Sleep Parameter Assessment Accuracy of a Consumer Home Sleep Monitoring Ballistocardiograph Beddit Sleep Tracker: A Validation Study.

Jarno Tuominen1, Karoliina Peltola1, Tarja Saaresranta2,3, Katja Valli1,4.   

Abstract

STUDY
OBJECTIVES: Growing interest in monitoring sleep and well-being has created a market for consumer home sleep monitoring devices. Additionally, sleep disorder diagnostics, and sleep and dream research would benefit from reliable and valid home sleep monitoring devices. Yet, majority of currently available home sleep monitoring devices lack validation. In this study, the sleep parameter assessment accuracy of Beddit Sleep Tracker (BST), an unobtrusive and non-wearable sleep monitoring device based on ballistocardiography, was evaluated by comparing it with polysomnography (PSG) measures. We measured total sleep time (TST), sleep onset latency (SOL), wake after sleep onset (WASO), and sleep efficiency (SE). Additionally, we examined whether BST can differentiate sleep stages.
METHODS: We performed sleep studies simultaneously with PSG and BST in ten healthy young adults (5 female/5 male) during two non-consecutive nights in a sleep laboratory.
RESULTS: BST was able to distinguish SOL with some accuracy. However, it underestimated WASO and thus overestimated TST and SE. Also, it failed to discriminate between non-rapid eye movement sleep stages and did not detect the rapid eye movement sleep stage.
CONCLUSIONS: These findings indicate that BST is not a valid device to monitor sleep. Consumers should be careful in interpreting the conclusions on sleep quality and efficiency provided by the device.
© 2019 American Academy of Sleep Medicine.

Entities:  

Keywords:  Beddit Sleep Tracker; polysomnography; sleep parameter accuracy; sleep research methods; validation

Mesh:

Year:  2019        PMID: 30853052      PMCID: PMC6411173          DOI: 10.5664/jcsm.7682

Source DB:  PubMed          Journal:  J Clin Sleep Med        ISSN: 1550-9389            Impact factor:   4.062


  21 in total

1.  Detecting REM sleep from the finger: an automatic REM sleep algorithm based on peripheral arterial tone (PAT) and actigraphy.

Authors:  Sarah Herscovici; Avivit Pe'er; Surik Papyan; Peretz Lavie
Journal:  Physiol Meas       Date:  2006-12-12       Impact factor: 2.833

2.  Direct comparison of two actigraphy devices with polysomnographically recorded naps in healthy young adults.

Authors:  Nicola Cellini; Matthew P Buman; Elizabeth A McDevitt; Ashley A Ricker; Sara C Mednick
Journal:  Chronobiol Int       Date:  2013-05-30       Impact factor: 2.877

3.  Validation of an automated wireless system to monitor sleep in healthy adults.

Authors:  John R Shambroom; Stephan E Fábregas; Jack Johnstone
Journal:  J Sleep Res       Date:  2011-08-22       Impact factor: 3.981

Review 4.  Consumer Sleep Technologies: A Review of the Landscape.

Authors:  Ping-Ru T Ko; Julie A Kientz; Eun Kyoung Choe; Matthew Kay; Carol A Landis; Nathaniel F Watson
Journal:  J Clin Sleep Med       Date:  2015-12-15       Impact factor: 4.062

5.  Unobtrusive online monitoring of sleep at home.

Authors:  Joonas Paalasmaa; Mikko Waris; Hannu Toivonen; Lasse Leppäkorpi; Markku Partinen
Journal:  Conf Proc IEEE Eng Med Biol Soc       Date:  2012

6.  Alternatives to polysomnography (PSG): a validation of wrist actigraphy and a partial-PSG system.

Authors:  Anastasi Kosmadopoulos; Charli Sargent; David Darwent; Xuan Zhou; Gregory D Roach
Journal:  Behav Res Methods       Date:  2014-12

Review 7.  Consumer sleep tracking devices: a review of mechanisms, validity and utility.

Authors:  Bhanu Prakash Kolla; Subir Mansukhani; Meghna P Mansukhani
Journal:  Expert Rev Med Devices       Date:  2016-04-18       Impact factor: 3.166

8.  Practice parameters for the use of actigraphy in the assessment of sleep and sleep disorders: an update for 2007.

Authors:  Timothy Morgenthaler; Cathy Alessi; Leah Friedman; Judith Owens; Vishesh Kapur; Brian Boehlecke; Terry Brown; Andrew Chesson; Jack Coleman; Teofilo Lee-Chiong; Jeffrey Pancer; Todd J Swick
Journal:  Sleep       Date:  2007-04       Impact factor: 5.849

9.  Inter-device reliability of an automatic-scoring actigraph for measuring sleep in healthy adults.

Authors:  Matthew Driller; Joseph McQuillan; Shannon O'Donnell
Journal:  Sleep Sci       Date:  2016-08-24

10.  Adaptive Heartbeat Modeling for Beat-to-Beat Heart Rate Measurement in Ballistocardiograms.

Authors:  Joonas Paalasmaa; Hannu Toivonen; Markku Partinen
Journal:  IEEE J Biomed Health Inform       Date:  2014-03-28       Impact factor: 5.772

View more
  10 in total

Review 1.  The future of sleep health: a data-driven revolution in sleep science and medicine.

Authors:  Ignacio Perez-Pozuelo; Bing Zhai; Joao Palotti; Raghvendra Mall; Michaël Aupetit; Juan M Garcia-Gomez; Shahrad Taheri; Yu Guan; Luis Fernandez-Luque
Journal:  NPJ Digit Med       Date:  2020-03-23

2.  A validation study of a ballistocardiograph sleep tracker against polysomnography.

Authors:  Mahnoosh Kholghi; Irene Szollosi; Mitchell Hollamby; DanaKai Bradford; Qing Zhang
Journal:  J Clin Sleep Med       Date:  2022-04-01       Impact factor: 4.062

Review 3.  Integrating sleep, neuroimaging, and computational approaches for precision psychiatry.

Authors:  Andrea N Goldstein-Piekarski; Bailey Holt-Gosselin; Kathleen O'Hora; Leanne M Williams
Journal:  Neuropsychopharmacology       Date:  2019-08-19       Impact factor: 7.853

4.  Evaluating consumer and clinical sleep technologies: an American Academy of Sleep Medicine update.

Authors:  Sharon Schutte-Rodin; Maryann C Deak; Seema Khosla; Cathy A Goldstein; Michael Yurcheshen; Ambrose Chiang; Dominic Gault; Joseph Kern; Daniel O'Hearn; Scott Ryals; Nitun Verma; Douglas B Kirsch; Kelly Baron; Steven Holfinger; Jennifer Miller; Ruchir Patel; Sumit Bhargava; Kannan Ramar
Journal:  J Clin Sleep Med       Date:  2021-11-01       Impact factor: 4.062

5.  Validation of a nonwearable device in healthy adults with normal and short sleep durations.

Authors:  David A Hsiou; Chenlu Gao; Robert C Matlock; Michael K Scullin
Journal:  J Clin Sleep Med       Date:  2022-03-01       Impact factor: 4.062

6.  A New Approach for Detecting Sleep Apnea Using a Contactless Bed Sensor: Comparison Study.

Authors:  Ibrahim Sadek; Terry Tan Soon Heng; Edwin Seet; Bessam Abdulrazak
Journal:  J Med Internet Res       Date:  2020-09-18       Impact factor: 5.428

7.  Prospective cohort study to evaluate the accuracy of sleep measurement by consumer-grade smart devices compared with polysomnography in a sleep disorders population.

Authors:  Claire M Ellender; Syeda Farah Zahir; Hailey Meaklim; Rosemarie Joyce; David Cunnington; John Swieca
Journal:  BMJ Open       Date:  2021-11-09       Impact factor: 2.692

8.  Evaluations of Commercial Sleep Technologies for Objective Monitoring During Routine Sleeping Conditions.

Authors:  Jason D Stone; Lauren E Rentz; Jillian Forsey; Jad Ramadan; Rachel R Markwald; Victor S Finomore; Scott M Galster; Ali Rezai; Joshua A Hagen
Journal:  Nat Sci Sleep       Date:  2020-10-27

Review 9.  The future of sleep health: a data-driven revolution in sleep science and medicine.

Authors:  Ignacio Perez-Pozuelo; Bing Zhai; Joao Palotti; Raghvendra Mall; Michaël Aupetit; Juan M Garcia-Gomez; Shahrad Taheri; Yu Guan; Luis Fernandez-Luque
Journal:  NPJ Digit Med       Date:  2020-03-23

Review 10.  The Home-Based Sleep Laboratory.

Authors:  Yael Hanein; Anat Mirelman
Journal:  J Parkinsons Dis       Date:  2021       Impact factor: 5.568