Chonghe Wang1, Xiaoshi Li2, Hongjie Hu2, Lin Zhang1, Zhenlong Huang1, Muyang Lin1,3, Zhuorui Zhang1, Zhenan Yin4, Brady Huang5, Hua Gong1, Shubha Bhaskaran4, Yue Gu2, Mitsutoshi Makihata6, Yuxuan Guo1, Yusheng Lei1, Yimu Chen1, Chunfeng Wang1,7, Yang Li1, Tianjiao Zhang1, Zeyu Chen8, Albert P Pisano6, Liangfang Zhang1, Qifa Zhou8, Sheng Xu9,10,11,12. 1. Department of Nanoengineering, University of California San Diego, La Jolla, CA, USA. 2. Materials Science and Engineering Program, University of California San Diego, La Jolla, CA, USA. 3. School of Precision Instrument and Optoelectronic Engineering, Tianjin University, Tianjin, China. 4. Department of Electrical and Computer Engineering, University of California San Diego, La Jolla, CA, USA. 5. Department of Radiology, School of Medicine, University of California San Diego, La Jolla, CA, USA. 6. Department of Mechanical and Aerospace Engineering, University of California San Diego, La Jolla, CA, USA. 7. The Key Laboratory of Materials Processing and Mold of Ministry of Education, School of Materials Science and Engineering, School of Physics & Engineering, Zhengzhou University, Zhengzhou, Henan, China. 8. Department of Ophthalmology and Biomedical Engineering, Viterbi School of Engineering, University of Southern California, Los Angeles, CA, USA. 9. Department of Nanoengineering, University of California San Diego, La Jolla, CA, USA. shengxu@ucsd.edu. 10. Materials Science and Engineering Program, University of California San Diego, La Jolla, CA, USA. shengxu@ucsd.edu. 11. Department of Electrical and Computer Engineering, University of California San Diego, La Jolla, CA, USA. shengxu@ucsd.edu. 12. Department of Bioengineering, University of California San Diego, La Jolla, CA, USA. shengxu@ucsd.edu.
Abstract
Continuous monitoring of the central-blood-pressure waveform from deeply embedded vessels, such as the carotid artery and jugular vein, has clinical value for the prediction of all-cause cardiovascular mortality. However, existing non-invasive approaches, including photoplethysmography and tonometry, only enable access to the superficial peripheral vasculature. Although current ultrasonic technologies allow non-invasive deep-tissue observation, unstable coupling with the tissue surface resulting from the bulkiness and rigidity of conventional ultrasound probes introduces usability constraints. Here, we describe the design and operation of an ultrasonic device that is conformal to the skin and capable of capturing blood-pressure waveforms at deeply embedded arterial and venous sites. The wearable device is ultrathin (240 μm) and stretchable (with strains up to 60%), and enables the non-invasive, continuous and accurate monitoring of cardiovascular events from multiple body locations, which should facilitate its use in a variety of clinical environments.
Continuous monitoring of the central-blood-pressure waveform from deeply embedded vessels, such as the carotid artery and jugular vein, has clinical value for the prediction of all-cause cardiovascular mortality. However, existing non-invasive approaches, including photoplethysmography and tonometry, only enable access to the superficial peripheral vasculature. Although current ultrasonic technologies allow non-invasive deep-tissue observation, unstable coupling with the tissue surface resulting from the bulkiness and rigidity of conventional ultrasound probes introduces usability constraints. Here, we describe the design and operation of an ultrasonic device that is conformal to the skin and capable of capturing blood-pressure waveforms at deeply embedded arterial and venous sites. The wearable device is ultrathin (240 μm) and stretchable (with strains up to 60%), and enables the non-invasive, continuous and accurate monitoring of cardiovascular events from multiple body locations, which should facilitate its use in a variety of clinical environments.
The variation of blood-pressure (BP) waveforms contains abundant information of
the dynamic cardiovascular status[1,2]. Each one of peaks and notches in the
arterial BP waveform represents a specific left heart activity. Likewise, the
characteristic morphology of the venous BP waveform is closely related to relevant right
heart events[3]. Therefore, continuous
monitoring of subtle changes of those vital signals can provide remarkable insights for
cardiovascular disease diagnosis and prognosis[4]. Although monitoring the vascular pulsation at the peripheral
sites is useful for specific symptoms, emerging evidence suggests that the central
arterial and venous BP waveforms possess significantly more relevance to cardiovascular
events than the peripheral BP (PBP)[5-8]. Firstly, major
organs, including the heart, kidneys, lungs, and the brain, are directly exposed to the
central arteries. Therefore the distending pressure in the large elastic arteries (such
as aorta and carotid) is a vital determinant of the degenerative changes that
characterize accelerated aging and hypertension[9]. Secondly, amplification and reflection effect caused by the
complexities of peripheral vascular resistance along the conduit artery, namely the
stiffness mismatch between the peripheral and central vessels, is tough to evaluate.
This uncertainty often creates irregular and unpredictable influence on the PBP
waveform, making it incapable of achieving reliable cardiovascular status
assessment[10]. Thirdly,
although the central-blood-pressure (CBP) waveform can sometimes be derived from the PBP
waveform by the translational equation, demographic results indicate that clinical
treatment, such as using BP lowering drugs, can exert different effects on PBP and CBP
waveforms[7,11], causing inaccurate recordings[12]. Such inaccuracy can cause errors in
the assessment of myocardial oxygen requirements[13] and ventricular load and hypertrophy[14], as well as disparities in the actions of
different vasodilator agents[15].
Therefore, treatment decisions for cardiovascular diseases should be based on CBP rather
than PBP waveforms[16].The gold standard for recording the CBP waveforms in the carotid artery and
jugular venous sites, cardiac catheterization (also known as cannulation), involves
implanting a fiber-based pressure sensor into the relevant vasculature[17] (Supplementary note 1). Despite its high
accuracy, it causes patient suffering and increases the risk of infection, and thus is
too invasive for routine inspections[1].
Although there are several non-invasive methods including the optical method
(photoplethysmography (PPG), or volume clamp)[18], tonometry[4,19], and ultrasound wall-tracking that can
potentially monitor the CBP waveform, they suffer from several technical challenges.
Specifically, the PPG has insufficient penetration depth (<8 mm) for measuring
central vasculature which is often embedded in the tissue thickness > 3
cm[20]. Other technical problems
of PPG can be summarized as signal aliasing from venous and arterial
pulsations[21], susceptibility
to heat, moisture[22], and high
dependence on constant blood constituent[23]. The tonometry involves using strain sensors to detect the vessel
pulsation. This method highly relies on the efficiency of the blood vessel flattening by
the tonometer. Therefore, this is only recommended for the PBP measurement where a
supporting bony structure is available, which can provide a solid mechanical
support[24]. Due to this reason,
its accuracy largely degrades for measuring central vasculatures that have no proximal
supporting skeletons. Also, this method is susceptible to the subject’s obesity
that highly dampens the pulse wave propagation (Supplementary notes 2 and 3; Supplementary Figs. 1 and 2).
With the high penetrating capability, the ultrasound wall-tracking technique utilizes
the high-speed imaging probe to track the pulsation of the vasculature embedded in deep
tissues[25]. However, the
imaging probe is highly sensitive to motion artifacts, which will bring significant
burden to its associated wall-tracking recognition algorithm (Supplementary note 4; Supplementary Fig. 3). Additionally,
current ultrasound imaging probes are heavy and bulky. A reliable acoustic coupling
interface requires a stable holding of the probe by the operator. This will introduce an
inevitable compression of local vasculatures, which will change their distending
behavior and lead to inaccurate recordings (Supplementary Fig. 4). Therefore, this
method is not suitable for long-term monitoring purposes.Wearable devices with the compliant mechanical properties similar to the skin
offer the capability of non-invasive, continuous monitoring of a variety of vital
signs[26], such as local field
potentials[27],
temperature[28], sweat
content[29,30], and skin hydration[31]. However, their applications have typically been
limited to recording signals on the skin or shallow tissue underneath the epidermis.
Here, we introduce an approach that allows integrating ultrasonic technology on the
wearable system. The ultrasonic waves can effectively penetrate the human tissues up to
4 cm, which opens up a third dimension to the sensing range of current state-of-the-art
wearable electronics. With similar mechanical properties to the skin and an ultrathin
profile, the wearable ultrasonic device can ensure a conformal intimate contact to the
curvilinear and time-dynamic skin surface, and continuously monitor CBP of deep
vasculatures without the operational difficulties or instabilities in the other
conventional approaches. This capability of non-invasive, continuous, and accurate
monitoring of deep biological tissues/organs opens up opportunities for diagnosing and
predicting a broad range of cardiovascular diseases in a wearable format.
Results:
Device design and working principle.
The device hybridizes high performance rigid 1–3 piezoelectric
composites with soft structural components (Fig.
1a and Supplementary Figs. 5 and 6). The anisotropic 1–3 composite
possesses better acoustic coupling with soft biological tissue than isotropic
piezoelectric materials. By balancing geometrical and electrical designs, our
device can reach an ultrathin thickness (240 μm, three orders of
magnitude thinner than existing medical ultrasonic probes) (Supplementary note 5). The elastic
and failure strain levels are up to 30% and 60%, respectively (Supplementary Figs. 7 and 8). The
functional material used in this study, 1–3 piezoelectric composite with
a thickness of 200 μm, has a working frequency of 7.5 MHz (Supplementary Fig. 9).
This material selection enables a 400 μm axial resolution (Methods
section and Supplementary Fig.
10) that is comparable with available medical ultrasonic probes at
the same working frequency. The 1–3 composite has piezoelectric
micro-rods embedded, in a periodic configuration in a passive epoxy matrix,
which substantially increases the longitudinal coupling coefficient
k33 by suppressing the shear vibrating modes. The rigid
piezoelectric transducer element is diced to have 0.9*0.9 mm2
footprint to allow sufficient penetration depth into the tissue, and also has
minimal mechanical loading to the entire device (Fig. 1b).
Figure | 1
Design and working principle of the stretchable ultrasonic device.
a, Schematics of the stretchable ultrasonic device, where key
components are labeled. The high-performance 1–3 composite with periodic
piezoelectric rods embedded in an epoxy matrix suppresses shear vibration modes
and enhances longitudinal ultrasonic penetration into the skin. The vertical
interconnect access (VIA) is used connect the top and bottom electrodes,
allowing the co-planar anisotropic conductive film (ACF) bonding to the
electrodes to enhance the robustness of the device. When mounted on the human
neck, the device enables monitoring of CBP by capturing the pulsating vessel
diameter of carotid artery, internal jugular vein (int jugular), and external
jugular vein (ext jugular) using the pulse-echo method, as illustrated as the
bottom left graph. The device can locate the dynamic anterior (ant) and
posterior (post) walls of the vessel using high-directivity ultrasonic beam, as
bottom middle graph shows. The corresponding shifting echo radio frequency
signals reflected from the anterior and posterior walls appear in the bottom
right. b, The device conforming to complex surfaces and under mixed
modes of stretching and twisting, demonstrating the mechanical compliance and
robustness of the device. The large contact angle of the water droplet on the
device in the left panel shows the hydrophobic properties of silicone
encapsulation materials that can be used as a barrier to moisture/sweat.
Bilayer stacking of polyimide (PI, 4 μm)/Cu (20 μm) (Fig. 1a top left) is utilized to fabricate
the stretchable electrodes that interconnect a 4×5 array of transducers
in the device (Detailed fabrication process appears in Methods). The transducers
can be individually addressed by 20 stimulating electrodes on the top and a
common ground at the bottom. The array design aims to map the vessels’
positions, thus enabling sensing and monitoring using the exact transducer
overlaying above the targeted vessel without tedious manual positioning (Supplementary note 6;
Supplementary Fig.
11). The top simulating electrodes and the bottom ground are routed
to the same plane by a vertical-interconnect-access (VIA) for optimized
mechanical robustness and ease of electrical bonding (Fig. 1a, Supplementary Fig. 12, and Supplementary note
7).The working principle is illustrated in Fig. 1a bottom. Technically, the device can continuously record the
pulsating blood vessel diameter, which will be translated into localized BP
waveforms[32]. Next, the
BP waveform can be calculated as: where is the diastolic pressure, which will be
acquired on the brachial artery using a BP cuff, and is the diastolic arterial cross-section and
is the vessel rigidity coefficient. Assuming
that artery is rotationally symmetrical, can be calculated as: where is the diameter waveform of the target artery.
Detailed working principle, resolution, calibration, and validation of our
device appear in Methods section, Supplementary notes 8 and 9. When
the device is softly laminated on the skin (Supplementary Fig. 13), each
transducer can be individually activated and controlled with a power consumption
of 23.6 mW. When the ultrasonic wave reaches interfaces, both transmission and
reflection will occur. The transmission wave with reduced intensity allows
penetrating into deeper layers of tissues. The reflection wave that carries
critical location information of the interfaces (e.g., the anterior and
posterior walls) can be sensed by the same transducer[33]. The vessel diameter measurement results
are validated by clinical ultrasonography with excellent correspondence (99.7%,
Supplementary Fig.
14). At a high pulse repetitive frequency (2000 Hz), the time of
flight (TOF) signals corresponding to the pulsating anterior and posterior walls
can be accurately recorded by an oscilloscope with 2 GHz sampling frequency,
which will appear as separate and shifting peaks in the amplitude mode (Fig. 1a right bottom). The device can capture
the pulsating blood vessel diameter dynamically with high spatial resolution
(axial resolution 0.77 μm) and temporal resolution (500 μs).The entire device is encapsulated by a silicone elastomer whose modulus
is on par with that of the human skin. The elastomer is only 15 μm thick
to balance a trade-off between mechanical robustness and sufficient acoustic
emission performance (Supplementary Fig. 15, Supplementary note 5). Hydrophobic
nature of the silicone elastomer provides a barrier for moisture, which protects
the device from possible sweat corrosion (Fig.
1b). Due to its soft mechanics, the as-fabricated ultrasound patch
allows conforming to both developable (Fig.
1b left) and non-developable surfaces (Fig. 1b middle). Also, the device is robust and can endure twisting
and stretching (Fig. 1b right), showing its
high potential for skin integration purposes.
Device characterization.
The piezoelectric transducer converts electrical potential between the
top and bottom electrodes to mechanical vibrations, and vice versa. The
efficiency of this process is exhibited by the measured impedance and phase
angle spectra (Fig. 2a), showing excellent
piezoelectricity whose k33 was measured to be 0.81, much higher than
that of bulk PZT (~0.58)[33] due
to its anisotropic high aspect ratio rod configuration compared with the
isotropic bulk PZT[34]. The
transducer performance is evaluated on the wrist ulnar artery of a healthy male.
The echo signal appears in Fig. 2b, in
which the TOF of the two peaks corresponds to the positions of the anterior and
posterior walls of the ulnar artery, respectively. Signal analysis in the time
and frequency domains of the posterior wall is illustrated in Fig. 2c, demonstrating that the material has a central
frequency of 7.5 MHz and possesses good sensitivity of 32% at −6 dB
bandwidth (Methods section) and a peak to peak voltage of 100 mV. Prediction of
the beam pattern (Matlab R2016b, TAC_GUI toolbox) of our stretchable ultrasonic
device appears in Fig. 2d. The results show
that in the longitudinal direction, the conformal probe has excellent beam
directivity and sufficient penetration for deep tissue detection, reaching a
penetration depth of up to 40 mm (with the piezoelectric transducer size of
0.9*0.9 mm2). The larger the piezoelectric material size, the deeper
the ultrasonic wave can penetrate (Supplementary Fig. 16). The
1–3 composite has a low acoustic impedance (17 MRayl), which provides
excellent acoustic coupling with the human skin. Additionally, the bottom
circular electrode diameter is designed to be 0.6 mm to balance the practical
bonding robustness and impedance matching (Supplementary Fig. 17).
Figure | 2
Electrical, mechanical, acoustical, and biocompatibility
characterizations of the conformal ultrasonic device. a, Impedance
and phase angle spectra of the 1–3 composite, showing the excellent
piezoelectricity. The resonant and anti-resonant frequency ranges are labeled in
shaded circles. The left inset is the equivalent RLC circuit diagram of the
piezoelectric transducer. At the resonant frequency, the impedance of the
equivalent circuit is at the minimum, which will be the most power efficient. At
the anti-resonant frequency, the impedance of the equivalent circuit is at the
maximum, and the transducer will have the largest damping. b,
Ultrasonic receiving signals on the ulnar artery, with two clear echo peaks from
the anterior (ant) and posterior (post) vessel walls. Inset is a schematic
diagram representing the transducer and the ulnar artery to show echo peaks
aligned with the ant and post vessel walls. Tx represents the transducer.
c, Time and frequency domain characterization of the signal in
b (post-wall peak), showing excellent signal quality and
bandwidth (dashed line), indicating the high sensitivity of the transducer.
d, Simulated acoustic emission profile of a piezoelectric
material size of 0.9 × 0.9 mm2 (inset) with excellent beam
directivity and penetration depth (>4 cm). e, Bi-axial
tensile test of the device with stretchability up to 60% in the x-direction and
50% in y-direction without fracture. The zoomed-in image of the dashed box shows
the slight plastic deformation when biaxial strain is larger than 30% in the
x-direction. f, Fluorescent images of the fibroblast cells before
(left panel) and after 16 hours (right panel) continuous exposure to the
ultrasound generated by the conformal ultrasonic device. The 100% survival rate
of the cells prove the excellent bio-compatibility of the conformal ultrasonic
device.
The elastomeric matrix with iterative stretchable circuit designs and
ultrathin encapsulation assemblies provides exceptionally conformal contacts to
the human skin under various deformation modes (Supplementary Fig. 18). The device
can be reversibly stretched up to 30% in the x-direction and 25% in the
y-direction. The maximum stretchability can reach up to 60% in the x-direction
and 50% in the y-direction (Fig. 2e). These
mechanical characteristics enable robust and seamless contact with the skin
(Supplementary note
10), given the fact that the human skin typically exhibits a linear
elastic response to tensile strain <20%[35]. The electrical performance of the
device remains stable under stretching and moist environment (Supplementary Fig. 19). Cell
viability testing under controlled ultrasound intensity is shown in Fig. 2f and Supplementary Fig. 20. The
fibroblast cells (HFF-1) are cultured under the ultrasonic wave emission from
our conformal probe with a 100% survival rate after 16 hours of continuous
exposure, showing excellent biocompatibility of our device (Detailed cell
information appears in the Methods section).
Performance validation.
A conformal and intimate contact between the device and the human skin
is paramount for the robust performance of the device. Continuous measurements
on the radial artery using our device and a commercial tonometer, the
noninvasive gold standard for BP waveform measurements, when the wrist is bent
to different angles are shown in Fig. 3a.
In this scenario, the tonometer needs to be held tightly by the operator to
remain stable on the wrist, resulting in not only great pressure (~100 Pa) to
skin, which causes severe irritation (Fig.
3b), but also significant waveform distortion and erroneous readings
(Fig. 3c). However, our device
self-adheres to the skin and adds minimal pressure (~5 Pa) to the skin due to
its ultra-lightweight (0.15 g) and skin-like modulus. This unique feature
enables the device to not only monitor in long-term without any discomfort but
also maintain relatively stable and continuous recording even during motion. The
correlation curves of these two devices are shown in Fig. 3d. Most importantly, our device has smaller
relative measurement uncertainty (1%), higher measurement precision (within 2
mmHg), and higher accuracy (grade A) than the commercial equipment (Supplementary note 11,
Methods section, and Supplementary Figs. 21 and 22). Furthermore, the tonometer is highly
operator-dependent, which is reflected by the fact that either tiny offset from
the central arterial axis or moderate holding forces of the tonometer probe will
introduce tremendous recording error of the BP waveform (Supplementary Figs. 23 and 24). Our
conformal ultrasonic array with the ultra-lightweight and vessel positioning
capability demonstrated above exhibits substantial advantages over the
applanation tonometry.
Figure | 3
Comparison between the conformal ultrasound (US) sensor and the
commercial tonometer, exercise hemodynamics monitoring, and central arterial and
venous pulse measurements. a, Continuous measurements of radial
pulse waveforms by both US sensor (top) and the commercial tonometer (bottom)
under the same condition. b, Comparison of applied pressure levels
to the skin during the measurement. Inset figures show the skin irritation
brought by the tonometer and the conformal US sensor. Dots represent all data
points. Error bars represent ± s.d. (N=4). c, Comparison of
the BP waveforms measured continuously when wrist is bent at a rate of
~15°/s, showing the robust performance of the conformal device. Different
postures from 0° to 30° are labeled in different shades.
d, Autocorrelation of the waveforms in c, showing
the conformal US device can maintain stable measurements in motion.
e, Pulse waveforms on the radial artery before (bf) and after
(aft) exercise, showing the changes in absolute pressure values and waveform
morphologies. f, Pulse waveforms averaged from 10 continuous
periods and normalized to the same diastolic and systolic pressure values to
demonstrate the change in morphologies caused by vasodilation. g,
Illustration of the US sensor measurement locations marked with arrows: the left
carotid artery (CA), external jugular vein (Ext JV) and internal jugular vein
(Int JV). The right jugular vein and carotid artery are also highlighted.
h, A typical pulse waveform measured from the carotid artery,
directly correlated to the left atrial and ventricular events. Different phases
and characteristic morphologies are marked. i, A typical pulse
waveform from the internal jugular vein, directly correlated to right atrial and
ventricular activities. Different phases and characteristic morphologies are
marked.
More interestingly, the conformal ultrasound probe enables a gel-free
working mode. Traditional ultrasonic transducer highly relies on ultrasound gel
to eliminate interfacial air gaps between the probe and the skin for good
acoustic coupling. The gel is unpleasantly cold and has to be frequently
reapplied to prevent it from drying out. In this device, we add a 15 μm
thick layer of silicone as the acoustic coupling layer. The silicone is sticky
and ultra-low modulus (~5 kPa) to ensure intimate contact with the skin without
applying any gel. The quality of the acquired signals and waveforms is
comparable to those acquired with the gel (Supplementary Fig. 25 and Methods
section). Moisture/human sweat that significantly influences the PPG
measurements does not affect the performance of the conformal ultrasound device
(Supplementary Fig.
26). The excellent design enables the device with remarkable
durability, allowing highly reproducible testing results after four weeks (Supplementary Fig.
27).
Dynamic hemodynamics monitoring.
Due to its excellent mechanical compliance and lightweight, our device
can maintain intimate and stable contact with the human skin both mechanically
and acoustically in different body postures with pure van der Waals force (Supplementary Fig. 28).
During exercise, the muscles require more delivery of nutrients and oxygen, so
the cardiac output increases to meet the need[36]. On the one hand, human vessels dilate
to increase the delivery, so the vascular resistance and reflection are reduced.
On the other hand, the heart rate and systolic strength increase to boost up the
cardiac output. The measured heart rates on radial artery during resting
(~75/min) and immediately after exercise (~112/min) are shown in Fig. 3e. The BP waveform has a higher systolic peak
due to the stronger ventricular systole to get more substantial cardiac
output[37]. The averaged
waveform morphology changes before and after intense exercise appear in Fig. 3f (normalized to same systolic and
diastolic pressure values), showing a steep drop of the systolic peak after
physical training due to the vasodilation induced vascular resistance decrease.
It is worth noting that we need to calibrate our device before and after
exercise that will significantly change the diastolic pressure. However, we can
observe the vessel dilation and vascular resistance decrease by comparing the
normalized waveforms, regardless of what the absolute BP values are.
CBP recording.
From the pathophysiological perspective, the CBP waveform is a crucial
and the most direct predictor of main cardiovascular events[5]. Monitoring those events in a continuous
and long-term manner can lead to remarkable advancements in cardiovascular
disease diagnosis and prevention[6], which are yet to be realized by existing medical tools
(Supplementary note
12). We demonstrate highly accurate direct measurements of the
central vasculatures pulsating behavior in deep tissues, including the carotid
artery and internal and external jugular veins. Schematic illustration of the
measurement configuration appears in Fig.
3g. The carotid artery (CA, ~25 mm underneath the skin with slight
individual variations[38], near
the central aorta) carries a significant amount of blood from the left ventricle
(LV) and left atrium (LA) to the rest of the body. A typical period of the
carotid artery BP waveform measured by our device shows a clear systolic peak
and dicrotic notch (Fig. 3h). The former
indicates the ventricular systole and the latter suggests the closure of the
aortic valve. The detailed CBP measurement calibration appears in Supplementary note 9. The
corresponding validation using the commercial tonometer exhibits remarkable
correspondence (Supplementary
Fig. 29). The blood flow sequence in the central cardiovascular
system and direct relationship between the central vessels and the heart appears
in Supplementary Fig.
30. Detailed clinical meaning of arterial BP waveforms is illustrated
in Supplementary Fig.
31 and discussed in Supplementary note 13.The internal jugular vein, carrying the venous blood to the right atrium
and right ventricle and finally to the lung, reflects the right heart
activities. A typical jugular venous pressure waveform measured by our device
appears in Fig. 3i. The normalized pressure
waveform is obtained by the volume assessment method[39]. Associated algorithms and equations of
this method are discussed in details in Supplementary note 9. The jugular
venous waveform comprises three characteristic peaks: A (Atrial contraction), C
(Tricuspid bulging, ventricular contraction), and V (Systolic filling of the
atrium); and two descents: X (Atrial relaxation) and Y (Early ventricular
filling). Those components correspond to various events during each cardiac
cycle. The jugular venous waveform measured by clinical color Doppler imaging
machine on the same subject appears in Supplementary Fig. 32, showing the
corresponding A, C, and X peaks and V and Y descents. The jugular venous
distension (JVD), seen as a vessel bulging on the neck created by deep
exhalation of the subject (Supplementary Fig. 33), can predict right-side heart
failure[40]. More
detailed discussions are included in Supplementary note 13.
BP waveform monitoring from central to peripheral.
Due to the amplification effect, namely the progressive vascular
resistance, stiffness, and impedance mismatch between central and peripheral
vessels, the arterial pressure waveform varies from central to
peripheral[6]. Although
the diastolic and mean arterial pressures are relatively constant, systolic
pressure can be up to 40 mmHg higher in the peripheral than the central
artery[41]. This
amplification effect on the various part of the body contains abundant
information related to age, gender, height, heart rate, and systematic diseases
affecting the vasculature[42].
Those data, if carefully collected, can be critical for improving the efficacy
of diagnosis and prognosis of cardiovascular diseases[43]. However, in current clinical settings,
those valuable data and signals can only be obtained by the professional
clinician in a considerably infrequent manner. The challenges for existing
approaches are discussed in Supplementary note 12.Our device allows observing this intriguing phenomenon. The
amplification effect will increase as we move from the large and highly elastic
central arteries (e.g., the carotid) to the small and stiff peripheral artery
(e.g., the radial and dorsalis pedis) (Fig.
4, the first row and Supplementary Fig. 34). This
phenomenon is due to the backward propagation of pulse waves generated at the
arterioles. At the central cite, those reflected pulses need to travel a long
distance so they adds to the CBP waveform in a misaligned manner. On the
contrary, at the peripheral site, they travel a much shorter distance so they
instantly contribute to the peripheral BP waveform in a way that is
time-aligned. Therefore, the more peripheral the artery is, the higher the
amplification/augmentation will be (Fig. 4,
the second row). To validate this amplification effect, we use a commercial
tonometer to test on the same subject. Remarkable corresponding results appear
in the third row of Fig. 4. An exciting
feature brought by this direct diameter measurement approach is that it allows
observing this amplification phenomenon and deriving accurate pulse pressure
values from multiple body parts simultaneously by the diastolic pressure at only
the brachial site, with no site mismatch (Supplementary note 14 and Supplementary Fig. 35).
The upstroke gradient increases as a result of the pressure amplification (Supplementary Fig. 36 and
Supplementary note
15).
Figure | 4
BP measurements from the central to peripheral arteries and validation
using the commercial tonometer. Measurement positions (top row), collected
arterial pressure waveforms (middle row), and the BP waveform of one period
compared with results from the tonometer (bottom row). From the 1st
column to 4th column are carotid artery (CA), brachial artery (BA),
radial artery (RA), and pedal artery (Dorsalis pedis), respectively, showing an
increase of the amplification effect by progressive vascular resistance, a
longer interval time between systolic peak and dicrotic notch (illustrated by
the grey area), a higher systolic pressure value, and a higher upstroke gradient
(the slope of the BP waveform at the beginning of the upstroke).
Another notable feature of the BP waveforms is the progressive time
interval between the systolic peak and the dicrotic notch, which is illustrated
by the grey area. When the pressure wave travels down from the central arteries,
its magnitude increases due to impedance mismatches encountered on the way,
which creates a reflected wave that travels back to the heart during late
systole and early diastole. This reflection wave takes a longer time to travel
from a location that is more distant from the heart, and thus contributes to an
increase in the time interval between the systolic peak and the dicrotic notch
(gray areas in the waveforms in Fig. 4, the
third row)[44]. The capability
of capturing those systemic variations demonstrates the device’s
potential for accurate clinically relevant diagnosis.
Electrocardiogram (ECG) correlation for arterial stiffness
calculation.
Characteristics of the arterial pulse propagation possess a strong
relationship with vascular stiffness, which is one of the key determinants of
cardiovascular risks[45]. Among
all vascular parameters, pulse wave velocity (PWV), is the most accessible and
reliable way to evaluate the arterial stiffness (Supplementary note 16) and can be
calculated as follows[46]:
where D is the distance between the ECG sensor
and the ultrasonic sensor. The measurement schematics are shown in Fig. 5a, with the simultaneous measurements
of ECG and pulse arrival times (PATs) at three different sites including
brachial, radial, and dorsalis pedis. Figures
5b-d present the ECG correlation results: Case 1, brachial artery
(Fig. 5b); Case 2, radial artery (Fig. 5c); and Case 3, pedal artery (Fig. 5d), where ECG is measured on the chest
for all three cases (Supplementary Fig. 37). As seen in Figs. 5b-d, PWV in Case I is 5.4 m/s, where D is 54
cm and time difference is 100 ms; PWV in Case II is 5.8 m/s, where
D is 104 cm and time difference is 180 ms; PWV in Case III
is 5.3 m/s, where the D is 159 cm and the time difference is
300 ms. The PWV measurements are highly reproducible (Supplementary Fig. 38) and have
been validated by the commercial equipment, shown in Fig. 5e and Supplementary Fig. 39 (Detailed
testing condition and measurement methods appears in the Methods section).
Figure | 5
ECG correlation based PWV calculation to evaluate the arterial
stiffness. a, Illustration of the measurement positions of the
three cases from the central to the peripheral arteries, with simultaneous ECG
measurements. BP waveforms are normalized based on the systolic pressure.
b, Case 1: simultaneous measurement of ECG and BP at the BA.
c, Case 2: simultaneous measurement of ECG and BP at the RA.
d, Case 3: simultaneous measurement of ECG and BP at the pedal
artery (Dorsalis pedis). The graphs on the right panel of b,
c, and d show the PAT, as indicated by the grey
area, for each case. e, PAT results compared between the conformal
device and the tonometer. Dots represent all data points. Error bars represent
± s.d. (N=10).
Discussion
He have shown a new class of conformal and stretchable ultrasonic devices
that offer noninvasive, accurate, and continuous monitoring of vital signs from well
below the human skin, adding a new dimension to the sensing range of conventional
stretchable electronics. This device exploits strategic material integration and
advanced microfabrication techniques to achieve both state-of-the-art functions and
suitable mechanical compliance that allows intimate coupling with the human skin.
This device can be used to capture a series of key features in the central blood
vessels with reliable performance, holding strong clinical implications.Device performance and functionality could be improved. The measurement of
absolute pressure using PAT is feasible by ECG correlation, bypassing the need for
re-calibration with the BP cuff (Supplementary note 16). Furthermore, ultrasonic imaging on the human
body could provide the accurate cross-sectional area of the vessel for calculating
the BP waveform, which is especially valuable for the CVP measurement on the
irregularly shaped jugular vein. Additionally, integrating the post-end functions
such as the electronic control, signal processing, waveform pattern recognition,
wireless communications, and power sources in a stretchable and lightweight format
would significantly enhance device wearability. To reduce the system-level power
consumption, a lower sampling rate analog-to-digital converter could be used to
decrease the power budget of data transmission. And a phased-array control algorithm
could be implemented to allow the focusing and steering of the ultrasonic beam
inside the human body.
Methods
Fabrication of the stretchable ultrasonic device.
The fabrication can be summarized into three parts: (1) stretchable
circuit patterning; (2) transfer printing; and (3) soft elastomeric packaging
(Supplementary Fig.
40). First, a Cu foil (20 μm thick, MicroThin, Oak-Mitsui
Inc.) was coated with PI from poly(pyromellitic dianhydride-co-4,
40-oxydianiline) amic acid solution. This process was performed at 4000 rpm for
60 s. And then the foil was soft baked on a hotplate at 110 °C for three
minutes and 150 °C for one min, and then cured in a nitrogen oven at 300
°C for 1 h. A glass slide coated with a layer of polydimethylsiloxane
(PDMS, Sylgard 184® silicone elastomer, 20:1) served as the
substrate to laminate the Cu foil with the PI layer in contact with the PDMS. UV
ozone surface activation for 3 min was used to increased the bonding between the
PI and PDMS. Then, a laser ablation system (Supplementary Fig. 41), 0.342 mJ
power, 900 kHz pulse repetition frequency, 300 mm/s laser cutting speed, and 241
ns pulse width, was utilized to create the circuit pattern with the highest
resolution (Supplementary Fig.
42). After that, using water-soluble tape (Aquasol), the circuit was
transferred on a 15 μm thick Eco-flex (0030®,
Smooth-On) substrate spin-coated on a glass slide (Supplementary Fig. 43). After
removing the water soluble tape, the circuit surface was cleaned using flux to
remove the surface oxidation (Supplementary Fig. 44) created during the laser ablation process to
increase the welding strength. The welding with the top and bottom electrodes
was achieved with the solder paste at 150 °C for 5 min. After that, the
device was encapsulated with Eco-flex. Curing was done at room temperature for 2
hours, and then the glass slides were peeled off. Finally, spin-coating an
additional layer of silbione on the Ecoflex substrate facilitates removing the
interfacial gap and thus the necessity of the gel during testing.
Poling of the 1–3 composite.
Poling the 1–3 composite (Smart Material Corp.) increased its
piezoelectric coefficient and electromechanical coupling factor of the
composite[47]. The
poling involved using an electric field to align the dipoles of the
piezoelectric materials, which enhances the piezoelectricity and performance of
the device[48]. The polarizing
hysteresis loop (Supplementary
Fig. 45) was measured in silicone oil. The poling of the device was
implemented at d.c. 1.2 kV/cm for 15 min. An excessive poling electric field
caused breakdown of the piezoelectric materials, thus reducing the signal
strength (Supplementary Fig.
46).
Measurement and data analysis of the BP waveforms.
The BP waveform measurement was carried out on a healthy male of age 22,
under the approval by the Institutional Review Board of the University of
California, San Diego (IRB no. 170812). Written informed consent was obtained
from all the human subjects. All measurements were carried out on the subject
when sitting. The measurement setup is illustrated in Supplementary Fig. 47. A layer of
Silbione® was applied at the bottom of the device surface
to enable gel-free measurements. The signal analysis was based on the TOF
analysis, which was the gauge of the time interval between the signal peak and
zero-point time. The TOF was used to calculate the propagation distance, by
multiplying the speed of ultrasound in the specimen. The device was activated by
an ultrasonic pulser (Olympus 5077 PR) at 100 V, using the transmit/receive
mode. The pulse repetitive frequency was 2000 Hz. The echo signal was received
by a Picoscope (Picoscope 6404) with a temporal resolution of 500 μs,
which allowed precise vessel wall tracking (Supplementary note 8). Discussions
of measurement principle, resolutions, uncertainty, accuracy, and precision are
illustrated in Supplementary
notes 8 and 11 and Supplementary Figs. 22, and 48–50. The circuit enabling
simultaneous measurement of pulse pressure in varies site is illustrated in
Supplementary note
14 and Supplementary Fig. 51. The clinical color Doppler machine (Mindray
DC 7) was used to confirm the characteristic peaks in the venous waveform. The
BP waveform measurement results were validated by a SphygmoCor
EM3® tonometer.
ECG correlation.
The ECG correlation to the BP waveforms at different locations was taken
on the same subject, when sitting, consecutively with a two-minute interval, to
guarantee the subject had a relatively constant BP value and arterial stiffness.
The longest duration of skin integration on the same skin region was two hours.
No allergic reactions, redness, or damage to the skin were observed in any of
our studies. The diastolic pressure was calibrated using a commercial BP cuff
(Smart Logic Technology, 6016) before each monitoring period. Additionally, the
testing subject maintained a stable physiological and psychological status to
guarantee stable levels of BP and vasculature stiffness. A detailed discussion
of PAT, pulse transit time (PTT), and PWV is illustrated in Supplementary Fig. 52 and Supplementary note
16.
Transducer selection, bandwidth, and resolution characterization.
In the experiments, we chose piezoelectric ultrasound transducers (PUTs)
instead of capacitive micromachined ultrasound transducers (cMUTs) or
piezoelectric micromachined ultrasound transducers (pMUTs), because of the low
cost and ease of fabrication of PUTs. More detailed considerations are presented
in Supplementary note 5
and Supplementary Fig.
53. The transducer bandwidth (32%) was calculated by using the
frequency range (2.4 MHz) divided by central frequency (7.5 MHz). The axial
resolution characterization exploited a thin metal wire suspended at the center
of a beaker filled with water. A 1×10 linear array of transducers was
fabricated and attached to the beaker wall parallel to the ground. All signals
of the ten transducer elements were acquired and combined with one transducer as
the transmitter and the other as the receiver (e.g., 1T2R, 1T3R…1T10R;
2T1R, 2T3R…2T10R; …; 10T1R, 10T2R….10T9R). The total 90
signals were used, by the delay-and-sum (DAS) reconstruction algorithm, to
reconstruct the image (Supplementary Fig. 10c). The obtained image with low-level side
lobes resulted from the reconstruction algorithm and the limited number of
elements used for imaging.
Cell viability assay under ultrasound exposure.
The human skin fibroblast cells (HFF-1) were first purchased from
American Type Culture Collection (ATCC) and cultured in Dulbecco’s
Modified Eagle’s Medium (DMEM, Gibco) supplemented with 10% fetal bovine
serum (FBS, Gibco) and 1% penicillin/streptomycin (Gibco) under 37 °C
within 5% CO2. Then the HFF-1 cells were subcultured and seeded into
24-well plate at the density of 1 × 104/ml and incubated for
another 24 hours. No antibodies were used in the experiment. Then a beam of
ultrasound at the frequency of 7.5 MHz was applied to the bottom of the culture
plate. After 2, 6, and 16 hours’ ultrasound exposure, the cells were
stained with calcein AM (Invitrogen, 3M, excitation/emission = 488 nm/525 nm)
and propidium iodide (PI, Invitrogen, 3M, excitation/emission = 530 nm/620 nm)
for 15 min, and then were imaged under fluorescence microscopy (EVOS,
Thermofisher Scientifics). For the positive control group, the cells were
treated with 75% ethanol for 10 min.
Authors: Juliane R Sempionatto; Muyang Lin; Lu Yin; Ernesto De la Paz; Kexin Pei; Thitaporn Sonsa-Ard; Andre N de Loyola Silva; Ahmed A Khorshed; Fangyu Zhang; Nicholas Tostado; Sheng Xu; Joseph Wang Journal: Nat Biomed Eng Date: 2021-02-15 Impact factor: 25.671
Authors: Sheng Shu; Jie An; Pengfei Chen; Di Liu; Ziming Wang; Chengyu Li; Shuangzhe Zhang; Yuan Liu; Jianzhe Luo; Lulu Zu; Wei Tang; Zhong Lin Wang Journal: Research (Wash D C) Date: 2021-06-21