| Literature DB >> 21974570 |
Philip Heimann1, Oleg Krupin, William F Schlotter, Joshua Turner, Jacek Krzywinski, Florian Sorgenfrei, Marc Messerschmidt, David Bernstein, Jaromir Chalupský, Vera Hájková, Stefan Hau-Riege, Michael Holmes, Libor Juha, Nicholas Kelez, Jan Lüning, Dennis Nordlund, Monica Fernandez Perea, Andreas Scherz, Regina Soufli, Wilfried Wurth, Michael Rowen.
Abstract
We present the x-ray optical design of the soft x-ray materials science instrument at the Linac Coherent Light Source, consisting of a varied line-spaced grating monochromator and Kirkpatrick-Baez refocusing optics. Results from the commissioning of the monochromator are shown. A resolving power of 3000 was achieved, which is within a factor of two of the design goal.Year: 2011 PMID: 21974570 DOI: 10.1063/1.3633947
Source DB: PubMed Journal: Rev Sci Instrum ISSN: 0034-6748 Impact factor: 1.523