Capnography and the Bain circuit II: Validation of a computer model.

J E Beneken, N Gravenstein, S Lampotang, J J van der Aa, J S Gravenstein.   

Abstract

Validation of a computer model is described. The behavior of this model is compared both with mechanical ventilation of a test lung in a laboratory setup that uses a washout method and with manual ventilation. A comparison is also made with results obtained from a volunteer breathing spontaneously through a Bain circuit and with results published in the literature. This computer model is a multisegment representation of the Bain circuit and connecting tubing. For each segment, gas pressure, gas volume flow, and partial pressure of carbon dioxide are calculated for any number of breaths wanted. As a result, the time course of these variables can be generated for any location or, conversely, the carbon dioxide distribution in the system can be calculated for any time instant. A test lung, the human lungs, the ventilator bellows, and the reservoir bag are each represented by a single segment. The shapes of pressure and flow curves and of the capnograms taken at different locations in the Bain tubing are in good agreement. The washout study permits measurement of the time delay between the first expiration and the arrival of carbon dioxide at a particular location. The carbon dioxide level in the test lung decreases during inspiration and is stable during expiration. Quantitative agreement between model and experimental transport delays and carbon dioxide levels is such that the differences can be explained by the inaccuracy of the measurement. This is concluded from a sensitivity analysis. The study of the effect of segment size shows an almost optimal agreement between model behavior and experimental results for a 36-segment model. Execution of a thorough validation is imperative before such models can be used for clinical management and decision making or for teaching.

Entities:  

Mesh:

Substances:

Year:  1987        PMID: 3112317     DOI: 10.1007/bf01695939

Source DB:  PubMed          Journal:  J Clin Monit        ISSN: 0748-1977


  6 in total

1.  Carbon dioxide elimination during spontaneous ventilation with a modified Mapleson D system: studies in a lung model.

Authors:  D K Rose; R J Byrick; A B Froese
Journal:  Can Anaesth Soc J       Date:  1978-09

2.  Rebreathing and end-tidal CO2 during spontaneous breathing with the Bain circuit.

Authors:  W E Spoerel
Journal:  Can Anaesth Soc J       Date:  1983-03

3.  Rebreathing and carbon dioxide elimination with the Bain circuit.

Authors:  W E Spoerel
Journal:  Can Anaesth Soc J       Date:  1980-07

4.  Factors influencing capnography in the Bain circuit.

Authors:  N Gravenstein; S Lampotang; J E Beneken
Journal:  J Clin Monit       Date:  1985-01

5.  Respiratory compensation during spontaneous ventilation with the Bain circuit.

Authors:  R J Byrick
Journal:  Can Anaesth Soc J       Date:  1980-03

6.  Capnography and the Bain circuit I: A computer model.

Authors:  J E Beneken; N Gravenstein; J S Gravenstein; J J van der Aa; S Lampotang
Journal:  J Clin Monit       Date:  1985-04
  6 in total
  2 in total

1.  A mass balance model for the Mapleson D anaesthesia breathing system.

Authors:  M A Lovich; B A Simon; J G Venegas; N M Sims; J B Cooper
Journal:  Can J Anaesth       Date:  1993-06       Impact factor: 5.063

2.  Reduction of fresh gas flow requirements by a circle-modified bain breathing circuit.

Authors:  J B Valdrighi; P N Nance
Journal:  J Clin Monit       Date:  1991-01
  2 in total