Literature DB >> 31139273

Effect of Non-fluoroscopic Catheter Tracking on Radiation Exposure during Pulmonary Vein Isolation: Comparison of Four ablation systems.

Yoshihisa Naruse1, Fehmi Keçe1, Marta de Riva1, Masaya Watanabe1, Adrianus P Wijnmaalen1, Reza Alizadeh Dehnavi1, Martin J Schalij1, Katja Zeppenfeld1, Serge A Trines1.   

Abstract

BACKGROUND: A novel non-fluoroscopic catheter tracking system (Mediguide) can be used in combination with a 3D mapping system for atrial fibrillation (AF) ablation. However, the benefit on radiation exposure of the Mediguide system compared to other ablation systems is unknown.
METHODS: We retrospectively enrolled consecutive 73 patients (51 men; 59±11 years; 60 paroxysmal AF) undergoing pulmonary vein isolation by the same operator. Radiation time, radiation effective dose, procedure time, AF recurrence after ablation, and procedure-related complications were compared among 4 different ablation systems.
RESULTS: Mediguide was used in 16 patients (group A), CARTO™ in 17 (group B), Cryoballoon in 30 (group C), and Multi-electrode Pulmonary Vein Ablation Catheter (PVAC) in 10 (group D). Although procedure time was shorter in patients with Cryoballoon (median 110 [interquartile range 99-120] min) and PVAC (123 [112-146] min) compared to those with Mediguide (181 [168-214] min) and CARTO (179 [160-195] min) (P<0.001), radiation exposure time and effective dose were decreased in patients with Mediguide compared to the other ablation systems (A: 5 [3-6] min; B: 14 [11-16] min; C: 14 [11-18] min; D: 20 [16-24] min, P<0.001 and A: 1.1 [0.8-2.0] mSv; B: 2.5 [1.3-3.8] mSv; C: 2.0 [1.4-2.5] mSv; D: 1.7 [1.4-3.6] mSv, P=0.015, respectively). AF recurrence rates and procedure-related complications were comparable among the 4 groups.
CONCLUSION: The Mediguide system reduces radiation exposure compared to other ablation systems without increasing AF recurrence or procedure-related complications.

Entities:  

Keywords:  Ablation; Atrial fibrillation; Non-fluoroscopic Catheter Tracking System; Radiation exposure

Year:  2018        PMID: 31139273      PMCID: PMC6533812          DOI: 10.4022/jafib.2068

Source DB:  PubMed          Journal:  J Atr Fibrillation        ISSN: 1941-6911


  21 in total

1.  2012 HRS/EHRA/ECAS Expert Consensus Statement on Catheter and Surgical Ablation of Atrial Fibrillation: recommendations for patient selection, procedural techniques, patient management and follow-up, definitions, endpoints, and research trial design.

Authors:  Hugh Calkins; Karl Heinz Kuck; Riccardo Cappato; Josep Brugada; A John Camm; Shih-Ann Chen; Harry J G Crijns; Ralph J Damiano; D Wyn Davies; John DiMarco; James Edgerton; Kenneth Ellenbogen; Michael D Ezekowitz; David E Haines; Michel Haissaguerre; Gerhard Hindricks; Yoshito Iesaka; Warren Jackman; Jose Jalife; Pierre Jais; Jonathan Kalman; David Keane; Young-Hoon Kim; Paulus Kirchhof; George Klein; Hans Kottkamp; Koichiro Kumagai; Bruce D Lindsay; Moussa Mansour; Francis E Marchlinski; Patrick M McCarthy; J Lluis Mont; Fred Morady; Koonlawee Nademanee; Hiroshi Nakagawa; Andrea Natale; Stanley Nattel; Douglas L Packer; Carlo Pappone; Eric Prystowsky; Antonio Raviele; Vivek Reddy; Jeremy N Ruskin; Richard J Shemin; Hsuan-Ming Tsao; David Wilber
Journal:  Europace       Date:  2012-03-01       Impact factor: 5.214

2.  Real-time integration of intracardiac echocardiography to facilitate atrial tachycardia ablation in a patient with a Senning baffle.

Authors:  Dennis W den Uijl; Nico A Blom; Adrianus P Wijnmaalen; Jeroen J Bax; Martin J Schalij; Katja Zeppenfeld
Journal:  Circ Arrhythm Electrophysiol       Date:  2009-10

3.  Nonfluoroscopic sensor-guided navigation of intracardiac electrophysiology catheters within prerecorded cine loops.

Authors:  Christopher Piorkowski; Gerhard Hindricks
Journal:  Circ Arrhythm Electrophysiol       Date:  2011-08

4.  Catheter ablation for paroxysmal atrial fibrillation: a randomized comparison between multielectrode catheter and point-by-point ablation.

Authors:  Alan Bulava; Jiří Haniš; David Sitek; Ondřej Ošmera; Dan Karpianus; Michal Snorek; Kateřina Rehoušková; František Toušek; Ladislav Pešl
Journal:  Pacing Clin Electrophysiol       Date:  2010-09       Impact factor: 1.976

5.  Radiation exposure during radiofrequency catheter ablation for atrial fibrillation.

Authors:  Laurent Macle; Rukshen Weerasooriya; Pierre Jais; Christophe Scavee; Florence Raybaud; Kee-Joon Choi; Meleze Hocini; Jacques Clementy; Michel Haissaguerre
Journal:  Pacing Clin Electrophysiol       Date:  2003-01       Impact factor: 1.976

6.  Visualization of multiple catheters with electroanatomical mapping reduces X-ray exposure during atrial fibrillation ablation.

Authors:  Marco Scaglione; Luigi Biasco; Domenico Caponi; Matteo Anselmino; Andrea Negro; Paolo Di Donna; Antonella Corleto; Antonio Montefusco; Fiorenzo Gaita
Journal:  Europace       Date:  2011-03-18       Impact factor: 5.214

7.  Radiation exposure during catheter ablation of atrial fibrillation.

Authors:  Lars Lickfett; Mahadevappa Mahesh; Chandra Vasamreddy; David Bradley; Vinod Jayam; Zayd Eldadah; Timm Dickfeld; Deborah Kearney; Darshan Dalal; Berndt Lüderitz; Ronald Berger; Hugh Calkins
Journal:  Circulation       Date:  2004-10-25       Impact factor: 29.690

8.  Prevalence of diagnosed atrial fibrillation in adults: national implications for rhythm management and stroke prevention: the AnTicoagulation and Risk Factors in Atrial Fibrillation (ATRIA) Study.

Authors:  A S Go; E M Hylek; K A Phillips; Y Chang; L E Henault; J V Selby; D E Singer
Journal:  JAMA       Date:  2001-05-09       Impact factor: 56.272

9.  Reduction of fluoroscopy exposure and procedure duration during ablation of atrial fibrillation using a novel anatomical navigation system.

Authors:  Martin Rotter; Yoshihide Takahashi; Prashanthan Sanders; Michel Haïssaguerre; Pierre Jaïs; Li-Fern Hsu; Fréderic Sacher; Jean-Luc Pasquié; Jacques Clementy; Mélèze Hocini
Journal:  Eur Heart J       Date:  2005-03-01       Impact factor: 29.983

10.  Obesity is a major determinant of radiation dose in patients undergoing pulmonary vein isolation for atrial fibrillation.

Authors:  Joris Ector; Octavian Dragusin; Bert Adriaenssens; Wim Huybrechts; Rik Willems; Hugo Ector; Hein Heidbüchel
Journal:  J Am Coll Cardiol       Date:  2007-06-29       Impact factor: 24.094

View more
  1 in total

1.  Metal interference alert guided septal approach with 3 catheter positions on intracardiac echocardiography for a near-zero fluoroscopy catheter ablation of atrial fibrillation.

Authors:  Yu-Ki Iwasaki; Yuhi Fujimoto; Kanako Ito-Hagiwara; Eiichiro Oka; Hiroshi Hayashi; Teppei Yamamoto; Hiroshige Murata; Kenji Yodogawa; Wataru Shimizu
Journal:  Int J Cardiol Heart Vasc       Date:  2021-10-25
  1 in total