Chloe L Chung1, Derek M Kelly2, Jack R Steele3, Denis J DiAngelo1. 1. Department of Orthopaedic Surgery and Biomedical Engineering, The University of Tennessee Health Science Center, Memphis, USA. 2. Campbell Clinic Orthopaedics and Le Bonheur Children's Hospital, Memphis, USA. 3. The Center for Orthotics and Prosthetics, Inc., Memphis, USA.
Abstract
INTRODUCTION: Thoracolumbar braces are used to treat Adolescent Idiopathic Scoliosis. The objective of this study was to design and validate a mechanical analog model of the spine to simulate a thoracolumbar, single-curve, scoliotic deformity in order to quantify brace structural properties and corrective force response on the spine. METHODS: The Scoliosis Analog Model used a linkage-based system to replicate 3D kinematics of spinal correction observed in the clinic. The Scoliosis Analog Model is used with a robotic testing platform and programmed to simulate Cobb angle and axial rotation correction while equipped with a brace. The 3D force and moment responses generated by the brace in reaction to the simulated deformity were measured by six-axis load cells. RESULTS: Validation of the model's force transmission showed less than 6% loss in the force analysis due to assembly friction. During simulation of 10° Cobb angle and 5° axial rotation correction, the brace applied 101 N upwards and 67 N inwards to the apical connector of the model. Brace stiffness properties were 0.5-0.6 N/° (anteroposterior), 0.5-2.3 N/° (mediolateral), 23.3-26.5 N/° (superoinferior), and 0.6 Nm/° (axial rotational). CONCLUSIONS: The Scoliosis Analog Model was developed to provide first time measures of the multidirectional forces applied to the spine by a thoracolumbar brace. This test assembly could be used as a future design and testing tool for scoliosis brace technology.
INTRODUCTION: Thoracolumbar braces are used to treat Adolescent Idiopathic Scoliosis. The objective of this study was to design and validate a mechanical analog model of the spine to simulate a thoracolumbar, single-curve, scoliotic deformity in order to quantify brace structural properties and corrective force response on the spine. METHODS: The Scoliosis Analog Model used a linkage-based system to replicate 3D kinematics of spinal correction observed in the clinic. The Scoliosis Analog Model is used with a robotic testing platform and programmed to simulate Cobb angle and axial rotation correction while equipped with a brace. The 3D force and moment responses generated by the brace in reaction to the simulated deformity were measured by six-axis load cells. RESULTS: Validation of the model's force transmission showed less than 6% loss in the force analysis due to assembly friction. During simulation of 10° Cobb angle and 5° axial rotation correction, the brace applied 101 N upwards and 67 N inwards to the apical connector of the model. Brace stiffness properties were 0.5-0.6 N/° (anteroposterior), 0.5-2.3 N/° (mediolateral), 23.3-26.5 N/° (superoinferior), and 0.6 Nm/° (axial rotational). CONCLUSIONS: The Scoliosis Analog Model was developed to provide first time measures of the multidirectional forces applied to the spine by a thoracolumbar brace. This test assembly could be used as a future design and testing tool for scoliosis brace technology.
Entities:
Keywords:
Scoliosis; analog model; brace; loading/response model; spinal orthosis
Scoliosis is a three-dimensional (3D) skeletal deformity consisting of a combination
of spinal axial rotation (AR) and lateral curvature, or Cobb angle (CA), of 10° or
greater.[1] Thoracolumbar braces are commonly used to treat Adolescent
Idiopathic Scoliosis (AIS) curves between 20° and 50° with the curve apex below T6
in patients who have significant growth remaining (Risser grade ≤ 2).[2,3] Braces serve to maintain, and,
in some cases, reduce the spinal curve to prevent progression of the deformity by
applying corrective forces.[2,4]
Until recently, the outcome of bracing treatment has been highly debated. Weinstein
et al. conducted a 5-year follow-up study (Bracing in Adolescent Idiopathic
Scoliosis Trial, BrAIST), providing evidence that braces are effective (72% success
rate) at treating scoliosis.[5]Scoliosis braces can be rigid, flexible, or composite and can utilize passive or
active corrective mechanisms. Many braces use a three-point pressure principle as
the method of correction, which involves fixation superior, inferior, and at the
apex level of the curve.[6] The apical pad is located slightly anterior and two to three
rib levels below the apex of the curve in order to engage the ribs and help control
AR correction of the spine. The magnitude and direction of corrective forces applied
by a brace to the spine remain unknown and are a common concern for clinicians and
orthotists who are tasked with making design alterations, such as the addition of
Velcro straps, extra pads, and section cut-outs, to improve fit and comfort for the
patient. There is no standard of application or common understanding of how these
alterations affect the structural properties or the corrective capacity of a brace.
The process is largely subjective and follows best judgment and practice, yet these
steps are critical and influence how well a brace will work and patient compliance.
In-brace imaging is commonly used to measure the degree of spinal correction and
determine brace efficacy, however, no force analysis is used during the design
phase. BrAIST co-author, Lori Dolan identified a need to determine which forces are
applied to the spine and to understand the effects brace alterations have on curve
correction and spinal loading.[7] A scientific method is needed to
determine the multidirectional force mechanism of scoliosis braces.Current experimental research of scoliosis bracing mechanics has largely been limited
to two-dimensional (2D) and 3D imaging techniques[8-15] and computational
models.[16-36] In vivo imaging methods
provide a measure of the angular deformity but no information of the brace mechanics
or structural properties. Computational models include empirical models, dynamic
musculoskeletal simulations, and finite element models (FEM). The recent literature
shows utilization of advanced 3D FEM techniques in combination with in vivo
techniques such as hi-resolution and low-dose biplanar radiographs, pressure mapping
systems, and compliance tracking devices. However, the majority of published FEM are
limited to a few patient-specific applications and ultimately lack validation from
empirical data.A great amount of effort has been put into the development and validation of one FEM
for simulation of brace action. Vergari et al.[35,36] created personalized
parametric finite element beam models of 42 patients diagnosed with AIS. These
patient-specific FEMs were generated using 3D reconstructions from bi-planar EOS
images (EOS Imaging, Paris, France) of the patient out-of-brace and in-brace in
order to simulate clinical indices of 3D spatial deformation related to the
deformity. Orthosis action was simulated in the model by applying local
displacements at each pad position corresponding to the difference in pad position
before treatment and in-brace. The relevant clinical indices of kyphosis, lordosis,
CA, vertebral AR, torsion index, and 3D rib hump were calculated in the deformed FEM
after simulation. These values were compared with the measured in-brace values for
each parameter to determine the error in the model. The FEM showed promising results
that support its capacity to simulate brace action with only 87% of the clinical
indices presenting error lower than the corresponding measurement uncertainty. This
work was focused on simulating the behavior of the trunk’s geometric deformation in
response to known brace pad displacements. The authors noted that the FEM was
limited by imposing known displacements to simulate the pad pressure, and they
expressed the need for explicit brace modeling and analysis of contact forces in
order to improve the model.A number of researchers have been using skin-to-brace interface force distribution
and pressures and strap fastener transducers as a method for understanding bracing
mechanics.[11-14] Mac-Thiong et al.[11] performed an
external biomechanical evaluation of the Boston Brace System for the treatment of
AIS to relate the brace interface forces at different regions of the trunk with the
internal brace strap tension. A custom pressure mapping system made of force-sensing
transducers was created for the purpose of this experiment. To measure the internal
strap tension, an experimental, in-line load cell was added to the brace design.
These techniques showed the relationship between strap tension magnitude and the
compressive forces applied by the brace to the critical trunk regions. However, this
study was not able to provide any estimation of the biomechanical loads on the spine
due to the brace.Ultimately, none of the existing research on this topic seems to answer the clinical
question of “how much force is applied to the spine by a brace?” Current FEM cannot
answer that question without generalized assumptions regarding the anatomical
stiffness properties, force transmission through the anatomy, and structural
properties of the custom brace. Pressure pads and strap fastener transducers can
give an estimate of the direct compressive loads applied to the torso, and they are
limited in that they do not have the sensitivity to capture multidirectional forces
applied by the brace. To our knowledge, no mechanical model of a scoliotic spine or
testing assembly is available to study bracing mechanics in scoliosis. The objective
of this study was to design and validate a mechanical analog model of the spine to
simulate a thoracolumbar, single-curve, scoliotic deformity in order to quantify
brace structural properties and force response on the spine.
Methods
An analog model of an adolescent scoliotic spine and a robotic testing platform (RTP)
were used to measure the force response of a scoliosis brace over a range of spinal
correction. A single-curve, thoracolumbar Boston brace (Avon, MA) was analyzed.
Model design
A mechanically equivalent analog model of the AIS condition was developed, and
the force output of the model was validated.[37] The Scoliosis Analog Model
(SAM) used a linkage-based system to simulate 3D spinal correction of CA and AR
deformity of a single-curve scoliotic spine. The SAM was designed to engage with
the interior surface of a scoliosis brace at each of the three spinal levels,
superior, inferior, and at the apex.Characteristics and dimensions from patient records and bi-planar EOS images (EOS
Imaging, Paris, France)[38,39] (IRB 14-03110-XP reviewed and approved by the University of
Tennessee Health Science Center IRB Committee) were used to customize the SAM
for use with a unique Boston brace, including height (1.7 m), “out of brace”
spinal measures CA (28°) and AR (5°), and “in brace” spinal measures of CA (18°)
and AR (0°) (Figure
1(a)). The spatial locations of the critical anatomy, including the
apical vertebral body and the superior and inferior junctional vertebral bodies,
corresponded to the connection points of the main linkages of the SAM (Figure 1(b)). Each of the
three vertebral bodies was represented in the SAM by a linkage assembly. The
linkage assembly consisted of a combination of steel linkages and connectors,
and plastic arms and shells (Figure 1(c)). Each linkage component had a clevis joint and ball or
pin joint at its ends held with a pin connector. The distance between the
critical vertebral bodies corresponded to the linkage lengths (D1: 8.1 cm, D2:
6.6 cm). The distance between the outer surface profile of the torso to the
center of the vertebral body (D3: 12.8 cm, D4: 10.5 cm, D5: 13.6 cm)
corresponded to the length of the arm component, which served to attach the
shell to the linkages using the pin connectors. Each shell had a specific
geometry that matched and interfaced with the internal contoured surface of the
superior, apical, and inferior critical regions of the brace.
Figure
1.
Steps in designing the scoliosis analog model.
(a) Coronal plane data, (b) critical anatomy corresponding to SAM
components, and (c) SAM components.
SAM: Scoliosis Analog Model.
Note: Example EOS scan and parameters are shown, not the actual
patient scan and data used for this study.
Steps in designing the scoliosis analog model.
(a) Coronal plane data, (b) critical anatomy corresponding to SAM
components, and (c) SAM components.SAM: Scoliosis Analog Model.Note: Example EOS scan and parameters are shown, not the actual
patient scan and data used for this study.The model utilized the three-point pressure principle to engage with the inside
of the brace at the apical shell and the endpoints of the curve across the
entire critical region, as defined by the orthotist.[6] These regions could be
directly lateral to the junctional bodies (within the coronal plane), offset
superior–inferior, or offset anterior–posterior (out of the coronal plane). The
arm-shell components were able to pivot about the pin connector to engage with
offset critical regions. The CA corresponded to the angular displacement of the
linkages relative to the vertical (caudal-cranial) axis. The AR of the apical
vertebral body was used to define the apical connector’s degree of offset from
the coronal plane. By using these critical anatomical parameters, the
experimental SAM was designed to closely replicate a clinical scoliosis
deformity.
Testing assembly
The upper and lower sections of the SAM were attached to the moving actuation
system and base of a RTP[40] (Figure 2(a)). The four main components of the RTP used for this
study were the Exlar linear actuator (Model GSX-30; Curtis-Wright, Chanhassen,
MN), rotary servo actuator (Model FHA-25 C; Harmonic Drive, Peabody, MA), upper
load cell (ULC) (Model 100M40 six-axis load cell; JR3 Inc., Woodland, CA), and
lower load cell (LLC) (Model 67M25S3; JR3 Inc., Woodland, CA). The upper
mounting plate of the SAM was centered and fixed to the ULC that was attached to
the vertical actuator and rotary actuator. The lower mounting plate of the SAM
was centered and fixed to the LLC that was attached to the RTP lower platform.
Since scoliosis is a 3D deformity of the spine, the SAM was aligned to the
“deformed” reference plane (offset from coronal plane to the same degree as AR
of the apical vertebral body). This reference plane was where the simulations
and analyses were performed (Figure 2(b,c)). The reference plane’s vertical axis corresponded to
the brace’s axial direction and RTP Z component, and the reference plane’s
horizontal axis corresponded to the brace’s radial direction and RTP Y
component.
Figure
2.
Test assembly. (a) The SAM mounted in the
programmable robotic testing platform, (b) front view of the SAM
showing anatomical coronal plane and reference plane, and (c)
transverse view of the SAM showing anatomical coronal plane and
reference plane.
SAM: Scoliosis Analog Model.
Test assembly. (a) The SAM mounted in the
programmable robotic testing platform, (b) front view of the SAM
showing anatomical coronal plane and reference plane, and (c)
transverse view of the SAM showing anatomical coronal plane and
reference plane.SAM: Scoliosis Analog Model.
Validation of SAM force output
SAM was tested with a RTP without any input load to ensure the links moved
freely. Less than 3 N of vertical force occurred over 40° of linkage rotation.
Validation of the transmission of forces acting on the SAM was performed by
applying a series of input forces with varying orientation (Figure 3) to the apical connector of the
SAM during simulated CA correction. Force components transmitted to the superior
connector of the linkage assembly were recorded and compared to results from a
computational model. Based on the rigid body design of the SAM, the force
response was further verified by theoretical calculations using the
trigonometric relation of the known orientations of the linkages and the applied
force. The percent error relative to the computational results was calculated
for each input force orientation.
Figure 3.
Setup for validation of
SAM force output.
SAM: Scoliosis Analog Model.
Setup for validation of
SAM force output.SAM: Scoliosis Analog Model.A 4.54 kg (10 lb) weight acted as the controlled force input for the experimental
tests. The weight was suspended from a cable that passed over a pulley and
connected to the apex of the SAM. A Brecknell digital handheld tension scale
(Brecknell, Fairmont, MN) was suspended in line between the attachment at the
SAM apex and the pulley and revealed a 10% reduction of the suspended weight (or
4.5 N). For this reason, all tests and theoretical calculations were performed
using 40 N as the input force, Fi. Relative to the apex of the SAM, the height
of the pulley assembly was adjustable, which allowed the input force to be
applied at a set orientation.The testing platform was programmed to displace along the vertical axis a given
distance and speed (4.8 mm/s). Downward movement caused the SAM to simulate a
40° CA change in the spinal curve. Five orientations of the input force relative
to a horizontal line along the Y-axis were tested, including −20°, −10°, 0°
(horizontal), +10°, +20°, and +30°. In these tests, negative angles are below
the horizontal reference and positive angles are above the horizontal reference
(Figure 3). Each
input load orientation was tested three times. Before each run, the suspended
weight was reset and stabilized. The force components, Fz and Fy, and moment,
Mx, applied by the weight were measured at the load cell at a sampling rate of
25 Hz. These measured components were transformed to the pinned joint of the
superior connector of the linkage assembly (Fz′ and Fy′, Mx′=0) for analysis.
Also, the linkage assembly angular displacements (simulated CA) were measured at
a sampling rate of 25 Hz. The tests were also performed using a 2D computer
aided design (CAD) SAM. An analysis was performed using Working Model 2D (Design
Simulation Technologies, Inc., Canton, MI).The force components measured during the experimental tests were compared to the
force components measured during the computer simulations. The axial force
components for the experimental and computational tests were plotted against the
simulated CA for each input force angle. Using Microsoft Excel (2007), a fourth
order polynomial equation was generated for each curve. The percent error
relative to the computational results was calculated for each input force
orientation. The mean percent error revealed that there was a 6% force reduction
within the SAM system over the testing range. This error was attributed to the
frictional losses between the components of the linkage assembly.
Test protocol
To measure the force response of the brace, coupled CA and AR motion was
simulated. The SAM assembly was mounted on the RTP and the weight of the SAM was
zeroed from the load cells. The SAM was first tested without any input load to
ensure the links moved freely and registered less than 3 N of force at the ULC
and LLC. As instructed by an orthotist, the brace was mounted onto the SAM
(Figure 4(a,b)).
Using tensiometer devices attached to the Velcro straps, each strap was
tightened between 22 N and 45 N (5–10 lb). The tensiometer devices (222 N
(50 lb) max capacity, 0.1 N resolution; Berkley Digital Scale) were powered by a
D.C. power supply (3 V; BK Precision) and controlled using a National
Instruments LabVIEW 2010 VI program. The testing platform was programmed to
either displace downward along the Z (vertical) axis or to rotate about the Z
axis. Programmed linear movement caused the SAM to transition from the
“corrected” alignment (18° CA) to the “deformed” alignment (28° CA), simulating
10° CA change in the spinal curve. Programmed rotation caused the SAM to axially
rotate within the brace, which was inferiorly fixed to the platform using a
brace fixture, effectively simulating AR correction (0° to 5°). Coupled CA and
AR correction was achieved by executing both motions simultaneously (Figure 5(a,b)). Strap tension
remained within ±4.5 N (1 lb) of the initially set value. The setup was cycled
five times before collecting data to ensure the brace settled onto the model.
Figure
4.
Testing assembly. (a) SAM with single-curve
thoracolumbar Boston brace and (b) tensiometer attached to
strap.
SAM: Scoliosis Analog Model.
Figure
5.
Methodology for simulating a changing spinal
curve using linkage components. (a) Corrected alignment and (b)
deformed alignment.
Testing assembly. (a) SAM with single-curve
thoracolumbar Boston brace and (b) tensiometer attached to
strap.SAM: Scoliosis Analog Model.Methodology for simulating a changing spinal
curve using linkage components. (a) Corrected alignment and (b)
deformed alignment.
Data management
The 3D force and moment components, which were generated by the brace as a
reaction to the simulated deformity, and the linkage assembly angular
displacements were measured at a sampling rate of 25 Hz. These 3D forces read at
both load cells and CA change were plotted in a force–displacement curve (FDC).
The moment about the vertical axis (Mz) read at the ULC and AR change were
plotted in a moment–displacement curve (MDC). The FDC and MDC provide a visual
representation of the loading behavior of the brace during the simulation. The
slope of the curves represented the structural stiffness properties of the brace
with respect to the specific loading axis. Stiffness was expressed as a
resistive force relative to the angular change of the linkage system. Linear
stiffness properties (X, Y, Z) were derived from the 3D force response relative
to CA change. This definition was preferred over the traditional linear
stiffness measure so that the resultant stiffness values were more easily
relatable to the clinical measurements (i.e. it provided a way to normalize the
measure relative to the CA and AR deformity). Axial rotational stiffness was
derived from the Z-moment relative to AR change. Apical shell loads in the 2D
reference plane were calculated by transforming the load cell readings to the
spatial location of the apical connector. Using the method of joints, the forces
were balanced to determine the magnitude and direction of the load applied by
the brace to the model apical connector. This calculation was done assuming that
the connectors were frictionless pins and that all linkages were rigid,
two-force members.
Results
The brace force response was measured by the ULC and LLC during the simulation of the
scoliosis deformity. The 3D force and moment responses are shown in Figure 6(a) (ULC Forces),
Figure 6(b) (LLC
Forces), and Figure 6(c)
(ULC Moment). Calculated brace structural stiffness properties are shown within each
figure. With X- and Y-stiffness values between 0.5–0.6N/° and 0.5–2.3N/°,
respectively, it can be seen that this brace offers little structural stiffness
along those axes. However, with a Z-stiffness between 23.3 and 26.5N/° and
AR-stiffness of 0.6 Nm/°, the brace has much greater stiffness along the Z-axis and
limited rotational control. The customized SAM replicated the 10° CA and 5° AR
correction observed in the patient's EOS X-ray images. Using the data from each load
cell, the apical shell loads were calculated for the deformed alignment of 28° CA
and 5° AR. The brace applied 101 N upwards and 67 N inwards to the apical connector
of the model (Figure 7).
Collectively, these results serve to quantify the mechanics and structural
properties of the scoliosis brace tested.
Figure 6.
Plotted brace force and
moment response with derived stiffness properties. (a)
Force–displacement curves of the upper load cell, (b) force–displacement
curves of the lower load cell, and (c) moment–displacement curves of the
upper load cell.
Figure
7.
Free-body diagram of the SAM at 28° CA and 5°
AR.
SAM: Scoliosis Analog Model; CA: Cobb angle; AR: axial
rotation.
Plotted brace force and
moment response with derived stiffness properties. (a)
Force–displacement curves of the upper load cell, (b) force–displacement
curves of the lower load cell, and (c) moment–displacement curves of the
upper load cell.Free-body diagram of the SAM at 28° CA and 5°
AR.SAM: Scoliosis Analog Model; CA: Cobb angle; AR: axial
rotation.
Discussion
Braces serve to reduce and prevent progression of the spinal curve by applying
multidirectional corrective forces. Orthotists generate an expected brace force
response through alterations in the brace design, such as apical pad placement and
orientation. Specifically, while considering the location of the apical pad of the
brace, the orthotist conceptualizes the force vector necessary to correct the spinal
deformity. Until now, there has been no experimental model capable of quantifying
the magnitude and direction of the corrective forces applied by a physical brace to
a torso.Structural properties provide a means to compare and classify bracing technology.
Interpretation of the measured 3D force and moment data provides a basic
understanding of the mechanics and structural properties of the particular brace
studied. During the fabrication of a brace, an orthotist may desire to change brace
structural stiffness to achieve better patient fit, more comfort, or better brace
performance. For example, a patient with poor muscle tone and excessive anatomical
spinal alignment may require a stiffer brace to achieve and maintain the desired
level of correction, and the design changes, such as addition of extra Velcro
straps, could provide the increase in brace stiffness required (i.e. increase brace
forces applied to the spine). Currently, these design changes are made purely based
on the professional judgment of the orthotist. The SAM could also be used as a tool
to quantify how these design changes affect the structural properties of the brace
and the achieved brace loads applied to the spine. These features include brace
pads, pad placement, pad geometry, material selection, material cut-outs, rigid
components, and other fitting elements.Since this research model and testing methodology are new, directly comparable data
are not available. Van den Hout et al. measured the magnitude of direct compressive
forces at the body brace interface.[13] The forces exerted by the
internal lumbar and thoracic pads of a Boston brace were measured in 16 patients
with AIS using an electronic pressure measuring system. The mean forces measured at
the lumbar brace pad was 214 N (range, 0–727 N) and at the thoracic brace pad was 66
N (range, 4–209 N). These force ranges are comparable to the magnitude of the forces
measured by the SAM.
Limitations of study
General limitations that may affect the validity of these results include the
simplification of the model compared with in vivo biomechanics, the 2D
analytical methods, the lack of directly comparable brace-applied force measures
for validation, and design approximations that were made during model
development. The model is able to capture the force components, but does not
consider the effective area of the applied forces beyond the defined critical
regions of the torso. This could have been addressed by measuring the
brace-model interface pressure and force distribution. The testing did not
account for loss of brace stiffness over longer periods of wear. The elastic
nature of the brace material allowed for the brace to return to its corrected
alignment after each test. The Boston brace used in this study was an older
brace design that provided limited AR correction. Newer brace designs that offer
more control of AR correction should be studied in future work.An important design feature of the SAM is that the “spine” is assumed to have
negligible or zero stiffness. In other words, the SAM linkages are rigid and do
not reflect the compliant nature of the anatomy. Also, no effect of the ribs was
simulated because they were assumed to be rigid. Thus, all reactant forces
measured can be attributed to the brace alone. Despite these limitations, this
research has resulted in a novel model, methodology, and baseline data for
future research on scoliosis bracing technologies.
Conclusions
The SAM and testing methodology enables study of scoliosis bracing technology through
the measure of brace structural properties and force response on the spine. Measured
force components provided the first evidence of multidirectional force transmission
by confirming an upward lifting effect. Future SAM design improvements will permit
the 3D study of complex curve types. Future testing may include variable pad and
strapping parameters such as orientation, spatial location, size, material, and
shape to determine the impact that they have on force transmission. By understanding
the effects of removing, adding, or altering the fitting pads and straps,
recommended brace design changes may result in improved in-brace correction and
treatment outcome.