Literature DB >> 31338402

Linear, planar and spherical tensor-valued diffusion MRI data by free waveform encoding in healthy brain, water, oil and liquid crystals.

Filip Szczepankiewicz1,2, Scott Hoge1,2, Carl-Fredrik Westin1,2.   

Abstract

Recently, several biophysical models and signal representations have been proposed for microstructure imaging based on tensor-valued, or multidimensional, diffusion MRI. The acquisition of the necessary data requires non-conventional pulse sequences, and data is therefore not available to the wider diffusion MRI community. To facilitate exploration and development of analysis techniques based on tensor-valued diffusion encoding, we share a comprehensive data set acquired in a healthy human brain. The data encompasses diffusion weighted images using linear, planar and spherical diffusion tensor encoding at multiple b-values and diffusion encoding directions. We also supply data acquired in several phantoms that may support validation. The data is hosted by GitHub: https://github.com/filip-szczepankiewicz/Szczepankiewicz_DIB_2019.

Entities:  

Keywords:  B-tensor encoding; Diffusion magnetic resonance imaging; Free waveform encoding; Human brain in vivo; Liquid crystal; Multidimensional diffusion encoding; Oil; Water

Year:  2019        PMID: 31338402      PMCID: PMC6626882          DOI: 10.1016/j.dib.2019.104208

Source DB:  PubMed          Journal:  Data Brief        ISSN: 2352-3409


Specifications table The data facilities design and testing of analysis techniques that require tensor-valued (or multidimensional) diffusion encoding. This provides value since acquisition of such data currently relies on a custom pulse sequence that is not widely available. The data includes repeated sampling of spherical b-tensors for analysis of noise characteristics. A subset of the data is matched with respect to the diffusion time spectrum for analysis of models of diffusion time dependency.

Data

This dataset contains diffusion weighted magnetic resonance imaging (MRI) data as well as fluid attenuated inversion recovery T2-weighted morphological images (T2-FLAIR) and can be found at https://github.com/filip-szczepankiewicz/Szczepankiewicz_DIB_2019. Experiments were performed in a healthy human brain in vivo, water, oil and hexagon phase liquid crystals (Fig. 1, Fig. 2).
Fig. 1

Signal versus b-value in brain and phantoms that contain HEX, oil or water. The circular markers are individual signal measurements (shifted along b to avoid overlap) and the crosses show the average signal for each combination of b-tensor shape and b-value in the ROIs (red outline in inset plots). The solid lines show the trend of the signal averaged over rotations of all b-tensors with a given b-value and shape, based on the q-space trajectory imaging representation [12]. Note that the scales of the y-axes are different across objects to compensate for the varying levels of diffusivity.

Fig. 2

Signal averaged over rotations for four b-values (columns) and three b-tensor shapes (rows). The color scale is equal within each column but differs within each row. There is an appreciable difference in the spherical and linear encoding at high b-values, where the difference is caused primarily by the presence of anisotropic diffusion [15], [21].

Signal versus b-value in brain and phantoms that contain HEX, oil or water. The circular markers are individual signal measurements (shifted along b to avoid overlap) and the crosses show the average signal for each combination of b-tensor shape and b-value in the ROIs (red outline in inset plots). The solid lines show the trend of the signal averaged over rotations of all b-tensors with a given b-value and shape, based on the q-space trajectory imaging representation [12]. Note that the scales of the y-axes are different across objects to compensate for the varying levels of diffusivity. Signal averaged over rotations for four b-values (columns) and three b-tensor shapes (rows). The color scale is equal within each column but differs within each row. There is an appreciable difference in the spherical and linear encoding at high b-values, where the difference is caused primarily by the presence of anisotropic diffusion [15], [21]. Data was exported from the scanner in the native DICOM format and anonymized using DicomCleaner (http://www.pixelmed.com/cleaner.html, access date 2019-02-10). We removed subject identity, replaced all UIDs and hardware and institution ID. All tags described by Newhauser et al. [1] were either removed or overwritten with anonymized variants. The vendor-specific private tags were unaltered. These contain information that is specific to the free waveform (FWF) sequence, such as waveform timing, gradient scaling and balance gradient parameters; necessary to reconstruct the experiment in detail. We also provide FWF-header tools that are capable of extracting this information, available at https://github.com/filip-szczepankiewicz/fwf_header_tools. DICOM images for each subject are stored in folders denoted ‘DICOM_zip’ in a compressed format (.zip). Anonymized DICOM images were converted to 4D NIfTI files in a compressed format (.nii.gz), along with b-value (.bval) and encoding direction files (.bvec), using MRIcroGL (v1.0.20180623, https://www.nitrc.org/projects/mricron). These files are stored under folders denoted ‘NII’ along with corresponding meta data files (.json). NIfTI images were also prepared to be compatible with the multidimensional diffusion MRI framework [2] available at https://github.com/markus-nilsson/md-dmri. As such, the diffusion-weighted data from each subject or phantom was merged into a single 4D-volume and an appropriate experimental parameter structure (_xps.mat) was constructed. These files are denoted ‘FWF_MERGED’ and are stored in folders denoted ‘MD-dMRI’. In addition to the original in vivo data, we provide a set of data that was corrected for motion and eddy-currents by registering the images to an extrapolated reference [3] using ElastiX [4]. These files are denoted with a suffix ‘_mc’. As a supplement to the protocol overview in Table 3, the repository also contains the gradient waveforms used in the FWF sequence (Table 1), the sampling schemes in terms of diffusion directions native to the scanner (.dvs, Table 2), a complete protocol description, and the ElastiX co-registration parameters.
Table 3

Order of acquisition with different b-tensor shapes. In the case of the water phantom, and oil phantom, the planar encoding was omitted. The time is given in minutes:seconds. The total acquisition time was approximately 23 minutes for the complete protocol.

ShapePartTime
Spherical1/52:21
Linear1/41:20
Planar1/41:20
Spherical2/52:21
Linear2/41:20
Planar2/41:20
Spherical3/52:21
Linear3/41:20
Planar3/41:20
Spherical4/52:21
Linear4/41:14
Planar4/41:14
Spherical5/52:21
Table 1

Gradient waveform samples along the time dimension for linear, planar and spherical encoding. Note that the linear encoding uses only g1 from the spherical encoding waveform (g2 and g3 are zero). The planar encoding was optimized separately and uses only g2 and g3 (g1 is zero). The waveforms are defined in the space that is employed by the gradient system such that the integral is zero only if the second part is multiplied by −1.

Linear and Spherical
Planar
g1g2g3g2g3
0.00000.00000.00000.00000.0000
−0.20050.93340.3029−0.73010.6840
−0.20500.93240.3031−0.72890.6853
−0.21460.93020.3032−0.72630.6880
−0.23130.92630.3030−0.72220.6924
−0.25890.91930.3019−0.71620.6986
−0.30590.90600.2980−0.70770.7072
−0.38920.87670.2883−0.69580.7189
−0.38500.71470.3234−0.67870.7350
−0.36870.52550.3653−0.65360.7575
−0.35090.32410.4070−0.61460.7894
−0.33230.11660.4457−0.55060.8353
−0.3136−0.09060.4783−0.44390.8274
−0.2956−0.29130.5019−0.32170.7803
−0.2790−0.47930.5139−0.19310.7293
−0.2642−0.64910.5118−0.05980.6745
−0.2518−0.79570.49390.07660.6164
−0.2350−0.87220.43290.21420.5553
−0.2187−0.91110.35410.35140.4915
−0.2063−0.94090.27470.48610.4255
−0.1977−0.96270.19330.61680.3576
−0.1938−0.97680.10800.74170.2883
−0.1967−0.98200.01590.85900.2178
−0.2114−0.9751−0.08830.96720.1467
−0.2292−0.9219−0.21500.99960.0406
−0.2299−0.8091−0.35610.9984−0.0639
−0.2290−0.6748−0.50110.9888−0.1521
−0.2253−0.5239−0.64600.9749−0.2247
−0.2178−0.3620−0.78680.9593−0.2840
−0.2056−0.1948−0.91940.9436−0.3326
−0.1391−0.0473−0.99080.9283−0.3730
−0.04760.0607−0.99870.9138−0.4074
0.02150.1452−0.99090.8998−0.4374
0.07250.2136−0.97590.8862−0.4642
0.11140.2709−0.95790.8728−0.4891
0.14260.3204−0.93830.8590−0.5128
0.16900.3641−0.91770.8447−0.5362
0.00000.00000.00000.00000.0000

Gradients are off during the refocusing pulse

0.00000.00000.00000.00000.0000
−0.3734−0.17680.9125−0.69630.7184
−0.3825−0.23100.8965−0.68610.7281
−0.3919−0.28950.8752−0.67480.7387
−0.4015−0.35430.8465−0.66190.7502
−0.4108−0.42900.8065−0.64720.7630
−0.4182−0.52020.7469−0.63010.7771
−0.4178−0.64230.6451−0.61020.7928
−0.3855−0.81730.4321−0.58650.8105
−0.3110−0.94180.1401−0.55780.8305
−0.2526−0.9669−0.0674−0.52260.8531
−0.2100−0.9541−0.2213−0.47800.8789
−0.1766−0.9227−0.3474−0.41990.9081
−0.1491−0.8788−0.4570−0.34090.9406
−0.1258−0.8239−0.5555−0.22540.9747
−0.1056−0.7583−0.6459−0.03031.0000
−0.0882−0.6809−0.72930.95490.2984
−0.0734−0.5900−0.80610.99840.0635
−0.0615−0.4830−0.87530.9997−0.0387
−0.0533−0.3556−0.93490.9953−0.1015
−0.0506−0.2005−0.98010.9899−0.1448
−0.0575−0.0019−1.00000.9848−0.1764
−0.09090.2976−0.95210.9802−0.2004
−0.30270.9509−0.08600.9762−0.2190
−0.27370.9610−0.06920.9728−0.2337
−0.25240.9675−0.05960.9699−0.2452
−0.23640.9719−0.05330.9676−0.2544
−0.22450.9749−0.04900.9657−0.2615
−0.21580.9770−0.04590.9642−0.2669
−0.20970.9785−0.04390.9631−0.2708
−0.20580.9794−0.04260.9624−0.2733
−0.20390.9798−0.04200.9621−0.2745
0.00000.00000.00000.00000.0000
Table 2

Sets of directions derived from platonic solids [12]. These directions can be arbitrarily combined with no co-linearity and retained isotropic sampling.

u1u2u3
6 Directions
0.00000.52570.8507
0.00000.5257−0.8507
0.52570.85070.0000
0.5257−0.85070.0000
0.85070.00000.5257
−0.85070.00000.5257
10 Directions
0.57740.57740.5774
0.00000.93420.3568
0.35680.00000.9342
−0.35680.00000.9342
−0.57740.57740.5774
0.00000.9342−0.3568
0.57740.5774−0.5774
0.93420.35680.0000
0.5774−0.57740.5774
0.9342−0.35680.0000
30 Directions
0.00000.20180.9794
0.00000.2018−0.9794
0.20180.97940.0000
0.2018−0.97940.0000
0.97940.00000.2018
−0.97940.00000.2018
0.40350.85470.3265
0.4035−0.85470.3265
0.40350.8547−0.3265
0.4035−0.8547−0.3265
0.85470.32650.4035
−0.85470.32650.4035
0.8547−0.32650.4035
−0.8547−0.32650.4035
0.32650.40350.8547
0.32650.4035−0.8547
−0.32650.40350.8547
−0.32650.4035−0.8547
0.20180.73000.6530
0.2018−0.73000.6530
0.20180.7300−0.6530
0.2018−0.7300−0.6530
0.73000.65300.2018
−0.73000.65300.2018
0.7300−0.65300.2018
−0.7300−0.65300.2018
0.65300.20180.7300
0.65300.2018−0.7300
−0.65300.20180.7300
−0.65300.2018−0.7300
Gradient waveform samples along the time dimension for linear, planar and spherical encoding. Note that the linear encoding uses only g1 from the spherical encoding waveform (g2 and g3 are zero). The planar encoding was optimized separately and uses only g2 and g3 (g1 is zero). The waveforms are defined in the space that is employed by the gradient system such that the integral is zero only if the second part is multiplied by −1. Sets of directions derived from platonic solids [12]. These directions can be arbitrarily combined with no co-linearity and retained isotropic sampling. Order of acquisition with different b-tensor shapes. In the case of the water phantom, and oil phantom, the planar encoding was omitted. The time is given in minutes:seconds. The total acquisition time was approximately 23 minutes for the complete protocol.

Experimental design, materials, and methods

Hardware and imaging

Imaging was performed on a MAGNETOM 3T Prisma (Siemens Healthcare, Germany) which has a maximum gradient amplitude of 80 mT/m at a slew rate of 200 T/m/s (software version VE11C), and a 20-channel head-coil array. We used a custom pulse sequence based on the diffusion weighted spin-echo (a_ep2d_diff) to support free waveform encoding (FWF, version 1.12a) [5]. Briefly, the FWF sequence replaces the trapezoidal diffusion encoding gradient pulses with waveforms that can be arbitrarily defined by the user (Fig. 3).
Fig. 3

The top row shows the gradient waveforms used for linear, planar and spherical b-tensor encoding; each waveform is scaled in amplitude to yield b = 2 ms/μm2. The durations of the pulses were identical for all encoding shapes, as shown in the top-left panel. Note that g1 is the same for both linear and spherical encoding. The bottom row shows the diffusion encoding directions (including antipodal points) of the symmetry axis (g1).

The top row shows the gradient waveforms used for linear, planar and spherical b-tensor encoding; each waveform is scaled in amplitude to yield b = 2 ms/μm2. The durations of the pulses were identical for all encoding shapes, as shown in the top-left panel. Note that g1 is the same for both linear and spherical encoding. The bottom row shows the diffusion encoding directions (including antipodal points) of the symmetry axis (g1). We used imaging parameters TR = 3.2 s, TE = 91 ms, FOV = 220 × 220 × 60 mm3, matrix = 92 × 92 × 25, resolution = 2.5 × 2.5 × 2.5 mm3, partial-Fourier = 7/8, bandwidth = 1940 Hz/pix, and echo spacing = 0.6 ms. We used interleaved slice excitation, strong fat saturation, in-plane acceleration iPAT = 2 with GRAPPA reconstruction and 30 reference lines. The imaging stack was transversal (axial) and centered on the corpus callosum. The phase encoding was done along the anterior-posterior direction. A 2D T2-FLAIR was acquired for anatomical reference. Here, we used imaging parameters TR = 5.5 s, TE = 95 ms, FOV = 256 × 256 × 120 mm3, matrix = 256 × 256 × 50, resolution = 1.0 × 1.0 × 2.4 mm3, bandwidth = 222 Hz/Px, echo spacing = 8.7 ms, and inversion time 1.9 s. We used interleaved slice excitation, no fat saturation, in-plane acceleration iPAT = 2 with GRAPPA reconstruction and 47 reference lines.

Gradient waveform optimization

Asymmetric gradient waveforms for tensor-valued diffusion encoding were tailored to the hardware by using the optimization framework by Sjölund et al. [6]. Each waveform was specified as a function of time in discrete steps for three orthogonal axes, denoted Spherical and planar tensor encoding (STE and PTE) were optimized using the Euclidean norm, gradient amplitude limit 80 mT/m, slew rate limit 100 T/m/s, heating coefficient 0.9 and 77 temporal samples. The STE waveform was designed to have its longest diffusion time along g1 [7]. The waveform used for linear tensor encoding (LTE) was defined as g1 from STE, making them perfectly matched with respect to the diffusion time spectrum along that direction [7], [8], [9]. For all waveforms, the duration of the waveforms before and after the refocusing pulse was 35.67 ms and 30.85 ms separated by 8.02 ms (Fig. 3). Since the waveforms were designed to be asymmetric, they were Maxwell-compensated to alleviate effects of concomitant gradients, as described by Szczepankiewicz et al. [10]. The waveforms are stated in Table 1, and the framework for Maxwell-compensated waveform optimization is available at https://github.com/jsjol/NOW. The waveform was resampled to match the gradient system raster time (10 μs) by linear interpolation. A balance gradient was used to correct for small timing and interpolation errors so that the zeroth moment of the final gradient waveform was zero [10], [11].

Sampling scheme

The signal was sampled at diffusion encoding strengths b = [.1, .7, 1.4, 2] ms/μm2. Spherical encoding was performed in 10 rotations for each b-value, and all measurements were repeated five times. Linear and planar encoding was performed in 10, 10, 16, 46 rotations for the same four b-values, respectively. The direction sets were derived from platonic solids (atoms of 6, 10 and 30 directions) so that multiple atoms could be combined with optimal directional coverage of the sphere with no colinear measurements within each shell [12]. Fig. 3 shows the sampling scheme for each shell and shape and all direction sets are defined in Table 2. We also included 13 images without diffusion encoding (b = 0). Each direction vector () defines the target direction of the first axis of the gradient waveform (). The rotated gradient waveform is therefore defined aswhere Note that the waveform amplitude is always scaled such that it yields the requested b-value, and that the timing or duration of the waveforms was never changed. In total, the signal was sampled 377 times per voxel for a total scan time of approximately 23 min. The design matrix is such that it supports inversion of the q-space trajectory imaging representation (QTI) [12], and all techniques that require equal, or lower, measurement rank, for example diffusional variance decomposition (DIVIDE and CODIVIDE) [13], [14], [15], diffusion tensor imaging (DTI) [16] and diffusion kurtosis imaging (DKI) [17]. The order of acquisition was permuted so that energy consumption and heating [18], as well as potential system drift, was alleviated [19]; the order of b-values and directions was randomly permuted for each shape. Table 3 shows the acquisition order, and an exact description of the shapes, b-values and directions is available in the metadata.

Subject and phantoms

In vivo experiments were performed in a single healthy brain (age 53 y, male). The study was approved by the local ethics committee and written consent was given before the study. Identical experiments were also performed in a liquid crystal phantom. The crystals are in a reverse hexagonal phase (HEX), which creates nanometer-scale tubes in which water diffusion is effectively one-dimensional [20]. The liquid crystals therefore exhibit a high microscopic diffusion anisotropy and variable macroscopic diffusion anisotropy due to variable orientation coherence on the voxel scale. Furthermore, the linear and spherical encoding was performed in water and oil. As a brief overview, Fig. 1 shows examples of the signal vs b-value in brain, HEX, water and oil, and Fig. 2 shows the signal averaged over direction in an axial slice through the brain for all included b-tensor shapes and b-values.

Specifications table

Subject areaMagnetic resonance imaging physics
More specific subject areaDiffusion magnetic resonance imaging
Type of dataDiffusion-weighted signal
How was data acquiredSpin-echo with echo-planar readout with free waveform diffusion encoding on clinical MRI hardware
Data formatRaw anonymized data in DICOM and NIfTI formats with native and complementary metadata
Experimental factorsSignal was diffusion-weighted with linear, planar and spherical b-tensor encoding in healthy human brain, oil, water and hexagonal phase liquid crystal at encoding strength up to 2 ms/μm2 for 10 to 46 rotations.
Experimental featuresLinear and spherical encoding are matched with respect to the diffusion time spectrum along one direction. Encoding tensor shapes, strengths and rotations were distributed over time to minimize effects of drift.
Data source locationBrigham and Women's Hospital, Boston, MA, USA
Data accessibilityMulti-format raw data with corresponding metadata is available at: https://github.com/filip-szczepankiewicz/Szczepankiewicz_DIB_2019
Value of the data

The data facilities design and testing of analysis techniques that require tensor-valued (or multidimensional) diffusion encoding. This provides value since acquisition of such data currently relies on a custom pulse sequence that is not widely available.

The data includes repeated sampling of spherical b-tensors for analysis of noise characteristics.

A subset of the data is matched with respect to the diffusion time spectrum for analysis of models of diffusion time dependency.

  15 in total

1.  elastix: a toolbox for intensity-based medical image registration.

Authors:  Stefan Klein; Marius Staring; Keelin Murphy; Max A Viergever; Josien P W Pluim
Journal:  IEEE Trans Med Imaging       Date:  2009-11-17       Impact factor: 10.048

2.  Anonymization of DICOM electronic medical records for radiation therapy.

Authors:  Wayne Newhauser; Timothy Jones; Stuart Swerdloff; Warren Newhauser; Mark Cilia; Robert Carver; Andy Halloran; Rui Zhang
Journal:  Comput Biol Med       Date:  2014-07-26       Impact factor: 4.589

3.  Q-space trajectory imaging for multidimensional diffusion MRI of the human brain.

Authors:  Carl-Fredrik Westin; Hans Knutsson; Ofer Pasternak; Filip Szczepankiewicz; Evren Özarslan; Danielle van Westen; Cecilia Mattisson; Mats Bogren; Lauren J O'Donnell; Marek Kubicki; Daniel Topgaard; Markus Nilsson
Journal:  Neuroimage       Date:  2016-02-23       Impact factor: 6.556

4.  Diffusional kurtosis imaging: the quantification of non-gaussian water diffusion by means of magnetic resonance imaging.

Authors:  Jens H Jensen; Joseph A Helpern; Anita Ramani; Hanzhang Lu; Kyle Kaczynski
Journal:  Magn Reson Med       Date:  2005-06       Impact factor: 4.668

5.  Neurite density imaging versus imaging of microscopic anisotropy in diffusion MRI: A model comparison using spherical tensor encoding.

Authors:  Björn Lampinen; Filip Szczepankiewicz; Johan Mårtensson; Danielle van Westen; Pia C Sundgren; Markus Nilsson
Journal:  Neuroimage       Date:  2016-11-27       Impact factor: 6.556

6.  The link between diffusion MRI and tumor heterogeneity: Mapping cell eccentricity and density by diffusional variance decomposition (DIVIDE).

Authors:  Filip Szczepankiewicz; Danielle van Westen; Elisabet Englund; Carl-Fredrik Westin; Freddy Ståhlberg; Jimmy Lätt; Pia C Sundgren; Markus Nilsson
Journal:  Neuroimage       Date:  2016-07-20       Impact factor: 6.556

7.  Quantification of microscopic diffusion anisotropy disentangles effects of orientation dispersion from microstructure: applications in healthy volunteers and in brain tumors.

Authors:  Filip Szczepankiewicz; Samo Lasič; Danielle van Westen; Pia C Sundgren; Elisabet Englund; Carl-Fredrik Westin; Freddy Ståhlberg; Jimmy Lätt; Daniel Topgaard; Markus Nilsson
Journal:  Neuroimage       Date:  2014-10-02       Impact factor: 6.556

8.  Constrained optimization of gradient waveforms for generalized diffusion encoding.

Authors:  Jens Sjölund; Filip Szczepankiewicz; Markus Nilsson; Daniel Topgaard; Carl-Fredrik Westin; Hans Knutsson
Journal:  J Magn Reson       Date:  2015-10-31       Impact factor: 2.229

9.  The importance of correcting for signal drift in diffusion MRI.

Authors:  Sjoerd B Vos; Chantal M W Tax; Peter R Luijten; Sebastien Ourselin; Alexander Leemans; Martijn Froeling
Journal:  Magn Reson Med       Date:  2016-01-29       Impact factor: 4.668

10.  Extrapolation-Based References Improve Motion and Eddy-Current Correction of High B-Value DWI Data: Application in Parkinson's Disease Dementia.

Authors:  Markus Nilsson; Filip Szczepankiewicz; Danielle van Westen; Oskar Hansson
Journal:  PLoS One       Date:  2015-11-03       Impact factor: 3.240

View more
  4 in total

1.  Probing tissue microstructure by diffusion skewness tensor imaging.

Authors:  Lipeng Ning; Filip Szczepankiewicz; Markus Nilsson; Yogesh Rathi; Carl-Fredrik Westin
Journal:  Sci Rep       Date:  2021-01-08       Impact factor: 4.379

2.  Towards unconstrained compartment modeling in white matter using diffusion-relaxation MRI with tensor-valued diffusion encoding.

Authors:  Björn Lampinen; Filip Szczepankiewicz; Johan Mårtensson; Danielle van Westen; Oskar Hansson; Carl-Fredrik Westin; Markus Nilsson
Journal:  Magn Reson Med       Date:  2020-03-06       Impact factor: 4.668

3.  Motion-compensated gradient waveforms for tensor-valued diffusion encoding by constrained numerical optimization.

Authors:  Filip Szczepankiewicz; Jens Sjölund; Erica Dall'Armellina; Sven Plein; Jürgen E Schneider; Irvin Teh; Carl-Fredrik Westin
Journal:  Magn Reson Med       Date:  2020-10-13       Impact factor: 4.668

4.  Validation of cardiac diffusion tensor imaging sequences: A multicentre test-retest phantom study.

Authors:  Irvin Teh; William A Romero R; Jordan Boyle; Jaume Coll-Font; Erica Dall'Armellina; Daniel B Ennis; Pedro F Ferreira; Prateek Kalra; Arunark Kolipaka; Sebastian Kozerke; David Lohr; François-Pierre Mongeon; Kévin Moulin; Christopher Nguyen; Sonia Nielles-Vallespin; Brian Raterman; Laura M Schreiber; Andrew D Scott; David E Sosnovik; Christian T Stoeck; Cyril Tous; Elizabeth M Tunnicliffe; Andreas M Weng; Pierre Croisille; Magalie Viallon; Jürgen E Schneider
Journal:  NMR Biomed       Date:  2022-02-08       Impact factor: 4.478

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.