Computational study of the blood flow in three types of 3D hollow fiber membrane bundles.

Jiafeng Zhang, Xiaobing Chen, Jun Ding, Katharine H Fraser, M Ertan Taskin, Bartley P Griffith, Zhongjun J Wu.   

Abstract

The goal of this study is to develop a computational fluid dynamics (CFD) modeling approach to better estimate the blood flow dynamics in the bundles of the hollow fiber membrane based medical devices (i.e., blood oxygenators, artificial lungs, and hemodialyzers). Three representative types of arrays, square, diagonal, and random with the porosity value of 0.55, were studied. In addition, a 3D array with the same porosity was studied. The flow fields between the individual fibers in these arrays at selected Reynolds numbers (Re) were simulated with CFD modeling. Hemolysis is not significant in the fiber bundles but the platelet activation may be essential. For each type of array, the average wall shear stress is linearly proportional to the Re. For the same Re but different arrays, the average wall shear stress also exhibits a linear dependency on the pressure difference across arrays, while Darcy's law prescribes a power-law relationship, therefore, underestimating the shear stress level. For the same Re, the average wall shear stress of the diagonal array is approximately 3.1, 1.8, and 2.0 times larger than that of the square, random, and 3D arrays, respectively. A coefficient C is suggested to correlate the CFD predicted data with the analytical solution, and C is 1.16, 1.51, and 2.05 for the square, random, and diagonal arrays in this paper, respectively. It is worth noting that C is strongly dependent on the array geometrical properties, whereas it is weakly dependent on the flow field. Additionally, the 3D fiber bundle simulation results show that the three-dimensional effect is not negligible. Specifically, velocity and shear stress distribution can vary significantly along the fiber axial direction.

Entities:  

Mesh:

Substances:

Year:  2013        PMID: 24141394      PMCID: PMC5413142          DOI: 10.1115/1.4025717

Source DB:  PubMed          Journal:  J Biomech Eng        ISSN: 0148-0731            Impact factor:   2.097


  31 in total

1.  Computational model-based design of a wearable artificial pump-lung for cardiopulmonary/respiratory support.

Authors:  Zhongjun J Wu; M Ertan Taskin; Tao Zhang; Katharine H Fraser; Bartley P Griffith
Journal:  Artif Organs       Date:  2011-12-06       Impact factor: 3.094

2.  Non-Newtonian effects of blood flow on hemodynamics in distal vascular graft anastomoses.

Authors:  Jie Chen; Xi-Yun Lu; Wen Wang
Journal:  J Biomech       Date:  2005-08-01       Impact factor: 2.712

3.  Uniformity of the fluid flow velocities within hollow fiber membranes of blood oxygenation devices.

Authors:  Ali R Mazaheri; Goodarz Ahmadi
Journal:  Artif Organs       Date:  2006-01       Impact factor: 3.094

4.  Pulsatile flow and mass transport over an array of cylinders: gas transfer in a cardiac-driven artificial lung.

Authors:  Kit Yan Chan; Hideki Fujioka; Robert H Bartlett; Ronald B Hirschl; James B Grotberg
Journal:  J Biomech Eng       Date:  2006-02       Impact factor: 2.097

5.  Computational fluid flow and mass transfer of a functionally integrated pediatric pump-oxygenator configuration.

Authors:  Brian Fill; Mark Gartner; Greg Johnson; Marc Horner; Jeff Ma
Journal:  ASAIO J       Date:  2008 Mar-Apr       Impact factor: 2.872

6.  Red blood cell damage by shear stress.

Authors:  L B Leverett; J D Hellums; C P Alfrey; E C Lynch
Journal:  Biophys J       Date:  1972-03       Impact factor: 4.033

7.  Effect of fluid flow on smooth muscle cells in a 3-dimensional collagen gel model.

Authors:  S Wang; J M Tarbell
Journal:  Arterioscler Thromb Vasc Biol       Date:  2000-10       Impact factor: 8.311

8.  Relationship of dose of hemodialysis and cause-specific mortality.

Authors:  W E Bloembergen; D C Stannard; F K Port; R A Wolfe; J A Pugh; C A Jones; J W Greer; T A Golper; P J Held
Journal:  Kidney Int       Date:  1996-08       Impact factor: 10.612

9.  Evaluation of shear-induced platelet activation models under constant and dynamic shear stress loading conditions relevant to devices.

Authors:  Jawaad Sheriff; João Silva Soares; Michalis Xenos; Jolyon Jesty; Marvin J Slepian; Danny Bluestein
Journal:  Ann Biomed Eng       Date:  2013-02-12       Impact factor: 3.934

10.  Computational design and in vitro characterization of an integrated maglev pump-oxygenator.

Authors:  Juntao Zhang; M Ertan Taskin; Andrew Koert; Tao Zhang; Barry Gellman; Kurt A Dasse; Richard J Gilbert; Bartley P Griffith; Zhongjun J Wu
Journal:  Artif Organs       Date:  2009-07-22       Impact factor: 3.094

View more
  1 in total

Review 1.  Physiological and Technical Considerations of Extracorporeal CO2 Removal.

Authors:  Christian Karagiannidis; Felix Hesselmann; Eddy Fan
Journal:  Crit Care       Date:  2019-03-09       Impact factor: 9.097

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.