| Literature DB >> 30813958 |
Simon Damian Steppacher1, Markus Simon Hanke2, Corinne Andrea Zurmühle2, Pascal Cyrill Haefeli2, Frank Michael Klenke2, Moritz Tannast2.
Abstract
BACKGROUND: Ultrasonography is a fast and patient-friendly modality to assess cartilage thickness. However, inconsistent results regarding accuracy have been reported. Therefore, we asked what are (1) the accuracy, (2) reproducibility, and (3) reliability of ultrasonographic cartilage thickness measurement using contrast-enhanced micro-CT for validation?Entities:
Keywords: Accuracy; Cartilage thickness; Micro-CT; Ultrasonography; Validation
Mesh:
Year: 2019 PMID: 30813958 PMCID: PMC6391750 DOI: 10.1186/s13018-019-1099-8
Source DB: PubMed Journal: J Orthop Surg Res ISSN: 1749-799X Impact factor: 2.359
Fig. 1a A pen-like ultrasonic transducer was used (DUB micro®, Rev. 2.14a, Taberna pro medicum, Lüneburg, Germany). b The ultrasonic signal fades while traveling through cartilage. At the cartilage–bone border, the signal is partially reflected (dashed lines). Cartilage thickness is calculated based on the time the reflected signal needs to travel through cartilage. Therefore, thickness directly relates to speed of sound in cartilage. c A-mode ultrasonic image: upper cartilage border (arrow); “leading interface” (LI), an interference pattern which occurs due to the great difference in acoustic impedance of cartilage and gel; cartilage–bone border (asterisk)
Fig. 2For validation of ultrasonic cartilage thickness measurement we used the micro-CT with iodine contrast agent (a) for comparison. For micro-CT measurements the cartilage-bone samples were harvested with a punch biopsy and embedded in a radiolucent tube. While the lower limit of the cartilage (c) adjacent to the bone (b) was clearly visible in the micro-CT, detection of the upper limit of the cartilage was enhanced by iodine contrast agent (a). This allowed exact determination of cartilage thickness
Fig. 3The Bland–Altman analysis [13] was performed by plotting the difference between the two measurement techniques against their average. The analysis showed that the means of the measurement pairs were spread evenly and randomly, and therefore, no systematic error existed for a native (1580 m/s) and b adjusted speed of sound (1696 m/s) in cartilage. CT = computed tomography, US = ultrasonography, CI = confidence interval
Fig. 4Correlation of micro-CT measurement of cartilage thickness with ultrasonic measurement a with native (1580 m/s) or b adjusted speed of sound (1696 m/s). A strong correlation between ultrasonic and micro-CT measurement was found for both native and adjusted speed of sound (both p < 0.001 and r = 0.973). CT = computed tomography, US = ultrasonograph
Results of reproducibility and reliability
| Parameter | ICC intraobserver 1 | ICC intraobserver 2 | ICC interobserver |
|---|---|---|---|
| Cartilage thickness | 0.992 (0.986–0.996) | 0.994 (0.990–0.997) | 0.993 (0.990–0.995) |
Values are expressed as mean with 95% confidence interval in parentheses
ICC intraclass correlation coefficient
Selected literature on accuracy of ultrasonic cartilage thickness measurement
| Author, year | Joint | Species | Setup |
| Ground truth | Ultrasonic measurement | Accuracy |
|---|---|---|---|---|---|---|---|
| Aisen et al., 1984 [ | Knee | Bovine | Cadaver a | 3 | Ruler | B-mode, 7.5 MHz, 1750 m/s speed of sound, immersed mineral oil bad | Mean difference ranging from 0.1 to 0.2 mm (cartilage thickness from 1.0 to 4.2 mm) |
| Modest et al., 1989 [ | Hip | Human | Cadavera | 24 | Microscope | A-mode, 7.5 MHz, 1760 m/s speed of sound, thawed plugs | Comparison of the average of 5 ultrasonic with 12 microscopic measurements; 88% of the ultrasonic values within one standard deviation of the microscopic measurements |
| Martino et al., 1993 [ | Knee | Human | In vivob | 18 | Microscope | B-mode, 7.5 MHz, fixed in 90% ethanol, transcutaneous | Significant difference of 0.3 mm for the minimum cartilage difference, no significant difference of 0.2 mm for the maximum cartilage thickness |
| Jurvelin et al., 1995 [ | Knee | Canine and bovine | Cadavera | 45 | Microscope/needle probe | A-mode, 10 MHz, temporarily stored at 4 °C Ringer solution, 1760 m/s speed of sound | A strong correlation was found between ultrasonic and microscopic measurement ( |
| Ostergaard et al., 1995 [ | Knee | Human | In vivob | 39 | MRI | B-mode, 7.5 MHz, 1750 m/s speed of sound, transcutaneous | Mean difference of 0.2 mm (95% confidence interval − 1.4–1.0 mm) and a strong correlation with |
| Chérin et al., 2001 [ | Patella | Murine | Cadavera | 48 | Microscope | B-mode, 50 MHz, measurement within minutes of sacrifice in 0.9% saline solution, 1600 m/s speed of sound | A strong correlation with |
| Töyräs et al., 2002 [ | Patella | Bovine | Cadavera | 60 | Needle probe | A-mode, 22 MHz, 1654 m/s speed of sound. Plugs obtained within few hours after slaughter | Mean difference of 0.02 ± 0.99 mm with a strong correlation of |
| Pellaumail et al., 2002 [ | Patella | Murine | Cadavera | 16 | Microscope | B-mode, 50 MHz, measurement within minutes of sacrifice in 0.9% saline solution, 1600 m/s speed of sound | Difference ranging from 0.09 to 0.79 mm depending on age and cartilage degeneration |
| Tarhan et al., 2003 [ | Knee | Human | In vivob | 206 | MRI | B-mode, 5–10 MHz, transcutaneous | Correlation coefficient |
| Mathiesen et al., 2004 [ | Knee | Human | Cadaverb | 24 | Caliper gauge | B-mode, 10 MHz, thawed at room temperature | Mean difference was 0.0 ± 0.4 mm. |
| Yoon et al., 2008 [ | Knee | Human | In vivob | 102 | MRI | B-mode, 12.5 MHz, transcutaneous | Correlation coefficient |
| Naredo et al., 2009 [ | Knee | Human | Cadaverb | 24 | Stereoscopic magnifying loupe | B-mode, 14 MHz, thawed knees, transcutaneous | Agreement of the two methods was assessed using the ICC (intraclass correlation coefficient). The ICC was 0.719 for the medial epicondyle, 0.285 for the lateral epicondyle, and 0.267 for the intercondyle region. Mean difference ranged from 0.266 to 0.326 mm. |
| Aula et al., 2010 [ | Knee | Bovine | Cadavera | 10 | Contrast agent CT | A-mode, 30 MHz, thawed and stored in phosphate-buffered saline | Correlation coefficient |
| Spannow et al., 2011 [ | Knee, ankle, wrist, finger | Human | In vivob | 143 | MRI | B-mode, 6-14 MHz, transcutaneous | Mean difference ranging from 0.08 to 0.48 mm. |
| Ohashi et al., 2012 [ | Knee | Porcine | In vitroa | 62 | Microscope | Conventional B-mode and real-time spatial compound B-mode, 10 MHz, 1488 and 1709 m/s speed of sound, thawed, stored in saline solution | Correlation coefficient |
| Ohashi et al., 2012 [ | Knee | Human | In vivob | 505 | MRI | B-mode, 10 MHz, transcutaneous | Correlation coefficient |
| Mandl et al., 2015 [ | Finger | Human | Cadaverb | 19 | Calibrated photos | B-mode, 7-15 MHz, 1696 m/s speed of sound | Correlation coefficient |
| Pradsgaard et al., 2015 [ | Knee | Human | In vivob | 69 | MRI | B-mode, 6–14 MHz, transcutaneous, 1696 m/s speed of sound | Correlation coefficient rho was 0.86 for the medial condyle, 0.71 for the lateral condyle, and 0.70 for the intercondylar region. The mean difference was 0.31 mm for the medial condyle, 0.23 mm for the lateral condyle, and 0.33 mm for the intercondylar region |
| Current study | Hip, knee, and shoulder | Bovine and porcine | Cadavera | 50 | Contrast agent micro-CT | A-mode, 20 MHz, 1580 m/s and 1696 m/s speed of sound, non-frozen, evaluated within few hours of sacrifice | A strong correlation with a correlation coefficient of |
n number of measurements, TKA total knee arthroplasty, MRI magnetic resonance imaging, CT computer tomography
aUltrasonic measurements were performed with the probe directly on cartilage surface without interfering soft tissue
bUltrasonic measurements were performed with soft tissue between the probe and cartilage potentially resulting in increased measurement error