| Literature DB >> 31410699 |
Sophie Murer1, Juergen Scheidler1,2, Ulrike L Mueller-Lisse3,4, Marissa Helling1, Michael Scherr1,5, Ullrich G Mueller-Lisse6.
Abstract
BACKGROUND: Application of an endorectal coil (ERC) for 3.0-T prostate magnetic resonance imaging (MRI) is contentious. We hypothesised that a multicoil phased-array protocol provides T2-weighted images (T2WI) and diffusion-weighted images (DWI) with reduced field-of-view (DWIreduced) and monoexponential apparent diffusion coefficient (ADC) maps that are technically equivalent with ERC or without ERC (noERC).Entities:
Keywords: Actinidia deliciosa; Diffusion-weighted imaging; Magnetic resonance imaging; Phantoms (imaging); Prostate
Mesh:
Year: 2019 PMID: 31410699 PMCID: PMC6692805 DOI: 10.1186/s41747-019-0111-8
Source DB: PubMed Journal: Eur Radiol Exp ISSN: 2509-9280
Fig. 2Axial images of a green Hayward kiwifruit, Actinidia deliciosa (a digital photograph of fresh cut section; b corresponding T2-weighted image [T2WI]; OP, outer pericarp; PP, peripheral placenta) and a human prostate (c T2WI: PZa, anterior region of peripheral zone; PZpl, posterior region of peripheral zone; TZa, anterior region of transitional zone; TZp, posterior region of transitional zone). Respective placement of elliptical regions of interest (ROIs) comprising at least ten adjacent voxels is demonstrated in axial images of a kiwifruit (d diffusion-weighted image, b value 800 s/mm2; e T2WI) and human prostate (f T2WI)
Fig. 1Setup of kiwifruit phantom and comparison with a male human pelvis. Respective axial MR-images of kiwifruit phantom (a, c) and human pelvis (b, d) examined with bi-parametric magnetic resonance imaging at 3.0 T without (a, b) and with an endo-rectal coil (c, d) included in a multichannel phased-array coil system show similar coil-load; “Bo” indicates copper-sulphate solution bottle, “Gel” indicates gel compress-packs, and “W” indicates open-spout plastic watering-can. Photographs (e, f) show a watering can (W) with endo-rectal coil handle fixed in the spout (arrow) with a wine-cork and rubber-band, and set-up complete with supporting sandbags (Sa) and abdominal phased-array surface coil (BPA)
Magnetic resonance sequences and parameters for axial imaging in kiwifruits and human prostates at two different institutions (site 1, site 2)
| Parameter | T2WI site 1 | T2WI site 2 | DWIreduced site 1 | DWIreduced site 2 |
|---|---|---|---|---|
| Field of view (mm2) | 180 × 180 | 160 × 160 | 149 × 73 | 179 × 71 |
| Imaging matrix | 320 × 320 | 320 × 320 | 90 × 44 | 140 × 42 |
| Slice thickness (mm) | 3.0 | 3.0 | 3.5 | 3.5 |
| Slice spacing (mm) | 3.0 | 3.3 | 3.5 | 3.5 |
| Voxel size (mm3) | 0.56 × 0.56 × 3.00 | 0.50 × 0.50 × 3.00 | 1.66 × 1.66 × 3.50 | 1.28 × 1.69 × 3.50 |
| No. of slices | 26 | 30 | 20 | 22 |
| PAT mode | GRAPPA | GRAPPA | n/a | n/a |
| Acceleration factor PE | 2 | 2 | n/a | n/a |
| Echo time (ms) | 101 | 98 | 74 | 74 |
| Echo train length | 25 | 17 | 38 | 42 |
| Time to repeat (ms) | 7500 | 7400 | 4500 | 4500 |
| Bandwidth (Hz/px) | 200 | 180 | 1500 | 1190 |
| n/a | n/a | 50, 400, 800 | 50, 800 | |
| Number of excitations | 3 | 2 | 2, 5, 12 (per | 4, 16 (per |
| Acquisition time (min:s) | 04:22 | 04:38 | 04:32 | 04:44 |
| Flip angle | 160° | 120° | 90° | 90° |
| Percent sampling | 100 | 100 | 100 | 75 |
| Percent phase field of view | 100 | 100 | 48.9 | 40 |
ADC apparent diffusion coefficient, DWI echo-planar diffusion-weighted imaging with reduced field of view and selective excitation, GRAPPA generalised autocalibrating partially parallel acquisition, n/a not applicable, PAT parallel acquisition technique, T2WI T2-weighted turbo-spin-echo imaging
Signal-to-noise ratios in kiwifruits examined with and without an endorectal coil at site 1 (twelve corresponding regions of interest in anterior and posterior locations) and at site 2 (three corresponding regions of interest in anterior and posterior locations)
| Kiwifruits, outer pericarp, site 1 | ERC, | noERC, | ERC versus noERC | |||||
| SNR, anterior (mean ± SD) | SNR, posterior (mean ± SD) | SNR, anterior (mean ± SD) | SNR, posterior (mean ± SD | SNR, anterior | SNR, posterior | |||
| T2WI | 14.2 ± 2.4 | 13.9 ± 3.7 | 13.3 ± 2.4 | 13.1 ± 2.7 | ||||
| DWIreduced
| 4.8 ± 2.4 | 7.8 ± 2.7 | 0.013 | 7.4 ± 4.0 | 8.2 ± 2.4 | 0.005 | ||
| DWIreduced
| 7.6 ± 7.1 | 7.3 ± 2.8 | 7.0 ± 3.2 | 10.7 ± 4.7 | 0.044 | 0.047 | ||
| DWIreduced ADC | 5.3 ± 2.2 | 21.9 ± 12.0 | < 0.001 | 10.5 ± 5.5 | 34.1 ± 13.3 | < 0.001 | 0.002 | 0.024 |
| Kiwifruits, outer pericarp, site 2 | ERC, | noERC, | ERC versus noERC | |||||
| SNR, anterior (mean ± SD) | SNR, posterior (mean ± SD) | SNR, anterior (mean ± SD) | SNR, posterior (mean ± SD | SNR, anterior | SNR, posterior | |||
| T2WI | 19.8 ± 5.1 | 17.4 ± 2.9 | 14.5 ± 6.8 | 18.5 ± 6.7 | ||||
| DWIreduced
| 8.8 ± 1.5 | 7.1 ± 1.4 | 6.6 ± 3.1 | 8.4 ± 1.5 | ||||
| DWIreduced
| 5.3 ± 4.2 | 8.2 ± 1.2 | 9.9 ± 0.5 | 7.8 ± 1.4 | ||||
| DWIreduced ADC | 10.8 ± 3.0 | 41.7 ± 13.1 | 5.2 ± 0.6 | 26.7 ± 9.8 | ||||
| Kiwifruits, peripheral placenta, site 1 | ERC, | noERC, | ERC versus noERC | |||||
| SNR, anterior (mean ± SD) | SNR, posterior (mean ± SD) | SNR, anterior (mean ± SD) | SNR, posterior (mean ± SD | SNR, anterior, | SNR, posterior | |||
| T2WI | 9.4 ± 3.9 | 8.6 ± 2.6 | 7.8 ± 2.6 | 8.1 ± 2.9 | ||||
| DWIreduced
| 7.6 ± 4.3 | 6.1 ± 1.9 | 9.1 ± 2.8 | 9.1 ± 4.9 | ||||
| DWIreduced
| 7.8 ± 3.1 | 7.0 ± 3.1 | 10.5 ± 2.9 | 11.5 ± 6.2 | 0.046 | |||
| DWIreduced ADC | 15.1 ± 8.2 | 15.2 ± 8.6 | 26.7 ± 16.1 | 25.5 ± 9.2 | 0.036 | 0.010 | ||
| Kiwifruits, peripheral placenta, site 2 | ERC, | noERC, | ERC versus noERC | |||||
| SNR, anterior (mean ± SD) | SNR, posterior (mean ± SD) | SNR, anterior (mean ± SD) | SNR, posterior (mean ± SD) | SNR, anterior | SNR, posterior, | |||
| T2WI | 9.0 ± 0.7 | 9.8 ± 3.5 | 9.0 ± 0.7 | 10.4 ± 2.9 | ||||
| DWIreduced
| 7.1 ± 2.0 | 6.8 ± 0.2 | 6.5 ± 1.8 | 8.7 ± 2.1 | 0.010 | |||
| DWIreduced
| 7.2 ± 1.3 | 6.9 ± 2.2 | 8.6 ± 0.4 | 8.4 ± 1.6 | ||||
| DWIreduced ADC | 24.7 ± 2.2 | 35.6 ± 14.0 | 22.5 ± 11.4 | 22.1 ± 4.3 | ||||
p values were obtained using Student’s t test for paired samples
ADC apparent diffusion coefficient, DWI echo-planar diffusion-weighted imaging with reduced field of view and selective excitation, ERC with endorectal coil, noERC without endorectal coil, ROI region-of-interest, SD standard deviation, SNR signal-to-noise ratio, T2WI T2-weighted turbo-spin-echo imaging
Signal-to-noise ratio in human prostates with benign hyperplasia examined with endorectal coil filled with air in site 2 (n = 66 corresponding regions of interests, in anterior and posterior locations, respectively), and without endorectal coil at site 1 (n = 60 corresponding regions of interest, in anterior and posterior locations, respectively)
| Prostates, peripheral zone | ERC, site 2, | noERC, site 1, | ||||
| SNR, PZa (mean ± SD) | SNR, PZpl (mean ± SD) | SNR, Pza (mean ± SD) | SNR, PZpl (mean ± SD) | |||
| T2WI | 6.2 ± 2.6 | 7.0 ± 3.0 | 4.4 ± 1.2 | 6.0 ± 2.8 | < 0.001 | |
| DWIreduced
| 4.9 ± 2.2 | 5.8 ± 3.0 | 0.026 | 5.8 ± 4.2 | 5.4 ± 2.7 | |
| DWIreduced
| 8.0 ± 4.3 | 11.0 ± 6.0 | 0.001 | 12.6 ± 7.1 | 12.6 ± 6.5 | |
| DWIreduced ADC | 6.0 ± 2.3 | 8.2 ± 4.1 | < 0.001 | 6.4 ± 4.4 | 9.3 ± 12.3 | |
| Prostates, transitional zone | ERC, site 2, | noERC, site 1, | ||||
| SNR, Tza (mean ± SD) | SNR, TZp (mean ± SD) | SNR, Tza (mean ± SD) | SNR, TZp (mean ± SD) | |||
| T2WI | 8.3 ± 3.5 | 7.6 ± 3.2 | 5.1 ± 1.9 | 6.1 ± 6.4 | ||
| DWIreduced
| 7.5 ± 2.8 | 7.1 ± 2.7 | 8.4 ± 3.7 | 8.6 ± 3.9 | ||
| DWIreduced
| 13.5 ± 5.4 | 13.7 ± 5.0 | 16.5 ± 5.8 | 16.5 ± 6.1 | ||
| DWIreduced ADC | 9.4 ± 3.5 | 8.7 ± 3.6 | 9.2 ± 3.5 | 11.1 ± 9.8 | ||
p values were obtained using Student’s t test for paired samples
ADC apparent diffusion coefficient, DWI echo-planar diffusion-weighted imaging with reduced field of view and selective excitation, ERC with endo-rectal coil, noERC without endo-rectal coil, Pza anterior region of peripheral zone of human prostate, PZpl posterior-lateral region of peripheral zone of human prostate, SNR signal-to-noise ratio, T2WI T2-weighted turbo-spin-echo imaging, Tza anterior region of transitional zone of human prostate; TZp posterior region of transitional zone of human prostate
Signal-to-noise ratios in one kiwifruit examined with an endorectal coil filled with air at both site 1 and site 2, and with perfluorocarbon in site 2
| Kiwifruit, outer pericarp ( | With air-filled ERC, site 1 | With air-filled ERC, site 2 | With PFC-filled ERC, site 2 | ||||||
| SNR, anterior (mean ± SD) | SNR, posterior (mean ± SD) | SNR, anterior (mean ± SD) | SNR, posterior (mean ± SD) | SNR, anterior (mean ± SD) | SNR, posterior (mean ± SD) | ||||
| T2WI | 16.0 ± 1.3 | 16.7 ± 2.2 | 15.4 ± 2.9 | 9.0 ± 1.3 | 0.047 | 12.2 ± 1.7 | 11.8 ± 0.4 | ||
| DWIreduced b = 50 | 4.0 ± 1.2 | 7.0 ± 2.1 | 6.2 ± 3.3 | 8.3 ± 1.6 | 8.2 ± 1.5 | 9.3 ± 3.0 | |||
| DWIreduced b = 800 | 4.6 ± 3.4 | 6.0 ± 1.0 | 9.9 ± 2.4 | 9.5 ± 2.3 | 9.7 ± 2.4 | 8.0 ± 1.6 | |||
| DWIreduced ADC | 7.2 ± 1.3 | 29.5 ± 12.9 | 12.9 ± 2.9 | 40.3 ± 22.5 | 13.6 ± 6.2 | 56.8 ± 9.1 | 0.040 | ||
| Kiwifruit, outer pericarp | ERC air site 1 versus site 2 | ERC air versus PFC site 2 | |||||||
| SNR, anterior | SNR, posterior, | SNR, anterior, | SNR, posterior, | ||||||
| T2WI | |||||||||
| DWIreduced
| |||||||||
| DWIreduced
| |||||||||
| DWIreduced ADC | |||||||||
| Kiwifruit, peripheral placenta ( | With air-filled ERC, site 2 | With air-filled ERC, site 2 | With PFC-filled ERC, site 2 | ||||||
| SNR, anterior (mean ± SD) | SNR, posterior (mean ± SD) | SNR, anterior (mean ± SD) | SNR, posterior (mean ± SD) | SNR, anterior (mean ± SD) | SNR, posterior (mean ± SD) | ||||
| T2WI | 9.8 ± 2.8 | 9.7 ± 3.3 | 7.5 ± 0.7 | 8.7 ± 0.7 | 6.0 ± 0.6 | 7.0 ± 2.0 | |||
| DWIreduced
| 5.9 ± 2.0 | 6.0 ± 0.6 | 9.2 ± 1.1 | 7.4 ± 1.4 | 8.9 ± 1.3 | 5.9 ± 0.8 | |||
| DWIreduced
| 7.0 ± 0.9 | 5.3 ± 1.4 | 9.9 ± 0.9 | 8.0 ± 2.8 | 11.8 ± 2.2 | 6.8 ± 1.2 | |||
| DWIreduced ADC | 22.5 ± 8.4 | 21.8 ± 7.5 | 28.5 ± 12.3 | 23.2 ± 9.7 | 25.0 ± 13.0 | 25.9 ± 3.9 | |||
| Kiwifruit, peripheral placenta | ERC air site 1 versus site 2 | ERC air versus PFC site 2 | |||||||
| SNR, anterior, | SNR, posterior, | SNR, anterior, | SNR, posterior | ||||||
| T2WI | |||||||||
| DWIreduced
| |||||||||
| DWIreduced
| |||||||||
| DWIreduced ADC | |||||||||
p values were obtained using Student’s t test for paired samples
ADC apparent diffusion coefficient, DWI echo-planar diffusion-weighted imaging with reduced field of view and selective excitation, ERC endorectal coil, PFC perfluorocarbon, ROI region of interest, SD standard deviation, SNR signal-to-noise ratio, T2WI T2-weighted turbo-spin-echo imaging
Diameters and volumes of kiwifruits (n = 5) and human prostates (without endorectal coil, n = 10; with endorectal coil, n = 11)
| Diameter (mean ± SD) | Volume (mean ± SD) (cm3) | |||
|---|---|---|---|---|
| AP (cm) | LR (cm) | CC (cm) | ||
| Kiwifruits, noERC | 4.9 ± 0.1 | 5.3 ± 0.2 | 6.5 ± 0.5 | 89.2 ± 11.2 |
| Kiwifruits,_ERC | 4.8 ± 0.1 | 5.4 ± 0.2 | 6.5 ± 0.4 | 89.7 ± 9.5 |
| Prostates,_noERC | 4.6 ± 1.0 | 6.0 ± 1.1 | 5.7 ± 1.3 | 90.8 ± 48.5 |
| Prostates, ERC | 4.2 ± 0.9 | 6.0 ± 0.5 | 6.5 ± 1.0 | 87.5 ± 35.7 |
| Kiwifruits_noERC/ERC | ||||
| Prostates,_noERC/ERC | ||||
| noERC, kiwifruits versus prostates | ||||
| ERC, kiwifruits versus prostates | 0.028 | 0.005 | ||
p values were obtained using Student’s t test
AP anterior-posterior, CC cranio-caudal, ERC with endorectal coil, LR left-to-right, noERC without endorectal coil, SD standard deviation
Fig. 3Biparametric magnetic resonance imaging at 3.0 T of a green Hayward kiwifruit (Actinidia deliciosa) was performed at site 1 both without (a–d) and with an endo-rectal coil (e–h) included in a multichannel phased-array coil system: axial (a, e) and sagittal (b, f) T2-weighted imaging (T2WI), axial diffusion-weighted imaging with a b value of 800 s/mm2 (c, g), and corresponding apparent diffusion coefficient map (d, h) obtained with a reduced field of view
Fig. 4Biparametric magnetic resonance imaging at 3.0 T in a green Hayward kiwifruit (Actinidia deliciosa) was performed at site 2 using an endorectal coil included in a multichannel phased-array coil system. Images in the upper panel (a–d) were obtained with air inside the endorectal coil balloon, while images in the lower panel (e–h) were obtained with perfluorocarbon inside the endorectal-coil balloon: axial (a, e) and sagittal (b, f) T2-weighted imaging, axial diffusion-weighted imaging with a b value of 800 s/mm2 (c, g ), and corresponding apparent diffusion coefficient map (d, h) obtained with a reduced field of view
Fig. 5Biparametric magnetic resonance imaging at 3.0 T in patients with benign prostate hyperplasia was performed without endorectal coil at site 1 (a–d) and with endorectal coil at site 2 (e–h) included in a multichannel phased-array coil system: axial (a, e) and sagittal (b, f) T2-weighted imaging, axial diffusion-weighted imaging with a b value of 800 s/mm2 (c, g), and corresponding apparent diffusion coefficient map (d, h) obtained with a reduced field of view
Diameters of kiwifruit phantom and human pelvis, and respective distances from the posterior surface of kiwifruit and human prostates to the scanner table
| Diameter (mean ± SD) | Distance from kiwifruit or prostate to scanner table (mean ± SD) (cm) | ||
|---|---|---|---|
| AP (cm) | LR (cm) | ||
| Kiwifruits,_noERC | 21.6 ± 2.3 | 37.5 ± 2.3 | 8.0 ± 0.3 |
| Kiwifruits, ERC | 23.1 ± 2.4 | 36.9 ± 1.9 | 9.1 ± 0.2 |
| Patients’_pelvis,_noERC | 23.0 ± 2.4 | 39.7 ± 2.8 | 9.9 ± 0.9 |
| Patients’_pelvis,_ERC | 20.8 ± 1.4 | >36.0 | 10.0 ± 0.8 |
| Kiwifruits,_noERC versus ERC | 0.002 | ||
| Patients, noERC versus ERC | 0.021 | n/a | |
| noERC, kiwifruits versus patients | 0.006 | ||
| ERC, kiwifruits versus patients | n/a | < 0.001 | |
p values were obtained using Student’s t test
AP anterior-posterior, CC cranio-caudal, ERC with endorectal coil, LR left-to-right, noERC without endorectal coil, SD standard deviation