Early First-Generation Trifecta Valve Failure: A Case Series and a Review of the Literature.

Daisuke Kaneyuki1, Hiroyuki Nakajima2, Toshihisa Asakura2, Akihiro Yoshitake2, Chiho Tokunaga2, Masato Tochii2, Jun Hayashi2, Akitoshi Takazawa2, Hiroaki Izumida2, Atsushi Iguchi2.   

Abstract

BACKGROUND: The Trifecta valve (Abbott, St Paul, MN) has excellent hemodynamic performance with acceptable rates of freedom from structural valve degeneration. However, some recent studies have reported early Trifecta valve failure. Here, we report a case series of seven Trifecta valve failures with a review of the literature.
METHODS: Of 107 implantations of Trifecta bioprostheses between 2012 and 2014, we encountered seven Trifecta valve failures (6.5%). Failure of a 19-mm Trifecta valve occurred in 1 patient, failure of a 21-mm Trifecta valve occurred in 5 patients, and failure of a 23-mm Trifecta valve occurred in 1 patient. The mean duration of valve durability was 51 ± 16 months. The mean effective orifice area index on the first echocardiogram after Trifecta valve implantation was 0.96 ± 0.26. The mode of presentation was prosthetic valve stenosis in 3 patients and severe aortic regurgitation in 4 patients.
RESULTS: Six patients underwent redo surgical aortic valve replacement. The common pathologic findings were circumferential pannus formation with noncoronary cusp tear and leaflet calcification. The rates of preoperative end-renal stage disease and postoperative prosthesis-patient mismatch were higher in patients with Trifecta valve failure. The incidence of early Trifecta valve failure was 3.1% at 48 months and 13.1% at 72 months.
CONCLUSIONS: In our experience, early Trifecta valve failure was caused by cusp tears or leaflet calcification. Patients with end-renal stage disease and prosthesis-patient mismatch should be closely followed. Some patients with cusp tears may require urgent surgery.
Copyright © 2020 The Society of Thoracic Surgeons. Published by Elsevier Inc. All rights reserved.

Entities:  

Year:  2019        PMID: 31336064     DOI: 10.1016/j.athoracsur.2019.05.073

Source DB:  PubMed          Journal:  Ann Thorac Surg        ISSN: 0003-4975            Impact factor:   4.330


  10 in total

1.  Mid-term clinical and health-related quality of life outcomes for the Trifecta bioprosthesis.

Authors:  Biswarup Purkayastha; Md Wasim Khan; Atanu Saha; Debasis Das; Lalit Kapoor; Mrinalendu Das; Pradeep Narayan
Journal:  Indian J Thorac Cardiovasc Surg       Date:  2021-04-27

2.  Impact of postoperative patient-prosthesis mismatch as a risk factor for early structural valve deterioration after aortic valve replacement with Trifecta bioprosthesis.

Authors:  Tatsuto Wakami; Shigeki Koizumi; Tadaaki Koyama
Journal:  J Cardiothorac Surg       Date:  2022-07-08       Impact factor: 1.522

3.  Early cusp tear of the second-generation Trifecta valve with Glide Technology.

Authors:  Akimasa Morisaki; Yosuke Takahashi; Shinsuke Nishimura; Toshihiko Shibata
Journal:  Interact Cardiovasc Thorac Surg       Date:  2021-10-04

4.  Early surgical aortic bioprosthesis failure caused by cusp tear: A word of caution.

Authors:  Thibault Schaeffer; Denis Berdajs; Oliver Reuthebuch
Journal:  Clin Case Rep       Date:  2020-11-28

5.  Different Rates of Bioprosthetic Aortic Valve Failure With Perimount™ and Trifecta™ Bioprostheses.

Authors:  Rüdiger Lange; Zahra Alalawi; Stephanie Voss; Johannes Boehm; Markus Krane; Keti Vitanova
Journal:  Front Cardiovasc Med       Date:  2022-01-20

6.  Modes of failure of Trifecta aortic valve prosthesis.

Authors:  Pietro Giorgio Malvindi; Hassan Kattach; Suvitesh Luthra; Sunil Ohri
Journal:  Interact Cardiovasc Thorac Surg       Date:  2022-07-09

7.  Trifecta versus Perimount Magna Ease aortic valves: Failure mechanisms.

Authors:  Ryo Suzuki; Toshiro Ito; Masato Suzuki; Shunsuke Ohori; Ryo Takayanagi; Shiro Miura
Journal:  Asian Cardiovasc Thorac Ann       Date:  2022-05-22

8.  Structural valve deterioration of a pericardial bioprosthesis.

Authors:  Hassan Kattach; Clifford W Barlow; Sunil K Ohri
Journal:  JTCVS Open       Date:  2021-12-14

9.  Structural valve deterioration after aortic valve replacement with the Trifecta valve.

Authors:  Paul Werner; Jasmin Gritsch; Sabine Scherzer; Christoph Gross; Marco Russo; Iuliana Coti; Alfred Kocher; Guenther Laufer; Martin Andreas
Journal:  Interact Cardiovasc Thorac Surg       Date:  2021-01-01

10.  A Retrospective Comparison of Hemodynamic and Clinical Outcomes between Two Differently Designed Aortic Bioprostheses for Small Aortic Annuli.

Authors:  Do Jung Kim; Sak Lee; Hyun-Chel Joo; Young-Nam Youn; Kyung-Jong Yoo; Seung Hyun Lee
Journal:  J Clin Med       Date:  2021-03-04       Impact factor: 4.241

  10 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.