| Literature DB >> 31028253 |
M Graeser1,2, F Thieben3,4, P Szwargulski3,4, F Werner3,4, N Gdaniec3,4, M Boberg3,4, F Griese3,4, M Möddel3,4, P Ludewig5, D van de Ven6, O M Weber7, O Woywode8, B Gleich9, T Knopp3,4.
Abstract
Determining the brain perfusion is an important task for diagnosis of vascular diseases such as occlusions and intracerebral haemorrhage. Even after successful diagnosis, there is a high risk of restenosis or rebleeding such that patients need intense attention in the days after treatment. Within this work, we present a diagnostic tomographic imager that allows access to brain perfusion quantitatively in short intervals. The device is based on the magnetic particle imaging technology and is designed for human scale. It is highly sensitive and allows the detection of an iron concentration of 263 pmolFe ml-1, which is one of the lowest iron concentrations imaged by MPI so far. The imager is self-shielded and can be used in unshielded environments such as intensive care units. In combination with the low technical requirements this opens up a variety of medical applications and would allow monitoring of stroke on intensive care units.Entities:
Mesh:
Substances:
Year: 2019 PMID: 31028253 PMCID: PMC6486595 DOI: 10.1038/s41467-019-09704-x
Source DB: PubMed Journal: Nat Commun ISSN: 2041-1723 Impact factor: 14.919
Fig. 1Implementation and concept of the presented system. Picture of the developed MPI head scanner. The system-calibration robot is placed to the left of the imager (top). Scanner concept (bottom). The central coil is used as a drive-field generator with a frequency of 25.699 kHz and an amplitude of 6 mT μ0−1. The cross section is formed by two half circles connected by straight parts forming an ellipsoid-like shape. The selection-field coils are driven by sinusoidal currents with opposite signs superimposed by an offset current. This generates an FFP oscillating on the y-axis in-between the selection-field coils. The superposition of the drive-field and the selection-field generates a Cartesian like FFP trajectory covering an FOV within the xy-plane of 100 × 140 mm2. The particle response is received by dedicated coils in x-direction and y-direction
Fig. 2Sensitivity of the developed MPI brain scanner. 50 μl samples of Perimag with varying dilutions are moved to three positions on the FOV diagonal (upper 6 images). The sample is considered to be detected if the movement of the sample correlates to the signal shift in the image. Below 2 μgFe this correlation fails. The same procedure is done with an ellipsoid of 134 ml filled with different concentrations (lower 3 images). At a concentration of 14.7 ngFe ml−1 (263 pmol Fe ml−1, 2 μgFe total) the position is still detectable while it fails at concentrations below that value
Fig. 3Spatial resolution of the developed MPI brain scanner. a–c Two samples of 250 μl Perimag with a concentration of 8.5 mgFe ml−1 were placed at varying distance along the x−direction, y−direction, and z-direction. The pixel spacing of the images is 5 mm and marks the lower boundary that can be reached. In x-direction the samples can be separated even for a distance of 5 mm. In y-direction the samples with 5 mm distance cannot be resolved such that the resolution limit in y-direction is 6 mm. In z-direction the samples can be resolved down to a distance of 26 mm. d For an iron mass of 125 μgFe the resolution in x-direction and y-direction remains below 10 mm
Fig. 4Reconstruction results of the 3D phantom. The space covered by the three letters is 90 × 90 mm2 for each layer. The distance between the individual layers is 32 mm. The letters consist of capillaries with an inner diameter of 1.3 mm filled with Perimag with a concentration of 8.5 mgFe ml−1. The reconstructed images show the three letters without interference of the other layers. The total image acquisition time of the 13 slices was 26 s
Fig. 5Results of the static stroke experiment. (left) Reconstruction results. In the control case both hemispheres and the stroke part were filled with a identical iron concentration of 965 ngFe ml−1 with a total amount of 938.8 µgFe. For the other cases the iron concentration in the stroke parts was less, which can be identified in the images as a lower particle concentration in all three cases. The stroke parts can be seen in the picture next to the head. (center). The lower concentration in the stroke area becomes even more prominent when subtracting the control image from the stroke images as can be seen in the second column of the images. As reference a cut of the CAD model is drawn next to the control image. (right) Photograph of the phantom mounted on a human head model
Fig. 6Dynamic imaging capabilities of the brain imager. The complete phantom can be seen in the picture on the right. It consists of two 50 ml tubes filled with glass spheres. To visualize the flow that is expected from the phantom a cross section sketch of the tubes is shown above the image. The flow enters the tubes and is evenly distributed via perforated rods inside. The flow exits the rods facing towards the center of the phantom and enters the drain rod facing the outer side of the phantom. In the control experiment the hoses remain untouched while in the stenosis setting the left inflow was pinched. The reconstructed images can be seen in gray scale on the left side. At the lower dosage of 85 μgFe both tubes were imaged and a stroke setting was detected. With the higher dosage of 850 μgFe the spatial resolution improved and the inflow and outflow became visible. For the control setting of the experiments with higher concentration, four regions with characteristic temporal behavior were selected and their temporal progression is shown in the graphs (top). The corresponding positions are marked in the reconstructed images of the high dose control experiment. The signal was first detected in the feeding hose (blue graph) with a lower duration compared to the perforated hollow rods (magenta graph). With a slightly higher temporal delay the signal increased in the tubes (green graph) and the discharge hose (purple graph). As the bolus diluted while passing through the phantom, the shape broadens and the signal duration was prolonged. From this time data, the time-to-peak (TTP), mean-transit-time (MTT), relative cerebral-blood-flow (rCBF), and relative cerebral-blood-volume (rCBV) perfusion maps were calculated. The rCBF and rCBV were normalized to the maximum value in the imaging volume. All time data were normalized to the arrival of the bolus