Or Perlman1, Kai Herz2,3, Moritz Zaiss2, Ouri Cohen4, Matthew S Rosen1,5, Christian T Farrar1. 1. Athinoula A. Martinos Center for Biomedical Imaging, Department of Radiology, Massachusetts General Hospital and Harvard Medical School, Charlestown, Massachusetts. 2. Magnetic Resonance Center, Max Planck Institute for Biological Cybernetics, Tübingen, Germany. 3. IMPRS for Cognitive and Systems Neuroscience, University of Tübingen, Tübingen, Germany. 4. Memorial Sloan Kettering Cancer Center, New York, New York. 5. Department of Physics, Harvard University, Cambridge, Massachusetts.
Abstract
PURPOSE: To understand the influence of various acquisition parameters on the ability of CEST MR-Fingerprinting (MRF) to discriminate different chemical exchange parameters and to provide tools for optimal acquisition schedule design and parameter map reconstruction. METHODS: Numerical simulations were conducted using a parallel computing implementation of the Bloch-McConnell equations, examining the effect of TR, TE, flip-angle, water T 1 and T 2 , saturation-pulse duration, power, and frequency on the discrimination ability of CEST-MRF. A modified Euclidean distance matching metric was evaluated and compared to traditional dot product matching. L-Arginine phantoms of various concentrations and pH were scanned at 4.7T and the results compared to numerical findings. RESULTS: Simulations for dot product matching demonstrated that the optimal flip-angle and saturation times are 30 ∘ and 1100 ms, respectively. The optimal maximal saturation power was 3.4 μT for concentrated solutes with a slow exchange rate, and 5.2 μT for dilute solutes with medium-to-fast exchange rates. Using the Euclidean distance matching metric, much lower maximum saturation powers were required (1.6 and 2.4 μT, respectively), with a slightly longer saturation time (1500 ms) and 90 ∘ flip-angle. For both matching metrics, the discrimination ability increased with the repetition time. The experimental results were in agreement with simulations, demonstrating that more than a 50% reduction in scan-time can be achieved by Euclidean distance-based matching. CONCLUSIONS: Optimization of the CEST-MRF acquisition schedule is critical for obtaining the best exchange parameter accuracy. The use of Euclidean distance-based matching of signal trajectories simultaneously improved the discrimination ability and reduced the scan time and maximal saturation power required.
PURPOSE: To understand the influence of various acquisition parameters on the ability of CEST MR-Fingerprinting (MRF) to discriminate different chemical exchange parameters and to provide tools for optimal acquisition schedule design and parameter map reconstruction. METHODS: Numerical simulations were conducted using a parallel computing implementation of the Bloch-McConnell equations, examining the effect of TR, TE, flip-angle, water T 1 and T 2 , saturation-pulse duration, power, and frequency on the discrimination ability of CEST-MRF. A modified Euclidean distance matching metric was evaluated and compared to traditional dot product matching. L-Arginine phantoms of various concentrations and pH were scanned at 4.7T and the results compared to numerical findings. RESULTS: Simulations for dot product matching demonstrated that the optimal flip-angle and saturation times are 30 ∘ and 1100 ms, respectively. The optimal maximal saturation power was 3.4 μT for concentrated solutes with a slow exchange rate, and 5.2 μT for dilute solutes with medium-to-fast exchange rates. Using the Euclidean distance matching metric, much lower maximum saturation powers were required (1.6 and 2.4 μT, respectively), with a slightly longer saturation time (1500 ms) and 90 ∘ flip-angle. For both matching metrics, the discrimination ability increased with the repetition time. The experimental results were in agreement with simulations, demonstrating that more than a 50% reduction in scan-time can be achieved by Euclidean distance-based matching. CONCLUSIONS: Optimization of the CEST-MRF acquisition schedule is critical for obtaining the best exchange parameter accuracy. The use of Euclidean distance-based matching of signal trajectories simultaneously improved the discrimination ability and reduced the scan time and maximal saturation power required.
Keywords:
chemical exchange rate; chemical exchange saturation transfer (CEST); magnetic resonance fingerprinting (MRF); optimization; pH; quantitative imaging
Authors: Moritz Zaiss; Goran Angelovski; Eleni Demetriou; Michael T McMahon; Xavier Golay; Klaus Scheffler Journal: Magn Reson Med Date: 2017-07-07 Impact factor: 4.668
Authors: Rachelle Crescenzi; Catherine DeBrosse; Ravi P R Nanga; Matthew D Byrne; Guruprasad Krishnamoorthy; Kevin D'Aquilla; Hari Nath; Knashawn H Morales; Michiyo Iba; Hari Hariharan; Virginia M Y Lee; John A Detre; Ravinder Reddy Journal: Hippocampus Date: 2017-02-03 Impact factor: 3.899
Authors: Moritz Zaiss; Zhongliang Zu; Junzhong Xu; Patrick Schuenke; Daniel F Gochberg; John C Gore; Mark E Ladd; Peter Bachert Journal: NMR Biomed Date: 2014-12-15 Impact factor: 4.044
Authors: Jinyuan Zhou; Jaishri O Blakeley; Jun Hua; Mina Kim; John Laterra; Martin G Pomper; Peter C M van Zijl Journal: Magn Reson Med Date: 2008-10 Impact factor: 4.668
Authors: Junzhong Xu; Moritz Zaiss; Zhongliang Zu; Hua Li; Jingping Xie; Daniel F Gochberg; Peter Bachert; John C Gore Journal: NMR Biomed Date: 2014-01-29 Impact factor: 4.044
Authors: Xiao-Yong Zhang; Feng Wang; Junzhong Xu; Daniel F Gochberg; John C Gore; Zhongliang Zu Journal: NMR Biomed Date: 2017-11-29 Impact factor: 4.044
Authors: Peter C M van Zijl; Kevin Brindle; Hanzhang Lu; Peter B Barker; Richard Edden; Nirbhay Yadav; Linda Knutsson Journal: Curr Opin Chem Biol Date: 2021-07-20 Impact factor: 8.972
Authors: Kai Herz; Sebastian Mueller; Or Perlman; Maxim Zaitsev; Linda Knutsson; Phillip Zhe Sun; Jinyuan Zhou; Peter van Zijl; Kerstin Heinecke; Patrick Schuenke; Christian T Farrar; Manuel Schmidt; Arnd Dörfler; Klaus Scheffler; Moritz Zaiss Journal: Magn Reson Med Date: 2021-05-07 Impact factor: 3.737