Our institution performs in vivo verification measurement for each of our total body irradiation (TBI) patients with optically stimulated luminescent dosimeters (OSLD). The lung block verification measurements were commonly higher than expected. The aim of this work is to understand this discrepancy and improve the accuracy of these lung block verification measurements. Initially, the thickness of the lung block was increased to provide adequate lung sparing. Further tests revealed the increase was due to electron contamination dose emanating from the lung block. The thickness of the bolus material covering the OSLD behind the lung block was increased to offset the electron contamination. In addition, the distance from the lung block to the dosimeter was evaluated for its effect on the OSLD reading and found to be clinically insignificant over the range of variability in our clinic. The results show that the improved TBI treatment technique provides for better accuracy of measured dose in vivo and consistency of patient setup.Entities:
Mesh:
Year: 2016 PMID: 27167290 PMCID: PMC5690919 DOI: 10.1120/jacmp.v17i3.6128
Source DB: PubMed Journal: J Appl Clin Med Phys ISSN: 1526-9914 Impact factor: 2.102
Figure 1TBI patient treatment geometry showing the lung block and spoiler in front of the patient for (a) AP‐PA and (b) LAT techniques.
Percentage deviation of OSLD reading with 5 mm and 10 mm bolus from the expected entrance dose with lung block shielding in two LAT TBI and two AP‐PA TBI patients. Entrance dose reading of the right lateral field in Patient #1 was not obtained with 10 mm bolus due to OSLD falling off the patient during treatment.
|
|
|
|
|
|---|---|---|---|
| 1 | Right Lateral | 15.9 % | 6.6% |
| Left Lateral | 25.5% | 8.5% | |
| 2 | Right Lateral | 51.6% | 7.9% |
| Left Lateral | 44.5% | 8.7% | |
| 3 | AP | 21.2% | 8.0% |
| PA | 10.4% | 6.8% | |
| 4 | AP | 29.7% | 7.7% |
| PA | 15.2% | 7.5% | |
| Mean difference | 22.5% | 6.4% | |
| SD | 0.19% | 0.03% |
Data from a RANDO phantom irradiation in the lateral TBI patient treatment geometry.