Kelly N Jahn1, Julie G Arenberg2. 1. 1 Department of Speech and Hearing Sciences, University of Washington, Seattle, WA, USA. 2. 2 Massachusetts Eye and Ear, Department of Otolaryngology, Harvard Medical School, Boston, MA, USA.
Abstract
Modeling data suggest that sensitivity to the polarity of an electrical stimulus may reflect the integrity of the peripheral processes of the spiral ganglion neurons. Specifically, better sensitivity to anodic (positive) current than to cathodic (negative) current could indicate peripheral process degeneration or demyelination. The goal of this study was to characterize polarity sensitivity in pediatric and adult cochlear implant listeners (41 ears). Relationships between polarity sensitivity at threshold and (a) polarity sensitivity at suprathreshold levels, (b) age-group, (c) preimplantation duration of deafness, and (d) phoneme perception were determined. Polarity sensitivity at threshold was defined as the difference in single-channel behavioral thresholds measured in response to each of two triphasic pulses, where the central high-amplitude phase was either cathodic or anodic. Lower thresholds in response to anodic than to cathodic pulses may suggest peripheral process degeneration. On the majority of electrodes tested, threshold and suprathreshold sensitivity was lower for anodic than for cathodic stimulation; however, dynamic range was often larger for cathodic than for anodic stimulation. Polarity sensitivity did not differ between child- and adult-implanted listeners. Adults with long preimplantation durations of deafness tended to have better sensitivity to anodic pulses on channels that were estimated to interface poorly with the auditory nerve; this was not observed in the child-implanted group. Across subjects, duration of deafness predicted phoneme perception performance. The results of this study suggest that subject- and electrode-dependent differences in polarity sensitivity may assist in developing customized cochlear implant programming interventions for child- and adult-implanted listeners.
Modeling data suggest that sensitivity to the polarity of an electrical stimulus may reflect the integrity of the peripheral processes of the spiral ganglion neurons. Specifically, better sensitivity to anodic (positive) current than to cathodic (negative) current could indicate peripheral process degeneration or demyelination. The goal of this study was to characterize polarity sensitivity in pediatric and adult cochlear implant listeners (41 ears). Relationships between polarity sensitivity at threshold and (a) polarity sensitivity at suprathreshold levels, (b) age-group, (c) preimplantation duration of deafness, and (d) phoneme perception were determined. Polarity sensitivity at threshold was defined as the difference in single-channel behavioral thresholds measured in response to each of two triphasic pulses, where the central high-amplitude phase was either cathodic or anodic. Lower thresholds in response to anodic than to cathodic pulses may suggest peripheral process degeneration. On the majority of electrodes tested, threshold and suprathreshold sensitivity was lower for anodic than for cathodic stimulation; however, dynamic range was often larger for cathodic than for anodic stimulation. Polarity sensitivity did not differ between child- and adult-implanted listeners. Adults with long preimplantation durations of deafness tended to have better sensitivity to anodic pulses on channels that were estimated to interface poorly with the auditory nerve; this was not observed in the child-implanted group. Across subjects, duration of deafness predicted phoneme perception performance. The results of this study suggest that subject- and electrode-dependent differences in polarity sensitivity may assist in developing customized cochlear implant programming interventions for child- and adult-implanted listeners.
Physiological and psychophysical responses vary considerably across stimulation sites
within individual cochlear implant (CI) listeners (e.g., DeVries & Arenberg, 2018b; DeVries, Scheperle, & Bierer,
2016; Pfingst &
Xu, 2004; Zhu, Tang,
Zeng, Guan, & Ye, 2012). Some of this variability may, in part,
result from within- and across-subject variation in spiral ganglion neuron (SGN)
integrity. Human postmortem temporal bone studies demonstrate that SGN density
varies widely across individuals and across the length of the cochlea within an
individual ear (Hinojosa &
Marion, 1983; Makary,
Shin, Kujawa, Liberman, & Merchant, 2011; Nadol, 1997; Nadol, Young, & Glynn, 1989; Otte, Schuknecht, & Kerr,
1978). In those temporal bone analyses, variability in SGN counts is
partially explained by demographic variables such as chronological age, duration of
deafness, and hearing loss etiology. Animal studies also indicate that long-term
auditory deprivation is associated with reduced SGN survival relative to normal
(e.g., Hall, 1990; Heffer et al., 2010; Ramekers et al., 2014;
Shepherd & Javel,
1997).Despite the observed relationships between SGN loss and multiple demographic
characteristics, the functional consequences of reduced SGN density remain
ambiguous. Signal detection theory analysis suggests that a dramatic loss of SGNs
(∼75%) may be required to detect measurable, albeit small, deficits in
psychoacoustic perception (Oxenham, 2016). Similarly, empirical data in humans do not demonstrate
consistent relationships between postmortem SGN density and speech perception scores
obtained during an individual’s lifetime (Fayad & Linthicum, 2006; Khan et al., 2005; Nadol et al., 2001; Otte et al., 1978).In vivo estimates of neural status may improve our ability to study how the integrity
of the SGNs relates to auditory perception with a CI during life. Over the years,
several proposed estimates of SGN density have been evaluated in animals and in
humans. In animal models, evoked potential responses vary systematically as a
function of SGN density; for instance, electrodes near cochlear regions with
relatively few surviving SGNs tend to have relatively small evoked potential
amplitudes, high evoked potential thresholds, and shallow evoked potential amplitude
growth functions (AGFs; Hall,
1990; Pfingst et al.,
2015; Ramekers
et al., 2014; Shepherd & Javel, 1997). Moreover, psychophysical temporal
integration abilities may depend, in part, on local SGN density (Pfingst et al., 2011; Zhou, Kraft, Colesa, &
Pfingst, 2015).In human CI listeners, evoked potential and temporal integration responses vary
widely across electrode sites (e.g., Bierer, Faulkner, & Tremblay, 2011;
Brown, Abbas, & Gantz,
1990; Cafarelli Dees
et al., 2005; DeVries
et al., 2016; Eisen
& Franck, 2004; Schvartz-Leyzac & Pfingst, 2016; Zhou & Pfingst, 2014). Some evidence
suggests that younger participants have larger electrically evoked compound action
potential (ECAP) amplitudes and steeper ECAP AGF slopes than older participants
(Brown, Abbas, Etler,
O’Brien, & Oleson, 2010; Cafarelli Dees et al., 2005). Relatively
shallow ECAP AGF slopes are also associated with relatively long durations of
hearing loss (Schvartz-Leyzac
& Pfingst, 2016) and poor speech perception scores (Brown et al., 1990; Kim et al., 2010). In
bilateral CI listeners, between-ear differences in phoneme perception are partially
explained by between-ear differences in ECAP responses (Schvartz-Leyzac & Pfingst, 2018) and
temporal integration abilities (Zhou & Pfingst, 2014). Moreover, deactivating CI channels with poor
temporal integration performance leads to improved speech perception scores for some
adult listeners (Zhou,
2017).Although indirect estimates of SGN density have been studied extensively in humans,
the number of remaining SGNs constitutes only one aspect of neural health.
Conceivably, the integrity of the peripheral processes could also influence the
fidelity of electrical stimulation by a CI. Recent computational modeling evidence
suggests that sensitivity to the polarity of an electrical stimulus may reflect
local peripheral process integrity (Joshi, Dau, & Epp, 2017; Rattay, Leao, & Felix,
2001; Rattay, Lutter,
& Felix, 2001; Resnick, O’Brien, & Rubinstein, 2018).Polarity sensitivity refers to the difference in psychophysical or physiological
responses to positive (anodic) and negative (cathodic) electrical current. Modeling
data suggest that better sensitivity to the anodic polarity than to the cathodic
polarity may indicate peripheral process degeneration or demyelination (Joshi et al., 2017; Rattay, Leao, et al., 2001;
Rattay, Lutter et al.,
2001; Resnick et al.,
2018). Differential sensitivity to each polarity is thought to reflect
differences in the site of spike initiation in response to anodic and cathodic pulse
shapes. Specifically, injecting cathodic current into the extracellular space
depolarizes nearby neural membrane while hyperpolarizing distant neural membrane
(Rubinstein, 1991).
Conversely, anodic current depolarizes distant neural membrane and hyperpolarizes
nearby neural membrane (Rubinstein, 1991). When the peripheral processes have degenerated and
the electrode is located distal to the soma, higher current levels for cathodic
relative to anodic polarities are required in order for cathodic stimuli to overcome
the unmyelinated cell body and generate an action potential near the central axon
(Joshi et al., 2017;
Macherey, Carlyon, Chatron,
& Roman, 2017; Rattay, Leao, et al., 2001; Rattay, Lutter et al., 2001; Resnick et al., 2018).Electrophysiological and psychophysical evidence in adult CI listeners suggests that
polarity sensitivity has the potential to provide insight into the status of the
electrode–neuron interface (Carlyon, Cosentino, Deeks, Parkinson, & Arenberg, 2018; Hughes, Choi, & Glickman,
2018; Hughes,
Goehring, & Baudhuin, 2017; Jahn & Arenberg, 2019; Macherey et al., 2017;
Macherey, Carlyon, van
Wieringen, Deeks, & Wouters, 2008; Macherey, van Wieringen, Carlyon, Deeks, &
Wouters, 2006; Spitzer & Hughes, 2017; Undurraga, Carlyon, Wouters, & van Wieringen,
2013; Undurraga, van
Wieringen, Carlyon, Macherey, & Wouters, 2010; van Wieringen, Macherey, Carlyon, Deeks, &
Wouters, 2008). At suprathreshold stimulation levels, evoked potential
amplitudes are generally larger (Bahmer & Baumann, 2013; Hughes et al., 2017, 2018; Macherey et al., 2008;
Undurraga et al.,
2010), and most comfortable listening levels (MCLs) are generally lower
(Macherey et al.,
2017), for anodic compared with cathodic stimulation. These findings
using suprathreshold stimulation levels suggest that anodic stimulation may be more
effective, in general, than cathodic stimulation in adult CI listeners.On the other hand, recent evidence using low-level stimulation shows that polarity
sensitivity at threshold is subject- and electrode-dependent in postlingually
deafened adults (Carlyon et al.,
2018; Jahn &
Arenberg, 2019; Macherey et al., 2017). Moreover, Jahn and Arenberg (2019) demonstrated that
the psychophysical polarity effect at threshold varies independently of electrode
position relative to the modiolus and intracochlear resistance. Taken together,
these studies provide increasing evidence that polarity sensitivity may reflect
neural integrity in CI listeners, and that the measure is relatively independent of
non-neural factors that influence the quality of the electrode–neuron interface in
humans.To comprehensively characterize polarity sensitivity in CI listeners and to determine
its utility in clinical interventions, it is important to evaluate polarity
sensitivity in individuals with diverse hearing histories and to determine whether
it relates to CI outcomes. To date, polarity sensitivity has been studied in small
samples of largely postlingually deafened adults. Children and adults with CIs
typically present with different demographic characteristics that may influence SGN
integrity. For instance, adult CI recipients are generally implanted at a more
advanced age and experience longer durations of preimplantation auditory deprivation
than children. Children and adults with CIs also tend to have different hearing loss
etiologies. Human histological studies demonstrate that chronological age, duration
of deafness, and hearing loss etiology are each predictive of SGN density (Makary et al., 2011; Nadol, 1997; Nadol et al., 1989; Otte et al., 1978). The
primary goal of this study was to expand upon previous literature by characterizing
polarity sensitivity and speech perception performance in a relatively large and
diverse sample of CI listeners. Two groups of participants with divergent hearing
histories were recruited: (a) adolescents who were deafened and implanted during
childhood and (b) older adult-implanted listeners.The primary outcome measure in this study was the polarity effect at threshold,
defined as the difference in the magnitude of single-channel behavioral thresholds
measured in response to anodic and cathodic polarities (cathodic threshold minus
anodic threshold; Carlyon et al., 2018; Jahn & Arenberg, 2019). The polarity
effect at threshold was chosen as the primary outcome measure because it has been
shown to vary independently of electrode position and tissue impedances in CI
listeners (Jahn & Arenberg,
2019). In this study, we also assessed the polarity effect at
suprathreshold levels by estimating individuals’ MCLs for each pulse shape and
calculating the polarity effect at MCL and for dynamic range (DR). Note that spike
timing differs between peripheral and central processes; however, the psychophysical
measures used in this study and in others are likely not sensitive to those latency
differences.Consistent with prior investigations, we predicted that the polarity effect at
threshold would be subject- and electrode-dependent, but that anodic stimulation
would generally result in lower MCLs than cathodic stimulation (Carlyon et al., 2018; Jahn & Arenberg, 2019;
Macherey et al.,
2017). We expected variability in polarity sensitivity to persist across the
diverse sample of pediatric and adult CI listeners included in this dataset.
Furthermore, we predicted that (a) long periods of preimplantation auditory
deprivation would be associated with lower (i.e., better) threshold sensitivity to
anodic stimulation than to cathodic stimulation (i.e., more peripheral
degeneration), (b) adult-implanted listeners would have larger polarity effects at
threshold (i.e., more peripheral degeneration) than child-implanted listeners, and
(c) relatively poor speech perception scores would be associated with relatively
large polarity effects at threshold (i.e., more peripheral degeneration). The
results of this investigation will provide insight into the characteristics of
polarity sensitivity in child- and adult-implanted listeners and may assist in
developing hypothesis-driven recommendations for the application of polarity
sensitivity to CI programming interventions.
Methods
Subjects
Demographic information for all subjects and ears tested in this study is
presented in Table
1. Data were obtained from a total of 41 ears (27 individual subjects, 13
males) implanted with Advanced Bionics HiRes 90K devices. Twenty ears (11
individual subjects) were deafened and implanted during childhood (prior to age
18 years). At the time of testing, subjects in the child-implanted group ranged
in age from 13 to 18 years (M = 15.2 years, standard deviation
[SD] = 1.4 years). Nine of the 11 child-implanted subjects
were bilaterally implanted, and data were collected from each ear. Subjects P11
and P12 are fraternal twins. No other subjects are related to one another.
Hereafter, this group of subjects will be referred to as the “child-implanted
group”, to signify that they were deafened and implanted as children and tested
during adolescence. In all figures, data from the child-implanted group are
denoted by green symbols.
Table 1.
Demographic Information.
ID
Etiology
Age (years)
First-implanted ear
Second-implanted ear
Age implanted (years)
Duration of deafness (years)
Electrode array
Age implanted (years)
Duration of deafness (years)
Electrode array
Child-implanted Listeners
P02
EVA
13.9
1.1
1.1
HF1J
3.1
3.1
HF1J
P03
Unknown
14.7
1.4
1.4
HF1J
5.6
5.6
HF1J
P04
Unknown
15.2
1.7
1.7
HF1J
4.7
4.7
HF1J
P06
Unknown
18.8
4.3
1.8
HF1J
10.9
8.5
HF1J
P07
Unknown
15.4
1.9
1.9
HF1J
4.9
4.9
HF1J
P09
Unknown
14.9
2.6
1.3
HF1J
3.9
2.7
HF1J
P11
DFNB1
15.3
1.4
1.2
HF1J
10.2
10.0
HF1J
P12
DFNB1
15.3
1.7
1.4
HF1J
10.2
10.0
HF1J
P13
EVA
13.4
9.2
6.4
HF1J
—
—
—
P16
DFNB1
14.6
1.0
1.0
HF1J
4.5
4.5
HF1J
P17
Unknown
15.5
1.3
1.3
HF1J
—
—
—
Mean (SD)
15.2 (1.4)
2.5 (2.4)
1.9 (1.5)
6.5 (3.1)
6.0 (2.8)
Adult-implanted Listeners
S22
Unknown
78.2
66.7
11.8
1J Helix
—
—
—
S23
Unknown
73.4
62.0
3.9
1J Helix
64.6
6.5
HF1J
S29
Noise exposure
87.8
76.8
30.3
HF1J
85.7
39.2
MS
S39
Genetic
54.4
30.1
8.0
HF1J
40.1
18.0
HF1J
S40
EVA
56.2
50.4
46.4
HF1J
—
—
—
S43
Noise exposure
72.5
67.9
18.7
MS
—
—
—
S45
Genetic
65.4
54.0
11.0
HF1J
61.0
18.0
MS
S46
Unknown
69.4
64.2
25.1
HF1J
—
—
—
S47
Unknown
40.4
36.4
10.3
MS
—
—
—
S49
Unknown
45.8
43.5
42.1
MS
44.2
42.8
MS
S50
Unknown
76.5
71.0
53.0
HF1J
—
—
—
S52
Unknown
71.2
66.0
6.1
HF1J
—
—
—
S53
Meningitis
56.0
44.1
42.9
1J Helix
—
—
—
S54
EVA
27.8
23.7
16.7
MS
—
—
—
S59
Ototoxicity
32.1
30.9
18.9
MS
—
—
—
S60
Meningitis
22.5
19.2
19.1
MS
—
—
—
Mean (SD)
59.9 (18.8)
50.4 (18.3)
22.8 (15.6)
59.1 (18.2)
24.9 (15.5)
Note. Demographic information for all
participants, including: hearing loss etiology (if known),
chronological age at time of testing (in years), age of
implantation for implanted each ear (in years), duration of
deafness for each implanted ear (in years), and electrode array
for each implanted ear. Duration of deafness is defined as the
time between diagnosis of severe-to-profound sensorineural
hearing loss and cochlear implant activation. Note that subjects
S47 and S59 are bilaterally implanted, but their
second-implanted ears were not tested as part of this study.
EVA = enlarged vestibular aqueduct; DFNB1 = genetic nonsyndromic
hearing loss. HF1J = HiFocus 1J electrode array; MS = mid-scala
electrode array; SD = standard deviation.
Demographic Information.Note. Demographic information for all
participants, including: hearing loss etiology (if known),
chronological age at time of testing (in years), age of
implantation for implanted each ear (in years), duration of
deafness for each implanted ear (in years), and electrode array
for each implanted ear. Duration of deafness is defined as the
time between diagnosis of severe-to-profound sensorineural
hearing loss and cochlear implant activation. Note that subjects
S47 and S59 are bilaterally implanted, but their
second-implanted ears were not tested as part of this study.
EVA = enlarged vestibular aqueduct; DFNB1 = genetic nonsyndromic
hearing loss. HF1J = HiFocus 1J electrode array; MS = mid-scala
electrode array; SD = standard deviation.Twenty-one ears (16 individual subjects) were implanted during adulthood (age 18
or older). Four of the adult-implanted subjects (S40, S49, S53, and S60) were
diagnosed with severe-to-profound sensorineural hearing loss as children, and
the remaining subjects became deaf as adults. At the time of testing,
adult-implanted participants ranged in age from 22 to 87 years
(M = 59.9 years, SD = 18.8 years). Seven
of the 16 adult-implanted participants presented with bilateral implants;
however, due to time constraints, only five of the bilaterally implanted adults
were tested in both ears. Hereafter, this group of subjects will be referred to
as the “adult-implanted group”, to signify that they were implanted and tested
as adults. In all figures, data from the adult-implanted group are denoted by
blue symbols.All subjects primarily used spoken language to communicate, and all but one
subject were native American English speakers. Subject S54 learned English as a
second language during early childhood. Each child provided written informed
assent, and his or her parents or legal guardians provided written informed
consent. Each adult provided written informed consent. All procedures were
approved by the University of Washington Human Subjects Division.
Electrical Stimuli
Electrical stimuli were controlled by the Bionic Ear Data Collection System
version 1.18.315 (Advanced Bionics, Valencia, CA) and custom MATLAB scripts
(MathWorks, Inc., Natick, MA). Stimuli were presented directly to the internal
device. Prior to testing, electrical stimuli were verified using a reference
implant and a digital storage oscilloscope.
Channel Selection for Polarity Testing
Advanced Bionics devices have 16 electrode contacts. Due to time and attention
constraints, it was not feasible to measure the polarity effect on every
electrode in each ear. Instead, four channels per ear were selected for polarity
effect testing. Within a subject, the four channels were selected based on the
relative magnitude of single-channel behavioral thresholds measured in response
to a spatially focused electrode configuration. Focused behavioral thresholds
are believed to reflect the overall quality of the electrode–neuron interface;
within a subject, higher threshold channels are thought to interface poorly with
the auditory nerve relative to lower threshold channels (for review, see Bierer, 2010). For
instance, channels with relatively high focused thresholds are often located
farther from the target neurons (DeVries & Arenberg, 2018b; DeVries et al., 2016;
Jahn & Arenberg,
2019; Long
et al., 2014) and have smaller evoked potential amplitudes (DeVries et al., 2016)
than lower threshold channels.In this study, the two lowest threshold channels and the two highest threshold
channels were selected for polarity effect assessment. None of the four channels
were directly adjacent to one another. Theoretically, selecting low- and
high-threshold channels within a subject allows for assessment of a subset of
electrodes that vary in the quality with which they interface with the auditory
nerve; low- and high-threshold channels represent “good” and “poor”
electrode–neuron interfaces, respectively.Single-channel-focused behavioral thresholds were assessed using a modified
Békésy-style sweep procedure that is 4 times faster than traditional adaptive
forced-choice methods (Bierer, Bierer, Kreft, & Oxenham, 2015; Sek, Alcántara, Moore, Kluk, & Wicher,
2005). Using current steering, stimuli were swept across the
electrode array by dividing the electrical current between two adjacent
intracochlear electrodes and varying the proportion of current directed to each
electrode.Stimuli were biphasic, cathodic-leading pulse trains (102 s/phase, 0-s interphase gap, 200.4 ms duration, 997.9 pulse per second)
presented in a steered quadrupolar (sQP) stimulation mode. A channel was
comprised of four adjacent intracochlear electrodes. The two middle electrodes
served as active electrodes, and the two outermost electrodes served as return
electrodes. The current focusing coefficient () was set to 0.9, indicating that 90% of the return current was
delivered through the intracochlear return electrodes (45% to each electrode)
and the remaining 10% was delivered through an extracochlear ground. Current
focusing coefficients can range from 0 to 1, with 1 representing the highest
possible degree of current focusing and, consequently, resulting in the most
spatially restrictive electrical field. The higher the current focusing
coefficient, the greater the observed channel-to-channel variability in focused
thresholds (Bierer &
Faulkner, 2010). A highly focused coefficient of 0.9 was selected to
capture as much within-subject variability in focused thresholds as possible
while remaining below the voltage compliance limits of the device.The modified sweep procedure has been described in detail in many other studies
from our laboratory (e.g., Bierer et al., 2015; DeVries & Arenberg, 2018b; DeVries et al., 2016;
Jahn & Arenberg,
2019). A brief review of the sweep procedure is provided here. To
sweep stimuli across the electrode array, current was steered between the two
active electrodes by varying the steering coefficient, alpha (). When = 0, all current is delivered through the more apical of the
active electrode pair. Conversely, when = 1, all current is delivered through the more basal active
electrode. Because sQP stimulation requires four adjacent intracochlear
electrodes, focused thresholds can only be obtained for Electrodes 2 to 15. Per
convention, on Channels 3–15, integer channel numbers refer to the number of the
basal active electrode when = 1; for Channel 2, an value of 0 is used to center the current on Electrode 2.The upper limit of stimulation on each electrode was set to each listener’s
electrode-specific MCL, which corresponded to a loudness rating of “6,” or
most comfortable, on the Advanced Bionics Clinical Loudness
Scale (Advanced Bionics, Valencia, CA). Pulse trains were presented starting at
a level 6 dB below each listener’s MCL and swept across the electrode array by
increasing alpha from 0 to 1 in step sizes of 0.1. The listener was instructed
to continuously depress the spacebar on a standard computer keyboard when he or
she could hear the stimulus and to release the spacebar when he or she could not
hear the stimulus. When the spacebar was depressed, the presentation level of
the stimulus decreased. Conversely, the presentation level increased when the
spacebar was released. The participants completed one forward sweep that
progressed basally (Channels 2–15) and one reverse sweep that progressed
apically (Channels 15–2). Final single-channel focused threshold estimates were
calculated as the weighted average of consecutive current levels at integer
channel numbers along the forward and reverse sweeps (as in Bierer et al.,
2015).Following the threshold measurement, the channels with the two lowest focused
thresholds and those with the two highest focused thresholds were identified for
each ear. If any of those channels were adjacent to one another, the channel
with the next-lowest or next-highest nonadjacent threshold was identified. These
four nonadjacent channels (two low-threshold and two high-threshold channels)
were used for subsequent polarity effect testing.
Polarity Effect Measurement
Polarity sensitivity was assessed on four nonadjacent channels within each ear:
the two channels with the lowest focused thresholds and the two channels with
the highest focused thresholds. Stimuli were 99 pulse per second trains
presented in a monopolar stimulation mode (43 µs/phase, 0 -µs interphase gap,
400 ms duration). A triphasic pulse shape was used, where the central
high-amplitude phase was twice the amplitude of the first and third phases. The
polarity of the central high-amplitude phase was either anodic (CAC) or cathodic
(ACA), depending on the experimental condition. Triphasic pulses concentrate the
charge of the polarity of interest into a brief time window while maintaining
the requisite charge balance for use in humans. The stimuli were identical to
those used in recent investigations of psychophysical polarity sensitivity
(Carlyon et al.,
2018; Jahn &
Arenberg, 2019).On the four selected channels, behavioral thresholds were measured in response to
each of the two triphasic pulses. The polarity effect (in dB) was defined as the
ACA threshold minus the CAC threshold (ACA − CAC). A positive polarity effect
value indicated that CAC thresholds were lower (i.e., better) than ACA
thresholds. Based on modeling data, lower anodic than cathodic thresholds may
indicate some degree of peripheral process degeneration (Joshi et al., 2017; Rattay, Leao, et al.,
2001; Rattay,
Lutter et al., 2001; Resnick et al., 2018). Conversely, a
negative polarity effect value indicated lower (i.e., better) ACA thresholds
compared with CAC thresholds and may reflect healthy peripheral processes (Joshi et al., 2017;
Rattay, Leao, et al.,
2001; Rattay,
Lutter et al., 2001; Resnick et al., 2018).To obtain suprathreshold polarity data and to set the upper limit of stimulation
for polarity assessment, MCLs were obtained on each of the four channels in
response to each polarity. To measure MCL, the current level was gradually
increased from a subthreshold level of 50 µA up until the subject reported a
loudness rating of “6,” or most comfortable on the Advanced
Bionics Clinical Loudness Scale (Advanced Bionics, Valencia, CA). For each
channel and polarity, the corresponding MCL was set as the upper limit of
stimulation for the threshold measurement procedure.For 27 of the 328 (8%) total channel/polarity combinations tested, MCL could not
be reached at stimulation levels below the voltage compliance limits of the
device. This tended to occur for certain subjects and was not related to CI
channel or polarity, as follows: P12 (both implanted ears, all
electrode/polarity combinations), P13 (first/only-implanted ear, both polarities
on Channel 15), P17 (first/only-implanted ear, all electrode/polarity
combinations except for the anodic polarity on Channel 2), and S29
(second-implanted ear, cathodic polarity on Channels 8 and 13). In those cases,
the upper limit of stimulation was set to the highest current level that could
be achieved without exceeding voltage compliance limits. In each case, the upper
stimulation level was equivalent to a subjective listening level of either 4
(medium soft) or 5 (medium), which was
sufficiently high for the subject to accurately perform the threshold
measurements. However, those 27 measurements were excluded from any subsequent
MCL and DR analyses.An adaptive one-up/one-down staircase tracking procedure was used to measure
single-channel behavioral thresholds for each polarity on each of the four
channels. For each adaptive track, the initial presentation level was set to 90%
of the MCL; for channel/polarity combinations where MCL could not be reached,
the initial presentation level was set to 98% of the upper stimulation level.
For subsequent tracks, the initial presentation level was set anywhere from 50
to 98% of the upper stimulation level. A lower starting level was used on
electrodes with large DRs to reduce the amount of time necessary to estimate
threshold. Higher starting levels were maintained for electrodes with small DRs
to ensure that the subject could comfortably hear the stimulus before the first
reversal.The subject was instructed to press the spacebar on the computer keyboard one
time whenever he or she heard a sound. The presentation level decreased if the
subject responded within 3 s after stimulus presentation and increased if the
subject did not respond within 3 s. The initial step size was 0.5 dB. After the
first reversal, the step size was reduced to 0.2 dB. Random delays of 0.1 to
0.6 s were incorporated prior to each stimulus presentation. After eight
reversals, the adaptive procedure terminated. Threshold was estimated as the
average of the final six reversals.The order of channels and polarities tested was randomized for each subject. Two
adaptive threshold tracks were completed for each polarity on each channel. The
two threshold estimates were averaged together to calculate a final threshold
value. A third and fourth run were completed if the thresholds estimated on the
first two runs differed by 1 dB or more. In those cases, threshold estimates
from each of the four runs were averaged together.
Speech Perception
Speech perception was assessed using medial vowel and consonant recognition
tasks. Phonemes were chosen as the speech perception tasks because they are
particularly sensitive to spectral and temporal distortions resulting from CI
processing and poor electrode–neuron interfaces (DiNino, Wright, Winn, & Bierer,
2016; Nie, Barco,
& Zeng, 2006; Shannon, Fu, & Galvin Iii, 2004; Xu, Thompson, & Pfingst, 2005).
Vowel stimuli were a closed set of 10 recorded vowels in /hVd/ context (/i/,
“heed”; /ɪ/, “hid”; /eɪ/, “hayed”; /ɛ/,”head”; /æ/,”had”; /ɑ/, “hod”; /u/,
“who’d”; /ʊ/, “hood”; /oʊ/, “hoed”; and /ʌ/, “hud”) that were spoken by one
female talker native to the Pacific Northwest region of the United States.
Consonant stimuli were a closed set of 16 recorded consonants presented in /aCa/
context (/p/, “aPa”; /t/, “aTa”; /k/, “aKa”; /b/, “aBa”; /d/, “aDa”; /ɡ/, “aGa”;
/f/, “aFa”; /θ/, “aTHa”; /s/, “aSa”; /ʃ/, “aSHa”; /v/, “aVa”; /z/, “aZa”; /dʒ/,
“aJa”; /m/, “aMa”; /n/, “aNa”; and /l/, “aLa”) and spoken by one male talker
(stimuli were the same as those used by Shannon, Jensvold, Padilla, Robert,
& Wang, 1999).
Testing was performed in a double-walled sound-treated booth (IAC RE-243).
Stimuli were presented through an external A/D device (SIIF USB SoundWave 7.1)
and a Crown D75 amplifier at a calibrated level of 60 dB-A through a Bose 161
speaker placed at 0° azimuth and 1 m from the participant’s head. Custom
software (ListPlayer2 version 2.2.11.52, Advanced Bionics, Valencia, CA) was
used to present the stimuli and to record responses.Participants were tested with one ear at a time using their everyday listening
programs. Unilateral CI users wore an earplug in the nonimplanted ear during
speech perception testing. After each speech token was presented, a graphical
user interface with the possible phoneme choices was displayed on a computer
screen. The participant selected his or her response using a computer mouse.
Participants completed one practice run consisting of three repetitions of each
speech token prior to beginning the experiment. Performance feedback was
provided during the practice run. During the experiment, two runs consisting of
three repetitions of each speech token were conducted, resulting in a total of
six presentations of each speech token. Feedback was not provided during the
experiment. Stimuli were pseudorandomly interleaved within each run. If scores
on the two runs differed by more than 10%, a third run consisting of three
additional repetitions was presented. Scores from each run were averaged
together to achieve a final percentage correct.
Statistical Analyses
Data were analyzed using R Version 3.3.1 (R Core Team, 2016). Linear
mixed-effects models were employed for all analyses to account for clustering of
electrode-specific data within ears and for clustering of two ears within the
same listener. “Subject” and “ear” were included as random effects in the
models, where appropriate. Models were fit using restricted maximum likelihood
parameter estimates to minimize small sample estimation bias (McNeish, 2017). An
unstructured covariance matrix was specified for each model. An Aikaike
information criterion with a bias correction for small samples (AICc) was used
for model selection (Hurvich
& Tsai, 1989).Note that traditional R2 values are invalid for
multilevel models. Instead, two pseudo-R2 values,
described by Nakagawa and
Schielzeth (2013), are presented where applicable: (a) marginal
R2
(R2marginal), representing the
proportion of the total variance explained by the fixed effects, and (b)
conditional R2
(R2conditional), representing the
proportion of the variance explained by both the fixed and random effects. The
difference between the R2marginal and
R2conditional reflects the
variability in the random effects; here, this would represent across-subject
variability. The lmerTest (Kuznetsova, Brockhoff, & Christensen, 2017), MuMIn (Bartón, 2018), and
Lattice (Sarkar,
2008) R packages were used to perform statistical analyses and to assess
the validity of model assumptions. Bonferroni corrections for multiple
comparisons were applied and are noted where appropriate.
Results
Electrode Array Considerations
Subjects presented with a variety of electrode array types (HiFocus 1 J, 1 J
Helix, and Mid-Scala; Table
1). Different electrode arrays are designed to achieve different
positions in the cochlea relative to the modiolus (Dhanasingh & Jolly, 2018)
and can influence absolute threshold measurements (DeVries & Arenberg, 2018b; DeVries et al., 2016;
Jahn & Arenberg,
2019; Long
et al., 2014). However, recent evidence suggests that the
psychophysical polarity effect varies independently of electrode position
relative to the inner wall of the cochlea in CI listeners (Jahn & Arenberg, 2019). This is
likely because the polarity effect is a difference score, rather than an
absolute threshold measurement. The same stimuli and analysis methods used in
Jahn and Arenberg
(2019) were used in this study. Prior to data analysis, we confirmed
that the polarity effect was not influenced by electrode array type,
F(2, 36.31) = 0.63, p = .63, or electrode
cochlear location, F(1, 129.83) =0.03,
p = .86, in this sample of subjects. Electrode cochlear
location was defined as apical (Electrodes 2–8) or basal (Electrodes 9–15).
Electrode position is not considered further.
Characterization of the Polarity Effect at Threshold and at Suprathreshold
Levels
The first analysis served to characterize the polarity effect at threshold and at
suprathreshold levels. Figure
1(a) to (d)
shows single-channel thresholds and MCLs measured in response to the ACA and CAC
pulses for two child-implanted participants (a and b) and two adult-implanted
participants (c and d). Black squares represent responses to the ACA pulse, and
red circles represent responses to the CAC pulse. Solid lines connect the
threshold responses, and dashed lines connect the MCLs. Based on modeling data,
lower thresholds or MCLs in response to anodic than to cathodic pulses may
reflect peripheral process degeneration (Joshi et al., 2017; Rattay, Leao, et al.,
2001; Rattay,
Lutter et al., 2001; Resnick et al., 2018).
Figure 1.
Single-channel auditory detection thresholds and MCLs measured in
response to the cathodic (ACA) and anodic (CAC) pulse shapes for two
child-implanted ears (a and b) and two adult-implanted ears (c and
d). Black squares represent responses to the ACA pulse shape, and
red circles represent responses to the CAC pulse shape. Solid lines
connect threshold measurements, and dotted lines connect MCL
measurements. MCL = most comfortable listening level.
Single-channel auditory detection thresholds and MCLs measured in
response to the cathodic (ACA) and anodic (CAC) pulse shapes for two
child-implanted ears (a and b) and two adult-implanted ears (c and
d). Black squares represent responses to the ACA pulse shape, and
red circles represent responses to the CAC pulse shape. Solid lines
connect threshold measurements, and dotted lines connect MCL
measurements. MCL = most comfortable listening level.Positive polarity effect values indicate lower thresholds or MCLs in response to
anodic than to cathodic pulses; negative polarity effect values indicate the
inverse. The magnitude of the polarity effect at threshold and at MCL varied
across- and within-subjects. Across electrodes, the polarity effect at threshold
ranged from −4.82 dB to 3.54 dB (M = 0.25 dB,
SD = 1.52 dB), and 102 of the 164 electrodes (62.20%) had
lower thresholds for anodic than for cathodic pulses. The polarity effect at MCL
was calculated on 149 of the 164 electrodes tested. At MCL, the polarity effect
ranged from −2.13 dB to 4.83 dB (M = 1.12 dB,
SD = 1.24 dB) and 133 of the 149 electrodes (89.26%) had
lower MCLs for anodic than for cathodic pulses. An electrode with lower
threshold sensitivity to anodic stimulation did not necessarily have lower
suprathreshold sensitivity to anodic stimulation, and vice versa.Assessing the polarity effect at threshold and at MCL allowed us to calculate the
difference in DR between the two polarities (i.e., DR polarity effect). DR was
calculated as the difference between MCL and threshold for each polarity on each
tested electrode. Like threshold and MCL, the DR polarity effect was calculated
as the difference in DR for the cathodic (ACA) pulse versus the anodic (CAC)
pulse (ACA − CAC). Therefore, a positive DR polarity effect value indicates that
the DR for the cathodic pulse is larger than the DR for the anodic pulse. A
negative DR polarity effect value indicates a larger DR for the anodic than for
the cathodic pulse.The DR polarity effect was assessed on 149 electrodes. The DR polarity effect
varied from −2.05 dB to 5.31 dB (M = 0.87 dB,
SD = 1.25 dB), and 115 of the 149 electrodes (77.18%)
demonstrated larger DRs for cathodic pulses than for anodic pulses. Like
threshold and MCL, the magnitude of the DR polarity effect was subject- and
electrode-dependent. Despite the finding that most electrodes had better
sensitivity to the anodic polarity at threshold and at MCL, the majority of
electrodes exhibited larger DRs for the cathodic polarity than for the anodic
polarity.Relationships between polarity sensitivity at threshold and at suprathreshold
levels were evaluated to determine whether polarity effects at low current
levels are predictive of those at higher current levels. Initially, “age-group”
was included in the models as an independent variable to account for potential
differences between child- and adult-implanted ears; however, in each case, more
parsimonious model fits (i.e., lower AICc values) were obtained when age-group
was excluded. Thus, the relationships between polarity sensitivity at threshold
and at suprathreshold levels did not differ between child- and adult-implanted
listeners, and “age-group” was not included in the final models. To elucidate
this finding, data from child- and adult-implanted ears are denoted by separate
colors in each figure (green and blue symbols, respectively).Figure 2(a) to (c) shows single-channel
data for the relationships between (a) polarity effect at threshold and at MCL,
(b) polarity effect at threshold and for DR, and (c) polarity effect at MCL and
for DR. Dashed lines split each panel into four quadrants at
y = 0 and x = 0. Data points that fall in the
upper right-hand quadrants and in the lower left-hand quadrants have the same
sign (positive or negative, respectively) for both of the measures represented
in the figure.
Figure 2.
Single-channel data for the relationships between (a) polarity effect
at threshold and at MCL, (b) polarity effect at threshold and for
dynamic range (DR), and (c) polarity effect at MCL and for DR.
Dashed lines split each panel into four quadrants at
y = 0 and x = 0. Data points
that fall in the upper right-hand quadrants and in the lower
left-hand quadrants have the same sign (positive or negative,
respectively) for both of the measures represented in the figure.
Circles indicate data from first-implanted ears, and triangles
represent data from second-implanted ears. Green symbols indicate
data from child-implanted ears, and blue symbols indicate data from
adult-implanted ears. CI = cochlear implant; MCL = most comfortable
listening level; ACA = cathodic; CAC = anodic.
Single-channel data for the relationships between (a) polarity effect
at threshold and at MCL, (b) polarity effect at threshold and for
dynamic range (DR), and (c) polarity effect at MCL and for DR.
Dashed lines split each panel into four quadrants at
y = 0 and x = 0. Data points
that fall in the upper right-hand quadrants and in the lower
left-hand quadrants have the same sign (positive or negative,
respectively) for both of the measures represented in the figure.
Circles indicate data from first-implanted ears, and triangles
represent data from second-implanted ears. Green symbols indicate
data from child-implanted ears, and blue symbols indicate data from
adult-implanted ears. CI = cochlear implant; MCL = most comfortable
listening level; ACA = cathodic; CAC = anodic.Polarity effects at threshold, MCL and DR were highly correlated. Therefore, to
statistically analyze the relationships between each measure, it was necessary
to specify separate mixed-effects models. A Bonferroni adjustment for multiple
comparisons was applied (adjusted = .017). The polarity effect at threshold was positively
correlated with the polarity effect at MCL,
R2marginal = .42,
R2conditional = .67,
F(1, 139.27) = 132.78, p < .001,
wherein positive polarity effect values at threshold were associated with
positive polarity effect values at MCL. This suggests that, in general,
electrodes that had lower threshold sensitivity to the anodic polarity than to
the cathodic polarity also had lower suprathreshold sensitivity to the anodic
polarity than to the cathodic polarity.Furthermore, the polarity effect at threshold was negatively correlated with the
DR polarity effect, R2marginal = .27,
R2conditional = .60,
F(1, 139.27) = 74.22, p < .001.
Specifically, better threshold sensitivity to one polarity was associated with
larger DRs for that polarity. However, oftentimes (on 42.28% of electrodes), DR
was larger for the cathodic polarity irrespective of polarity sensitivity at
threshold. This can be seen in the upper right-hand quadrant of Figure 2(b); many
electrodes have larger DRs for the cathodic polarity despite having lower
threshold sensitivity to the anodic polarity.Finally, the polarity effect at MCL was positively correlated with the DR
polarity effect, R2marginal = .06,
R2conditional = .23,
F(1, 119.02) = 8.33, p = .005, indicating
that lower suprathreshold sensitivity to anodic stimulation was generally
associated with smaller DRs for the anodic than for the cathodic polarity.
Overall, the DR in response to each polarity was influenced by both the
threshold and the MCL. However, better sensitivity at threshold or at MCL for a
particular polarity did not necessarily mean that the DR was larger in response
to that polarity. On some electrodes, low-amplitude anodic pulses at MCL may
restrict the anodic DR relative to the cathodic DR.
Polarity Sensitivity as a Function of Channel Classification
Recall that channels were selected for polarity testing based on the relative
within-subject magnitude of single-channel focused behavioral thresholds. The
second analysis evaluated differences in polarity sensitivity on
low-focused-threshold channels compared with high-focused-threshold channels.
Once again, during the model selection procedure, it was determined that
polarity sensitivity as a function of channel classification did not differ
between child-implanted and adult-implanted listeners, and that including
“age-group” in the models resulted in poorer fits (i.e., higher AICc values).
Thus, the final models for this analysis did not include “age-group” as an
independent variable.Figure 3 shows
single-channel polarity effect as a function of channel classification
(low-focused-threshold vs. high-focused-threshold channels) for (a) the polarity
effect at threshold, (b) the polarity effect at MCL, and (c) the DR polarity
effect. As the polarity effects at threshold, MCL and DR are highly correlated,
separate mixed-effects models were specified to assess differences in each
measure as a function of channel classification. A Bonferroni adjustment for
multiple comparisons was applied (adjusted = .017). On high-focused-threshold channels, the polarity
effect at threshold ranged from −3.20 dB to 2.90 dB
(M = 0.61 dB, SD = 1.21 dB). On
low-focused-threshold channels, the polarity effect at threshold ranged from
−4.82 dB to 3.54 dB (M = −0.11 dB,
SD = 1.71 dB). High-focused-threshold channels had
significantly larger polarity effects, on average, than low-focused-threshold
channels (Figure 3(a)),
R2marginal = .07,
R2conditional = .29,
F(1, 122) = 12.97, p < .001. This
suggests that better threshold sensitivity to anodic stimulation than to
cathodic stimulation was more common on the high-focused-threshold channels than
on the low-focused-threshold channels.
Figure 3.
Single-channel polarity effects as a function of channel
classification (high-focused-threshold vs. low-focused-threshold
channels) for (a) the polarity effect at threshold, (b) the polarity
effect at MCL, and (c) the polarity effect for dynamic range.
Circles indicate data from first-implanted ears, and triangles
represent data from second-implanted ears. Green symbols indicate
data from child-implanted ears, and blue symbols indicate data from
adult-implanted ears. CI = cochlear implant. *** denotes a
statistically significant difference in polarity effect between
high- and low-threshold channels at the p < .001
level. ACA = cathodic; CAC = anodic; MCL = most comfortable
listening level.
Single-channel polarity effects as a function of channel
classification (high-focused-threshold vs. low-focused-threshold
channels) for (a) the polarity effect at threshold, (b) the polarity
effect at MCL, and (c) the polarity effect for dynamic range.
Circles indicate data from first-implanted ears, and triangles
represent data from second-implanted ears. Green symbols indicate
data from child-implanted ears, and blue symbols indicate data from
adult-implanted ears. CI = cochlear implant. *** denotes a
statistically significant difference in polarity effect between
high- and low-threshold channels at the p < .001
level. ACA = cathodic; CAC = anodic; MCL = most comfortable
listening level.However, the polarity effect at MCL did not differ between high- and
low-focused-threshold channels (Figure 3(b)),
R2marginal < .01,
R2conditional = .44,
F(1, 110.71) = 0.05, p = .83. On
high-focused-threshold channels, the polarity effect at MCL ranged from −1.70 dB
to 3.52 dB (M = 1.14 dB, SD = 1.16 dB). On
low-focused-threshold channels, the polarity effect at MCL ranged from −2.13 dB
to 4.83 dB (M = 1.11 dB, SD = 1.33 dB). This
suggests that suprathreshold polarity sensitivity did not differ as a function
of channel classification.Finally, the DR polarity effect was larger for low-focused-threshold channels
than for high-focused-threshold channels (Figure 3(c)),
R2marginal = .05,
R2conditional = .30,
F(1, 110.4) = 11.11, p = .001. In other
words, low-focused-threshold channels tended to have larger DRs for the cathodic
polarity than for the anodic polarity. On high-focused-threshold channels, the
DR polarity effect ranged from −1.52 dB to 3.23 dB
(M = 0.58 dB, SD = 1.00 dB). On
low-focused-threshold channels, the DR polarity effect ranged from −2.04 dB to
5.30 dB (M = 1.15 dB, SD = 1.40 dB). Although
the polarity effects at threshold and DR differed significantly between high-
and low-focused-threshold channels, note that there is substantial variability
in both outcome measures, irrespective of channel classification.
Across-Site Average Polarity Effect, Demographics, and Speech
Perception
The final analyses assessed the relationships between the across-site average
polarity effect at threshold, demographic characteristics (age and duration of
deafness), and speech perception. Table 2 shows the across-site average
polarity effects at threshold for each ear tested. The across-site average
polarity effect was calculated by averaging the polarity effects at threshold
across the four tested electrodes within each ear. This across-site averaging
method has been used in several studies to quantify and relate
electrode-specific measures to demographic characteristics (e.g., DeVries et al., 2016;
Jahn & Arenberg,
2019; Scheperle,
2017; Schvartz-Leyzac & Pfingst, 2016, 2018). During the model
selection procedure, it was determined that more parsimonious model fits (i.e.,
lower AICc values) were obtained when “age-group” (child-implanted vs.
adult-implanted) was included in the model instead of “chronological age.” This
is likely because chronological age was bimodally distributed in this sample.
Table 2.
Across-Site Average Polarity Effects for Individual Participants.
ID
First CI
Second CI
Low-threshold channels M
(SD)
High-threshold channels M
(SD)
All channels M
(SD)
Low-threshold channels M
(SD)
High-threshold channels M
(SD)
All channels M
(SD)
Child-implanted Listeners
P02
0.80 (0.29)
1.93 (0.80)
1.36 (0.81)
2.08 (0.19)
1.53 (0.42)
1.81 (0.42)
P03
0.56 (0.93)
2.47 (0.15)
1.51 (1.23)
1.36 (0.25)
0.45 (1.59)
0.91 (1.07)
P04
0.86 (0.67)
−0.21 (0.83)
0.32 (0.87)
−0.06 (0.99)
−1.86 (0.59)
−0.96 (1.23)
P06
1.02 (0.32)
−1.01 (0.02)
0.00 (1.19)
−2.49 (0.99)
0.83 (1.70)
−0.82 (2.22)
P07
−2.05 (1.03)
1.05 (0.53)
−0.50 (1.91)
−2.18 (0.82)
−0.04 (0.32)
−1.11 (1.34)
P09
−0.57 (0.21)
0.03 (0.09)
−0.27 (0.37)
1.01 (1.29)
−1.05 (0.05)
−0.02 (1.40)
P11
−2.72 (0.49)
−0.46 (0.57)
−1.59 (1.37)
−0.30 (1.23)
1.47 (0.95)
0.58 (1.36)
P12
0.43 (0.11)
2.34 (0.43)
1.38 (1.13)
−3.41 (1.51)
0.07 (0.90)
−1.67 (2.25)
P13
0.23 (0.04)
−0.66 (1.11)
−0.21 (0.83)
—
—
—
P16
0.15 (0.18)
2.49 (0.27)
1.32 (1.36)
−0.07 (0.74)
0.55 (0.29)
0.24 (0.58)
P17
0.36 (0.05)
0.41 (0.61)
0.38 (0.35)
—
—
—
Mean (SD)
−0.08 (1.22)
0.76 (1.35)
0.34 (0.99)
−0.45 (1.87)
0.21 (1.11)
−0.12 (1.12)
Adult-implanted Listeners
S22
−2.70 (1.73)
0.97 (0.66)
−0.87 (2.37)
—
—
—
S23
−0.69 (0.54)
−1.74 (2.07)
−1.22 (1.37)
−0.98 (0.22)
0.05 (0.37)
−0.47 (0.64)
S29
0.41 (1.06)
0.95 (0.50)
0.68 (0.74)
−0.03 (0.20)
2.69 (0.30)
1.33 (1.58)
S39
0.55 (0.32)
0.62 (0.44)
0.59 (0.31)
1.27 (0.32)
1.41 (0.30)
1.34 (0.27)
S40
2.86 (0.96)
1.58 (0.13)
2.22 (0.93)
—
—
—
S43
−0.03 (0.66)
−0.39 (0.54)
−0.21 (0.53)
—
—
—
S45
0.21 (0.06)
1.36 (0.06)
0.78 (0.67)
2.19 (0.71)
0.85 (0.13)
1.52 (0.87)
S46
0.28 (0.47)
0.12 (0.23)
0.20 (0.32)
—
—
—
S47
2.22 (0.54)
0.17 (0.79)
1.19 (1.31)
—
—
—
S49
−2.45 (0.42)
1.03 (0.38)
−0.71 (2.03)
0.37 (0.71)
1.63 (0.26)
1.00 (0.85)
S50
−1.97 (2.04)
0.98 (1.10)
−0.49 (2.16)
—
—
—
S52
−0.64 (0.51)
−0.06 (0.14)
−0.35 (0.45)
—
—
—
S53
1.26 (0.99)
0.93 (1.06)
1.09 (0.86)
—
—
—
S54
−1.91 (4.12)
0.29 (0.53)
−0.81 (2.71)
—
—
—
S59
−0.98 (3.56)
1.84 (0.11)
0.43 (2.62)
—
—
—
S60
1.22 (0.16)
−0.57 (0.59)
0.33 (1.09)
—
—
—
Mean (SD)
−0.15 (1.61)
0.50 (0.91)
0.17 (0.92)
0.56 (1.21)
1.33 (0.98)
0.94 (0.81)
Note. Across-site average polarity effects at
threshold (in dB) for all ears tested. Average polarity effects
are shown for the two low-threshold channels, the two
high-threshold channels, and all four channels combined. Low-
threshold channels refer to the two nonadjacent channels with
the lowest focused behavioral thresholds within an individual
ear. High-threshold channels refer to the two nonadjacent
channels with the highest focused behavioral thresholds within
an individual ear. Focused thresholds were measured with a
steered quadrupolar electrode configuration (focusing
coefficient = 0.9). CI = cochlear implant;
SD = standard deviation.
Across-Site Average Polarity Effects for Individual Participants.Note. Across-site average polarity effects at
threshold (in dB) for all ears tested. Average polarity effects
are shown for the two low-threshold channels, the two
high-threshold channels, and all four channels combined. Low-
threshold channels refer to the two nonadjacent channels with
the lowest focused behavioral thresholds within an individual
ear. High-threshold channels refer to the two nonadjacent
channels with the highest focused behavioral thresholds within
an individual ear. Focused thresholds were measured with a
steered quadrupolar electrode configuration (focusing
coefficient = 0.9). CI = cochlear implant;
SD = standard deviation.Figure 4 shows the
across-site average polarity effects at threshold for the child-implanted and
adult-implanted listeners. A mixed-model analysis
(R2marginal = 04,
R2conditional = .24) revealed that
the polarity effect at threshold did not vary systematically as a function of
age-group, F(1, 27.00) = 0.03, p > .05, or
duration of deafness, F(1, 32.14) = 0.99,
p > .05. There was substantial variability in the
across-site average polarity effect for both groups of subjects. For the
child-implanted group, the polarity effect at threshold ranged from −4.48 dB to
2.68 dB (M = 0.13 dB, SD = 1.51 dB). For the
adult-implanted group, the polarity effect at threshold ranged from −4.82 dB to
3.54 dB (M = 0.36 dB, SD = 1.53 dB).
Figure 4.
Across-site average polarity effects at threshold (in dB) for the
child-implanted ears and the adult-implanted ears. Circles indicate
data from first-implanted ears, and triangles represent data from
second-implanted ears. CI = cochlear implant; ACA = cathodic;
CAC = anodic.
Across-site average polarity effects at threshold (in dB) for the
child-implanted ears and the adult-implanted ears. Circles indicate
data from first-implanted ears, and triangles represent data from
second-implanted ears. CI = cochlear implant; ACA = cathodic;
CAC = anodic.Notably, duration of deafness varied widely for the adult-implanted listeners
(range = 3.9–53.0 years) and less so for the child-implanted listeners
(range = 1.0–10.1 years). We also previously showed that the polarity effect at
threshold differs between low- and high-focused-threshold channels. So, we
subsequently evaluated the relationship between duration of deafness and the
across-site average polarity effect separately for each age-group and for each
channel classification.Figure 5(a) and (b) shows the relationship
between duration of deafness and the across-site average polarity effect on
high-focused-threshold channels for (a) the child-implanted group and (b) the
adult-implanted group. On high-focused-threshold channels, the average polarity
effect increased with increasing durations of deafness for the adult-implanted
listeners, R2marginal = .28,
R2conditional = .46;
F(1, 15.53) = 7.31, p = .016;
Bonferroni-adjusted = .025, but not for the child-implanted listeners,
R2marginal = .02,
R2conditional = .15;
F(1, 13.43) = 0.50, p = .49. On
low-focused-threshold channels, the polarity effect did not vary as a function
of duration of deafness for either group (ps > .05).
Figure 5.
Across-site average polarity effect at threshold (in dB) as a
function of duration of deafness (in years) on
high-focused-threshold channels for (a) the child-implanted ears and
(b) the adult-implanted ears. Circles indicate data from
first-implanted ears, and triangles represent data from
second-implanted ears. CI = cochlear implant; ACA = cathodic;
CAC = anodic.
Across-site average polarity effect at threshold (in dB) as a
function of duration of deafness (in years) on
high-focused-threshold channels for (a) the child-implanted ears and
(b) the adult-implanted ears. Circles indicate data from
first-implanted ears, and triangles represent data from
second-implanted ears. CI = cochlear implant; ACA = cathodic;
CAC = anodic.As many participants were bilaterally implanted, we performed a supplementary
demographic analysis to determine whether polarity sensitivity differed between
the first- and second-implanted ears. There was no between-ear difference in the
polarity effects for threshold, F(1, 161.24) < 0.01,
p = .96, MCL, F(1, 136.83) = 0.46,
p = .50, or DR, F(1, 145.27) = 0.06,
p = .81.Finally, relationships between phoneme perception, the across-site average
polarity effect at threshold, and demographic variables were evaluated. Table 3 shows phoneme
perception scores in percentage correct for each ear tested. Percentage correct
scores were converted to rationalized arcsine units prior to statistical
analysis to normalize error variance (Studebaker, 1985). The mixed-effects
models predicting phoneme perception included fixed effects for across-site
average polarity effect, age-group, and duration of deafness (vowels:
R2marginal = .23,
R2conditional = .74; consonants:
R2marginal = .22,
R2conditional = .71). Neither vowel
nor consonant perception were predicted by the across-site average polarity
effect, ps > .05; vowels: F(1,
26.96) = 2.22; consonants: F(1, 26.24) = 1.29, or age-group,
ps > .05; vowels: F(1, 27.21) = 1.83;
consonants: F(1, 24.52) = 2.12. However, duration of deafness
was inversely related to vowel, F(1, 33.82) = 9.93,
p = .003, and consonant, F(1,
30.74) = 9.36, p = .005, perception (Figure 6). Specifically, phoneme
perception decreased with increasing duration of preimplantation auditory
deprivation.
Table 3.
Speech Perception Scores.
ID
Vowel scores (percentage correct)
Consonant scores (percentage
correct)
First CI
Second CI
First CI
Second CI
Child-implanted Listeners
P02
97
100
79
84.5
P03
93.5
100
82
82
P04
96.5
97
77
86.5
P06
100
93
—
—
P07
98.5
96
95
72
P09
83
91.5
83
80
P11
63.5
22
60.5
17
P12
53.5
21.5
52
21.7
P13
98.5
—
77
—
P16
90
95
95
75
P17
93.5
—
63.3
—
Mean (SD)
88.0 (15.5)
79.4 (32.8)
76.5 (14.3)
64.9 (28.4)
Adult-implanted Listeners
S22
96.5
—
80
—
S23
95
87.7
74
75
S29
92
92.3
79
74
S39
100
95
92
91
S40
48
—
35
—
S43
67.7
—
80
—
S45
100
100
80
80
S46
55
—
31
—
S47
100
—
90
—
S49
41
88
41
44
S50
53.3
—
51
—
S52
82
—
55
—
S53
90
—
81
—
S54
98.5
—
85.5
—
S59
90
—
77
—
S60
45
—
41
—
Mean (SD)
78.4 (22.5)
92.6 (5.1)
67.0 (20.9)
72.8 (17.5)
Note. Vowel and consonant perception scores (in
percentage correct) for all ears tested. Stimuli were presented
in quiet at a level of 60 dB-A. CI = cochlear implant;
SD = standard deviation.
Figure 6.
(a) Vowel perception (RAUs) and (b) consonant perception (RAUs) as a
function of duration of deafness (in years). Circles indicate data
from first-implanted ears, and triangles represent data from
second-implanted ears. Green symbols indicate data from
child-implanted ears, and blue symbols indicate data from
adult-implanted ears. CI = cochlear implant; RAU = rationalized
arcsine unit.
(a) Vowel perception (RAUs) and (b) consonant perception (RAUs) as a
function of duration of deafness (in years). Circles indicate data
from first-implanted ears, and triangles represent data from
second-implanted ears. Green symbols indicate data from
child-implanted ears, and blue symbols indicate data from
adult-implanted ears. CI = cochlear implant; RAU = rationalized
arcsine unit.Speech Perception Scores.Note. Vowel and consonant perception scores (in
percentage correct) for all ears tested. Stimuli were presented
in quiet at a level of 60 dB-A. CI = cochlear implant;
SD = standard deviation.As before, relationships between speech perception and the across-site average
polarity effect were also assessed separately for high- and
low-focused-threshold channels; however, the relationship between speech
perception and polarity sensitivity did not differ as a function of channel
classification (ps > .05).
Discussion
Modeling evidence suggests that sensitivity to electrical stimulus polarity may
reflect the health of the SGN peripheral processes in CI listeners (Joshi et al., 2017; Rattay, Leao, et al., 2001;
Rattay, Lutter et al.,
2001; Resnick et al.,
2018). Specifically, better sensitivity to anodic (positive) current than
to cathodic (negative) current may indicate some degree of peripheral process
degeneration. The primary aim of this study was to characterize polarity sensitivity
in child- and adult-implanted listeners and to determine the relationship between
polarity sensitivity and traditional CI outcome measures. Results indicated that
polarity sensitivity varied widely within- and across-ears and did not differ
between child-implanted and adult-implanted participants. Interestingly, although
most electrodes showed better threshold and suprathreshold sensitivity to the anodic
polarity than to the cathodic polarity, the psychophysical DR of the cathodic
polarity was often larger than that of the anodic polarity. Across subjects, phoneme
perception performance was predicted by duration of deafness but not by polarity
sensitivity. Moreover, polarity sensitivity at threshold was related to duration of
deafness in the adult-implanted ears but not in the child-implanted ears. Subject-
and electrode-dependent differences in polarity sensitivity may be useful in
customizing programming interventions for CI listeners with a variety of hearing
histories.
Polarity Sensitivity at Threshold and at Suprathreshold Levels
The primary outcome measure in this study was the psychophysical polarity effect
at threshold, which has been hypothesized to reflect local peripheral process
integrity (Carlyon et al.,
2018; Jahn &
Arenberg, 2019; Macherey et al., 2017) and varies independently of electrode
position and intracochlear resistance in CI listeners (Jahn & Arenberg, 2019). The
polarity effect was defined as the difference in single-channel behavioral
thresholds measured in response to cathodic (ACA) and anodic (CAC) pulse shapes
(ACA − CAC). Our data suggest that the polarity effect at threshold is subject-
and electrode-dependent in a relatively large sample of child-implanted and
adult-implanted listeners (n = 41 ears). These findings are
consistent with previous studies in postlingually deafened adults with CIs
(Carlyon et al.,
2018; Jahn &
Arenberg, 2019; Macherey et al., 2017).In the present sample, the majority of electrodes (62.20%) showed lower threshold
sensitivity to the anodic polarity than to the cathodic polarity. Based on
modeling data, the proportion of electrodes with lower threshold sensitivity to
anodic than to cathodic stimulation reflects the proportion of fibers with some
degree of peripheral process degeneration or demyelination (Joshi et al., 2017;
Rattay, Leao, et al.,
2001; Rattay,
Lutter et al., 2001; Resnick et al., 2018). This theory
assumes that the site of spike initiation differs for anodic and cathodic
current. At low current levels, maximum depolarization occurs near the periphery
in response to cathodic stimulation, leading to relative hyperpolarization in
more central regions of the neuron. The inverse occurs for anodic stimulation,
wherein maximal depolarization occurs closer to the central axon.Thus, in a peripherally degenerated neuron, relatively high current levels are
required for cathodic stimulation to overcome the unmyelinated cell body and a
region of central hyperpolarization in order to generate an action potential.
Theoretically, lower thresholds are expected for anodic stimulation because the
action potential would not need to overcome the unmyelinated cell body or a
region of strong hyperpolarization to excite the central axon (Joshi et al., 2017;
Rattay, Leao, et al.,
2001; Rattay,
Lutter et al., 2001; Resnick et al., 2018). If polarity
sensitivity does reflect peripheral process integrity in humans, then our data
indicate that some degree of peripheral degeneration is likely to occur near
most electrode sites in child-implanted and adult-implanted listeners.Our results also indicated that channels with relatively high focused behavioral
thresholds are more likely to have lower threshold sensitivity to anodic
stimulation than to cathodic stimulation compared with channels with lower
focused thresholds. Focused thresholds are believed to reflect the overall
quality of the electrode–neuron interface, which is influenced by electrode
position (DeVries &
Arenberg, 2018b; DeVries et al., 2016; Long et al., 2014), intracochlear bone
and tissue growth (Spelman,
Clopton, & Pfingst, 1982), and the integrity of the auditory
neurons (Goldwyn, Bierer,
& Bierer, 2010). Channels with high focused thresholds are
believed to interface relatively poorly with the auditory nerve, as they are
often located far from the modiolus (DeVries & Arenberg, 2018b; DeVries et al., 2016;
Jahn & Arenberg,
2019; Long
et al., 2014) and have small evoked potential amplitudes and steep
evoked potential growth functions (Bierer et al., 2011; DeVries et al.,
2016).It is unlikely that the polarity effect at threshold was significantly influenced
by electrode position relative to the modiolus or intracochlear resistance in
this study. In a sample of 11 adult CI listeners, Jahn and Arenberg (2019) demonstrated
that the polarity effect at threshold varies independently of electrode position
relative to the modiolus and intracochlear resistance. They also showed that,
across listeners, the polarity effect at threshold predicted focused behavioral
thresholds (sQP; focusing coefficient = 0.9); specifically, channels with
relatively high focused thresholds were more likely to have better sensitivity
to anodic than to cathodic stimulation than channels with lower focused
thresholds.The present results agree with those of Jahn and Arenberg (2019), demonstrating
that channels with high focused thresholds, on average, had better sensitivity
to anodic stimulation relative to cathodic stimulation than
low-focused-threshold channels. We also show that this finding is consistent
across child-implanted and adult-implanted listeners. However, it should be
noted that although the polarity effect at threshold differed significantly
between high- and low-focused-threshold channels, there was substantial
variability in the magnitude of the polarity effect, irrespective of channel
classification.Some of the observed variability in polarity sensitivity likely results from the
channel selection procedure. Channels were selected based on the relative
magnitude of within-subject focused thresholds. It is likely that focused
thresholds on many of the selected electrodes were low or high as a result of
electrode position and intracochlear bone and tissue growth and not necessarily
because of the local status of the auditory nerve (Bierer et al., 2015; DeVries & Arenberg,
2018b; DeVries
et al., 2016; Jahn & Arenberg, 2019; Long et al., 2014). Regardless, the
present results suggest that channels that are estimated to interface poorly
with the auditory nerve are more likely to have lower threshold sensitivity to
anodic than to cathodic stimulation than channels with better interfaces. This
provides additional evidence that polarity sensitivity reflects an underlying
characteristic of the electrode–neuron interface that may be related to SGN
integrity in child- and adult-implanted listeners.If polarity sensitivity reflects peripheral process degeneration, it may have
utility in customizing CI programming parameters. Taken together, the results
from this study and from Jahn and Arenberg (2019) imply that a comprehensive approach that
considers electrode position and polarity sensitivity may help in selecting
channels for deactivation or current focusing. DeVries and Arenberg (2018a)
demonstrated that some listeners receive speech perception benefit when a subset
of electrodes that are located far from the modiolus are stimulated using a
spatially focused electrode configuration; however, some listeners did not
benefit, or performed worse, with that type of listening strategy. Others have
shown that deactivating channels that are estimated to interface poorly with the
auditory nerve improve speech perception scores for some listeners but not
others (Bierer & Litvak,
2016; Noble,
Gifford, Hedley-Williams, Dawant, & Labadie, 2014; Noble et al., 2016;
Noble, Labadie, Gifford,
& Dawant, 2013; Zhou, 2017).An individualized CI programming approach that considers estimates of electrode
position and neural integrity may be ideal. For example, it may be desirable to
implement current focusing on an electrode that is located far from the
modiolus, but that has better threshold sensitivity to cathodic stimulation than
to anodic stimulation (possibly indicating that the target neurons are healthy).
On the other hand, it may be best to deactivate or employ an anodic pulse shape
on an electrode with better threshold sensitivity to anodic stimulation
(possibly indicating local peripheral degeneration). Our results, and others
(e.g., Noble et al.,
2016), indicate that this combined programming approach could be
assessed in older children and adults with CIs. For infants and toddlers, it may
be possible to assess polarity sensitivity using objective measures such as the
ECAP. However, electrophysiological estimates of polarity sensitivity have not
yet been evaluated in young children.In addition to the primary outcome measure (polarity effect at threshold), we
assessed the polarity effect at MCL and differences in DR as a function of
polarity. A vast majority of electrodes tested (89.26%) had better
suprathreshold sensitivity to the anodic than to the cathodic polarity. This is
consistent with prior studies showing that MCLs are often lower for anodic than
for cathodic pulse shapes in postlingually deafened adults (Carlyon, Deeks, &
Macherey, 2013; Macherey et al., 2006, 2008, 2017). These findings also align with
modeling data suggesting a peripheral-to-central shift in the site of spike
initiation for cathodic stimulation at high current levels (Joshi et al., 2017;
Rattay, Leao, et al.,
2001; Rattay,
Lutter et al., 2001; Resnick et al., 2018). Moreover, the
present results demonstrate that variability in suprathreshold polarity
sensitivity is observed in both child- and adult-implanted listeners.Elevated MCLs for cathodic compared with anodic triphasic pulse shapes may
suggest that the anodic phase is generally more effective at exciting the SGNs
in human CI listeners. In psychophysical polarity sensitivity studies, lower
MCLs have been elicited in response to both pseudomonophasic and triphasic
electrical pulse shapes (Carlyon et al., 2013; Macherey et al., 2006, 2008, 2017). However, it
should be noted that psychophysical polarity effects in response to triphasic
pulses are often smaller in magnitude than those measured in response to
pseudomonophasic pulses (Carlyon et al., 2013; Macherey et al., 2006). The difference
in polarity effect magnitude between the two pulse shapes may be related to
differences in the ratio between the duration of the short and long phases of
each stimulus. Yet, Carlyon
et al. (2013) demonstrated that the size of the duration effect does
not differ markedly between the two waveforms. Moreover, electrophysiological
evidence suggests that suprathreshold neural responses are phase locked to the
anodic phase of pseudomonophasic and symmetric biphasic electrical pulses (Undurraga et al., 2010,
2013). The
present results are consistent with prior behavioral and electrophysiological
evidence showing that, for a variety of pulse shapes, anodic stimulation may be
more effective than cathodic stimulation at suprathreshold levels in human CI
listeners.If the majority of electrodes are more sensitive at threshold and at MCL to the
anodic polarity than to the cathodic polarity, it may be assumed that the DR is
simply shifted downward for anodic stimulation. However, although correlated,
the magnitude and sign of the polarity effect at threshold did not necessarily
correspond to that at MCL. This suggests that DR may also differ between anodic
and cathodic polarities in an electrode-dependent manner. In fact, despite
generally lower sensitivity for anodic pulses at threshold and at MCL, DR
remained larger for cathodic pulses on most electrodes (77.18%). Macherey et al. (2017)
demonstrated that loudness tends to grow less steeply as a function of current
level for cathodic pulse shapes than for anodic pulse shapes. Thus, if loudness
grows less steeply for cathodic than for anodic stimulation, larger DRs would
generally be expected for cathodic pulses.Notably, CI listeners programmed with larger electrical DRs have better speech
perception scores (Bento
et al., 2005; Fu
& Shannon, 2000; Loizou, Dorman, & Fitzke, 2000;
Zeng & Galvin,
1999) and better binaural sensitivity (Todd, Goupell, & Litovsky, 2017)
than those with smaller electrical DRs. We performed an exploratory regression
analysis to determine whether the DRs for cathodic or anodic stimuli were
correlated with vowel identification performance in this sample of CI listeners.
Linear mixed-effects models were used to account for clustering of two ears
within the bilaterally implanted listeners. Results indicated that better vowel
identification scores were associated with larger DRs for both the cathodic,
F(1, 132.19) = 8.02, p = .005, and the
anodic, F(1, 136.46) = 4.38, p = .04, pulse
shapes. The subjects’ clinical MAPs were not available for analysis, but it is
likely that subjects with larger DRs for the experimental stimuli also had
larger DRs in their everyday CI programs.The results of this investigation suggest that electrode-dependent differences in
DR as a function of polarity may have applications to CI programming. It is
possible that a pulse shape can be tailored on an electrode-specific basis to
maximize DR. For instance, anodic pulse shapes could be implemented on channels
with large anodic DRs and cathodic pulse shapes on electrodes with large
cathodic DRs. Our results suggest that novel programming interventions based on
polarity sensitivity may be attempted in both child- and adult-implanted
listeners. Future investigations should also assess the stability of polarity
sensitivity at MCL and DR over time.
Polarity Sensitivity as a Function of Age and Duration of Deafness
Another primary hypothesis of this study was that child-implanted listeners would
have smaller polarity effects at threshold than adult-implanted listeners. This
prediction was based on human temporal bone literature and limited behavioral
and electrophysiological evidence. Postmortem temporal bone studies demonstrate
a reduction in SGN density with increasing age and duration of hearing loss
(Makary et al.,
2011; Nadol,
1997; Nadol
et al., 1989; Otte et al., 1978). Histopathological data also show that hearing
loss etiology, which often differs between child- and adult-deafened
individuals, is a strong predictor of SGN density in humans (Nadol, 1997; Nadol et al., 1989;
Otte et al.,
1978). Available behavioral and electrophysiological evidence supports
the temporal bone findings, suggesting that children with CIs have lower focused
behavioral thresholds (DiNino, O’Brien, Bierer, Jahn, & Arenberg, 2019) and steeper
ECAP AGFs (Brown et al.,
2010) than adults. For these reasons, we predicted that the
child-implanted listeners would experience less peripheral process degeneration
than adult-implanted listeners, and that this would manifest as larger polarity
effects in the adults.However, results demonstrated that the polarity effect at threshold did not
differ between the child- and adult-implanted listeners in this study. Instead,
substantial variability in the polarity effect was observed across all subjects
and electrodes. If polarity sensitivity does reflect peripheral process
integrity, this may suggest that both child- and adult-implanted listeners
experience some degree of peripheral degeneration, even if they are implanted
early in life. It should also be noted that we did not test infants or young
children, and most of the adolescents in this study had been deaf since early
childhood. As SGN degeneration begins at the peripheral processes and progresses
centrally, some degree of peripheral degeneration should be expected in older
children and adults with profound deafness (Johnsson, 1974). Moreover, the initial
hypothesis was informed by human temporal bone and animal literature that had
assessed SGN density. It is possible that some degree of peripheral process
degeneration has occurred in both groups, but that SGN density remains higher in
the child-implanted listeners than in the adult-implanted listeners. Future
studies will investigate this distinction.We also hypothesized that individuals with longer periods of auditory deprivation
prior to implantation would have larger polarity effects than those with shorter
durations of deafness. This hypothesis was confirmed on high-focused-threshold
channels for adult-implanted listeners but not for child-implanted listeners.
Duration of deafness did not correlate with polarity sensitivity on
low-focused-threshold channels in either group of subjects. This finding is
somewhat consistent with Jahn and Arenberg (2019) who demonstrated that the across-site
average polarity effect was larger in adults with relatively long
preimplantation durations of deafness. However, in that study, the polarity
effect was averaged across all 16 electrodes in each ear. In this study, the
polarity effect was only assessed on four electrodes per ear, which may have
limited our ability to detect a relationship with duration of deafness.It is also possible that limited variability in duration of deafness among the
child-implanted listeners obscured a relationship with polarity sensitivity. The
longest duration of deafness in the child-implanted group was 10.1 years. In
contrast, the adult-implanted ears ranged in duration of deafness from 3.9 to 53
years. However, despite a limited range of preimplantation auditory deprivation,
the child-implanted listeners still demonstrated substantial variability in
polarity sensitivity. It is possible that hearing loss etiology, which is
strongly associated with SGN survival in human temporal bone analyses (Nadol, 1997; Nadol et al., 1989;
Otte et al.,
1978), is a more robust predictor of peripheral degeneration than age or
duration of deafness. Unfortunately, the majority of our participants presented
with unknown etiologies, so its relationship with polarity sensitivity could not
be evaluated here. Future investigations should attempt to recruit participants
with known hearing loss etiologies.
Phoneme Perception Is Related to Duration of Deafness, But Not to Polarity
Sensitivity
A final goal of this study was to evaluate the relationship between polarity
sensitivity and phoneme perception scores. We hypothesized that individuals with
large polarity effects at threshold would have poorer phoneme perception scores
than individuals with smaller polarity effects at threshold. Contrary to this
prediction, polarity sensitivity at threshold was not related to either vowel or
consonant perception. Instead, duration of deafness predicted phoneme
perception. Individuals with relatively short preimplantation periods of
auditory deprivation tended to have better phoneme perception scores than
individuals with longer durations of deafness. The observed relationship between
duration of deafness and speech perception is consistent across many
investigations (e.g., Blamey
et al., 2013; Green et al., 2007; Holden et al., 2013; Lazard et al., 2012).
Changes in the central auditory system as a consequence of auditory deprivation
that are not captured by peripheral measures likely play an important role in CI
outcomes.Moreover, signal detection theory analyses suggest that a dramatic loss of SGNs
may be necessary in order to detect measurable deficits in auditory
discrimination of intensity, frequency, and interaural time differences (Oxenham, 2016). In line
with signal detection theory, relationships between indirect estimates of SGN
integrity and speech perception are largely inconsistent across studies.
Histopathological studies have failed to demonstrate consistent, positive
relationships between postmortem SGN density and speech perception scores
assessed during life (Fayad
& Linthicum, 2006; Khan et al., 2005; Nadol et al., 2001;
Otte et al.,
1978). A few ECAP analyses have demonstrated that larger peak amplitudes
(DeVries et al.,
2016; Scheperle,
2017) and steeper AGF slopes (Brown et al., 1990; Kim et al., 2010) are
associated with better speech perception scores. For bilateral CI listeners,
between-ear differences in ECAP amplitude and slope measures (Schvartz-Leyzac & Pfingst,
2018) and in temporal integration abilities (Zhou & Pfingst, 2014) may be
predictive of between-ear differences in speech perception abilities. However,
there are also many studies that have not observed relationships between ECAP
responses and speech perception outcomes (e.g., Cosetti et al., 2010; Franck & Norton,
2001; Turner,
Mehr, Hughes, Brown, & Abbas, 2002).If SGN integrity plays a role in CI performance, the relationship is likely
complex and extends beyond measuring speech perception scores (Oxenham, 2016). In
fact, modeling data suggest that speech perception tasks may not be the optimal
psychophysical tools for assessing the effects of modest peripheral degeneration
on auditory perception with a CI. Resnick et al. (2018) proposed that
mild-to-moderate peripheral degeneration would not influence perception of
relevant speech features, which are generally longer than 20 ms in duration.
Instead, modest demyelination is expected to alter coding of fine temporal cues,
such as those needed to detect interaural timing differences. An interaural
timing difference-based sound localization task might be more appropriate than
speech perception for probing the behavioral implications of peripheral
degeneration.
Concluding Remarks
Variability in CI outcomes may be related, in part, to within- and
between-listener variation in the quality of the electrode–neuron interface. SGN
integrity may contribute to the efficacy with which a CI electrode interfaces
with the auditory nerve; however, it is difficult to estimate neural health in
humans. This study characterized polarity sensitivity, a proposed estimate of
SGN peripheral process integrity, in child- and adult-implanted listeners with
CIs. We demonstrated that, if polarity sensitivity reflects neural integrity in
CI listeners, then both child-implanted and adult-implanted listeners likely
experience some degree of peripheral degeneration, even if they are implanted
early in life. Future endeavors to apply polarity sensitivity to the study of
individualized programming strategies should incorporate both pediatric and
adult listeners.Subsequent investigations may attempt to use polarity sensitivity, in conjunction
with estimates of electrode position, to select appropriate CI electrodes for
deactivation and current focusing. It may also be possible to selectively
stimulate electrode sites with pulse shapes that optimize a listener’s DR. We
also demonstrated that polarity sensitivity may not predict phoneme perception
scores for CI listeners. Instead, future studies, especially those that intend
to implement programming adjustments, should consider utilizing psychophysical
tasks that assess one’s ability to process fine temporal cues.