Eun Jin Lee1,2, Hyun-Ki Kim1, Sunyoung Ahn1, Woochang Lee1, Hyun Soo Kim2, Sail Chun1, Won-Ki Min1.
Abstract
BACKGROUND: We evaluated the analytical performance of a newly developed electrochemiluminescence immunoassay for everolimus and sirolimus compared to that of liquid chromatography-tandem mass spectrometry (LC-MS/MS).Entities:
Keywords: everolimus; immunoassay; liquid chromatography-tandem mass spectrometry; sirolimus; therapeutic drug monitoring
Mesh:
Substances:
Year: 2019 PMID: 31197901 PMCID: PMC6757180 DOI: 10.1002/jcla.22941
Source DB: PubMed Journal: J Clin Lab Anal ISSN: 0887-8013 Impact factor: 2.352
Precisions of the Elecsys® Everolimus and Elecsys® Sirolimus assays
| Elecsys® Assay | Level | Mean (ng/mL) | Within‐run imprecision | Total imprecision | ||
|---|---|---|---|---|---|---|
| SD (ng/mL) | CV (%) | SD (ng/mL) | CV (%) | |||
| Everolimus | Low | 2.86 | 0.132 | 4.5 | 0.185 | 6.4 |
| Middle | 9.84 | 0.235 | 2.3 | 0.458 | 4.5 | |
| High | 18.52 | 0.486 | 2.6 | 0.918 | 4.9 | |
| Sirolimus | Low | 3.81 | 0.187 | 4.8 | 0.320 | 8.1 |
| Middle | 10.72 | 0.452 | 4.1 | 0.644 | 5.8 | |
| High | 19.75 | 0.658 | 3.3 | 0.937 | 4.7 | |
Abbreviations: CV, coefficient of variation; SD, Standard deviation.
Figure 1Linearities (left) and %recoveries (right) in Elecsys® Everolimus assays (A) and Elecsys® Sirolimus assays (B). The solid lines are Y = X, and dotted lines represent linear regression
Figure 2Deming regression (left) and Bland‐Altman plots (right) between electrochemiluminescence immunoassay and liquid chromatography‐tandem mass spectrometry for evaluating everolimus assay (A) and sirolimus assay (B). In the left figures, the solid lines are Y = X and dotted lines represent the Deming regression. In the right figures, the solid lines are Y = X and dotted lines represent the mean value (middle line) and ± 2 SD (upper and lower lines)
Summary of comparison between electrochemiluminescence immunoassay and liquid chromatography‐tandem mass spectrometry
| Elecsys® Everolimus | Elecsys® Sirolimus | |
|---|---|---|
| Deming | ||
| N | 60 | 60 |
|
Slope |
1.216 |
1.286 |
|
Intercept |
0.423 |
0.188 |
| Std Err Est | 1.258 | 1.008 |
| Passing‐Bablok | ||
| N | 60 | 60 |
|
Slope |
1.156 |
1.254 |
|
Intercept |
0.788 |
0.392 |
| Std Err Est | 1.240 | 1.000 |
|
| 0.959 | 0.980 |
Abbreviations: CI, confidence interval; N, Number; Std Err Est, standard error of the estimate.
Correlation coefficient (R) was calculated by Pearson's correlation analysis.