Literature DB >> 31032195

Back-projection algorithm in generalized form for circular-scanning-based photoacoustic tomography with improved tangential resolution.

Bo Wang1, Tianning Su1, Weiran Pang2, Ningning Wei1, Jiaying Xiao2, Kuan Peng2.   

Abstract

BACKGROUND: The back-projection algorithm is the most common method for the reconstruction of circular-scanning-based photoacoustic tomography (CSPAT) due to its simplicity, computational efficiency, and robustness. It usually can be implemented in two models: one for ideal point detector, and the other for planar transducer with infinite element size. However, because most transducers in CSPAT are planar with a finite size, the off-center targets will be blurred in the tangential direction with these two reconstruction models.
METHODS: Here in this paper, we put forward a new model of the back projection algorithm for the reconstruction of CSPAT with finite size planar transducer, in which the acoustic spatial temporal response of the employed finite size transducer is approximated with a virtual detector placed at an optimized distance behind the transducer, and the optimized distance is determined by a phase square difference minimization scheme. Notably, this proposed method can also be suitable for reconstruction with the ideal point detector and infinite planar detector, and thus is a generalized form of the back-projection algorithm.
RESULTS: Compared with the two conventional models of the back-projection method and the modified back-projection method, the proposed method in this work can significantly improve the tangential resolution of off-center targets, thus improving the reconstructed image quality. These findings are validated with both simulations and experiments.
CONCLUSIONS: We propose a generalized model of the back projection algorithm to restore the elongated tangential resolution in CSPAT in case of a planar transducer of finite size, which can also be applicable for point and large-size planar transducers. This proposed method may also guide the design of CSPAT scanning configurations for potential applications such as human breast imaging for cancer detection.

Entities:  

Keywords:  Photoacoustic tomography (PAT); back-projection algorithm; tangential resolution; virtual point detector

Year:  2019        PMID: 31032195      PMCID: PMC6462567          DOI: 10.21037/qims.2019.03.12

Source DB:  PubMed          Journal:  Quant Imaging Med Surg        ISSN: 2223-4306


  13 in total

1.  Analytic explanation of spatial resolution related to bandwidth and detector aperture size in thermoacoustic or photoacoustic reconstruction.

Authors:  Minghua Xu; Lihong V Wang
Journal:  Phys Rev E Stat Nonlin Soft Matter Phys       Date:  2003-05-09

2.  Model-based correction of finite aperture effect in photoacoustic tomography.

Authors:  Meng-Lin Li; Yi-Chieh Tseng; Chung-Chih Cheng
Journal:  Opt Express       Date:  2010-12-06       Impact factor: 3.894

3.  Thermoacoustic tomography with integrating area and line detectors.

Authors:  Peter Burgholzer; Christian Hofer; Günther Paltauf; Markus Haltmeier; Otmar Scherzer
Journal:  IEEE Trans Ultrason Ferroelectr Freq Control       Date:  2005-09       Impact factor: 2.725

4.  Image reconstruction for photoacoustic scanning of tissue structures.

Authors:  C G Hoelen; F F de Mul
Journal:  Appl Opt       Date:  2000-11-01       Impact factor: 1.980

5.  Design and evaluation of a hybrid photoacoustic tomography and diffuse optical tomography system for breast cancer detection.

Authors:  Lei Xi; Xiaoqi Li; Lei Yao; Stephen Grobmyer; Huabei Jiang
Journal:  Med Phys       Date:  2012-05       Impact factor: 4.071

6.  Liquid acoustic lens for photoacoustic tomography.

Authors:  Chaolong Song; Lei Xi; Huabei Jiang
Journal:  Opt Lett       Date:  2013-08-01       Impact factor: 3.776

7.  Deblurring algorithms accounting for the finite detector size in photoacoustic tomography.

Authors:  Heinz Roitner; Markus Haltmeier; Robert Nuster; Dianne P O'Leary; Thomas Berer; Guenther Paltauf; Hubert Grün; Peter Burgholzer
Journal:  J Biomed Opt       Date:  2014-05       Impact factor: 3.170

8.  Integrated photoacoustic and diffuse optical tomography system for imaging of human finger joints in vivo.

Authors:  Lei Xi; Huabei Jiang
Journal:  J Biophotonics       Date:  2015-10-02       Impact factor: 3.207

9.  Tangential resolution improvement in thermoacoustic and photoacoustic tomography using a negative acoustic lens.

Authors:  Manojit Pramanik; Geng Ku; Lihong V Wang
Journal:  J Biomed Opt       Date:  2009 Mar-Apr       Impact factor: 3.170

10.  Non-invasive imaging of epileptic seizures in vivo using photoacoustic tomography.

Authors:  Qizhi Zhang; Zhao Liu; Paul R Carney; Zhen Yuan; Huanxin Chen; Steven N Roper; Huabei Jiang
Journal:  Phys Med Biol       Date:  2008-03-11       Impact factor: 3.609

View more
  1 in total

1.  Image reconstruction for endoscopic photoacoustic tomography including effects of detector responses.

Authors:  Zheng Sun; Huifeng Sun
Journal:  Exp Biol Med (Maywood)       Date:  2022-03-01
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.