Jianguo Jia1, Qun Zhao1, Pei Lu2, Guiyong Fan3, Hao Chen3, Chaoqun Liu3, Jinlian Liu3, Sheng Chen4, Zhengshuai Jin4. 1. Department of Orthopedics, SuZhou WuJiang Hospital of Integrated Traditional Chinese and Western Medicine. 2. Department of Orthopedics, the Affiliated Wuxi No. 2 People's Hospital of Nanjing Medical University. 3. Department of Orthopaedics, Suzhou Kowloom Hospital, Medical College of Shanghai Jiaotong University. 4. Department of Orthopaedics, Jiangsu Shengze Hospital, Nanjing Medical University, Suzhou, Jiangsu, China.
Abstract
BACKGROUND: To compare the clinical efficacy between Orthopilot navigation system and conventional manual surgery in total hip arthroplasty (THA). METHODS: Electronic databases were searched to identify randomized controlled trials (RCTs) investigating Orthopilot navigation system versus conventional manual in patients undergoing THA. Outcome measurements include anteversion angle, inclination angle, preoperative leg length discrepancy, postoperative leg length discrepancy and femoral offset. Statistical software Stata 12.0 was used for data-analysis. RESULTS: A total of 5 studies were finally included in this meta-analysis. The results showed that the conventional manual group have a less anteversion angle than that in Orthopilot navigation system group (weighted mean difference (WMD) = 4.67, 95% confidence interval (CI) = 3.53, 5.82, P = .000). And pooled analysis showed that the inclination angle in Orthopilot navigation group was less than that in conventional manual group (WMD = -4.19, 95% CI = -8.00, -0.37, P = .031). There was no significant difference between the preoperative leg length discrepancy and postoperative leg length discrepancy (P > .05). Orthopilot navigation system compared with conventional manual procedure was associated with decreased of femoral offset by 2.76 (WMD = -2.76, 95%CI = -3.90, -1.62, P = .000). CONCLUSION: Both Orthopilot navigation system and conventional THA result in significant improvements in patient function with similar overall complication rates and have their own edges in cup position.
BACKGROUND: To compare the clinical efficacy between Orthopilot navigation system and conventional manual surgery in total hip arthroplasty (THA). METHODS: Electronic databases were searched to identify randomized controlled trials (RCTs) investigating Orthopilot navigation system versus conventional manual in patients undergoing THA. Outcome measurements include anteversion angle, inclination angle, preoperative leg length discrepancy, postoperative leg length discrepancy and femoral offset. Statistical software Stata 12.0 was used for data-analysis. RESULTS: A total of 5 studies were finally included in this meta-analysis. The results showed that the conventional manual group have a less anteversion angle than that in Orthopilot navigation system group (weighted mean difference (WMD) = 4.67, 95% confidence interval (CI) = 3.53, 5.82, P = .000). And pooled analysis showed that the inclination angle in Orthopilot navigation group was less than that in conventional manual group (WMD = -4.19, 95% CI = -8.00, -0.37, P = .031). There was no significant difference between the preoperative leg length discrepancy and postoperative leg length discrepancy (P > .05). Orthopilot navigation system compared with conventional manual procedure was associated with decreased of femoral offset by 2.76 (WMD = -2.76, 95%CI = -3.90, -1.62, P = .000). CONCLUSION: Both Orthopilot navigation system and conventional THA result in significant improvements in patient function with similar overall complication rates and have their own edges in cup position.
Total hip arthroplasty (THA) is one of the most effective and frequent performed operations worldwide to relieve pain and restore function to the hip joint osteoarthritis.[ Every year, more than 1 million people worldwide undergo THA for severe hip osteoarthritis (OA) with intractable or permanent pain and dysfunctions, and this number is expected to double within the next 20 years.[ THA completely changed the treatment method of previous hip disease with arthritis in the 1960s and achieved excellent long-term efficacy.[ The scholars have devoted themselves to prolonging the service life of artificial joints by accurately positioning the prosthesis, reducing wear, and reducing the fretting of the prosthesis in THA.[In recent years, with the breakthrough of computer and artificial intelligence technology, imageless navigation system has become an important application in clinic,[ and used successfully in total knee arthroplasty, unicompartmental knee arthroplasty, THA, and shoulder arthroplasty, etc.[ Imageless navigation system is used to detect the anatomic data, mechanical axis of the limbs or position of the joint intraoperatively through a 3-dimensional optical positioning device and without the preoperative or intraoperative images of patients (computed tomography (CT), magnetic resonance imaging (MRI), C-arm fluoroscopy, and X-ray image).[ In January 1997, Saragaglia et al first introduced the Orthopilot imageless navigation system into total knee arthroplasty.[ Imageless navigation uses optical sensors as 3D position sensors to track the target bones and surgical tools or implants. In cup placement, the imageless navigation system measures the anteversion and inclination angles relative to the anterior pelvic plane. Since then, a large number of clinical trials have showed that navigation operation of such systems improves precision and accuracy over conventional manual surgery.[ However, it remains controversial on the choice of imageless navigation or traditional operation in THA.[This study aimed to systematically compare the clinical efficacy between the 2 methods through meta-analysis so as to give some theoretical guidance for clinical practice.
Methods
The current meta-analysis was performed according to the recommendations of the Cochrane Handbook for Systematic Reviews of Interventions and was reported in compliance with the Preferred Reporting Items for Systematic Reviews and Meta-Analyses (PRISMA) statement guidelines. This is a meta-analysis; no ethics approval and consent to participate exist.
Search strategy
Following the recommendations of the Cochrane collaborations, the retrieval was performed in the online databases include Embase, Pubmed, Web of Science from inception to September 2018. Manually search the magazine catalog and references, and strive to find gray literature, such as unpublished academic papers and chapters in monographs. Searching all pertinent papers without restricting the language and translating if necessary. Keywords both for Chinese and English search were: total hip arthroplasty, THA, imageless, navigation, conventional, manual, and freehand. Search strategy was: total hip arthroplasty OR THA AND imageless OR navigat∗ AND conventional OR manual OR freehand.
Inclusion criteria
Inclusion criteria for the study wereAdults with severe hip disease (Osteoarthritis, developmental dysplasia of the hip (DDH), adult avascular necrosis (AVN), rheumatoid arthritis (RA), Paget's disease etc.);controlled trials, prospective studies, retrospective studies, and cohort studies;all patients underwent for THA;study compared clinical efficacy of Orthopilot navigation system and conventional manual approaches.
Exclusion criteria
The exclusion criteria were:duplicates publications;letters, comments, editorials, case reports, proceedings, personal communications, or reviews;cadaveric study;study objective or intervention measures failed to meet the inclusion criteria;the original documents of experimental design being not precise;studied with incomplete data.
Data extraction and quality assessment
Data in included trials were extracted by 2 independent investigators (Jianguo Jia and Qun Zhao). Disagreement between the 2 reviewers was settled by discussion and consulting to a third reviewer. The extracted information included:the basic characteristics of the included studies, including the titles, authors, journals, volume, publication date;research methodological characteristics: random, control, blindness, etc.;research object characteristics: The general condition of patients, such as gender, age, race, etc., and the baseline situation of patients, disease course, severity, etc.;sample size, Intervention methods, follow-up time, outcome measurements.The risk-of-bias assessment tool outlined in Cochrane Handbook was used to assess the methodological quality of controlled clinical trials (CCTs). Six domains are evaluated:random sequence generation;allocation concealment;blinding of patients and personal;blinding of outcome assessment;incomplete outcome data;and selective reporting risk.Relevant data were recorded in this analysis, including: first author's name, published year, sample size of Orthopilot navigation system and conventional manual approaches for THA etc.
Outcome measures
Outcome measurements include 5 main aspects:anteversion angle,inclination angle,preoperative leg length discrepancy,postoperative leg length discrepancy,femoral offset.
Statistical analysis
All data analyses were conducted by using the Stata 12.0 software. Dichotomous outcomes were expressed in terms of risk ratio (RR) and the weighted mean difference (WMD) or the weighted mean difference (WMD) was used for continuous outcomes, both with 95% confidence intervals (CI).[ Heterogeneity was tested using both the chi-square test and I2 test. A fixed-effects model was chosen when there was no statistical evidence of heterogeneity (I2 < 50%) and random-effects model was adopted if significant heterogeneity was found (I2 ≥ 50%).[ If the heterogeneity was found, we checked the study population, treatment, outcome and methodologies to determine the source of heterogeneity. If it could not be quantitatively synthesized or the event rate was too low to be measured, we used qualitative evaluation.[ By eliminating some of the studies for sensitivity analysis and making funnel plots to assess the bias. The difference was statistically significant when P < .05.
Results
Search results
A total of 115 articles were retrieved based on the search strategy through online database searching and manual searching. After removing the duplication, a total of 59 studies were included for overview. By reading the titles and abstracts, 54 articles that are not related to the purpose of the study were excluded. Finally, 5 studies[ involving 719 patients were included. The literature screening process and the results shown in Figure 1. Basic characteristics of included literature presented in Table 1.
Figure 1
The flow diagram of study selection.
Table 1
General characteristic of the included studies.
The flow diagram of study selection.General characteristic of the included studies.Figure 2 and Figure 3 shows the risk of bias summary and risk of bias graph respectively. Only 1 study did not describe the random sequence generation and classified as unclear risk of bias. Allocation concealment was low risk of bias in 2 studies, the rest studies were classified as unclear risk of bias.
Figure 2
Risk of bias summary for included studies. +, no bias; –, bias; ?, bias unknown.
Figure 3
Risk of bias graph: review authors’ judgments about each risk of bias item presented as percentages across all included studies.
Risk of bias summary for included studies. +, no bias; –, bias; ?, bias unknown.Risk of bias graph: review authors’ judgments about each risk of bias item presented as percentages across all included studies.
Anteversion angle
Three studies involving 468 patients compared the anteversion angle between the Orthopilot navigation system and conventional manual procedure. Fixed effect model was employed in meta-analysis with the absence of heterogeneity (I2 = 0.0%, P = .715) among the 3 studies. The results showed that the conventional manual group have a less anteversion angle than that in Orthopilot navigation system group (WMD = 4.67, 95%CI = 3.53, 5.82, P = .000, Fig. 4).
Figure 4
Forest plot for comparing Orthopilot navigation system and conventional manual procedure in terms of anteversion angle.
Forest plot for comparing Orthopilot navigation system and conventional manual procedure in terms of anteversion angle.
Inclination angle
Three studies included have compared inclination angle between Orthopilot navigation system and conventional manual procedure. Random effect model was employed in meta-analysis because the heterogeneity between the studies was significant (I2 = 91%). The meta-analysis showed that the inclination angle in Orthopilot navigation group was less than that in conventional manual group (WMD = −4.19, 95%CI = −8.00, −0.37, P = .031, Fig. 5).
Figure 5
Forest plot for comparing Orthopilot navigation system and conventional manual procedure in terms of inclination angle.
Forest plot for comparing Orthopilot navigation system and conventional manual procedure in terms of inclination angle.
Complications
We compared preoperative leg length discrepancy, result was shown in Figure 6. There was no significant difference between the preoperative leg discrepancy (WMD = −1.09, 95%CI = −5.01, 2.82, P = .584).
Figure 6
Forest plot for comparing Orthopilot navigation system and conventional manual procedure in terms of preoperative leg length discrepancy.
Forest plot for comparing Orthopilot navigation system and conventional manual procedure in terms of preoperative leg length discrepancy.We then compared postoperative leg length discrepancy between Orthopilot navigation system and conventional manual procedure. Results shown that there was no significant difference between Orthopilot navigation system and conventional manual procedure in terms of the postoperative leg length discrepancy (WMD = −1.36, 95% CI = −4.36, 1.63, P = .372, Fig. 7).
Figure 7
Forest plot for comparing Orthopilot navigation system and conventional manual procedure in terms of potoperative leg length discrepancy.
Forest plot for comparing Orthopilot navigation system and conventional manual procedure in terms of potoperative leg length discrepancy.Two trials involving 386 participants reported femoral offset. Orthopilot navigation system compared with conventional manual procedure was associated with decreased of femoral offset by 2.76 (WMD = −2.76, 95%CI = −3.90, −1.62, P = .000, Fig. 8).
Figure 8
Forest plot for comparing Orthopilot navigation system and conventional manual procedure in terms of femoral offset.
Forest plot for comparing Orthopilot navigation system and conventional manual procedure in terms of femoral offset.
Discussion
Total hip arthroplasty (THA) is a relatively mature and widely used in orthopaedic surgery.[ The key to the long-term success rate of THA is the accurate placement of the prosthesis.[ Inaccurate placement of the prosthesis will lead to false femoral head and acetabulum impact and limited mobility.[ Accurate mounting of the prosthesis requires the surgeon to accurately position the patient's pelvic and femoral locations.[ In traditional surgery, preoperative imaging, and template measurements assist in the intraoperative placement of the prosthesis.[ However, due to the large number of factors affecting preoperative imaging, and the unstable position of the intraoperative patient, the position of pelvis and femur changes,[ causing deviations between the intraoperative and preoperative results, which ultimately affect the placement of the prosthesis.[The core of imageless navigation technology is to use computer technology to accurately correlate the preoperative image data with the anatomy of the intraoperative patient, track the surgical instruments in real time by detecting markers and display them in a virtual scene, and transmit the information to the surgeon in real time. A number of clinical studies have shown that imageless navigation can improve the accuracy of prosthesis placement, thereby reducing the occurrence of complications such as postoperative dislocation, compared with traditional THA surgery.Imageless navigation (Orthopilot system) can obtain 3-dimensional image data and simulate the degree of hip movement before surgery, which facilitates to the preparation of planning; it can track and guide the operation during surgery, and it can also assess the placement of the prosthesis. However, due to the high cost of computer navigation equipment, complex application of the system, and navigation takes time, there continues to be controversial for THA under navigation. We selected 3 outcomes including anteversion, Inclination, complication and strive to fully compare the efficacy between the 2 procedures. By comparison, statistically significant differences on femoral offset, overall complications, anteversion and inclination between 2 groups were found. The results indicating that using Orthopilot navigation system can achieve less complication of femoral offset and better inclination compared with conventional THA, but conventional manual method showed its edge on anteversion.This systematic review included 5 trials, and the methodological quality evaluation results were all middle quality. But a slight risk of bias remained in some studies, which may be related to that the patients should be informed which surgical method can be chose and medical ethical problems are involved. Incomplete outcome data, insufficient cases number of trials, and the various level between clinical surgeons may also lead to low methodological quality evaluation results and affect reliability of meta-analysis results. In the same outcome measurements system, we include into the maximum of 5 articles least of 2, therefore, the heterogeneity between groups will increase. Therefore, the above conclusions still need to further verify depends on the emergence of more randomized controlled trials with higher quality and larger sample sizes in the future.
Conclusion
This meta-analysis demonstrates that acetabular implant positioning can be significantly improved by the use of Orthopilot navigation system. However, studies with larger sample sizes and long-term results are needed in the future.