Precision of the virtual occlusal record.

Keyan P Botsford, Michael C Frazier, Ahmed A M Ghoneima, Achint Utreja, Surya S Bhamidipalli, Kelton T Stewart.   

Abstract

OBJECTIVES: To evaluate the precision of the virtual occlusal record using the Carestream CS3600 Intraoral Scanner (Carestream Dental, Atlanta, Ga).
MATERIALS AND METHODS: A total of 20 participants were recruited for this prospective study using preestablished inclusion/exclusion criteria. A complete intraoral scan and two bite registrations were obtained. The participants were instructed to bite with normal pressure when bite registrations were acquired. Contact locations, size (circumference), and intensity were identified on the maxillary first molars and canines. Agreement between contact size and intensity was assessed with intraclass correlation coefficients. Kappa statistics evaluated agreement in contact locations. Statistical significance was set at P < .05.
RESULTS: All participant data were included for statistical analysis. Between the two bite registrations, nonstatistically significant differences were observed in the proportion of locations with contacts (P = .7681). A nonstatistically significant difference (-0.25 mm, P = .8416) in mean contact circumference size was observed. A statistically significant difference in mean contact intensity was observed (P = .0448). When evaluating agreement between the bite registrations, a weak correlation for size (intraclass correlation coefficient = 0.35) and intensity (intraclass correlation coefficient = 0.32) was observed as well as a moderate agreement for contact location (κ coefficient = 0.67).
CONCLUSIONS: The findings suggest that the Carestream intraoral scanner software possesses adequate precision when acquiring the location and size of the contacts in bite registrations. The scanner failed to demonstrate adequate precision when acquiring contact intensities in bite registrations. Additional research is warranted to further investigate the precision of virtual occlusal records with currently available software systems.

Entities:  

Keywords:  Bite registration; Intraoral scanner; Precision; Virtual occlusal record

Mesh:

Year:  2019        PMID: 30920871      PMCID: PMC8111830          DOI: 10.2319/092018-684.1

Source DB:  PubMed          Journal:  Angle Orthod        ISSN: 0003-3219            Impact factor:   2.079


  12 in total

1.  Accuracy of contacts calculated from 3D images of occlusal surfaces.

Authors:  R DeLong; S Knorr; G C Anderson; J Hodges; M R Pintado
Journal:  J Dent       Date:  2007-04-05       Impact factor: 4.379

2.  Dental impressions using 3D digital scanners: virtual becomes reality.

Authors:  Nathan S Birnbaum; Heidi B Aaronson
Journal:  Compend Contin Educ Dent       Date:  2008-10

3.  The accuracy of virtual interocclusal registration during intraoral scanning.

Authors:  Faraj Edher; Alan G Hannam; David L Tobias; Chris C L Wyatt
Journal:  J Prosthet Dent       Date:  2018-06-28       Impact factor: 3.426

4.  Virtual bite registration using intraoral digital scanning, CT and CBCT: In vitro evaluation of a new method and its implication for orthognathic surgery.

Authors:  Johanna Nilsson; Robert Geoff Richards; Andreas Thor; Lukas Kamer
Journal:  J Craniomaxillofac Surg       Date:  2016-06-23       Impact factor: 2.078

5.  Reproducibility of occlusal marking techniques.

Authors:  E Gazit; S Fitzig; M A Lieberman
Journal:  J Prosthet Dent       Date:  1986-04       Impact factor: 3.426

6.  Three-Dimensional Accuracy of Digital Static Interocclusal Registration by Three Intraoral Scanner Systems.

Authors:  Kuan Yee Wong; Roxanna Jean Esguerra; Vanessa Ai Ping Chia; Ying Han Tan; Keson Beng Choon Tan
Journal:  J Prosthodont       Date:  2017-11-21       Impact factor: 2.752

7.  Comparative analysis on reproducibility among 5 intraoral scanners: sectional analysis according to restoration type and preparation outline form.

Authors:  Ji-Man Park
Journal:  J Adv Prosthodont       Date:  2016-10-21       Impact factor: 1.904

8.  A comparison of the precision of three-dimensional images acquired by 2 digital intraoral scanners: effects of tooth irregularity and scanning direction.

Authors:  Ji-Won Anh; Ji-Man Park; Youn-Sic Chun; Miae Kim; Minji Kim
Journal:  Korean J Orthod       Date:  2016-01-25       Impact factor: 1.372

9.  Comparison of digital and conventional impression techniques: evaluation of patients' perception, treatment comfort, effectiveness and clinical outcomes.

Authors:  Emir Yuzbasioglu; Hanefi Kurt; Rana Turunc; Halenur Bilir
Journal:  BMC Oral Health       Date:  2014-01-30       Impact factor: 2.757

10.  Trueness and Precision of Four Intraoral Scanners in Oral Implantology: A Comparative in Vitro Study.

Authors:  Francesco G Mangano; Giovanni Veronesi; Uli Hauschild; Eitan Mijiritsky; Carlo Mangano
Journal:  PLoS One       Date:  2016-09-29       Impact factor: 3.240

View more
  4 in total

1.  Reproducibility of digital indirect bonding technique using three-dimensional (3D) models and 3D-printed transfer trays.

Authors:  Maria Eduarda Assad Duarte; Bruno Frazão Gribel; Alice Spitz; Flavia Artese; José Augusto Mendes Miguel
Journal:  Angle Orthod       Date:  2019-08-14       Impact factor: 2.079

2.  Integration of digital maxillary dental casts with 3D facial images in orthodontic patients.

Authors:  Zhuoxing Xiao; Zijin Liu; Yan Gu
Journal:  Angle Orthod       Date:  2020-05-01       Impact factor: 2.079

3.  Multifunctional Templates for Minimized Osteotomy, Implantation, and Palatal Distraction with a Mini-Screw-Assisted Expander in Schizodontism and Maxillary Deficit.

Authors:  Manfred Nilius; Katrin Hess; Dominik Haim; Bernhard Weiland; Guenter Lauer
Journal:  Case Rep Dent       Date:  2020-09-09

4.  The Effect of Scanning Strategy on Intraoral Scanner's Accuracy.

Authors:  Nikolaos A Gavounelis; Chrysoula-Maria C Gogola; Demetrios J Halazonetis
Journal:  Dent J (Basel)       Date:  2022-07-04
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.