Literature DB >> 31012768

Dynamic current focusing for loudness encoding in cochlear implants: a take-home trial.

Monique A M de Jong1, Jeroen J Briaire1, Séline F S van der Woude1, Johan H M Frijns1,2.   

Abstract

Objective: This study aimed to evaluate a more energy-efficient dynamic current focussing (DCF) speech-processing strategy after long-term listening experience. In DCF, tripolar stimulation is used near the threshold and loudness is controlled by the compensation coefficient σ. A recent acute pilot study showed improved spectral-temporally modulated ripple test (SMRT) scores at low loudness levels, but battery life was reduced to 1.5-4 hours. Design: Within-subject comparisons were made for the clinical versus. DCF strategy after 5 weeks of at-home usage. Speech intelligibility in noise, spectral ripple discrimination, temporal modulation detection, loudness growth, and subjective ratings were assessed. Study sample: Twenty HiRes90K (Advanced Bionics, Valencia, USA) cochlear implant (CI) users.
Results: Average battery life was 9 hours with the newly implemented DCF compared to 13.4 hours with the clinical strategy. Compared with measurements made at the beginning of the study, SMRT-scores and speech intelligibility in noise were significantly improved with DCF. However, both measures suffered from unexpected learning effects over time. The improvement disappeared and speech intelligibility in noise declined significantly relative to the final control measurement with the clinical strategy.
Conclusion: Most CI users can adapt to the DCF strategy in a take-home setting. Although DCF has the potential to improve performance on the SMRT test, learning effects complicate the interpretation of the current results.

Entities:  

Keywords:  Cochlear implants; current focussing; dynamic current focussing; loudness coding; spectral resolution; speech intelligibility

Mesh:

Year:  2019        PMID: 31012768     DOI: 10.1080/14992027.2019.1601270

Source DB:  PubMed          Journal:  Int J Audiol        ISSN: 1499-2027            Impact factor:   2.117


  4 in total

1.  Combining current focusing and steering in a cochlear implant processing strategy.

Authors:  Xin Luo; Ching-Chih Wu; Kathryn Pulling
Journal:  Int J Audiol       Date:  2020-09-23       Impact factor: 2.117

2.  A Longitudinal Comparison of Environmental Sound Recognition in Adults With Hearing Aids Before and After Cochlear Implantation.

Authors:  Michael S Harris; Aaron C Moberly; Ben L Hamel; Kara Vasil; Christina L Runge; William J Riggs; Valeriy Shafiro
Journal:  J Speech Lang Hear Res       Date:  2021-03-02       Impact factor: 2.674

Review 3.  Cochlear Implant Research and Development in the Twenty-first Century: A Critical Update.

Authors:  Robert P Carlyon; Tobias Goehring
Journal:  J Assoc Res Otolaryngol       Date:  2021-08-25

4.  Evaluating and Comparing Behavioural and Electrophysiological Estimates of Neural Health in Cochlear Implant Users.

Authors:  Tim Brochier; François Guérit; John M Deeks; Charlotte Garcia; Manohar Bance; Robert P Carlyon
Journal:  J Assoc Res Otolaryngol       Date:  2020-11-04
  4 in total