| Literature DB >> 31119690 |
Fritz H Hengerer1,2, Gerd U Auffarth3, Christoffer Riffel3, Ina Conrad-Hengerer3.
Abstract
INTRODUCTION: To evaluate long-term outcomes following stand-alone implantation of two second-generation trabecular micro-bypass stents (iStent inject®, Glaukos Corp., San Clemente, CA, USA) in eyes with predominantly primary open-angle glaucoma (POAG) and considerable preoperative disease burden.Entities:
Keywords: Glaucoma; IStent inject; Intraocular pressure; Medication; Microinvasive glaucoma surgery (MIGS); Second generation; Stand alone; Stent; Trabecular
Mesh:
Year: 2019 PMID: 31119690 PMCID: PMC6824385 DOI: 10.1007/s12325-019-00984-9
Source DB: PubMed Journal: Adv Ther ISSN: 0741-238X Impact factor: 3.845
Fig. 1iStent inject® and iStent inject® Trabecular Micro-Bypass Stent System. Provided by Glaukos Corp.
Fig. 2iStent inject® implantation: placement of two stents in the anterior chamber angle. Provided by Glaukos Corp.
Demographics and preoperative characteristics (n = 44 eyes of 31 patients)
| iStent | |
|---|---|
| Gender M/F | 23/21 |
Age (years) Mean ± SD (range) | 71.3 ± 10.5 (40–88) |
| Race | 100% Caucasian |
C:D ratio Mean ± SD (range) | 0.8 ± 0.1 (0.5–1.0) |
VF MD (dB) Mean ± SD (range) | – 6.4 ± 6.9 dB (− 24.4 to − 0.3) |
| Type of glaucoma | |
POAG PXG | 38 (86.4%) 4 (9.1%) |
| Appositional narrow anglea | 1 (2.3%) |
| Secondary (neovascular) | 1 (2.3%) |
| Mean medicated IOP (mmHg); mean ± SD | 25.3 ± 6.0 mmHg |
| Medicated IOP level (mmHg) | |
| ≤ 15 mmHg | 1 (2.3%) |
| ≤ 18 mmHg | 4 (9.1%) |
| Mean # medications; mean ± SD | 2.98 ± 0.88 |
| Eyes with prior glaucoma surgery | 22 (50.0%) |
| Prior glaucoma surgeries, | |
| Trabeculectomy | 13 |
| CPC | 14 |
| ALT/SLT | 3 |
| Laser iridotomy | 2 |
| Single iStent | 1 |
| XEN implantation | 2 |
| Eyes on 3–5 preoperative medications, | 33 (75%) |
SD standard deviation, C:D cup to disc, VF MD visual field mean deviation, CPC cyclophotocoagulation, ALT argon laser trabeculoplasty, SLT selective laser trabeculoplasty
aEyes had open-angle configuration in the area of iStent inject implantation
bEyes could have more than one surgery (there were 35 surgeries in 22 eyes)
Fig. 3Reduction in mean intraocular pressure through 36 months (p < 0.0001); eyes with available data at each visit. IOP intraocular pressure, SD standard deviation, M month, Preop preoperative
Fig. 4Proportional analysis of eyes with IOP ≤ 18 mmHg and ≤ 15 mmHg through 36 months*; eyes with available data at each visit. IOP intraocular pressure, M month, Preop preoperative. *Highly significant (p < 0.0001) increase in proportion of eyes with IOP ≤ 18 mmHg and IOP ≤ 15 mmHg at M36 vs. preoperatively
Fig. 5Reduction in medication burden at 36 months vs. preoperative (p < 0.0001); eyes with available data at each visit. M month, Preop preoperative
Fig. 6Proportional analysis of medication use through 36 months; eyes with available data at each visit. M month, Preop preoperative, Med medication