| Literature DB >> 31191932 |
Lisa Tedesco Triccas1,2, Ann-Marie Hughes1, Jane H Burridge1, Amy E Din1, Martin Warner1, Simon Brown1, Mahalekeshmi Desikan3, John Rothwell3, Geert Verheyden2.
Abstract
PURPOSE: Reliability of motor-evoked potential threshold and amplitude measurement of upper limb muscles is important when detecting changes in cortical excitability. The objective of this study was to investigate intra-rater, test-retest reliability and minimal detectable change of resting motor threshold and amplitude of a proximal and distal upper limb muscles, anterior deltoid and distal extensor digitorum communis in healthy adults.Entities:
Keywords: Transcranial magnetic stimulation; distal upper limb muscles; minimal detectable change; neurorehabilitation; outcome measurement; proximal upper limb muscles; reliability; upper limb
Year: 2018 PMID: 31191932 PMCID: PMC6453085 DOI: 10.1177/2055668318765406
Source DB: PubMed Journal: J Rehabil Assist Technol Eng ISSN: 2055-6683
Figure 1.Experimental set-up of the TMS, Brainsight® and EMG equipment – the figure of eight magnetic coil was placed on the participant’s head to measure MEP responses of the AD and extensor digitorum muscles.
Mean (SD), ICC with 95% confidence intervals, SEM and MDC of resting motor threshold and motor-evoked potential area of anterior deltoid and extensor digitorum communis input/output curves between tests 1–2 and test 1–3.
| Means (SD) | ICC | 95% CI | CV (%) | SEM | MDC %/mV | |
|---|---|---|---|---|---|---|
| RMT (%MSO) | ||||||
| Tests 1 and 2 (EDC) | 58.38 (9.87) 59.29 (11.82) | 0.91 | 0.75–0.95 | 18 | 2.96 | 8.20 |
| Tests 1 and 3 (EDC) | 58.38 (9.87) 60.20 (9.81) | 0.84 | 0.64–0.93 | 33 | 3.95 | 10.95 |
| Tests 1 and 2 (AD) | 77.17 (10.59) 73.38 (11.34) | 0.94 | 0.80–0.98 | 15 | 2.59 | 7.18 |
| Tests 1 and 3 (AD) | 77.17 (10.59) 76.45 (9.13) 76.81 | 0.69 | 0.15–0.91 | 26 | 5.90 | 16.35 |
| EDC MEP area (mV) of I/O curves from 100 to 130% | ||||||
| Tests 1 and 2 | 1.34 (0.80) 1.39 (1.04) | 0.50 | 0.03–0.79 | 68 | 0.57 | 1.58 |
| Tests 1 and 3 | 1.34 (0.80) 1.37 (0.66) | 0.65 | 0.22–0.87 | 54 | 0.47 | 1.30 |
| EDC MEP area of I/O curves from 100 to 150% | ||||||
| Tests 1 and 2 | 2.44 (1.44) 2.44 (1.82) | 0.59 | 0.06–0.86 | 67 | 0.92 | 2.55 |
| Tests 1 and 3 | 2.44 (1.44) 2.56 (1.44) | 0.69 | 0.24–0.89 | 58 | 0.80 | 2.22 |
AD: anterior deltoid; CI: confidence interval; CV: coefficient of variation; EDC: extensor digitorum communis; ICC: intraclass correlation coefficient; I/O: input/output; MDC: minimal detectable change; MEP: motor-evoked potential; mV: millivolt; MSO: Maximum Stimulator Output; RMT: resting motor threshold at maximum stimulator output; SEM: standard error of mean.