| Literature DB >> 31086512 |
Melike Korkmaz Toker1, Basak Altıparmak2, Ayse Gul Karabay3.
Abstract
OBJECTIVE: In obstetric patients' airway, guidelines have recommended the availability of advanced airway equipment. Our aim was to compare the larynx visualization provided by the Macintosh direct laryngoscope and McGrath video laryngoscope and the intubation time of patients undergoing cesarean section.Entities:
Keywords: Cesarean section; Direct Laryngoscopy; General anesthesia; McGrath Video laryngoscope
Year: 2019 PMID: 31086512 PMCID: PMC6500838 DOI: 10.12669/pjms.35.2.646
Source DB: PubMed Journal: Pak J Med Sci ISSN: 1681-715X Impact factor: 1.088
Demographic data and airway evaluation.
| DL (n = 50) | VL (n = 50) | p | |
|---|---|---|---|
| Age | 27.5 (24–31) | 26 (22–35) | 0.65 |
| Weight (kg) | 81.1±6.9 | 80.3±7.6 | 0.59[ |
| Height (cm) | 165.9±6.1 | 165.6±5.8 | 0.81[ |
| BMI | 29.6±2.2 | 29.3±2.5 | 0.45[ |
| Mallampati | 1.86±0.4 | 1.98±0.6 | 0.21[ |
Independent t-test results are expressed as mean ± standard deviation or as numbers of patients.
Mann-Whitney U-test results are expressed as the median (25th–75th interquartile range)
DL: direct laryngoscope, VL: video laryngoscope, BMI: Body Mass Index.
Time taken for glottic view, tracheal tube placement, and intubation time.
| DL (n =50) | VL (n = 50) | p[ | |
|---|---|---|---|
| Glottic view (s) | 19.5±3.9 | 17.7±4.4 | 0.028 |
| Tracheal tube placement (s) | 32.6±4.7 | 29.8±5.1 | 0.007 |
| Intubation time (s) | 40.1±5.4 | 34.7±5.2 | <0.001 |
Independent t-test results are expressed as mean ± standard deviation
DL: direct laryngoscope, VL: video laryngoscope
Glottic view assessment.
| DL (n = 50) | VL (n = 50) | p | |
|---|---|---|---|
| N (I/II/III/IV) | 3/34/9/4 | 16/32/2/0 | 0.003[ |
| % (I/II/III/IV) | 6/68/18/8 | 32/64/4/0 | |
| POGO (%) | 90 (86.75–92) | 94.5 (90–96) | < 0.001 |
Independent t-test results are expressed as mean ± standard deviation or as numbers of patients.
Mann-Whitney U test results are expressed as the median (25th–75th interquartile range)
DL: direct laryngoscope, VL: video laryngoscope, POGO: percentage of glottic opening
Hemodynamic parameters before blade insertion and 3 minutes after intubation.
| DL (n = 50) | VL (n = 50) | p[ | |
|---|---|---|---|
| Baseline | 79.3±8 | 77.9±7.4 | 0.3 |
| Post-intubation | 89.6±6.7 | 82.3±7.5 | <0.001 |
| Baseline | 76.7±7.4 | 77.6±7.6 | 0.5 |
| Post-intubation | 93.4±8.5 | 88.7±7.5 | 0.005 |
Independent t-test results are expressed as mean ± standard deviation
DL: direct laryngoscope, VL: video laryngoscope.