Recurrent Exposure of XEN Gel Stent Implant and Conjunctival Erosion.

Louis Arnould1, Vincent Theillac2, Sarah Moran3, Damien Gatinel3, Alice Grise-Dulac3.   

Abstract

PURPOSE: The purpose of this study was to report a case of recurrent conjunctival erosion and XEN Gel Stent exposure after implantation in a primary open-angle glaucoma patient. Minimally invasive glaucoma surgery (MIGS) is a new approach in glaucoma surgery, and there is a lack of published data with regard to the management of postoperative ocular surface complications.
METHODS: This is a case report of a patient with a persistent corneal erosion after XEN Gel Stent implantation. A standalone MIGS surgery was planned for both eyes in this case of refractory glaucoma. Despite a first surgical revision, the patient was suffering from a chronic leaking bleb, an exposed XEN Gel Stent, and severe hypotonia in the left eye.
RESULTS: A free conjunctival autografting technique covered with a complementary amniotic membrane graft was performed under local anesthesia. One month following bleb revision, the intraocular pressure was 13 mm Hg on medical treatment, and slit-lamp examination showed a well-formed moderate bleb with no leakage.
CONCLUSIONS: This is the first described case of the use of a free conjunctival autograft to treat a persistently leaking bleb after XEN Gel Stent implantation. This case highlights the potential complication of conjunctival erosion due to XEN Gel Stent implantation. It emphasizes the management of complicated filtration blebs after MIGS surgery.

Entities:  

Mesh:

Year:  2019        PMID: 30676470     DOI: 10.1097/IJG.0000000000001146

Source DB:  PubMed          Journal:  J Glaucoma        ISSN: 1057-0829            Impact factor:   2.503


  8 in total

1.  Complications and post-operative interventions in XEN45 gel stent implantation in the treatment of open angle glaucoma-a systematic review and meta-analysis.

Authors:  Bjorn Kaijun Betzler; Sheng Yang Lim; Boon Ang Lim; Vivien Cherng Hui Yip; Bryan Chin Hou Ang
Journal:  Eye (Lond)       Date:  2022-03-28       Impact factor: 3.775

Review 2.  Amniotic Membrane Transplantation in Ophthalmology: An Updated Perspective.

Authors:  Andrew Walkden
Journal:  Clin Ophthalmol       Date:  2020-07-22

3.  Is the Xen® Gel Stent really minimally invasive?

Authors:  Juliette Buffault; Christophe Baudouin; Antoine Labbé
Journal:  Am J Ophthalmol Case Rep       Date:  2020-08-04

Review 4.  [Glaucoma surgery today].

Authors:  Esther M Hoffmann; Fritz Hengerer; Karsten Klabe; Marc Schargus; Hagen Thieme; Bogomil Voykov
Journal:  Ophthalmologe       Date:  2021-03       Impact factor: 1.059

5.  PreserFlo MicroShunt® exposure: a case series.

Authors:  Roxane Bunod; Mathieu Robin; Juliette Buffault; Chafik Keilani; Antoine Labbé; Christophe Baudouin
Journal:  BMC Ophthalmol       Date:  2021-07-10       Impact factor: 2.209

6.  Outcomes of XEN45 gel stent using posterior small incision sub-tenon ab interno insertion (Semi-open) technique.

Authors:  Yu Xiang George Kong; In Young Chung; Ghee Soon Ang
Journal:  Eye (Lond)       Date:  2021-07-01       Impact factor: 4.456

7.  Cataract and glaucoma combined surgery: XEN® gel stent versus nonpenetrating deep sclerectomy, a pilot study.

Authors:  Vincent Theillac; Esther Blumen-Ohana; Jad Akesbi; Pascale Hamard; Alexandre Sellam; Emmanuelle Brasnu; Christophe Baudouin; Antoine Labbe; Jean-Philippe Nordmann
Journal:  BMC Ophthalmol       Date:  2020-06-16       Impact factor: 2.209

8.  Novel Use of Double-Layer Amniotic Membrane Technique in Tube Erosion in a Pediatric Patient.

Authors:  Altan Atakana Ozcan; Burak Ulas
Journal:  Beyoglu Eye J       Date:  2021-12-17
  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.