Nadine Amelung1, Volker Maus2, Daniel Behme2, Ismini E Papageorgiou3,4, Johanna Rosemarie Leyhe2, Michael Knauth2, Marios Nikos Psychogios2.
Abstract
BACKGROUND: Flat detector CT - angiography (FDCTA) has become a valuable imaging tool in post- and peri-interventional imaging after neurovascular procedures. Metal artifacts produced by radiopaque implants like clips or coils still impair image quality.Entities:
Keywords: Angiography; Brain/ brain stem; CT metal artifacts reduction; Coiling; Flat-detector CT angiography
Mesh:
Year: 2019 PMID: 31412810 PMCID: PMC6694691 DOI: 10.1186/s12880-019-0352-2
Source DB: PubMed Journal: BMC Med Imaging ISSN: 1471-2342 Impact factor: 1.930
Indication and type of examination are listed for each patient
| Patient | Target Vessels | Metal Implants | Protocol |
|---|---|---|---|
| 1 | ACom | clip, coiling | iv FDCTA 10s |
| 2 | ACom | coiling | 5 s rotational angiography |
| 3 | ICA | flow diverter, clip | iv FDCTA 10s |
| flow diverter, clip | iv FDCTA 20s | ||
| 4 | ICA | clip, stent-assisted coiling | iv FDCTA 10s |
| 5 | ACom | coiling | 5 s rotational angiography |
| 6 | ACom | coiling | 5 s rotational angiography |
| 7 | VA | coiling | 5 s rotational angiography |
| 8 | ICA | coiling | 5 s rotational angiography |
| 9 | MCA | clip | 5 s rotational angiography |
| 10 | MCA | 3 clips, coiling | iv FDCTA 10s |
| 11 | ACom | coiling | 5 s rotational angiography |
| 12 | ACom | coiling | 5 s rotational angiography |
| 13 | ICA | coiling | 5 s rotational angiography |
| 14 | ACom | coiling | 5 s rotational angiography |
| 15 | ACom | stent-assisted coiling | 5 s rotational angiography |
| stent assisted coiling | iv FDCTA 20s | ||
| 16 | ICA | 2 clips, coiling | 5 s rotational angiography |
| 17 | ACom | coiling | 5 s rotational angiography |
| 18 | ACom | coiling | 5 s rotational angiography |
| 19 | ACom | coiling | 5 s rotational angiography |
| 20 | ACom | coiling | 5 s rotational angiography |
| 21 | ICA and PCA | 3 clips | 5 s rotational angiography |
| 3 clips | iv FDCTA 10s | ||
| 3 clips | 5 s rotational angiography |
Abbreviations: ACom anterior communicating artery, ICA internal carotid artery, VA vertebral artery, MCA middle cerebral artery, PCA posterior cerebral artery
Fig. 1ROC curve of the results of both raters concerning detection or exclusion of reperfusion before and after application of MARA. Application of MARA improved sensitivity
Fig. 2In those cases, where reperfusion was diagnosed via DSA, application of MARA increased the detection of reperfusion. Eight out of ten cases of reperfusion were diagnosed correctly after application of MARA, whereas only 6 cases were diagnosed in the images without MARA. Additionally there were 2 cases, in which reperfusion should have been detected but images without MARA were not diagnostic
Fig. 3In image a multiple streak artifacts impair evaluation in the level of implant, whereas in image b after application of MARA the image quality is improved significantly and the streak artifacts are eliminated
Fig. 4In image a a statement concerning residual perfusion of the aneurysm was not possible due to streak artifacts and image hardening caused by the coil material. After application of MARA the residual perfusion could be detected as seen in image b, congruent to the result of the digital subtraction angiography as seen in image c