Literature DB >> 30834870

Perioperative comparison of the agreement between a portable fingertip pulse oximeter v. a conventional bedside pulse oximeter in adult patients (COMFORT trial).

R N Smith1, R Hofmeyr.   

Abstract

BACKGROUND: Low-cost, portable fingertip pulse oximeters are widely available to health professionals and the public. They are often not tested to International Organization for Standardization standards, or only undergo accuracy studies in healthy volunteers under ideal laboratory conditions.
OBJECTIVES: To pragmatically evaluate the agreement between one such device and a conventional bedside pulse oximeter in a clinical setting, in patients with varied comorbidities and skin pigmentations.
METHODS: A single-centre equipment comparison study was conducted. Simultaneous measurements were obtained in 220 patients with both a Contec CMS50D Fingertip Pulse Oximeter and a Nihon Kohden Life Scope MU-631 RK conventional bedside monitor. Peripheral oxygen saturations (SpO2) and pulse rates were documented, and patients' skin tone was recorded using the Fitzpatrick scale. Data were assessed using a Bland-Altman analysis with bias, precision and limits of agreement (LOA) calculated with 95% confidence intervals (CIs). A priori acceptability for LOA was determined to be 3%, in keeping with international standards.
RESULTS: The mean difference (therefore bias) between the conventional and fingertip oximeters for all data was -0.55% (95% CI -0.73 - -0.36). Upper and lower limits of agreement were 2.16% (95% CI 1.84 - 2.47) and -3.25% (95% CI -3.56 - -2.94). Regression analysis demonstrated worsening agreement with decreasing SpO2. When samples were separated into 'normal' (SpO2≥93%) and 'hypoxaemic' (SpO2 <93%) groups, the normal range displayed acceptable agreement between the two oximeters (bias -0.20% with LOA 2.20 - -2.27), while the hypoxaemic group fell outside the study's a priori limits. Heart rate measurements had a mean difference of -0.43 bpm (LOA -5.61 - 4.76). The study was not powered to detect differences among the skin tones, but demonstrated no trend for this parameter to alter the SpO2measurements.
CONCLUSIONS: During normoxia, portable fingertip pulse oximeters are reliable indicators of SpO2and pulse rates in patients with various comorbidities in a pragmatic clinical context. However, they display worsening agreement with conventional pulse oximeters during hypoxaemia. Skin tones do not appear to affect measurements adversely.

Entities:  

Mesh:

Year:  2019        PMID: 30834870     DOI: 10.7196/SAMJ.2019.v109i3.13633

Source DB:  PubMed          Journal:  S Afr Med J


  6 in total

1.  Preclinical evaluation of noncontact vital signs monitoring using real-time IR-UWB radar and factors affecting its accuracy.

Authors:  Jun-Young Park; Yonggu Lee; Ran Heo; Hyun-Kyung Park; Seok-Hyun Cho; Sung Ho Cho; Young-Hyo Lim
Journal:  Sci Rep       Date:  2021-12-08       Impact factor: 4.379

2.  The influence of orthopedic surgery on the incidence of post-operative delirium in geriatric patients: results of a prospective observational study.

Authors:  Krishnan Sircar; Ayla Yagdiran; Jan Bredow; Thorsten Annecke; Peer Eysel; Max Joseph Scheyerer
Journal:  J Clin Orthop Trauma       Date:  2022-08-27

Review 3.  Pulse Oximetry for Monitoring Patients with COVID-19 at Home. Potential Pitfalls and Practical Guidance.

Authors:  Andrew M Luks; Erik R Swenson
Journal:  Ann Am Thorac Soc       Date:  2020-09

Review 4.  The Use of Pulse Oximetry in the Assessment of Acclimatization to High Altitude.

Authors:  Tobias Dünnwald; Roland Kienast; David Niederseer; Martin Burtscher
Journal:  Sensors (Basel)       Date:  2021-02-10       Impact factor: 3.576

5.  Portable, consumer-grade pulse oximeters are accurate for home and medical use: Implications for use in the COVID-19 pandemic and other resource-limited environments.

Authors:  Walter A Schrading; Ben McCafferty; Jordan Grove; David B Page
Journal:  J Am Coll Emerg Physicians Open       Date:  2020-10-20

6.  Exploring the impact of pulse oximeter selection within the COVID-19 home-use pulse oximetry pathways.

Authors:  David Stell; Jonathan James Noble; Rebecca Hazell Kay; Man Ting Kwong; Michael John Russell Jeffryes; Liam Johnston; Guy Glover; Emmanuel Akinluyi
Journal:  BMJ Open Respir Res       Date:  2022-02
  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.