Li Huang1, Radhouene Neji1,2, Muhummad Sohaib Nazir1, John Whitaker1, Fiona Reid3, Filippo Bosio1, Amedeo Chiribiri1, Reza Razavi1, Sébastien Roujol1. 1. School of Biomedical Engineering and Imaging Sciences, Faculty of Life Sciences and Medicine, King's College London, United Kingdom. 2. MR Research Collaborations, Siemens Healthcare Limited, Frimley, United Kingdom. 3. School of Population Health and Environmental Sciences, Faculty of Life Sciences and Medicine, King's College London, United Kingdom.
Abstract
BACKGROUND: Myocardial T1 mapping shows promise for assessment of cardiomyopathies. Most myocardial T1 mapping techniques, such as modified Look-Locker inversion recovery (MOLLI), generate one T1 map per breath-held acquisition (9-17 heartbeats), which prolongs multislice protocols and may be unsuitable for patients with breath-holding difficulties. PURPOSE: To develop and characterize novel shortened inversion recovery based T1 mapping schemes of 2-5 heartbeats. STUDY TYPE: Prospective. POPULATION/PHANTOM: Numerical simulations, agarose/NiCl2 phantom, 16 healthy volunteers, and 24 patients. FIELD STRENGTH/SEQUENCE: 1.5T/MOLLI. ASSESSMENT: All shortened T1 mapping schemes were characterized and compared with a conventional MOLLI scheme (5-(3)-3) in terms of accuracy, precision, spatial variability, and repeatability. STATISTICAL TESTS: Kruskal-Wallis, Wilcoxon rank sum tests, analysis of variance, Student's t-tests, Bland-Altman analysis, and Pearson correlation analysis. RESULTS: All shortened schemes provided limited T1 time variations (≤2% for T1 times ≤1200 msec) and limited penalty of precision (by a factor of ~1.4-1.5) when compared with MOLLI in numerical simulations. In phantom, differences between all schemes in terms of accuracy, spatial variability, and repeatability did not reach statistical significance (P > 0.71). In healthy volunteers, there were no statistically significant differences between all schemes in terms of native T1 times and repeatability for myocardium (P = 0.21 and P = 0.87, respectively) and blood (P = 0.79 and P = 0.41, respectively). All shortened schemes led to a limited increase of spatial variability for native myocardial T1 mapping with respect to MOLLI (by a factor of 1.2) (P < 0.0001). In both healthy volunteers and patients, the two-heartbeat scheme and MOLLI led to highly linearly correlated T1 times (correlation coefficients ≥0.83). DATA CONCLUSION: The proposed two-heartbeat T1 mapping scheme yields a 5-fold acceleration compared with MOLLI, with highly linearly correlated T1 times, no significant difference of repeatability, and limited spatial variability penalty at 1.5T. This approach may enable myocardial T1 mapping in patients with severe breath-holding difficulties and reduce the examination time of multislice protocols. LEVEL OF EVIDENCE: 1 Technical Efficacy Stage: 3 J. Magn. Reson. Imaging 2019;50:641-654.
BACKGROUND: Myocardial T1 mapping shows promise for assessment of cardiomyopathies. Most myocardial T1 mapping techniques, such as modified Look-Locker inversion recovery (MOLLI), generate one T1 map per breath-held acquisition (9-17 heartbeats), which prolongs multislice protocols and may be unsuitable for patients with breath-holding difficulties. PURPOSE: To develop and characterize novel shortened inversion recovery based T1 mapping schemes of 2-5 heartbeats. STUDY TYPE: Prospective. POPULATION/PHANTOM: Numerical simulations, agarose/NiCl2 phantom, 16 healthy volunteers, and 24 patients. FIELD STRENGTH/SEQUENCE: 1.5T/MOLLI. ASSESSMENT: All shortened T1 mapping schemes were characterized and compared with a conventional MOLLI scheme (5-(3)-3) in terms of accuracy, precision, spatial variability, and repeatability. STATISTICAL TESTS: Kruskal-Wallis, Wilcoxon rank sum tests, analysis of variance, Student's t-tests, Bland-Altman analysis, and Pearson correlation analysis. RESULTS: All shortened schemes provided limited T1 time variations (≤2% for T1 times ≤1200 msec) and limited penalty of precision (by a factor of ~1.4-1.5) when compared with MOLLI in numerical simulations. In phantom, differences between all schemes in terms of accuracy, spatial variability, and repeatability did not reach statistical significance (P > 0.71). In healthy volunteers, there were no statistically significant differences between all schemes in terms of native T1 times and repeatability for myocardium (P = 0.21 and P = 0.87, respectively) and blood (P = 0.79 and P = 0.41, respectively). All shortened schemes led to a limited increase of spatial variability for native myocardial T1 mapping with respect to MOLLI (by a factor of 1.2) (P < 0.0001). In both healthy volunteers and patients, the two-heartbeat scheme and MOLLI led to highly linearly correlated T1 times (correlation coefficients ≥0.83). DATA CONCLUSION: The proposed two-heartbeat T1 mapping scheme yields a 5-fold acceleration compared with MOLLI, with highly linearly correlated T1 times, no significant difference of repeatability, and limited spatial variability penalty at 1.5T. This approach may enable myocardial T1 mapping in patients with severe breath-holding difficulties and reduce the examination time of multislice protocols. LEVEL OF EVIDENCE: 1 Technical Efficacy Stage: 3 J. Magn. Reson. Imaging 2019;50:641-654.
Native myocardial longitudinal relaxation time (T1) is sensitive to a wide range of cardiomyopathies.1 This biomarker is commonly estimated on a per‐voxel basis, which is referred to as myocardial T1 mapping.2 Myocardial T1 maps can also be performed before and after administration of a gadolinium‐based contrast agent. The combination of native and postcontrast myocardial and blood T1 times enables the estimation of the extracellular volume (ECV) fraction,3 which has important diagnostic and prognostic value.4A variety of imaging sequences have been proposed for myocardial T1 mapping and often use magnetization preparation pulses such as inversion,2, 5, 6 saturation,7, 8, 9, 10, 11 or hybrid pulses.12, 13 In these techniques, a series of images with different T1‐weightings is acquired and followed by voxel‐wise fitting to a model of the measured signal to generate a T1 map.2 The modified Look–Locker inversion recovery (MOLLI) sequence2 and its variations, such as the shortened MOLLI (ShMOLLI)5 and other modified MOLLI schemes,16 are inversion recovery based techniques. Although MOLLI T1 times have been shown to be dependent on several parameters including T2,14, 18 magnetization transfer,19 off‐resonance,16 inversion factor,20 and heart rate (HR),16 this approach is commonly used for myocardial T1 mapping due to its high reproducibility/high repeatability, high precision/low spatial variability, and high map quality/low artifact level.14, 15, 16, 17Typical MOLLI sequences consist of several inversion pulses, each followed by a series of electrocardiogram (ECG)‐triggered single‐shot acquisitions. A variety of MOLLI schemes have been proposed using different amounts of T1‐weighted images and inversion pulses.2, 5, 16 T1 map reconstruction of MOLLI sequences commonly uses a three‐parameter (3P) fitting model of the inversion recovery signal2 followed by a Look–Locker correction.21 During the fitting process, the signal polarity can be restored using a multifitting approach2 or a phase‐sensitive inversion recovery (PSIR) reconstruction.22 Alternative MOLLI reconstructions have been proposed using more complex models23 or Bloch equations simulation of the sequences.24, 25, 26The common acquisition manner in MOLLI sequences is a single T1 map per breath‐hold, thus limiting the total measurement time. Most MOLLI schemes acquire data over 9–17 heartbeats.2, 5, 16 However, breath‐holding capabilities may be as low as 2 seconds in patients with cardiac or respiratory disease.27 Therefore, shortened breath‐held acquisition may be beneficial to such patients. Furthermore, the required spatial coverage of myocardial T1 mapping (from single slice to full ventricular coverage) may also depend on the pathology being assessed.1 MOLLI T1 mapping with full ventricular coverage requires repeated breath‐held acquisitions, each for a single slice, thus increasing patient discomfort and prolonging scan time. Therefore, shorter breath‐holding requirement for myocardial T1 mapping would be advantageous for multislice T1 mapping protocols.In this work, we sought to develop and characterize novel shortened inversion recovery based T1 mapping schemes of 2–5 heartbeats.
Materials and Methods
All imaging was performed using a 1.5T MR scanner (Magnetom Aera, Siemens Healthcare, Erlangen, Germany). This work was conducted according to the Declaration of Helsinki and Good Clinical Practice guidelines and was approved by a local Research Ethics Committee (approval number 01/11/12 for the healthy volunteer study and 15/NS/0030 for the patient study). Informed consent was obtained from all participants.
T
Several shortened T1 mapping schemes were evaluated using two, three, four, or five ECG‐triggered single‐shot images following a single inversion pulse (see Supplementary Material 1). Both magnitude and phase images were reconstructed from these acquisitions. A short inversion time (TI) of TImin = 100 msec was used for the first image.Two T1 fitting reconstructions using a novel two‐parameter (2P) fitting model and a standard 3P fitting model were evaluated. 2P‐n (n = 2–5) and 3P‐n (n = 3–5), hereafter referred to as T1 mapping using n images following a single inversion pulse with the 2P and 3P fitting models, respectively. These schemes were compared with a conventional 5‐(3)‐3 MOLLI scheme (i.e., 3P‐8). Note that 3P‐5 can be seen as an approximation of ShMOLLI for native myocardial T1 mapping.16
PROPOSED 2P FITTING MODEL
For T1 fitting, an exhaustive search was performed over a normalized signal dictionary created using the proposed following model:where δ ≤ 1 is a constant term representing the inversion factor of the inversion pulse. δ was determined by Bloch equations simulation of the employed nonselective tuned inversion pulse (phase‐modulated hyperbolic tangent, duration 2.56 msec, frequency sweep 9.5 KHz, ζ = 10, tanκ = 22 with a flip angle of 300°, i.e., a peak B1 strength of 14.4 μT) over typical native and postcontrast myocardial T1 ranges ([400,1600] msec), B0 field inhomogeneity ([–150,+150] Hz), B1 field inhomogeneity ([80%,100%]) and a typical myocardial T2 time of 45 msec.20 The effective flip angle was approximated based on the average longitudinal magnetization over the slice profile and all simulated T1/T2/B0/B1 regimes. The average inversion factor δ was estimated as 0.9633. The signal dictionary Sdict(TI) was created for a 1‐msec‐step T1 range of [100,2200] msec, which covers the entire range of native and postcontrast myocardial and blood T1 times.Before dictionary matching, the polarity‐restored measured signal Srestored(TIj) (j = 1,2,…8) was computed from the measured signal Smeas(TIj) (j = 1,2,…8) using a modified PSIR approach. The first image with the shortest TI (TI1 = TImin = 100 msec), which is one of the only two common images among all schemes, was chosen as the reference phase image with "negative" polarity (i.e., Srestored(TI1) = –Smeas(TI1)). Using Bloch equations simulation of the sequence, this assumption was valid for any T1 time > 172 msec (in the presence of any T2 times ≥30 msec and imaging flip angles ≤85°), thus including the entire physiological ranges of native/postcontrast T1 times in myocardium, blood, and fat (see Supplementary Material 2 for more details). As the signal dictionary is normalized, the polarity‐restored measured signal was individually scaled to each dictionary entry Sdict(TIj) as:where n is the amount of T1‐weighted images in the 2P‐n scheme, and are the signal amplitude averages over all TIs (TI1‐TIn) of a dictionary entry and the polarity‐restored measured signal, respectively. Dictionary matching was finally performed by minimizing the L2‐norm between and each dictionary entry.This reconstruction was implemented on an affordable graphics processing unit (GPU) (NVIDIA, Quadro K620 2GB) using the compute unified device architecture (CUDA) to enable high‐performance computing. The parallelization level was set to the pixel level. This implementation was compared with a central processing unit (CPU)‐only implementation entirely developed in C++. The resulting T1 times were subsequently HR‐corrected as described in the second next section.
3P FITTING MODEL
The signal of the 3P fitting model is defined as:where A, B, and are the model parameters.2 Note that is often referred to as the apparent T1 time. A PSIR reconstruction was employed to restore the signal polarity as described above. A Levenberg–Marquardt solver, provided previously,28 was used for simultaneous estimation of A, B, and . T1 times were then approximated as:as proposed previously.2 Finally, a correction for imperfect inversion was performed by dividing T1 times with δ, the inversion factor of the inversion pulse as described above.20 A subsequent HR correction on the resulting T1 times was performed as described in the next section.
Heart Rate Correction
Myocardial T1 times using MOLLI have been shown to be HR‐dependent.16, 29, 30 In this work, a novel approach for correction of HR‐dependent T1 errors is proposed for each of the eight evaluated T1 mapping scheme 2P‐n (n = 2–5) and 3P‐n (n = 3–5) as well as MOLLI. This correction approach was created using phantom experiments in nine agarose/NiCl2 vials with different T1/T2 times representing typical T1/T2 ranges of native and postcontrast myocardium and blood (T1MES, Resonance Health, Burswood, WA, Australia). Imaging parameters are described in the next section. The T1 dependence on HR of each T1 mapping scheme is shown in Supplementary Material 3 for each vial. Different linear dependence of measured T1 over physiological HR (40–120 bpm) was observed for different T1 mapping schemes, which is vial‐ and thus T1‐dependent. Individual linear regressions were performed for each T1 mapping scheme and each vial, leading to different slopes and offsets for each T1 mapping scheme and vial (see Supplementary Material 3), as described below:Note that this relationship is also T2‐dependent. Therefore, two different correction models were developed for myocardium (using short‐T2 vials with T2~45 msec) and blood (using long‐T2 vials with T2 > 150 msec), respectively. Each correction model was created as follows. An empirical method was established to correct the HR‐dependent T1 errors by aligning measured T1 times to the value at a theoretical HR of 60 bpm (T1
(corr)) based on:To make this model applicable to any T1 times (and not limited to the ones corresponding to the phantom), a parabolic relationship between slope, offset, and T1 was empirically defined as:where a1, a2, and a3 are the coefficients of the parabolic function and were obtained from least square fitting (see Supplementary Material 4). Using Eq. (6), Eq. (7) can then be rewritten as:In such case, the offset can be derived as:where the positive root (" + " instead of " ± ") was found to provide a physiologically reasonable offset. Then the corrected T1 (T1
(corr)) can be computed from Eqs. 6, 9 as:
Experimental Validation
Numerical Simulations
Numerical simulations were used to study the T1 accuracy and precision of the proposed shortened T1 mapping schemes and the conventional 5‐(3)‐3 MOLLI scheme. The Bloch equations were used to simulate the signal of each sequence by measuring the simulated transverse magnetization at the k‐space center of each imaging readout. All numerical simulations used a simulated HR of 60 bpm and the following imaging parameters: TR/TE/TI1/TI2 = 2.7/1.1/100/180 msec, 62 phase encoding lines in linear ordering, partial Fourier factor = 7/8, and five start‐up pulses. The slice profile of the employed excitation pulse (Hann‐filtered sinc pulse without phase modulation, duration 0.48 msec, bandwidth 4660 Hz, time‐bandwidth product 1.6, prescribed flip angle 35°, peak strength 10.9 μT) was estimated by Bloch equations simulation over the same T1/T2 ranges and B0/B1 inhomogeneities as used for the inversion pulse. A resulting average excitation flip angle of 26° was obtained and used for the simulations. An average inversion flip angle of 164° corresponding to δ = 0.9633 was also used for the simulations.Numerical simulations were performed over a range of typical myocardial T1 times (300–1500 msec in steps of 50 msec) and myocardial T2 times (30–70 msec in steps of 5 msec). Monte‐Carlo simulation (N = 50,000) were performed for each pair of simulated T1/T2 times using random noise corresponding to a signal‐to‐noise ratio (SNR) of 50 in the longest‐TI image of the conventional MOLLI scheme (TImax = 4100 msec). Accuracy was assessed as the average over the N repetitions of the difference between the simulated and estimated T1 times. Precision was defined as the standard deviation (SD) of the estimated T1 times over the N repetitions.To evaluate the influence of SNR on T1 accuracy and precision, additional numerical simulations were performed for different SNR ([10,25,50,100]) and T1 range (300–1500 msec in steps of 50 msec), and a fixed T2 times of 45 msec.
PHANTOM STUDY
The proposed 2P fitting model was characterized and compared with the conventional 3P fitting model using different shortened T1 mapping schemes and the conventional 5‐(3)‐3 MOLLI scheme in a phantom with nine agarose/NiCl2 vials of different T1/T2 times in the ranges for native and postcontrast myocardium and blood (T1MES, Resonance Health). To this end, the conventional 5‐(3)‐3 MOLLI acquisition scheme was used. The first two to five ECG‐triggered single‐shot images following the first inversion pulse were used for 2P‐n (n = 2–5) and 3P‐n (n = 3–5). The conventional MOLLI reconstruction using all images (i.e., 3P‐8) was also performed. The 2D balanced steady‐state free precession (bSSFP) imaging readout used the following parameters: TR/TE/flip angle = 2.7 msec/1.1 msec/35°, field of view (FOV) = 360 × 306 mm2, voxel size = 1.4 × 2.1 mm2, three slices, slice gap = 8 mm, slice thickness = 8 mm, GRAPPA factor = 2, partial Fourier factor = 7/8, bandwidth = 1085 Hz/px, 62 phase‐encoding lines in linear ordering, and five start‐up pulses.Experiment #1: Characterization of T1 accuracy, spatial variability, and repeatability. The 5‐(3)‐3 MOLLI acquisition scheme with a simulated HR of 60 bpm was repeated five times for assessment of T1 accuracy, spatial variability, and repeatability of all schemes. The reference T2 times were obtained from the manufacturer. The reference T1 times were obtained using inversion recovery based spin echo T1 mapping (TI = [50, 100, 200, 300, 400, 500, 600, 700, 800, 900, 1000, 2000, 3000, 4000, 5000] msec, TE/TR = 15/15000 msec). A region of interest (ROI) was manually drawn for each vial. Measured T1 times were obtained for each vial as the averages over the five repetitions of the mean T1 times in the corresponding ROI. T1 accuracy was measured as the difference between measured and reference T1 times. T1 spatial variability was measured for each vial as the average over the five repetitions of the SD of T1 times in the corresponding ROI. T1 repeatability was estimated for each vial as the SD over the five repetitions of the mean T1 times in the corresponding ROI.Experiment #2: Characterization of the proposed HR correction. The performance of the proposed HR correction was evaluated using a second dataset of measurements where the 5‐(3)‐3 MOLLI scheme was acquired with different simulated HRs ([40-120] bpm in steps of 10 bpm). All 2P‐n and 3P‐n reconstructions were performed without and with HR correction using the two different correction models for short‐T2 and long‐T2 vials. Note that the data used for creating the HR correction models were obtained from a separated study performed earlier on a different day. T1 variation as the average absolute differences with respect to the value at the reference HR of 60 bpm, described as:was calculated for pre‐ and post‐HR‐correction on each vial, in order to indicate the T1 mapping sensitivity to HR of each evaluated T1 mapping schemes.
HEALTHY VOLUNTEER STUDY
In vivo characterization was performed in 16 healthy volunteers (seven male, 28 ± 3 years). Native myocardial T1 mapping was performed using the 5‐(3)‐3 MOLLI acquisition scheme and the imaging parameters described in the phantom study. This protocol was modified to acquire three slices in the short axis orientation, each in a separated breath‐hold. This acquisition was repeated twice for each healthy volunteer. All 2P‐n and 3P‐n reconstructions were performed without and with HR correction using the short T2 and long T2‐based correction models for myocardial and blood T1 analyses, respectively.Myocardial T1 analysis was based on a 16‐myocardial‐segment model,31 while blood T1 analysis was based on a single ROI drawn inside the left ventricular blood pool in the basal slice with careful exclusion of the papillary muscles. A representative example of ROIs used for myocardial and blood T1 quantification is shown in Supplementary Material 5. All data were visually inspected to detect the presence of severe artifacts or motion among the T1‐weighted images. Myocardial segments with apparent severe artifacts in the MOLLI T1 maps were discarded from quantitative myocardial T1 analysis of all schemes. Myocardial and blood T1 times, spatial variability, and repeatability were assessed for each subject. A segment‐wise T1 time was calculated as the average over the two repetitions of the T1 mean in each myocardial segment and blood pool. Segment‐wise T1 spatial variability was measured as the average over the two repetitions of the T1 spatial SD in each myocardial segment and blood pool. Segment‐wise T1 repeatability was estimated as the absolute difference between the two repetitions of the T1 mean in each myocardial segment and blood pool. The corresponding subject‐wise T1 time, spatial variability, and repeatability were computed as the averages over all nondiscarded segments, respectively.
Patient Study
Twenty‐four consecutive patients (17 male, 53 ± 17 years) referred for clinical cardiac MRI in our center were recruited. Native myocardial T1 mapping was performed in all patients. Eighteen of these patients (13 male, 53 ± 19 years) received an injection of 0.1 mmol/kg of gadobutrol (Gadovist, Bayer Vital, Leverkusen, Germany) as part of the clinical protocols. Postcontrast T1 mapping was thus also performed in these patients. Native and postcontrast myocardial T1 mapping were performed using the same 5‐(3)‐3 MOLLI acquisition scheme and imaging parameters described in the healthy volunteer study. Three slices were acquired in the short axis orientation, each in a separated breath‐hold. All 2P‐n and 3P‐n reconstructions were performed without and with HR correction using the short T2 and long T2‐based correction models for myocardial and blood T1 analyses, respectively. Subject‐wise myocardial and blood T1 times were measured as described in the healthy volunteers section.
Statistical Analysis
Kruskal–Wallis test and a one‐way analysis of variance (ANOVA) test were used to compare all T1 mapping schemes in phantom and in vivo, respectively. A result was considered statistically significant at the 5% significance level (i.e., P < 0.05) and all tests were two‐tailed. When the Kruskal–Wallis or one‐way ANOVA test demonstrated statistical significance, Wilcoxon rank sum tests or Student's t‐tests were performed for each pair of T1 mapping schemes using Bonferroni correction, which resulted in a statistical significance threshold of 0.05/C2
8 ≈ 0.0018. Correlation and agreement analyses in the form of Pearson correlation analysis and Bland–Altman plots with limits of agreement, respectively, were performed between each shortened T1 mapping scheme and MOLLI in terms of subject‐wise native/postcontrast myocardial/blood T1 times. Bland–Altman limits of agreement were calculated as the mean difference between methods (also called bias) ± 1.96 × (SD of differences); ~95% of differences between methods should lie within these limits.
Results
Computational cost of the 2P model‐based reconstruction
2P‐2, 2P‐3, 2P‐4, and 2P‐5 reconstruction times for one T1 map (256 × 256 matrix size) were 7, 8, 11, and 13 seconds using a CPU‐based implementation, respectively. These reconstruction times were reduced to 0.2 seconds for all 2P reconstructions using the proposed GPU‐based implementation. Reconstruction times for all CPU‐based implementations increased linearly with the number of slices, while GPU‐based reconstruction times increased at a slower rate. For example, 2P‐2, 2P‐3, 2P‐4, and 2P‐5 reconstruction times for 10 T1 maps (256 × 256 × 10 matrix size) were 65, 83, 104, and 130 seconds using the CPU‐based implementation, and were reduced to 1, 1.3, 1.7, and 2 seconds using the proposed GPU‐based implementation, respectively.
Numerical Simulations
Accuracy and precision of all evaluated T1 mapping schemes are shown in Fig. 1. All 3P‐n schemes led to limited T1 time variation (≤2%) with respect to MOLLI for the entire range of T1 times ([300,1500] msec). All 2P‐n schemes provided limited T1 time variations (≤2%) with respect to MOLLI for T1 times ≤1200 msec but resulted in reduced accuracy for longer T1 times. All shortened T1 mapping schemes led to a precision penalty with respect to MOLLI by a factor of ~1.4–1.5. All studied schemes were T2‐dependent. Lower T2 times were associated with decreased accuracy for all schemes.
Figure 1
Numerical simulations of T1 accuracy and precision of all T1 mapping schemes. T1 accuracy (a) and precision (b) are shown as a function of T1 using a typical myocardial T2 time of 45 msec for all T1 mapping schemes. Impact of T2 times on T1 accuracy (c) and precision (d) are shown for 2P‐2 and MOLLI.
Numerical simulations of T1 accuracy and precision of all T1 mapping schemes. T1 accuracy (a) and precision (b) are shown as a function of T1 using a typical myocardial T2 time of 45 msec for all T1 mapping schemes. Impact of T2 times on T1 accuracy (c) and precision (d) are shown for 2P‐2 and MOLLI.SNR had limited influence on the T1 time estimates of all schemes (variation ≤2% with respect to T1 estimations with an SNR of 100, see Supplementary Material 6). Lower SNR resulted in a reduced T1 precision of all schemes. However, SNR had limited influence on the relative precision penalty of all shortened T1 mapping schemes with respect to MOLLI, which was by a factor of 1.4–1.5 for the entire SNR range.
Phantom Study
Experiment #1: Characterization of T1 accuracy, spatial variability, and repeatability. T1 accuracy, spatial variability, and repeatability in phantom using all evaluated T1 mapping schemes (conventional MOLLI and shortened T1 mapping schemes: 2P‐n [n = 2–5] and 3P‐n [n = 3–5]) are shown in Fig. 2. All schemes were in good agreement with the reference T1 times for long‐T2 vials (i.e., T2 > 150 msec) with an average error of <11 msec for all schemes. All schemes led to underestimated T1 times for short‐T2 vials (i.e., T2~45 msec) with respect to the reference T1 times. Although MOLLI tended to provide slightly lower underestimation than shortened T1 mapping schemes (especially for short‐T2 vials), these differences were not statistically significant (P = 1.00). For a typical native myocardial T1 range (the vial with T1/T2 1160/48 msec), 2P‐2 and 2P‐5 led to an underestimation of 25 msec and 22 msec with respect to MOLLI, while the other shortened T1 mapping schemes led to an underestimation of <10 msec. The 2P‐n (n = 2–5) schemes tended to provide lower spatial variability than the 3P‐n (n = 3–5) schemes for typical postcontrast T1 range (<450 msec), while 2P‐2 and 3P‐3 tended to show higher spatial variability than other schemes for long T1 times (>1400 msec). Although all schemes tended to provide higher spatial variability than MOLLI (7–8 msec vs. 5 msec, respectively), and lower repeatability than MOLLI (1.2–1.5 msec vs. 1.0 msec, respectively), these differences were not statistically significant (P = 0.71 and P = 0.75, respectively).
Figure 2
T1 accuracy (a), spatial variability (b), and repeatability (c) of all T1 mapping schemes in phantom experiments. There were no statistically significant differences between all schemes in terms of accuracy (P = 1.00), spatial variability (P = 0.71) and repeatability (P = 0.75).
T1 accuracy (a), spatial variability (b), and repeatability (c) of all T1 mapping schemes in phantom experiments. There were no statistically significant differences between all schemes in terms of accuracy (P = 1.00), spatial variability (P = 0.71) and repeatability (P = 0.75).Experiment #2: Characterization of the proposed HR correction. Supplementary Material 7 shows the impact of the proposed HR correction for measured T1 times. After the proposed HR correction, T1 variation over all HR was reduced from a maximum of 55 msec to a maximum of 7 msec for all vials and T1 mapping schemes.
Healthy Volunteer Study
Example native myocardial T1 maps of a healthy volunteer using all T1 mapping schemes are shown in Fig. 3. All schemes provided similar visual image quality across all slices and segments, as well as similar native T1 ranges for myocardium and blood. The perceived noise, however, was higher in the left ventricular blood pool for all shortened T1 mapping schemes.
Figure 3
Example native T1 maps of a 29‐year‐old male healthy volunteer (HR 52 bpm) along the short axis using all T1 mapping schemes. Similar image quality and native T1 range were obtained across all slices using all schemes.
Example native T1 maps of a 29‐year‐old male healthy volunteer (HR 52 bpm) along the short axis using all T1 mapping schemes. Similar image quality and native T1 range were obtained across all slices using all schemes.The average HR over all healthy volunteers was 68 ± 12 bpm (51–90 bpm). On average, over all healthy volunteers, the magnitude of HR correction for native myocardium ranged from 0.03 ± 0.05 msec (≤0.17 msec) using 2P‐5 to 12 ± 9 msec (≤30 msec) using 2P‐2, while the magnitude of HR correction for native blood ranged from 3 ± 2 msec (≤8 msec) using 2P‐5 to 14 ± 12 msec (≤41 msec) using 2P‐2.Over all healthy volunteers, only one of 256 myocardial segments (0.4%) was discarded from the analysis. There were no statistically significant differences between all schemes in terms of native myocardial T1 times (P = 0.21), which were all in the range of 977–997 msec (Fig. 4a). There were no statistically significant differences between all shortened T1 mapping schemes in terms of myocardial T1 spatial variability (P = 0.87). However, they all had increased spatial variability by a factor of 1.2 with respect to MOLLI (56–59 msec vs. 48 msec, respectively, P < 0.0001) (Fig. 4b). There were no statistically significant differences between all schemes in terms of myocardial T1 repeatability, which were in the range of 14–18 msec (P = 0.87) (Fig. 4c).
Figure 4
Native T1 times, spatial variability, and repeatability for myocardium (a–c, respectively) and blood (d–f, respectively) obtained using all T1 mapping schemes in 16 healthy volunteers. Average (bar plots) and SD (error bars) over all healthy volunteers are presented. There were no statistically significant differences between native myocardial and blood T1 times (P = 0.21 and P = 0.79, respectively) and repeatability (P = 0.87 and 0.41, respectively) obtained using all schemes. All shortened T1 mapping schemes led to increased myocardial and blood T1 spatial variability with respect to MOLLI (P < 0.0001).
Native T1 times, spatial variability, and repeatability for myocardium (a–c, respectively) and blood (d–f, respectively) obtained using all T1 mapping schemes in 16 healthy volunteers. Average (bar plots) and SD (error bars) over all healthy volunteers are presented. There were no statistically significant differences between native myocardial and blood T1 times (P = 0.21 and P = 0.79, respectively) and repeatability (P = 0.87 and 0.41, respectively) obtained using all schemes. All shortened T1 mapping schemes led to increased myocardial and blood T1 spatial variability with respect to MOLLI (P < 0.0001).Over all healthy volunteers, no statistically significant differences were found between all schemes in terms of native blood T1 times, which were in the range of 1583–1623 msec (P = 0.79) (Fig. 4d). 2P‐2 and 3P‐3 yielded higher spatial variability in the blood pool (85 msec and 105 msec, respectively) than the other shortened T1 mapping schemes (63 msec, P ≤ 0.0025), which were all inferior to MOLLI (49 msec, P < 0.0001) (Fig. 4e). There were no statistically significant differences between all schemes in terms of repeatability of native blood T1 times (P = 0.41), which were in the range of 7–15 msec for all schemes (Fig. 4f).Segment‐wise assessment of native myocardial T1 times, spatial variability, and repeatability of 2P‐2 and MOLLI are shown in Fig. 5. The segmental variation (SD over all myocardial segments) of native myocardial T1 times, spatial variability, and repeatability was of similar range between 2P‐2 and MOLLI [13/6/3 msec vs. 12/8/3 msec, respectively]).
Figure 5
Segment‐wise native myocardial T1 times, spatial variability, and repeatability using 2P‐2 and MOLLI in 16 healthy volunteers. Data are shown as average ± SD over all healthy volunteers. No statistically significant differences were found between segmental values of native T1 times, spatial variability, and repeatability obtained using both methods.
Segment‐wise native myocardial T1 times, spatial variability, and repeatability using 2P‐2 and MOLLI in 16 healthy volunteers. Data are shown as average ± SD over all healthy volunteers. No statistically significant differences were found between segmental values of native T1 times, spatial variability, and repeatability obtained using both methods.
Patient Study
Example native and postcontrast myocardial T1 maps obtained in two patients using all the evaluated T1 mapping schemes are shown in Figs. 6 and 7, respectively. Similar visual image quality and native myocardial T1 range were obtained for all schemes, although a higher perceived noise level can be observed in the left ventricular blood pool using shortened T1 mapping schemes.
Figure 6
Example native myocardial T1 maps of a 38‐year‐old male patient (HR 84 bpm) admitted with syncope using all T1 mapping schemes. All schemes provided similar T1 map image quality and similar characteristics for native myocardial T1 times. Shortened schemes tended to have lower spatial homogeneity than MOLLI in the blood pool.
Figure 7
Example postcontrast myocardial T1 maps obtained in a 32‐year‐old female patient (HR 61 bpm) with severe left ventricular systolic dysfunction and pericardial effusion using all T1 mapping schemes. All schemes provided similar T1 map image quality as well as similar myocardial and blood T1 ranges across all slices.
Example native myocardial T1 maps of a 38‐year‐old male patient (HR 84 bpm) admitted with syncope using all T1 mapping schemes. All schemes provided similar T1 map image quality and similar characteristics for native myocardial T1 times. Shortened schemes tended to have lower spatial homogeneity than MOLLI in the blood pool.Example postcontrast myocardial T1 maps obtained in a 32‐year‐old female patient (HR 61 bpm) with severe left ventricular systolic dysfunction and pericardial effusion using all T1 mapping schemes. All schemes provided similar T1 map image quality as well as similar myocardial and blood T1 ranges across all slices.Over all patients, the average HR was 68 ± 14 bpm (36–98 bpm). In all, 34 of 384 myocardial segments (9%) from five patients for native T1 mapping and 11 of 288 myocardial segments (4%) from two patients for postcontrast T1 mapping were discarded from the quantitative analysis due to substantial artifacts and/or motion. HR correction of native myocardial T1 times led to changes from 0.04 ± 0.06 msec (≤0.25 msec) using 2P‐5 to 13 ± 12 msec (≤43 msec) using 2P‐2, while HR correction of native blood T1 times led to changes from 3 ± 3 msec (≤11 msec) using 2P‐5 to 14 ± 13 msec (≤47 msec) using 2P‐2. The magnitude of HR correction for postcontrast myocardial and blood T1 times was <4 msec using all schemes.Subject‐wise native and postcontrast T1 times for myocardium and blood using all schemes are shown in Fig. 8. There were no statistically significant differences between all schemes for each of these four T1 ranges (native/postcontrast myocardial/blood) (P ≥ 0.19).
Figure 8
Native/postcontrast myocardial/blood T1 times (a–d, respectively) in 24 patients using all T1 mapping schemes. Average (bar plots) and SD (error bars) over all patients are presented. All methods led to similar range of native/postcontrast myocardial/blood T1 times (P ≥ 0.19).
Native/postcontrast myocardial/blood T1 times (a–d, respectively) in 24 patients using all T1 mapping schemes. Average (bar plots) and SD (error bars) over all patients are presented. All methods led to similar range of native/postcontrast myocardial/blood T1 times (P ≥ 0.19).The Pearson correlation and Bland–Altman analyses in terms of subject‐wise native myocardial/blood T1 times (healthy volunteers and patients) and postcontrast myocardial/blood T1 times (patients only) are shown in Fig. 9 (only 2P‐2 vs. MOLLI) and Table 1 (each shortened scheme vs. MOLLI). Strong correlation was observed between 2P‐2 and MOLLI. The Pearson correlation coefficient between 2P‐2 and MOLLI for native blood T1 times was 0.83, and was ≥0.96 for all other T1 ranges. All other shortened T1 mapping schemes were also strongly correlated with MOLLI for each T1 range (Pearson correlation coefficient ≥ 0.90). For native blood T1 times, 2P‐2 and MOLLI were in moderate agreement (bias of 40 msec, 95% limits of agreement: –51 msec to 130 msec). For other T1 ranges, 2P‐2 and MOLLI were in good agreement with limited bias magnitude (≤17 msec) and narrow width of 95% limits of agreement (<43 msec). All other shortened T1 mapping schemes were in good agreement with MOLLI for native myocardial T1 mapping, with limited bias magnitude (≤19 msec) and narrow width of 95% limits of agreement (<39 msec).
Figure 9
Correlation and agreement tests between T1 times obtained using 2P‐2 and MOLLI in all subjects including 16 healthy volunteers and 24 patients. Subfigure (a–d) illustrate Pearson correlation analysis and Bland–Altman plot of native myocardial T1 times, native blood T1 times, postcontrast myocardial T1 times, and postcontrast blood T1 times, respectively. Strong correlation and good agreement were found between T1 times obtained using 2P‐2 and MOLLI. In Pearson correlation analysis plots, confidence interval (solid lines) and identity line (y = x, dashed line) are also plotted besides the linear regression line (solid line). Correlation information including the Pearson correlation coefficient (r‐value), linear regression relationship (y as a function of x), and coefficient of determination (r2) is also displayed in Pearson correlation plots. In Bland–Altman plots, "average" stands for (T12P‐2 + T1MOLLI)/2 and "difference" stands for (T12P‐2‐T1MOLLI).
Table 1
Pearson Correlation Analysis and Bland–Altman Plot Results
Native myocardial T1
Native blood T1
Postcontrast myocardial T1
Postcontrast blood T1
2P‐2
0.96
0.83
0.98
1.00
–17 ± 11 (–38,4)
40 ± 46 (–51,130)
–13 ± 11 (–34,8)
–3 ± 4 (–10,5)
2P‐3
0.97
0.94
0.97
1.00
0 ± 8 (–17,16)
17 ± 26 (–33,67)
–13 ± 11 (–34,9)
–3 ± 4 (–10,4)
2P‐4
0.98
0.98
0.97
1.00
–6 ± 7 (–20,8)
9 ± 15 (–21,39)
–12 ± 11 (–34,9)
–3 ± 4 (–10,5)
2P‐5
0.98
0.99
0.97
1.00
–19 ± 7 (–33,–6)
6 ± 12 (–18,30)
–12 ± 12 (–34,11)
–2 ± 4 (–10,5)
3P‐3
0.97
0.96
0.97
0.99
–11 ± 10 (–31,8)
–18 ± 23 (–62,26)
–10 ± 11 (–31,11)
15 ± 14 (–12,42)
3P‐4
0.98
0.99
0.98
0.99
–8 ± 7 (–23,6)
–8 ± 13 (–33,18)
–9 ± 9 (–27,8)
12 ± 13 (–13,38)
3P‐5
0.98
1.00
0.98
0.98
–7 ± 8 (–23,9)
–1 ± 8 (–16,15)
–9 ± 9 (–27,8)
12 ± 13 (–14,38)
Measured between each shortened T1 mapping scheme and MOLLI for native/postcontrast myocardial/blood T1 times. Data shown are as follows: first row, Pearson correlation coefficient as the r‐value; second row, bias ± SD (95% limits of agreement) in msec. All P‐values in Pearson correlation analysis are <0.0001.
Correlation and agreement tests between T1 times obtained using 2P‐2 and MOLLI in all subjects including 16 healthy volunteers and 24 patients. Subfigure (a–d) illustrate Pearson correlation analysis and Bland–Altman plot of native myocardial T1 times, native blood T1 times, postcontrast myocardial T1 times, and postcontrast blood T1 times, respectively. Strong correlation and good agreement were found between T1 times obtained using 2P‐2 and MOLLI. In Pearson correlation analysis plots, confidence interval (solid lines) and identity line (y = x, dashed line) are also plotted besides the linear regression line (solid line). Correlation information including the Pearson correlation coefficient (r‐value), linear regression relationship (y as a function of x), and coefficient of determination (r2) is also displayed in Pearson correlation plots. In Bland–Altman plots, "average" stands for (T12P‐2 + T1MOLLI)/2 and "difference" stands for (T12P‐2‐T1MOLLI).Pearson Correlation Analysis and Bland–Altman Plot ResultsMeasured between each shortened T1 mapping scheme and MOLLI for native/postcontrast myocardial/blood T1 times. Data shown are as follows: first row, Pearson correlation coefficient as the r‐value; second row, bias ± SD (95% limits of agreement) in msec. All P‐values in Pearson correlation analysis are <0.0001.Supplementary Material 8 shows native myocardial T1 maps obtained in a patient who was unable to sustain a stable breath‐hold for the entire duration of the acquisition. MOLLI led to substantial T1 map artifacts in both mid‐ventricular and apical slices, which were then discarded for all schemes from the quantitative analysis. All shortened T1 mapping schemes provided improved map quality in this patient.
Discussion
In this work, we proposed and evaluated shortened T1 mapping schemes combined with a novel 2P fitting model for myocardial T1 mapping. These methods were successfully evaluated in numerical simulations, phantom, healthy volunteers, and patients. Compared with the conventional MOLLI 5‐(3)‐3 scheme, shortened T1 mapping schemes (down to two heartbeats only) combined with the proposed 2P fitting model resulted in no significant differences in terms of T1 estimates and repeatability and had similar T1 ranges as well as limited reduction of precision/increase of spatial variability. Importantly, the resulting native/postcontrast myocardial/blood T1 times measured by all shortened T1 mapping schemes were highly linearly correlated with the corresponding values measured using MOLLI. Finally, the proposed GPU implementation of the exhaustive search‐based optimization of the 2P fitting model enables fast T1 map reconstruction, which is suitable for clinical application.In vivo myocardial T1 times, precision, and repeatability of MOLLI were in good agreement with previous works.6, 14, 32 All evaluated shortened T1 mapping schemes provided T1 times in a similar range as MOLLI for all native/postcontrast myocardial/blood T1 ranges. Moreover, all proposed shortened T1 mapping schemes have the same acquisition manner as 5‐(3)‐3 MOLLI. These observations suggest that, similar to MOLLI, these shortened T1 mapping schemes are also sensitive to T2 relaxation,14, 18 magnetization transfer,19 and off‐resonance effects.16 Furthermore, this work was performed at 1.5T. The potential of these shortened T1 mapping schemes at higher fields such as 3T remains to be demonstrated and will be the focus of future work.The feasibility of the shortened T1 mapping schemes for postcontrast myocardial T1 mapping was demonstrated in numerical simulations, phantom, and patients. The spatial variability penalty of shortened T1 mapping schemes was more pronounced for typical short postcontrast T1 times than typical native myocardial T1 times, which could be interpreted as a consequence of lacking a second short‐TI image. For short T1 times (i.e., typical postcontrast T1 times), all 2P‐n schemes had higher precision and lower spatial variability than the shortened 3P‐n schemes. Therefore, 2P‐n schemes may be advantageous over shortened 3P‐n schemes in the context of ECV mapping.The healthy volunteer study demonstrated that all proposed shortened T1 mapping schemes resulted in an increase of T1 spatial variability (by a factor of 1.2) for native myocardial T1 times when compared with MOLLI. This increase in spatial variability between 3P‐5 (approximated ShMOLLI for native T1) and MOLLI is in good agreement with previous comparison of ShMOLLI and MOLLI at 1.5T.5 Importantly, the use of fewer images for both 2P‐n and 3P‐n (i.e., n = 2/3/4) did not result in a further increase of spatial variability (i.e., precision loss) of native myocardial T1 time estimates when compared with 3P‐5. This suggests that the impact of each T1‐weighted image in the fitting process is dependent on its corresponding TI, similar to findings observed for saturation recovery‐based techniques.33, 34 Long TI images (i.e., with TI >> T1 range of interest) have reduced T1‐weighted contrast and may thus have reduced contributions to the precision of T1 estimates.For long T1 times such as in vivo native blood T1 times, 2P‐2 led to a larger increase of spatial variability by a factor of 1.7 with respect to MOLLI, while other 2P‐n schemes still maintained a limited increase of T1 spatial variability by a factor < 1.28. This could be explained by the lack of sampling of long TI times in the 2P‐2 scheme. Therefore, 2P‐3 could be a valuable alternative to 2P‐2 for native T1 mapping, as it offers lower spatial variability for blood T1 quantification. However, blood T1 time is usually measured as the spatial average over a large ROI, which may mitigate this effect for 2P‐2, as no statistically significant differences were found in terms of repeatability for blood T1 quantification among all methods.Myocardial T1 mapping based on the acquisition of only two images has been previously proposed using the AIR technique, which is based on saturation recovery.9 However, this technique was shown to considerably increase spatial variability of native myocardial T1 mapping by a factor of 2.5 when compared with MOLLI.35 The proposed 2P‐2 approach resulted in a limited increase of spatial variability for native myocardial T1 mapping by a factor of 1.2 when compared with MOLLI, and may thus be a valuable alternative for myocardial T1 mapping, as the acquisition can be performed in just two heartbeats.The proposed 2P model‐based fitting technique enables correction for surface coil sensitivity variations using the normalization step of the dictionary matching. Exhaustive search‐based optimization over the entire range of physiological myocardial and blood native/postcontrast T1 times guarantees finding the global minimum of the cost function over a least‐square optimization, and was successfully used with the 2P model. Although such an approach is more computationally intensive, the proposed GPU‐based implementation substantially reduced the computation time to the subsecond scale per map, which is suitable for clinical application. Our results demonstrate that GPUs are particularly well suited for the reconstruction of T1 maps. This finding is in good agreement with prior studies where GPU‐based reconstruction substantially reduced the computation time of standard MOLLI reconstructions.36 Further reduction of the computation time may be achieved using GPU cards with higher performance or advanced dictionary search approaches such as fast group matching algorithms.37Although 2P‐2 provides a 5‐fold acceleration with respect to MOLLI, the overall acceleration rate in multislice protocols may be reduced by the required rest‐periods between breath‐holds. However, the repetition of two‐heartbeat breath‐holds may enable the use of shortened recovery periods, improve patient comfort, and increase the probability of successful breath‐holds.Bloch equations simulation was only performed on the employed inversion pulse to determine the inversion factor in the proposed 2P fitting model. However, Bloch equations simulation could also be used for the whole pulse sequence to generate the signal dictionary.24, 25, 26 Such an approach could be used to model the effect of the 2D readouts and provide improved accuracy of the T1 estimates, which will be the focus of future work.In this work, the slice profile of the inversion pulses was approximated by one flip angle. Alternatively, the use of subslice‐based simulations may improve the accuracy of the slice profile correction.25 Estimation of the inversion factor could include a larger B0 range if fatty tissues are considered. The use of a weighted average over the B0 and B1 ranges could also be considered to further improve the accuracy of the inversion factor estimates.In this work, PSIR was employed for all reconstructions. Alternatively, a multifitting approach could have been used for all these reconstructions.2 However, in our preliminary results (data not shown), we observed that the multifitting approach tended to fail to recover the correct signal polarity for the 3P fitting model in the presence of short T1 times. Although this technique was found robust for the 2P fitting model with all T1 ranges, we decided to use the PSIR approach for uniformity consideration.No in‐plane motion correction was employed in this work. Image registration algorithms may provide different performance based on the amount and contrast of images used during the registration process. Therefore, to prevent such bias during in vivo evaluation of T1 spatial variability and repeatability, we decided to discard the registration step of the reconstruction and discard datasets with inappropriate breath‐hold. Nevertheless, retrospective image registration has been shown to improve the robustness of myocardial T1 mapping.38, 39 Therefore, the design of tailored image registration algorithms for the proposed shortened T1 mapping schemes will be the focus of future work. Finally, the use of shortened T1 mapping schemes has the potential to improve the native registration of the T1‐weighted images in patients unable to sustain long stable breath‐hold. This will be evaluated in future work in a larger patient cohort.The evaluated shortened T1 mapping schemes showed varying degrees of HR dependence. These results are aligned with previous studies that demonstrated the HR dependence of MOLLI T1 times.16, 29, 30 The proposed HR correction models were found successful in reducing HR‐induced T1 variation to <10 msec for the entire T1 ranges. Alternative HR correction models have been proposed previously using a linear correction model based on measured T1 and HR.29, 30 In those studies, the slope and offset of the linear correction were assumed to be T1‐independent. Although we showed that a linear relationship between HR and measured T1 times is valid for a given T1 range, the linear regression slope and offset are also T1‐dependent. Therefore, T1‐dependent correction models were found more accurate than a simple T1‐independent linear model.The HR dependence of T1 times is mainly due to the inaccuracies of the employed fitting models, partly caused by their T2 dependence.16 In this work, we found that the HR dependence of myocardial and blood T1 times were different, which could be explained by their large T2 difference (>150 msec vs. ~45 msec). Therefore, we decided to reconstruct two differently HR‐corrected T1 maps per slice: one with myocardium‐based HR correction and one with blood‐based HR correction. Automatic segmentation of the blood pool based on thresholding of T1 maps has been previously proposed for ECV quantification.40 Such an approach could allow the selection of the appropriate HR correction model on a per‐voxel basis and may be used to generate a single HR corrected T1 map for both the myocardium and blood.This work has some limitations. First, the shortened T1 mapping schemes have been evaluated from a subset of a conventional 5‐(3)‐3 MOLLI scheme. This choice was made to minimize the number of required breath‐holds per subject. Second, the spatial variability measured as the SD over an ROI was used as a surrogate of the T1 precision, as commonly reported in prior studies.5, 6, 14, 17 However, this approach is susceptible to partial volume effects as well as artifacts, and thus may not fully represent the impact of noise in T1 time estimates. Third, trends were visually observed between MOLLI and the proposed shortened T1 mapping schemes in terms of spatial variability and repeatability in phantom. However, the differences did not reach statistical significance, which may be due to insufficient statistical power related to the limited number of vials available in our phantom. Fourth, the patient study was based on a small cohort of consecutive patients referred for clinical cardiac MRI. The benefit of these techniques in a larger patient cohort including patients with breath‐holding difficulties remains to be demonstrated and will be the focus of future studies. Furthermore, evaluation of this technique would be required in a cohort of patients with proven cardiac disease where mapping has clinical utility, such as with hypertrophic cardiomyopathy and Anderson‐Fabry's disease.In conclusion, the proposed two‐heartbeat T1 mapping scheme yields a 5‐fold acceleration compared with MOLLI, with highly linearly correlated native/postcontrast myocardial/blood T1 times, no significant difference of repeatability, and a limited spatial variability penalty at 1.5T. This approach may be a valuable alternative for myocardial T1 mapping in patients with severe breath‐holding difficulties and reduce examination time of multislice protocols.Supplementary Material 1 5‐(3)‐3 MOLLI pulse sequence and PSIR T1 map reconstruction schemes. In the pulse sequence, ECG‐triggered 2D bSSFP multi‐TI imaging is performed for each slice. 2P‐n/3P‐n represent the T1 map reconstruction schemes using the proposed 2P/3P fitting models with the first n images in the acquisition order, respectively.Supplementary Material 2. Simulated inversion recovery signal polarity as a function of T1 and T2 times for different flip angles (FAs). Bloch equations simulation of the sequence were used to determine the signal polarity of the shortest inversion time (TI) image (approximated as the signal polarity at the readout time of the k‐space center). Simulations were performed for different T1 times (range: 100‐300 ms), T2 times (range: 30‐300 ms) and FAs (range: 10‐85°). Red and blue regions indicate "positive" and “negative” polarities of the shortest TI image, respectively. Areas with negative polarity (blue) indicate conditions where the proposed PSIR assumption is valid. The minimum T1 times satisfying the proposed PSIR assumption increases with higher FAs and shorter T2 times. A T1 time of 172 ms was the lower bound ensuring the validity of the proposed PSIR assumption for the entire ranges of studied FAs/T2 times, which is smaller than the lower limit of the physiological ranges of native/post‐contrast T1 times in myocardium, blood and fat.Supplementary Material 3. HR dependence of T1 estimates using 2P‐2 and MOLLI in phantom experiments. Each subfigure represents a different vial. The reference T1 and T2 values are given for each vial. Vial‐wise linear regression (dashed lines) of T1 vs. HR was performed for 2P‐2 (green) and MOLLI (black). Individual linear dependence of T1 on HR was observed for each vial. A stronger HR dependence was observed in the presence of long T1 times and short T2 times.Supplementary Material 4. Dependence of T1‐HR linearity on T1 for 2P‐2 and MOLLI in phantom experiments. Parabolic regression (dashed curves) of slopes vs. offsets (cf. Supplementary Material 3) was performed for 2P‐2 (green) and MOLLI (black) on both short‐T2 and long‐T2 vials mimicking myocardium and blood, respectively.Supplementary Material 5. Representative example of ROIs used for myocardial and blood T1 quantification. The blue contours represent the segmented myocardial region while the black contour represents the area used for blood T1 analysis.Supplementary Material 6. Simulated T1 accuracy and precision as a function of T1 for different SNR. Although SNR had limited impact on T1 accuracy of all techniques, lower SNR resulted in precision penalty for all techniques as expected. However, SNR had limited influence on the relative precision penalty of all shortened T1 mapping schemes with respect to MOLLI which remained by a factor of 1.4‐1.5 for all SNR.Supplementary Material 7. T1 variation over different HRs (40‐120 bpm) before and after the proposed HR‐correction using all T1 mapping schemes. Each subfigure represents a different vial. The reference T1 and T2 values are given for each vial. The proposed HR correction reduced T1 variations to a maximum of 7 ms in all cases.Supplementary Material 8. Example native myocardial T1 maps obtained in a 32 yr female patient (HR 61 bpm) with severe left ventricular systolic dysfunction and pericardial effusion using all T1 mapping schemes (same patient as shown in Figure 7). This patient was unable to sustain a long stable breathhold for the entire duration of the acquisition of the mid‐ventricular and apical slices, which resulted in substantial artifacts in the corresponding MOLLI T1 maps (see black arrows). The shortened schemes provided good T1 map image quality for all slices.Click here for additional data file.
Authors: Daniel R Messroghli; Aleksandra Radjenovic; Sebastian Kozerke; David M Higgins; Mohan U Sivananthan; John P Ridgway Journal: Magn Reson Med Date: 2004-07 Impact factor: 4.668
Authors: Mehmet Akçakaya; Sebastian Weingärtner; Tamer A Basha; Sébastien Roujol; Steven Bellm; Reza Nezafat Journal: Magn Reson Med Date: 2015-09-29 Impact factor: 4.668
Authors: James C Moon; Daniel R Messroghli; Peter Kellman; Stefan K Piechnik; Matthew D Robson; Martin Ugander; Peter D Gatehouse; Andrew E Arai; Matthias G Friedrich; Stefan Neubauer; Jeanette Schulz-Menger; Erik B Schelbert Journal: J Cardiovasc Magn Reson Date: 2013-10-14 Impact factor: 5.364
Authors: Sebastian Weingärtner; Nadja M Meßner; Johannes Budjan; Dirk Loßnitzer; Uwe Mattler; Theano Papavassiliu; Frank G Zöllner; Lothar R Schad Journal: J Cardiovasc Magn Reson Date: 2016-11-18 Impact factor: 5.364
Authors: Joong H Kim; Stephen Dodd; Frank Q Ye; Andrew K Knutsen; Duong Nguyen; Haitao Wu; Shiran Su; Simone Mastrogiacomo; Thomas J Esparza; Rolf E Swenson; David L Brody Journal: Sci Rep Date: 2021-01-08 Impact factor: 4.996