Neal A Shah1, Eric Catlin2, Navdeep Jassal3, Osama Hafez4, Devang Padalia5.
Abstract
To date, no case studies specifically describing a curved kyphoplasty needle becoming lodged in the vertebral body with the inability to be withdrawn have been reported. We describe a case involving a single level balloon kyphoplasty with a curved coaxial needle during which the cement delivery device could not be removed after cavity filling. In this case, a board-certified interventional pain management specialist was performing balloon kyphoplasty for an L2 osteoporotic vertebral compression fracture. The tools utilized in this procedure included flexible curved instruments designed to traverse the vertebral body and achieve uniform cement distribution through a unipedicular approach. Cannulation and cavity formation were completed without issue. Upon conclusion of cement filling, the curved cement delivery device was unable to be removed. After several attempts to remove the needle and consultation with both the device company and local spine surgeons, it was agreed that the device should be cut at the level of entry into the pedicle and left as a retained foreign object. The involved area was surgically exposed, the retained instrument was cut flush to the pedicle, and the free portion was removed without further complication. The patient followed up in clinic several months later without evidence of neurologic complications. Considerations when using a curved kyphoplasty device and a method of resolution when faced with the inability to remove such an instrument are discussed.Entities:
Keywords: complication; foreign body; kyphoplasty; lumbar pain; vertebral augmentation; vertebral compression fracture
Year: 2019 PMID: 31192072 PMCID: PMC6551195 DOI: 10.7759/cureus.4367
Source DB: PubMed Journal: Cureus ISSN: 2168-8184
Figure 1Sagittal short-TI inversion recovery (A) and standard T2-weighted (B) MRI scans show an acute L2 compression fracture with inferior endplate involvement
Figure 2Cutting the handle of the stuck curved coaxial needle to allow the removal of the surrounding access cannula
Figure 3Cutting the stuck needle at the level of entry into the pedicle after the access cannula has been removed
Figure 4Anteroposterior (A) and lateral (B) views of retained device