| Literature DB >> 30736796 |
Young-Eun Hwang1, Sang-Hoon Kang2,3, Hang-Keun Kim4,5.
Abstract
BACKGROUND: The aim of this study was to evaluate the accuracy of navigation according to the number of markers in terms of target registration errors (TREs) at each anatomical location during the registration process of the navigation system for the mandible.Entities:
Keywords: Image-guided surgery; Mandible; Navigation surgery; Registration
Mesh:
Year: 2019 PMID: 30736796 PMCID: PMC6368779 DOI: 10.1186/s13005-019-0190-z
Source DB: PubMed Journal: Head Face Med ISSN: 1746-160X Impact factor: 2.151
Fig. 1Mandibular models with markers placed. Anatomical sites are marked as holes for TRE measurements
Fig. 2Locations of the five registration reference markers. a 3-point markers. b 4-point markers. c 5-point markers. d 6-point markers. e 7-point markers
Fig. 3The registration process using Cbyon navigation system. a registration in the mandibular model. b registration in the Cbyon navigation system (e.g., 3-point)
Fig. 4TRE measurement of mandible model in the Cbyon navigation system. a Probing for TRE in the mandibular model. b TRE measurement in the Cbyon navigation system (e.g., antilingula)
Fig. 5Mandibular model with an RF for the Polaris optical tracker
Fig. 6Registration within 3D Slicer with Polaris optical tracker. a registration in the mandibular model. b registration in the Polaris optical tracker and 3D Slicer software (e.g., 3-point)
Fig. 7TRE measurement of mandible model on 3D Slicer with Polaris optical tracker. a Probing for TRE in the mandibular model. b TRE measurement in the 3D Slicer software (e.g., mandibular angle)
Target registration error of mandible in the Cybion navigation surgery device according to the number of registration marker (Mean ± SD)
| Anatomic landmarks for measurement | Number of registration marker (mm) | |||||
|---|---|---|---|---|---|---|
| 3 | 4 | 5 | 6 | 7 | ||
| Condyle head ( | 1.66 ± 0.62 | 1.70 ± 0.63 | 1.45 ± 0.75 | 1.68 ± 0.69 | 1.50 ± 0.58 | 0.015* |
| Condyle neck ( | 1.41 ± 0.60 | 1.51 ± 0.65 | 1.14 ± 0.66 | 1.61 ± 0.71 | 1.32 ± 0.60 | < 0.001* |
| Coronoid process ( | 1.18 ± 0.54 | 1.24 ± 0.46 | 1.04 ± 0.39 | 1.28 ± 0.55 | 1.23 ± 0.48 | 0.007* |
| Sigmoid notch ( | 1.15 ± 0.60 | 1.17 ± 0.47 | 0.97 ± 0.48 | 1.30 ± 0.57 | 1.05 ± 0.60 | < 0.001* |
| Posterior border ( | 1.25 ± 0.49 | 1.32 ± 0.57 | 1.11 ± 0.53 | 1.26 ± 0.51 | 1.17 ± 0.49 | 0.046* |
| Antilingula ( | 1.13 ± 0.53 | 1.10 ± 0.47 | 1.13 ± 0.46 | 1.30 ± 0.57 | 1.05 ± 0.43 | 0.007* |
| Mandibular anlge ( | 1.06 ± 0.52 | 1.01 ± 0.40 | 1.12 ± 0.50 | 1.26 ± 0.51 | 1.13 ± 0.49 | 0.009* |
p < .05*
Target registration error of mandible in the device using 3DSlicer software and Polaris optical tracker according to the number of registration marker (Mean ± SD)
| Anatomic landmarks for measurement | Number of registration marker (mm) | |||||
|---|---|---|---|---|---|---|
| 3 | 4 | 5 | 6 | 7 | ||
| Condyle head ( | 1.00 ± 0.42 | 0.92 ± 0.43 | 1.07 ± 0.43 | 0.80 ± 0.31 | 1.10 ± 0.53 | 0.065 |
| Condyle neck ( | 1.09 ± 0.47 | 0.99 ± 0.40 | 0.95 ± 0.49 | 0.87 ± 0.41 | 1.01 ± 0.33 | 0.361 |
| Coronoid process ( | 1.04 ± 0.47 | 0.95 ± 0.40 | 0.99 ± 0.36 | 1.04 ± 0.41 | 1.17 ± 0.43 | 0.353 |
| Sigmoid notch ( | 0.85 ± 0.43 | 0.76 ± 0.44 | 1.01 ± 0.47 | 0.96 ± 0.33 | 0.93 ± 0.46 | 0.213 |
| Posterior border ( | 1.02 ± 0.32 | 0.68 ± 0.26 | 0.98 ± 0.37 | 0.80 ± 0.37 | 0.92 ± 0.36 | 0.001* |
| Antilingula ( | 1.11 ± 0.45 | 0.83 ± 0.45 | 0.97 ± 0.30 | 0.96 ± 0.43 | 0.91 ± 0.41 | 0.134 |
| Mandibular anlge ( | 1.00 ± 0.37 | 0.90 ± 0.49 | 0.92 ± 0.35 | 0.91 ± 0.39 | 1.05 ± 0.38 | 0.545 |
p < .05*