J Bashford1, A Wickham2, R Iniesta3, E Drakakis2, M Boutelle2, K Mills4, C Shaw4.
Abstract
OBJECTIVES: Fasciculations are a clinical hallmark of amyotrophic lateral sclerosis (ALS). Compared to concentric needle EMG, high-density surface EMG (HDSEMG) is non-invasive and records fasciculation potentials (FPs) from greater muscle volumes over longer durations. To detect and characterise FPs from vast data sets generated by serial HDSEMG, we developed an automated analytical tool.Entities:
Keywords: Amyotrophic lateral sclerosis; Biomarker; Fasciculation; High-density surface EMG
Mesh:
Year: 2019 PMID: 31078984 PMCID: PMC6553680 DOI: 10.1016/j.clinph.2019.03.032
Source DB: PubMed Journal: Clin Neurophysiol ISSN: 1388-2457 Impact factor: 3.708
Patient characteristics. ALS, amyotrophic lateral sclerosis; MMN, multifocal motor neuropathy; BFS, benign fasciculation syndrome. Muscle powers were assessed at baseline by the same clinician (JB).
| Patient No. | Age (years) | Gender | Diagnosis | Site of symptom onset | Duration since symptom onset (months) | Biceps power (MRC scale, 5 = normal) | Gastrocnemius power (MRC scale, 5 = normal) | Number of assessments undertaken | ||
|---|---|---|---|---|---|---|---|---|---|---|
| R | L | R | L | |||||||
| 1 | 59 | M | ALS | Left leg | 24 | 5 | 5 | 5 | 5 | 4 |
| 2 | 57 | M | ALS | Right arm | 19 | 4 | 4 | 4 | 4 | 5 |
| 3 | 50 | M | ALS | Left arm | 23 | 2 | 2 | 5 | 5 | 4 |
| 4 | 58 | M | ALS | Right arm | 60 | 1 | 1 | 5 | 5 | 6 |
| 5 | 59 | F | ALS | Bulbar | 27 | 5 | 5 | 5 | 5 | 6 |
| 6 | 61 | M | ALS | Left arm | 10 | 5 | 5 | 5 | 5 | 7 |
| 7 | 68 | F | MMN | Left arm | 204 | 4 | 5 | 5 | 5 | 6 |
| 8 | 57 | M | BFS | Both legs | 8 | 5 | 5 | 5 | 5 | 4 |
Fig. 1Data processing. (a) Probability threshold. The spike detection threshold was based on the probability of a given amplitude occurring in the thirty-minute recording. A spike’s duration was found by tracking the spike to its nearest zero point (orange circles). Adjacent phases (positive-negative transition) were joined and recorded as a single spike. (b) Principle of the super-channel (SC). For spikes A and B, the channel (1–4) with the highest peak-trough amplitude (shown in red) was transferred into the SC. The channel of origin for each spike was stored. (For interpretation of the references to colour in this figure legend, the reader is referred to the web version of this article.).
Fig. 2Noise level analysis (phase one). (a) Distribution of mean noise bands for biceps and gastrocnemius. Each data point represents one minute of recording. There was a significant difference in noise levels between biceps and gastrocnemius (p < 0.0001, Mann-Whitney test). (b). Relationships between mean noise band and optimal AT. ATinc was calculated from manual counts using 10 s windows. The best-fit lines were calculated with weighted least-squares regression due to heteroscedasticity of the data. For biceps (red), n was 295 and r2 was 0.648. For gastrocnemius (blue), n was 304 and r2 was 0.402. (For interpretation of the references to colour in this figure legend, the reader is referred to the web version of this article.).
Fig. 3Receiver operating characteristic (ROC) curves for FP identification (phase three). (a) Comparison of two analytical models. For each muscle (biceps in red and gastrocnemius in blue), 40 one-minute representative recordings were analysed. Model one (M1, triangle symbols) took the form Y = A1 and model two (M2, circular symbols) took the form Y = A2X, where Y was the optimal ATinc (μV), X was the mean noise band (μV) and A1/2 was a positive value. From left-to-right, threshold values for A1 were 100, 60, 40, 35, 30, 25, 20, 15, 10 and 5, and values for A2 were 14, 12, 10, 9, 8, 7, 6, 5, 4 and 2. Median sensitivity and specificity are displayed (non-parametric distributions). Area-under-the curve (AUC, 2 s.f.) represents the accuracy of each model for each muscle. * indicates the performance of M1 when A1 = 40; ∼ indicates performance of M1 when A1 = 20; # indicates the performance of M2 when A2 = 7. (b) Pooled results for model two. The results from biceps and gastrocnemius were pooled to produce a total of 80 one-minute recordings. Threshold values for A2 are displayed. Curves for median sensitivity and specificity are plotted alongside lower and upper limits for 95% confidence interval (CI). AUC represents the accuracy of this model. (For interpretation of the references to colour in this figure legend, the reader is referred to the web version of this article.).
Performance of two analytical models at FP identification. Model 1: Y = A1; Model 2: Y = A2X, where Y = optimal ATinc (μV), X = mean noise band (μV) and A2 = positive value. AUC, area under the curve.
| Model | Muscle | Optimal A1/2 | Median sensitivity (%) | Median specificity (%) | AUC (%) |
|---|---|---|---|---|---|
| 1 | Biceps | 30 | 79.8 | 90.2 | 84.1 |
| Gastrocnemius | 20 | 82.0 | 89.1 | 84.8 | |
| 2 | Biceps | 8 | 85.4 | 91.1 | 88.1 |
| Gastrocnemius | 7 | 84.3 | 88.8 | 88.4 | |
Fig. 4Data exclusion. (a) Finding the optimal amplitude exclusion threshold (AT The AT was varied to calculate the sensitivity and specificity for 80 pooled (biceps and gastrocnemius) one-minute recordings. Boxes represent median and inter-quartile range (IQR). Whiskers represent (upper quartile + 1.5 * IQR) and (lower quartile − 1.5 * IQR) according to Tukey’s method. Data points beyond this range are plotted individually. (b) AT One minute of amplitude inclusion thresholds (linearly related to noise levels), showing a burst in noise levels between 1535–1540 s. Amplitude exclusion threshold of 100 μV (red line) applied to exclude shaded region. (c) AT Corresponding one minute of super-channel recording, showing exclusion of noisy period. Fasciculation count (369), time analysed (0.917 m) and fasciculation frequency (403/m) have been automatically adjusted. (For interpretation of the references to colour in this figure legend, the reader is referred to the web version of this article.)
Fig. 5Application of the analytical pipeline. (a) FPs from raw data. Black bar indicates 5 ms duration. (b) Example raw data. 60 s of raw data from gastrocnemius of ALS patient 3. (c) Example SPiQE output. Super-channel of corresponding 60 s of data in ‘b’ after application of optimal analytical pipeline. (d) Model performance. The optimal model (Y = 8X) has been applied to 80 one-minute recordings and compared with high precision manual counts. Medians and 95% confidence intervals plotted. Wilcoxon signed-rank test confirms no significant difference (p = 0.394).