negar Rezaei; Zahra Esfahani; Sina Azadnajafabad; Naser Ahmadi; Sarvenaz Shahin; Ameneh Kazemi; Alireza Namazi Shabestari; Ardeshir Khosravi; Ali H Mokdad; Bagher Larijani; Farshad Farzadfar
Journal:  PLoS One       Date:  2022-04-21       Impact factor: 3.752

7.  The feasibility and safety of disposable endoscope vs. conventional endoscope for upper gastrointestinal tract examination: a multicenter, randomized, parallel, non-inferiority trial.

Authors:  De-Feng Li; Rui-Yue Shi; Yan-Hui Tian; Zheng-Lei Xu; Ying-Sheng Zhou; Xian-Jiu Sun; Jin-Wei Cai; Yuan-Yuan Fang; Huan Peng; Jia-Min Wang; Tao Dong; Yong-de Cai; Jun Yao; Li-Sheng Wang
Journal:  Z Gastroenterol       Date:  2021-11-12       Impact factor: 1.769

Review 8.  Robotic, self-propelled, self-steerable, and disposable colonoscopes: Reality or pipe dream? A state of the art review.

Authors:  Conchubhair Winters; Venkataraman Subramanian; Pietro Valdastri
Journal:  World J Gastroenterol       Date:  2022-09-21       Impact factor: 5.374

9.  Enabling the future of colonoscopy with intelligent and autonomous magnetic manipulation.

Authors:  James W Martin; Bruno Scaglioni; Joseph C Norton; Venkataraman Subramanian; Alberto Arezzo; Keith L Obstein; Pietro Valdastri
Journal:  Nat Mach Intell       Date:  2020-10-12

Review 10.  Endorobots for Colonoscopy: Design Challenges and Available Technologies.

Authors:  Luigi Manfredi
Journal:  Front Robot AI       Date:  2021-07-14
  10 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.