| Literature DB >> 30662257 |
Gitansha Sachdev1, Dandapani Ramamurthy1.
Abstract
PURPOSE: To review the safety of the implantable phakic contact lens (IPCL V1, Caregroup Sight Solutions, India) for correction of myopia and myopic astigmatism.Entities:
Keywords: ICL; IPCL; implantable collamer lens; implantable phakic contact lens; myopia; myopic astigmatism; phakic intraocular lens; refractive correction
Year: 2019 PMID: 30662257 PMCID: PMC6327895 DOI: 10.2147/OPTH.S185304
Source DB: PubMed Journal: Clin Ophthalmol ISSN: 1177-5467
Preoperative demographics of patients undergoing implantable posterior contact lens implantation
| Characteristics | Mean ± SD | Range |
|---|---|---|
| Age (years) | 25.28±4.24 | 21 to 41 |
| Sphere (D) | −9.45±4.31 | −1.0 to −22.0 |
| Cylinder (D) | −1.72±1.26 | 0 to −6.0 |
| Manifest spherical equivalent (D) | −10.31±4.31 | −1.25 to −25.0 |
| Anterior chamber depth (mm) | 3.21±0.28 | 2.81 to 3.96 |
| Flat keratometry (D) | 43.60±1.44 | 40.5 to 46.75 |
| Steep keratometry (D) | 45.15±1.61 | 41.0 to 49.0 |
| Central corneal thickness (µm) | 509.81±41.32 | 423 to 606 |
| White to white (mm) | 11.59±1.06 | 11.03 to 12.55 |
| Endothelial cell density (mm2) | 2,754.58±235.94 | 2,510 to 3,353 |
| Axial length (mm) | 26.56±1.78 | 23.23 to 32.8 |
Figure 1Refractive outcomes following IPCL implantation.
Notes: (A) UDVA. (B) UDVA vs CDVA. (C) Change in CDVA. (D) Spherical equivalent refraction attempted vs achieved. (E) Spherical equivalent refraction accuracy. (F) Spherical equivalent refraction stability. (G) Refractive astigmatism.
Abbreviations: CDVA, corrected distance visual acuity; IPCL, implantable phakic contact lens; preop, preoperative; postop, postoperative; SEQ, spherical equivalent; UDVA, uncorrected distance visual acuity.