| Literature DB >> 30855243 |
Philip Heimann1, Alexander Reid1, Yiping Feng1, David Fritz1.
Abstract
For LCLS-II, a fluorescence intensity monitor for the non-invasive, pulse-by-pulse normalization of experiments has been developed. A prototype diagnostic was constructed with a microchannel plate assembly and two photodiodes. The diagnostic was then installed in the LCLS SXR instrument Kirkpatrick-Baez mirror chamber with the detectors located above the vertically reflecting mirror. The linearity, noise and position sensitivity of the detectors have been characterized. The photodiode responsivity is suitable for high pulse energies. The microchannel plate detector shows sufficient responsivity over a wide range of pulse energies. The relative signal from the two photodiodes provides a sensitive measure of the X-ray beam position. The fluorescence intensity monitor provides intensity normalization while being compatible with high incident power, a 0.93 MHz repetition rate and ultra-high vacuum. open access.Entities:
Keywords: X-ray free-electron laser; X-ray intensity monitor; diagnostics
Mesh:
Year: 2019 PMID: 30855243 PMCID: PMC6412172 DOI: 10.1107/S1600577519001802
Source DB: PubMed Journal: J Synchrotron Radiat ISSN: 0909-0495 Impact factor: 2.616
Figure 1(a) Photograph of the assembled FIM prototype. (b) Photograph of the FIM prototype installed above the LCLS SXR vertical KB mirror.
Figure 2The responsivity of photodiodes 1 and 2 compared with the X-ray pulse energy calculated from the LCLS upstream gas detector and gas attenuator at a photon energy of 1540 eV. The solid lines represent linear fits: y = c x. The colors correspond to individual runs, during which the gas attenuation is increased by a factor of ten.
Figure 3The responsivity of the MCP detector compared with the X-ray pulse energy calculated from the LCLS upstream gas detector and gas attenuator. The solid lines represent linear fits: y = cx. (a) At a photon energy of 1540 eV, a negative bias was applied with the MCP input voltage varied from −800 V to −1500 V. (b) At a photon energy of 1480 eV, a positive bias was applied with the anode voltage varied from 1100 V to 1550 V.
Evaluation of the noise of the photodiodes and MCP detector
| Pulse energy (J) | Photodiode 1 (σ/ | Photodiode 2 (σ/ | MCP (σ/ |
|---|---|---|---|
| 8.0 × 10−4 | 0.044 | 0.055 | |
| 1.2 × 10−4 | 0.058 | ||
| 1.5 × 10−5 | 0.024 | ||
| 1.1 × 10−6 | 0.027 | ||
| 1.2 × 10−7 | 0.028 |
Figure 4(a) At 800 eV photon energy, the intensities of photodiodes 1 and 2 and the MCP detector as the KB chamber is translated vertically. (b) Derived from the photodiode intensities in Fig. 4 ▸(a), f(y) from equation (2) is shown. The error bars are calculated from the σ/I values in Table 1 ▸ at 8.0 × 10−4 J pulse energy.