Accuracy evaluation of automated electrochemiluminescence immunoassay for everolimus and sirolimus compared to liquid chromatography-tandem mass spectrometry.

Eun Jin Lee1,2, Hyun-Ki Kim1, Sunyoung Ahn1, Woochang Lee1, Hyun Soo Kim2, Sail Chun1, Won-Ki Min1.   

Abstract

BACKGROUND: We evaluated the analytical performance of a newly developed electrochemiluminescence immunoassay for everolimus and sirolimus compared to that of liquid chromatography-tandem mass spectrometry (LC-MS/MS).
METHODS: According to Clinical and Laboratory Standards Institute guidelines, the analytical performance including precision, recovery, linearity, and carryover was evaluated. For correlation evaluation, the results of Elecsys® analysis of everolimus and sirolimus were compared with those of LC-MS/MS using 120 samples from patients treated with everolimus or sirolimus.
RESULTS: The within-run and total imprecision values were as follows: 2.3%-4.5% and 4.5%-6.4% for the everolimus assay; 3.3%-4.8% and 4.7%-8.1% for the sirolimus assay, respectively. The measured concentration was linear over the range of 0.718-27.585 ng/mL for everolimus analysis and 0.789-26.880 ng/mL for sirolimus analysis (all R2  > 0.99). Recovery was 93.5%-105.5% for the everolimus assay and 99.2%-109.1% for the sirolimus assay (except lowest levels). Carryover was -1.09% for the everolimus assay and -0.12% for the sirolimus assay. The results of the two chemiluminescence immunoassays showed acceptable correlations with those of LC-MS/MS (R = 0.9585 and R = 0.9799, respectively). The two immunoassays showed slightly proportional biases compared to LC-MS/MS.
CONCLUSION: Elecsys® Everolimus and Sirolimus assays showed acceptable analytical performance in precision, linearity, and correlation compared to LC-MS/MS These methods can be adopted in the clinical laboratory for rapid therapeutic drug monitoring of patients who require treatment with immunosuppressants.
© 2019 The Authors. Journal of Clinical Laboratory Analysis Published by Wiley Periodicals, Inc.

Entities:  

Keywords:  everolimus; immunoassay; liquid chromatography-tandem mass spectrometry; sirolimus; therapeutic drug monitoring

Mesh:

Substances:

Year:  2019        PMID: 31197901      PMCID: PMC6757180          DOI: 10.1002/jcla.22941

Source DB:  PubMed          Journal:  J Clin Lab Anal        ISSN: 0887-8013            Impact factor:   2.352


INTRODUCTION

Everolimus and sirolimus are widely used as immunodepressants to prevent rejection in patients who have undergone solid organ transplantation.1, 2, 3 These drugs have narrow therapeutic ranges of 3‐8 ng/mL for everolimus and 4‐12 ng/mL for sirolimus.4 Additionally, immunodepressants show pharmacokinetic variability and can cause severe toxicity including infection, bone marrow suppression, and metabolic effects.4, 5 Therefore, both accurate measurement of their concentrations in the blood and prompt therapeutic drug monitoring are crucial for maximizing their therapeutic effects and minimizing toxicity.6, 7, 8, 9, 10 Because liquid chromatography‐tandem mass spectrometry (LC‐MS/MS) is known to have low interference, LC‐MS/MS is currently used as a gold standard method for determining the concentrations of everolimus and sirolimus.11, 12, 13 However, broad adoption of LC‐MS/MS in clinical laboratories has some limitations including difficulties in automating the test process and high cost of the instrument.14, 15 Therefore, LC‐MS/MS must be combined with other methods such as automated immunoassays, which are cost‐effective, simple, and rapid. The Elecsys® Everolimus and Sirolimus assays manufactured by Roche Diagnostics (Basel, Switzerland) were recently developed for measuring the concentration of everolimus and sirolimus based on automated electrochemiluminescence immunoassays. In this study, we evaluated the analytical performance of the Elecsys® Everolimus and Sirolimus assays. Particularly, we determined whether Elecsys® Everolimus and Sirolimus assays can be used as an alternative method to LC‐MS/MS for measuring immunodepressant levels. By comparing the results of the Elecsys® Everolimus and Sirolimus assays to those of LC‐MS/MS, we confirmed the accuracy of the automated electrochemiluminescence immunoassays using samples collected from patients who underwent organ transplantation and were treated with everolimus or sirolimus.

MATERIALS AND METHODS

The ethics committee of Asan Medical Center approved the use of patient data (2018‐0125).

Samples

A total of 120 EDTA whole blood samples were collected from September to November 2017 in Asan Medical Center. For everolimus evaluation, study subjects consisted of 60 patients (51 males and 9 females) with a mean age of 56.1 years. They had received a liver transplantation (n = 51) or heart transplantation (n = 9). For sirolimus evaluation, study subjects consisted of 60 patients (33 males and 27 females) with a mean age of 52.3 years. They had received a kidney transplantation (n = 50), heart transplantation (n = 5), or other organ transplantations (n = 5). After collection, the samples were stored at −20°C until the time of analysis. Drug concentrations were equally distributed based on clinically significant concentrations including the therapeutic range: everolimus: 3‐8 ng/mL; sirolimus: 5‐15 ng/mL.4, 16 Quality control (QC) materials (PreciControl ISD Elecsys, Roche Diagnostics) were used to evaluate precision.

Electrochemiluminescence immunoassays

To measure everolimus and sirolimus, the samples were pretreated with Elecsys® Everolimus or Elecsys® Sirolimus ISD sample pretreatment, respectively. Calibration was performed using the Elecsys® Everolimus CalSet or Elecsys® Sirolimus CalSet, which consists of two calibrators at concentrations of 0.6 and 23.8 ng/mL for everolimus and 0.5 and 23.6 ng/mL for sirolimus. The immunoassays were performed using the Cobas e602 module (Roche Diagnostics) according to the manufacturers' instructions.

LC‐MS/MS assays

LC‐MS/MS assays were performed using a Waters ACQUITY UPLC I‐Class/Waters Xevo TQ‐S (Waters Corporation) equipped with an ACQUITY UPLC® HSS C18 SB Column (2.1 × 30 mm, 1.8 µm pore size, Waters Corporation). As an internal standard solution, whole blood samples were mixed with ascomycin (Cerilliant Corporation), d3‐sirolimus (IsoSciences), and d4‐everolimus (Cerilliant). 6PLUS1® Multilevel Calibrator Set Immunosuppressants (Chromsystems Instruments & Chemicals, Gräfelfing, Germany) were used as calibrators, and MassCheck® Immunosuppressants Whole Blood Control (Chromsystems Instruments & Chemicals) was used as a QC material. The samples were eluted and graduated with a mobile phase composed of ammonium acetate, formic acid, and water or methanol. Everolimus and sirolimus were detected in positive‐ion multiple‐reaction‐monitoring mode, and the following precursor/production pairs were detected as 1,220.8 > 1,203.8 m/z for everolimus and 821.5 > 768.5 m/z for sirolimus. The data were quantified using a six‐point standard curve. The limit of detection values were 0.2 and 0.3 ng/mL for everolimus and sirolimus, respectively.17 The limit of quantitation values were 1.2 and 1.1 ng/mL for everolimus and sirolimus, respectively.17

Precision

For precision evaluation, low, medium, and high concentrations of QC materials were used. Each QC material was measured in duplicate on the same day at more than 2‐hour intervals, and the measurements were performed sequentially over 10 days. The coefficient of variation (CV) for the within‐run assay and total CV were calculated according to the Clinical and Laboratory Standards Institute (CLSI) EP05‐A2.18 The acceptance criteria were based on the recommendations of the International Association of Therapeutic Drug Monitoring and Clinical Toxicology expert consensus group.19

Linearity and recovery

According to CLSI EP06‐A, linearity and recovery were evaluated.20 Patient samples at high and low concentrations were serially diluted to produce five different concentrations. The Elecsys® Everolimus and Sirolimus assays were performed four times on the same day. The means of the concentrations were compared to the expected values.

Comparison

Each measurement was performed within 4 hour in duplicate according to the CLSI EP9‐A3.21 Using Pearson's correlation analysis, the results determined by Elecsys® Everolimus or Sirolimus assays were compared to those determined by LC‐MS/MS The regression equation between the electrochemiluminescence immunoassays and LC‐MS/MS method was obtained by Deming regression and Passing‐Bablok analysis.22 Additionally, the difference in the two assays was identified using a Bland‐Altman plot.

Carryover

After patient samples at high concentrations (H1, H2, H3, and H4) were measured, low‐concentration samples (L1, L2, L3, and L4) were sequentially measured. Carryover (%) was calculated according to following equation: .23

Statistics

Microsoft Excel 2010 (Microsoft), EP Evaluator 8 software package (Data Innovations), and IBM SPSS 23 software (SPSS, Inc) were used for statistical analysis.

RESULTS

The results of precision evaluation are presented in Table 1. The within‐run CV of low, medium, and high levels of QC materials was 4.5%, 2.3%, and 2.6% in the everolimus assay and 4.8%, 4.1%, and 3.3% in the sirolimus assay, respectively. The total run CV of the three QC materials was 6.4%, 4.5%, and 4.9% in the everolimus assay and 8.1%, 5.8%, and 4.7% in the sirolimus assay, respectively. While CVs at low levels were higher than those at other levels, all evaluated levels of QC materials showed acceptable CVs within 8.1%.
Table 1

Precisions of the Elecsys® Everolimus and Elecsys® Sirolimus assays

Elecsys® AssayLevelMean (ng/mL)Within‐run imprecisionTotal imprecision
SD (ng/mL)CV (%)SD (ng/mL)CV (%)
EverolimusLow2.860.1324.50.1856.4
Middle9.840.2352.30.4584.5
High18.520.4862.60.9184.9
SirolimusLow3.810.1874.80.3208.1
Middle10.720.4524.10.6445.8
High19.750.6583.30.9374.7

Abbreviations: CV, coefficient of variation; SD, Standard deviation.

Precisions of the Elecsys® Everolimus and Elecsys® Sirolimus assays Abbreviations: CV, coefficient of variation; SD, Standard deviation. The results of linearity evaluation are presented in Figure 1. Linearity was acceptable, showing correlation coefficients (R 2) of 0.9943 for everolimus at 0.718‐27.585 ng/mL and 0.9958 for sirolimus at 0.789‐26.880 ng/mL. Except at the lowest level for both everolimus and sirolimus, the recovery also satisfied the acceptance criteria, which was within ± 10% of the expected value; we observed values of 93.5%‐105.5% in the everolimus assay and 99.2%‐109.1% in the sirolimus assay. At the lowest levels, the differences between the measured and expected values were 0.172 ng/mL for everolimus (mean of the measured value: 0.890 ng/mL; expected value: 0.718 ng/mL) and 0.129 ng/mL for sirolimus (mean of the measured value: 0.916 ng/mL; expected value: 0.789 ng/mL).
Figure 1

Linearities (left) and %recoveries (right) in Elecsys® Everolimus assays (A) and Elecsys® Sirolimus assays (B). The solid lines are Y = X, and dotted lines represent linear regression

Linearities (left) and %recoveries (right) in Elecsys® Everolimus assays (A) and Elecsys® Sirolimus assays (B). The solid lines are Y = X, and dotted lines represent linear regression

Correlation with LC‐MS/MS

The results of correlation analysis between the automated electrochemiluminescence immunoassays and LC‐MS/MS are shown in Figure 2. The concentrations determined by the Elecsys® Everolimus and Elecsys® Sirolimus assays showed significant correlation with the results determined by LC‐MS/MS (R = 0.9585 and R = 0.9799, respectively). The differences between electrochemiluminescence immunoassays and LC‐MS/MS were evaluated; the results are shown in Table 2. There were no constant differences in the immunoassays compared to LC‐MS/MS (intercept in everolimus assay: 0.423, 95% CI −0.027‐0.873; intercept in sirolimus assay: 0.188, 95% CI −0.162‐0.538 in Deming regression test). However, a low proportional difference was observed in both immunoassays (slope in everolimus assay: 1.216, 95% CI 1.152‐1.280; slope in sirolimus assay: 1.286, 95% CI 1.239‐1.333).
Figure 2

Deming regression (left) and Bland‐Altman plots (right) between electrochemiluminescence immunoassay and liquid chromatography‐tandem mass spectrometry for evaluating everolimus assay (A) and sirolimus assay (B). In the left figures, the solid lines are Y = X and dotted lines represent the Deming regression. In the right figures, the solid lines are Y = X and dotted lines represent the mean value (middle line) and ± 2 SD (upper and lower lines)

Table 2

Summary of comparison between electrochemiluminescence immunoassay and liquid chromatography‐tandem mass spectrometry

 Elecsys® EverolimusElecsys® Sirolimus
Deming
N6060
Slope (95% CI) 1.216 (1.152‐1.280) 1.286 (1.239‐1.333)
Intercept (95% CI) 0.423 (−0.027‐0.873) 0.188 (−0.162‐0.538)
Std Err Est1.2581.008
Passing‐Bablok
N6060
Slope (95% CI) 1.156 (1.093‐1.219) 1.254 (1.208‐1.301)
Intercept (95% CI) 0.788 (0.344‐1.231) 0.392 (0.045‐0.740)
Std Err Est1.2401.000
R a 0.9590.980

Abbreviations: CI, confidence interval; N, Number; Std Err Est, standard error of the estimate.

Correlation coefficient (R) was calculated by Pearson's correlation analysis.

Deming regression (left) and Bland‐Altman plots (right) between electrochemiluminescence immunoassay and liquid chromatography‐tandem mass spectrometry for evaluating everolimus assay (A) and sirolimus assay (B). In the left figures, the solid lines are Y = X and dotted lines represent the Deming regression. In the right figures, the solid lines are Y = X and dotted lines represent the mean value (middle line) and ± 2 SD (upper and lower lines) Summary of comparison between electrochemiluminescence immunoassay and liquid chromatography‐tandem mass spectrometry Abbreviations: CI, confidence interval; N, Number; Std Err Est, standard error of the estimate. Correlation coefficient (R) was calculated by Pearson's correlation analysis. Carryover analysis showed a value of −1.09% in the everolimus assay and −0.12% in the sirolimus assay.

DISCUSSION

In this study, we evaluated the overall analytic performance of the Elecsys® Everolimus and Sirolimus assays. Precision evaluation revealed that the total CVs were acceptable with values of 6.4% and 8.1% in the everolimus and sirolimus assays, respectively. The results showed excellent linearity with an R 2 > 0.99 in the clinically significant range. Carryover was also acceptable within ± 1.1% in both the everolimus and sirolimus assays. Comparison with LC‐MS/MS as a gold standard method showed significant accuracy with R > 0.95 in both the everolimus and sirolimus assays. However, a low level of positive proportional difference was observed in both immunoassays. Similar results were reported in previous studies. According to Shipkova et al, a positive proportional difference in everolimus immunoassay was observed with a slope of 1.131 (95% CI 1.074‐1.186) in Deming regression analysis.1 According to Fung et al, the intercept for the sirolimus assay results was 2.4 ng/mL (95% CI 1.6‐3.3 ng/mL), indicating a positive constant difference.23 Matrix effect, known to be caused by other components present in samples, can greatly affect the accuracy of immunoassays.24 Additionally, bias may be also caused by inaccurate calibration before measurement, non‐specific binding, or cross‐reactivity between immunoassays.25, 26 For example, the major metabolites of everolimus such as 46‐hydroxy, 24‐hydroxy everolimus, and 25‐hydroxy everolimus can cause approximately 2%‐72% cross‐reactivities in the everolimus immunoassay.27, 28 Moreover, sirolimus and its major metabolites can occur cross‐reactivity in the everolimus immunoassay because of their 46% structural similarity.29 We confirmed that the Elecsys® Everolimus and Sirolimus assays show a generally acceptable correlation with the LC‐MS/MS method. However, some significant differences were observed at a high concentration (>2‐fold the therapeutic range), which has not been frequently observed clinically. Therefore, clinicians may use LC‐MS/MS to confirm accurate drug concentrations if patient samples exceed the therapeutic range. And efforts to maintain qualified test results should be made by the manufacturer. The Elecsys® Everolimus and Sirolimus assays are conducted using samples from adult patients who have undergone transplant and have been co‐administered with other immunodepressants such as cyclosporine and corticosteroids. In this study, we could not compare patients with various conditions. The correlations of immunoassay compared to LC‐MS/MS may be different according to the type of organ transplantation because of the different metabolism.1, 26, 30, 31 However, because the clinical uses of the everolimus and the sirolimus are different, we cannot collect the same subdivided samples according to the type of organ transplantation. Clinically, the sirolimus has been more frequently used than everolimus in kidney transplantation due to its longer elimination half‐life, which makes clinicians easier to manage in medical practice.32 On the other hand, the everolimus has been mainly used for liver transplantation because the use of sirolimus in the early post‐transplant period has been influenced by the high incidence of hepatic artery thrombosis and decreased graft survival.33 Fortunately, the previous report showed that unlike other immunoassays, the Elecsys® Everolimus assay showed similar results in kidney, liver, and heart transplantations.1 In conclusion, the Elecsys® Everolimus and Sirolimus assays are useful for determining the concentrations of immunodepressants based on automatized electrochemiluminescence immunoassays. We confirmed the good analytical performance of the assays including precision, linearity, and carryover effects. Particularly, high correlations with the gold standard method (LC‐MS/MS) suggest that these assays can be used as alternatives to LC‐MS/MS in clinical laboratories without mass spectrometry instruments. The Elecsys® Everolimus and Sirolimus assays are useful and practical methods for monitoring the concentrations of everolimus and sirolimus in the blood of patients who have undergone organ transplantation.

CONFLICT OF INTEREST

The authors declare that they have no competing interests.
  25 in total

1.  Liquid chromatography-tandem mass spectrometry or automated immunoassays: what are the future trends in therapeutic drug monitoring?

Authors:  Gunnar Brandhorst; Michael Oellerich; Gregory Maine; Paul Taylor; Gerhard Veen; Pierre Wallemacq
Journal:  Clin Chem       Date:  2012-01-17       Impact factor: 8.327

2.  A direct method for the measurement of everolimus and sirolimus in whole blood by LC-MS/MS using an isotopic everolimus internal standard.

Authors:  Phillip E Morgan; Nigel W Brown; J Michael Tredger
Journal:  Ther Drug Monit       Date:  2014-06       Impact factor: 3.681

Review 3.  Analytical error and interference in immunoassay: minimizing risk.

Authors:  Catharine M Sturgeon; Adie Viljoen
Journal:  Ann Clin Biochem       Date:  2011-07-12       Impact factor: 2.057

4.  A simple and robust LC-MS/MS method for measuring sirolimus and everolimus in whole blood.

Authors:  Chao Yuan; Drew Payto; Jessica Gabler; Sihe Wang
Journal:  Bioanalysis       Date:  2014       Impact factor: 2.681

Review 5.  Review of the proliferation inhibitor everolimus.

Authors:  Björn Nashan
Journal:  Expert Opin Investig Drugs       Date:  2002-12       Impact factor: 6.206

Review 6.  Everolimus: a review of its use in renal and cardiac transplantation.

Authors:  Christopher Dunn; Katherine F Croom
Journal:  Drugs       Date:  2006       Impact factor: 9.546

Review 7.  Sirolimus: the evidence for clinical pharmacokinetic monitoring.

Authors:  Sunita Bond Stenton; Nilufar Partovi; Mary H H Ensom
Journal:  Clin Pharmacokinet       Date:  2005       Impact factor: 6.447

8.  Everolimus therapeutic concentration range defined from a prospective trial with reduced-exposure cyclosporine in de novo kidney transplantation.

Authors:  John M Kovarik; Helio Tedesco; Julio Pascual; Giovanni Civati; Marie-Noelle Bizot; Johanna Geissler; Heinz Schmidli
Journal:  Ther Drug Monit       Date:  2004-10       Impact factor: 3.681

9.  Everolimus pharmacokinetics and its exposure-toxicity relationship in patients with thyroid cancer.

Authors:  D de Wit; T C Schneider; D J A R Moes; C F M Roozen; J den Hartigh; H Gelderblom; H J Guchelaar; J J van der Hoeven; T P Links; E Kapiteijn; N P van Erp
Journal:  Cancer Chemother Pharmacol       Date:  2016-05-11       Impact factor: 3.333

10.  A review on therapeutic drug monitoring of immunosuppressant drugs.

Authors:  Niloufar Mohammadpour; Sepideh Elyasi; Naser Vahdati; Amir Hooshang Mohammadpour; Jamal Shamsara
Journal:  Iran J Basic Med Sci       Date:  2011-11       Impact factor: 2.699

View more
  2 in total

1.  Wide electrochemical window of screen-printed electrode for determination of rapamycin using ionic liquid/graphene composites.

Authors:  Sudkate Chaiyo; Sakda Jampasa; Natnicha Thongchue; Eda Mehmeti; Weena Siangproh; Orawon Chailapakul; Kurt Kalcher
Journal:  Mikrochim Acta       Date:  2020-03-24       Impact factor: 5.833

2.  Comparison of LC-MS/MS and EMIT methods for the precise determination of blood sirolimus in children with vascular anomalies.

Authors:  Yue-Tao Zhao; Hao-Ran Dai; Yue Li; Yuan-Yuan Zhang; Hong-Li Guo; Xuan-Sheng Ding; Ya-Hui Hu; Feng Chen
Journal:  Front Pharmacol       Date:  2022-09-06       Impact factor: 5.988

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.