Retained needle after cement injection during vertebral augmentation and its management strategy.

Angelika Kosse1, Jonathan Pishoi Nakhla2, Reza Yassari3, Apolonia Elisabeth Abramowicz4, Allan Brook5.   

Abstract

A middle aged patient with multiple myeloma resulting in numerous pathological fractures underwent an L2, L3, and L5 vertebral cement augmentation for pain relief. After injection, the trocar at L2, the final level, could not be withdrawn despite several attempts of needle rotation, a second needle inserted to distract on, and rocking the needle on the pedicle. After a neurosurgical consultation, the patient was transferred to the operating room for open removal. As the needle protruded approximately 3 inches from the patient's back, the patient could not be positioned supine, and was anesthetized and intubated in the right semi-prone position prior to being placed prone on the operating room table. The needle was surgically exposed, cut off at the pedicular bone edge, and its free component was removed. 2016 BMJ Publishing Group Ltd.

Entities:  

Keywords:  Complication; Degenerative; Intervention; Lumbosacral; Technique

Mesh:

Substances:

Year:  2016        PMID: 26891698      PMCID: PMC5483563          DOI: 10.1136/bcr-2015-012180

Source DB:  PubMed          Journal:  BMJ Case Rep        ISSN: 1757-790X


  3 in total

1.  Percutaneous vertebroplasty: complication avoidance and technique optimization.

Authors:  John M Mathis
Journal:  AJNR Am J Neuroradiol       Date:  2003-09       Impact factor: 3.825

Review 2.  Vertebroplasty.

Authors:  Bryan Jay; Sun Ho Ahn
Journal:  Semin Intervent Radiol       Date:  2013-09       Impact factor: 1.513

Review 3.  Vertebral augmentation in patients with multiple myeloma: a pooled analysis of published case series.

Authors:  O A Khan; W Brinjikji; D F Kallmes
Journal:  AJNR Am J Neuroradiol       Date:  2013-07-18       Impact factor: 3.825

  3 in total
  2 in total

1.  Retained Curved Needle After Balloon Kyphoplasty: A Complication with a Novel Device and Its Management.

Authors:  Neal A Shah; Eric Catlin; Navdeep Jassal; Osama Hafez; Devang Padalia
Journal:  Cureus       Date:  2019-04-02

2.  A Smart Technique to Remove Ruptured Inflatable Bone Tamp From the Vertebral Body in Balloon Kyphoplasty.

Authors:  Ioannis Papaioannou; Vasileios K Mousafeiris; Georgia Pantazidou; Thomas Repantis; Panagiotis Korovessis
Journal:  Cureus       Date:  2021-05-18
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.