| Literature DB >> 31092257 |
Shuai Jiang1, Weishi Li2.
Abstract
PURPOSE: To investigate the biomechanical changes in the proximal adjacent segment with different grades of degeneration after posterior lumbar interbody fusion (PLIF).Entities:
Keywords: Adjacent segment degeneration; Finite element analysis; Intervertebral disc pressure; Range of motion
Mesh:
Year: 2019 PMID: 31092257 PMCID: PMC6521416 DOI: 10.1186/s13018-019-1150-9
Source DB: PubMed Journal: J Orthop Surg Res ISSN: 1749-799X Impact factor: 2.359
Fig. 1Four finite element models of the L3-L5 lumbar spine (a) Healthy disc with no fusion (b) Healthy disc with L4-L5 posterior lumbar interbody fusion (c) Mild degenerated disc and L4-L5 posterior lumbar interbody fusion (d) Moderate degenerated disc and L4-L5 posterior lumbar interbody fusion
Material properties used in the finite element models
| Materials | Young’s modulus (MPa) | Poisson’s ratio | Cross-sectional area (mm2) | Element type |
|---|---|---|---|---|
| Vertebra | ||||
| Cortical bone | 120,000 | 0.3 | – | S3 |
| Cancellous bone | 100 | 0.2 | – | C3D4 |
| Posterior bone | 3500 | 0.25 | – | C3D4 |
| Endplate | 23.8 | 0.4 | – | S3 |
| Intervertebral disc | ||||
| Annulus ground substance (Mooney-Rivlin law) | – | – | C3D8 | |
| Nucleus pulpous (Mooney-Rivlin law) | – | – | C3D8 | |
| Annulus fiber layers | 360–550 | 0.3 | 0.7 | T3D2 |
| Ligaments | ||||
| Anterior longitudinal | 7.8 (< 12.0%), 20 (> 12.0%) | 0.3 | 63.7 | T3D2 |
| Posterior longitudinal | 10.0 (< 11.0%), 20 (> 11.0%) | 0.3 | 20 | T3D2 |
| Ligamentum flavum | 15.0 (< 6.2%), 19.5 (> 6.2%) | 0.3 | 40 | T3D2 |
| Supraspinous | 8.0 (< 20.0%), 15 (> 20.0%) | 0.3 | 30 | T3D2 |
| Interspinous | 10.0 (< 14.0%), 11.6 (> 14.0%) | 0.3 | 40 | T3D2 |
| Intertransverse | 10.0 (< 18.0%), 58.7 (> 18.0%) | 0.3 | 1.8 | T3D2 |
| Capsular | 7.5 (< 25.0%), 32.9 (> 25.0%) | 0.3 | 30 | T3D2 |
S3 3-node triangular element, C3D4 4-node tetrahedral element, C3D8 8-node hexahedral element, T3D2 2-node truss element
Ranges of motion predicted by the intact FE model compared with reported ROMs from in vitro study
| L3–4 | Flexion | Extension | Lateral bending | Torsion (Left) |
|---|---|---|---|---|
| Yamamoto | 7.5 ± 0.8 | 3.7 ± 0.3 | 5.7 ± 0.3 (Left) | 2.7 ± 0.4 (Left) |
| Present study | 8.11 | 3.97 | 5.61 | 2.30 |
| L4–5 | Flexion | Extension | Lateral bending | Torsion |
| Yamamoto | 8.9 ± 0.7 | 5.8 ± 0.4 | 5.5 ± 0.5 (Left) | 1.7 ± 0.3 (Left) |
| Present study | 8.26 | 6.02 | 5.14 | 2.69 |
The lateral bending and torsion values in the present study are expressed as averages
Fig. 2Comparison of the ROM for the L3-L4
Fig. 3Comparison of the pressure for the L3-L4 nucleus
Fig. 4Comparison of the pressure for the L3-L4 annulus fibrosus