Elastomeric pump malfunction resulting in over-infusion of local anesthetic.

Andrew Koogler1, Ganiyu Amusa1, Michael Kushelev1, Alec Lawrence1, Laurah Carlson1, Kenneth Moran1.   

Abstract

A 70 year-old female patient presented for a right humeral head replacement. Preoperatively an interscalene catheter was placed and postoperatively connected to an elastomeric pump for continuous infusion at 8 mL/h of Ropivacaine 0.2% with an additional 5 mL patient activated bolus available every 30 min. About 17 h after the elastomeric pump was connected to the catheter, the 550 mL reservoir was found to be empty, indicating the pump's infusion rate was more than 32 mL/h despite the pump still being set at an infusion rate of 8 mL/h with a possible 5 mL bolus every 30 min. There was no visible damage or leak in the pump system, and the insertion site was dry. The patient denied any changes to the pump settings. She was alert and oriented and denied any signs of local anesthetic toxicity. The catheter was immediately pulled and the manufacturer notified. The manufacturer found a red tab broken inside the patient-controlled bolus remote resulting in the over-infusion. Despite the dependability of elastomeric pumps, healthcare providers must be aware of their possible complications and malfunctions.

Entities:  

Keywords:  Elastomeric pump; interscalene catheter; pump malfunction

Year:  2019        PMID: 30728977      PMCID: PMC6350123          DOI: 10.1177/2050313X18823928

Source DB:  PubMed          Journal:  SAGE Open Med Case Rep        ISSN: 2050-313X


Introduction

Continuous interscalene perineural infusions have been shown to provide excellent postoperative analgesia in patients undergoing shoulder surgery.[1,2] A variety of infusion pumps, including elastomeric pumps, can be connected to perineural catheters and deliver a steady flow of local anesthetics to the targeted nerve structures. Perineural infusions have been shown to offer longer time to first pain and longer sleep time. Patients also experience less pain 1 week postoperatively and are able to be discharged home sooner when compared to general anesthesia alone. All of these benefits of perineural catheters can significantly reduce healthcare costs.[1,3] Elastomeric pumps are particularly popular due to their ability to be refilled, low cost, portability, independence from a power supply, disposability, as well as high patient satisfaction.[3-5] However, it has been suggested that these pumps may deliver inconsistent volume which could lead to inadequate analgesia or unintended side effects.[6,7] Such disadvantages can be caused by a multitude of factors such as temperature changes, viscosity of infused fluid, and even the height of the elastomeric pump relative to the catheter site.[8,9] Iliev et al.[6] found that the On-Q pump (I-Flow Corporation, Halyard Health, Irvine, CA, USA) delivered 22%–65% higher volume than the set rate during the initial 8 h of the infusion, followed by volume infusion within 20% of the intended delivery rate thereafter. According to another study, the ACTion pump (Ambu USA, Columbia, MD, USA) was the only studied pump that tested within the acceptable flow rate ranges.[7] Significant inaccuracies of elastomeric pumps resulting in total reservoir infusion in the first 24 h is extremely rare, and to our knowledge, only one reported case exists in the literature.[10] We present a case of an On-Q Pump (I-Flow Corporation, Halyard Health, Irvine, CA, USA) malfunction resulting in complete administration of the 550 mL local anesthetic reservoir within 17 h.

Case

A 70-year-old female patient presented to our institution for a right humeral head replacement. Preoperatively, at 1445, with the patient in a semi-sitting position, a right interscalene catheter was placed as described by Antonakakis et al.[11] under ultrasound guidance. Incremental injections of 30 mL Ropivacaine 0.5% allowed visualization of the catheter at the appropriate location near the brachial plexus. The skin at the site of catheter insertion was sealed with DermaBond (Ethicon, Somerville, NJ, USA), and the catheter was secured with a Tegaderm (3M Company, St. Paul, MN, USA). The exposed end of the catheter was capped during the patient’s surgery. The patient’s surgical procedure was performed uneventfully with general anesthesia in the beach chair position. Postoperatively, at 1615, the catheter was connected to a primed On-Q Pump programmed to deliver 8 mL/h of Ropivacaine 0.2% with an additional 5 mL patient-activated bolus available every 30 min. Once the catheter and pump were securely connected, the pump’s continuous line was unclamped. After achieving adequate postoperative anesthesia care unit (PACU) discharge criteria, the patient was transferred from the PACU to an inpatient room. The following morning at 0930, the elastomeric pump’s 550 mL reservoir was found to be empty, after approximately 17 h, indicating the pump’s infusion rate was more than 32 mL/h. There were no visible leaks in the pump system or moisture around the insertion site. The patient and her nurse denied any leakage or changes to the settings during the night. She was alert and oriented, denied dizziness, lightheadedness, metallic taste, or other signs of local anesthetic toxicity. There were no arrhythmias noted on telemetry. The catheter was immediately removed without any damage noted to the catheter or the metallic tip. On further inspection of the pump, the rate was still set to 8 mL/h and the patient-activated bolus every 30 min at 5  mL as originally programmed. The manufacturer was notified and the pump delivered to them.

Discussion

Elastomeric pumps are composed of a distensible bulb inside of a protective bulb with a built-in filling port, delivery tubing, and a bacterial filter. When the pump is filled, it expands and generates potential energy from the strain of the elastomeric membranes. This potential energy then drives the pressure to infuse the fluid. Flow is controlled with flow restrictor tubing of a fixed diameter and length. Elastomeric pumps may have a variety of delivery options: a fixed flow rate that cannot be altered; a basal infusion with on-demand boluses; or an adjustable flow rate controller and bolus device.[4] The pump in this case had an adjustable flow rate controller and bolus device. Elastomeric pumps have gained wide acceptance because of their simple design and ease of use. A study by Weisman et al. evaluated five different elastomeric pumps for flow rate accuracy, consistency, and effect of positioning on output while controlling for temperature. It was found that five commonly available pumps were reliably accurate if utilized within the manufacturers’ recommendations. However, there are various important factors that must be taken into account as it relates to pump output, such as height and temperature.[7] Despite being at room temperature and near level with the patient’s catheter insertion site, an over-infusion occurred with the presented pump. In our case specifically, we utilized the I-flow On-Q PainBuster (I-Flow Corporation, Halyard Health, Irvine, CA, USA) postoperative pain relief pump system. We reviewed this specific device’s manufacturer’s adverse event reports for incidents of “fast flow,” which demonstrated that in previous instances, the malfunction rested within the bolus remote itself. The On-Q pain relief system technical bulletin from Halyard Health states the volume of fluid in the pump, fluid temperature, and the relative height of the pump compared to the catheter site can affect flow rate accuracy, similar to what was described by Weisman et al.[7] After eliminating obvious reasons for over-infusion such as patient or staff tampering, an investigation of the physical device took place. On inspection of the device, a remnant of the red priming tab remained inside its keyhole located in the side of the patient-controlled bolus remote (Figure 1). The red tab allows for continuous flow and tube priming during pump setup. After priming, the red tab should be pulled straight out while avoiding twisting or pulling in another direction in order to avoid breaking it off in the keyhole. The manufacturer’s instruction guide specifically states, “If red tab is not removed or breaks while removing, continuous delivery will occur.” Possibly in the setup of the device, the red tab was incorrectly pulled upward while being removed. As a result, a small portion of the broken key remained inside the keyhole, permitting the continuous total flow of both the bolus and basal rate being infused into the patient. Fortunately, despite the resultant high flow rate, no harm befell the patient and appropriate steps were taken to rectify the situation. An incident report was created and the pump, along with the bolus device, was sent to Halyard who confirmed the red tab has broken during its removal.
Figure 1.

The red priming tab remaining inside the patient-controlled bolus remote keyhole.

The red priming tab remaining inside the patient-controlled bolus remote keyhole.

Conclusion

Overall, elastomeric pumps have been shown to be reliable and consistent if operated within manufacturer recommendations and the appropriate steps followed.[7] In one study, Remerand et al.[9] suggest that nursing staff should weigh pumps before connecting the catheter and during the first hours of infusion to verify that their weights are decreasing and local anesthetic is being infused appropriately. In any incident involving malfunction of an elastomeric pump, cooperation between the primary regional anesthetic team, pharmacy, and manufacturer is of the utmost importance in maintaining safe and secure practices for continuous regional anesthesia. At our institution, the pharmacy fills and primes these pumps with local anesthetics. Despite best practice instructions from the manufacturer, our department was unaware of this potential complication and have since learned how to prevent such an over-infusion. We are looking into the implementation of further safety mechanisms given the number of hands the pumps touch prior to being attached to the patient. Currently, we have developed a double check to ensure the key is undamaged and has been completely removed according to the manufacturer’s instructions. This report underscores the importance of adequately training individuals involved in the preparation, placement, and follow up of elastomeric pumps, as well as strictly following manufacturer recommendations and checking for potential malfunctions, such as a broken and retained priming key.
  11 in total

1.  Portable infusion pumps used for continuous regional analgesia: delivery rate accuracy and consistency.

Authors:  Brian M Ilfeld; Timothy E Morey; F Kayser Enneking
Journal:  Reg Anesth Pain Med       Date:  2003 Sep-Oct       Impact factor: 6.288

2.  A comparison of the hourly output between the Ambu® Smart-Infuser™ Pain Pump and the On-Q Pump® with Select-A-Flow™ Variable Rate Controller with standard and overfill volumes.

Authors:  Peter Iliev; Tarun Bhalla; Joseph D Tobias
Journal:  Paediatr Anaesth       Date:  2016-01-20       Impact factor: 2.556

Review 3.  Disposable infusion pumps.

Authors:  Elena A Skryabina; Teresa S Dunn
Journal:  Am J Health Syst Pharm       Date:  2006-07-01       Impact factor: 2.637

4.  Improved safety needed in handling elastomeric reservoir balls used for pain relief.

Authors:  Matthew Grissinger
Journal:  P T       Date:  2013-05

5.  Accuracy and consistency of modern elastomeric pumps.

Authors:  Robyn S Weisman; Andres Missair; Phung Pham; Juan F Gutierrez; Ralf E Gebhard
Journal:  Reg Anesth Pain Med       Date:  2014 Sep-Oct       Impact factor: 6.288

6.  A cost analysis of orthopedic foot surgery: can outpatient continuous regional analgesia provide the same standard of care for postoperative pain control at home without shifting costs?

Authors:  Andrea Saporito; Stefano Calciolari; Laura Gonzalez Ortiz; Luciano Anselmi; Alain Borgeat; José Aguirre
Journal:  Eur J Health Econ       Date:  2015-10-14

7.  Continuous interscalene block in patients having outpatient rotator cuff repair surgery: a prospective randomized trial.

Authors:  Emine Aysu Salviz; Daquan Xu; Ashton Frulla; Kwesi Kwofie; Uma Shastri; Junping Chen; Ali Nima Shariat; Sanford Littwin; Emily Lin; Jason Choi; Paul Hobeika; Admir Hadzic
Journal:  Anesth Analg       Date:  2013-12       Impact factor: 5.108

8.  Elastomeric pump reliability in postoperative regional anesthesia: a survey of 430 consecutive devices.

Authors:  Francis Remerand; Anne Sophie Vuitton; Michel Palud; Sylvie Buchet; Xavier Pourrat; Annick Baud; Marc Laffon; Jacques Fusciardi
Journal:  Anesth Analg       Date:  2008-12       Impact factor: 5.108

9.  Ultrasound-guided posterior approach for the placement of a continuous interscalene catheter.

Authors:  John G Antonakakis; Brian D Sites; Jeffrey Shiffrin
Journal:  Reg Anesth Pain Med       Date:  2009 Jan-Feb       Impact factor: 6.288

10.  The flow Rate Accuracy of Elastomeric Infusion Pumps After Repeated Filling.

Authors:  Masood Mohseni; Amin Ebneshahidi
Journal:  Anesth Pain Med       Date:  2014-04-07
View more
  2 in total

1.  Efficacy of a local anesthetic gel infusion kit for pain relief after minimally invasive colorectal surgery: an open-label, randomized clinical trial.

Authors:  Jung Kyong Shin; Heejoon Jeong; Woo Yong Lee; Seong Hyeon Yun; Yong Beom Cho; Jung Wook Huh; Yoon Ah Park; Woo Seog Sim; Hee Cheol Kim
Journal:  Sci Rep       Date:  2022-10-19       Impact factor: 4.996

2.  Characterization of Device-Related Malfunction, Injury, and Death Associated with Using Elastomeric Pumps for Delivery of Local Anesthetics in the US Food and Drug Administration MAUDE Database.

Authors:  Richard Teames; Andrew Joyce; Richard Scranton; Catherine Vick; Nayana Nagaraj
Journal:  Drug Healthc Patient Saf       Date:  2020-12-23
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.