| Literature DB >> 30849995 |
Christian Karagiannidis1, Felix Hesselmann2, Eddy Fan3.
Abstract
This article is one of ten reviews selected from the Annual Update in Intensive Care and Emergency Medicine 2019. Other selected articles can be found online at https://www.biomedcentral.com/collections/annualupdate2019 . Further information about the Annual Update in Intensive Care and Emergency Medicine is available from http://www.springer.com/series/8901 .Entities:
Mesh:
Substances:
Year: 2019 PMID: 30849995 PMCID: PMC6408850 DOI: 10.1186/s13054-019-2367-z
Source DB: PubMed Journal: Crit Care ISSN: 1364-8535 Impact factor: 9.097
Fig. 1Operating points of rotary blood pumps with corresponding hemolysis index (HI) (panel a), shear stress (panel b) and exposure time (t) (panel c) RPM revolutions per minute
Fig. 2Gap flow (panels a and b) and hydraulic efficiency (panel c) of rotary blood pumps
Priming volume and theoretical washout (data for commercially available devices were obtained from manufacturers’ manuals)
| Device | Priming volume (mL) | Flow range (L/min) | Theoretical washout (/min) |
|---|---|---|---|
| Novalung iLA | 175 | 0.5–4.5 | 2.86–25.71 |
| Quadrox-i Pediatric | 99 | 0.2–2.8 | 2.02–28.28 |
| Quadrox-i Small Adult | 175 | 0.5–5 | 2.86–28.57 |
| Quadrox-i Adult | 215 | 0.5–7 | 2.33–32.56 |
| Capiox RX05 | 43 | 0.1–1.5 | 2.36–34.88 |
| Hilite 800 LT | 55 | 0.1–0.8 | 1.82–14.55 |
| Hilite 2400 LT | 95 | 0.35–2.4 | 3.68–25.26 |
| Hilite 7000 LT | 320 | 1–7 | 3.13–21.88 |
Advantages and disadvantages of most common configurations of membrane lungs for CO2 retention applications
| Stacked | Wound | |
|---|---|---|
| Advantages | −Low pressure drop through higher permeability in flow direction | −Countercurrent flow possible |
| Disadvantages | −Inhomogeneous wash out of the corners and higher risk of thrombus formation | −Prone to clogging through washed in debris due to its smaller cross-sectional area |
| Examples | −iLA (Novalung, Germany) | −Hilite product line (Medos Medizintechnik AG, Germany) |