Literature DB >> 30659577

Intraoperative neuromonitoring versus visual nerve identification for prevention of recurrent laryngeal nerve injury in adults undergoing thyroid surgery.

Roberto Cirocchi1, Alberto Arezzo, Vito D'Andrea, Iosief Abraha, Georgi I Popivanov, Nicola Avenia, Chiara Gerardi, Brandon Michael Henry, Justus Randolph, Marcin Barczyñski.   

Abstract

BACKGROUND: Injuries to the recurrent inferior laryngeal nerve (RILN) remain one of the major post-operative complications after thyroid and parathyroid surgery. Damage to this nerve can result in a temporary or permanent palsy, which is associated with vocal cord paresis or paralysis. Visual identification of the RILN is a common procedure to prevent nerve injury during thyroid and parathyroid surgery. Recently, intraoperative neuromonitoring (IONM) has been introduced in order to facilitate the localisation of the nerves and to prevent their injury during surgery. IONM permits nerve identification using an electrode, where, in order to measure the nerve response, the electric field is converted to an acoustic signal.
OBJECTIVES: To assess the effects of IONM versus visual nerve identification for the prevention of RILN injury in adults undergoing thyroid surgery. SEARCH
METHODS: We searched CENTRAL, MEDLINE, Embase, ICTRP Search Portal and ClinicalTrials.gov. The date of the last search of all databases was 21 August 2018. We did not apply any language restrictions. SELECTION CRITERIA: We included randomised controlled trials (RCTs) comparing IONM nerve identification plus visual nerve identification versus visual nerve identification alone for prevention of RILN injury in adults undergoing thyroid surgery DATA COLLECTION AND ANALYSIS: Two review authors independently screened titles and abstracts for relevance. One review author carried out screening for inclusion, data extraction and 'Risk of bias' assessment and a second review author checked them. For dichotomous outcomes, we calculated risk ratios (RRs) with 95% confidence intervals (CIs). For continuous outcomes, we calculated mean differences (MDs) with 95% CIs. We assessed trials for certainty of the evidence using the GRADE instrument. MAIN
RESULTS: Five RCTs with 1558 participants (781 participants were randomly assigned to IONM and 777 to visual nerve identification only) met the inclusion criteria; two trials were performed in Poland and one trial each was performed in China, Korea and Turkey. Inclusion and exclusion criteria differed among trials: previous thyroid or parathyroid surgery was an exclusion criterion in three trials. In contrast, this was a specific inclusion criterion in another trial. Three trials had central neck compartment dissection or lateral neck dissection and Graves' disease as exclusion criteria. The mean duration of follow-up ranged from 6 to 12 months. The mean age of participants ranged between 41.7 years and 51.9 years.There was no firm evidence of an advantage or disadvantage comparing IONM with visual nerve identification only for permanent RILN palsy (RR 0.77, 95% CI 0.33 to 1.77; P = 0.54; 4 trials; 2895 nerves at risk; very low-certainty evidence) or transient RILN palsy (RR 0.62, 95% CI 0.35 to 1.08; P = 0.09; 4 trials; 2895 nerves at risk; very low-certainty evidence). None of the trials reported health-related quality of life. Transient hypoparathyroidism as an adverse event was not substantially different between intervention and comparator groups (RR 1.25; 95% CI 0.45 to 3.47; P = 0.66; 2 trials; 286 participants; very low-certainty evidence). Operative time was comparable between IONM and visual nerve monitoring alone (MD 5.5 minutes, 95% CI -0.7 to 11.8; P = 0.08; 3 trials; 1251 participants; very low-certainty evidence). Three of five included trials provided data on all-cause mortality: no deaths were reported. None of the trials reported socioeconomic effects. The evidence reported in this review was mostly of very low certainty, particularly because of risk of bias, a high degree of imprecision due to wide confidence intervals and substantial between-study heterogeneity. AUTHORS'
CONCLUSIONS: Results from this systematic review and meta-analysis indicate that there is currently no conclusive evidence for the superiority or inferiority of IONM over visual nerve identification only on any of the outcomes measured. Well-designed, executed, analysed and reported RCTs with a larger number of participants and longer follow-up, employing the latest IONM technology and applying new surgical techniques are needed.

Entities:  

Mesh:

Year:  2019        PMID: 30659577      PMCID: PMC6353246          DOI: 10.1002/14651858.CD012483.pub2

Source DB:  PubMed          Journal:  Cochrane Database Syst Rev        ISSN: 1361-6137


  83 in total

1.  Clinical value of intraoperative neuromonitoring of the recurrent laryngeal nerves in improving outcomes of surgery for well-differentiated thyroid cancer.

Authors:  Marcin Barczyński; Aleksander Konturek; Małgorzata Stopa; Alicja Hubalewska-Dydejczyk; Piotr Richter; Wojciech Nowak
Journal:  Pol Przegl Chir       Date:  2011-04

2.  The Efficacy of Intraoperative Neuromonitoring During Robotic Thyroidectomy: A Prospective, Randomized Case-Control Evaluation.

Authors:  Hye Yoon Lee; Ju-Young Lee; Gianlorenzo Dionigi; Jeoung Won Bae; Hoon Yub Kim
Journal:  J Laparoendosc Adv Surg Tech A       Date:  2015-11       Impact factor: 1.878

Review 3.  Full publication of results initially presented in abstracts.

Authors:  R W Scherer; P Langenberg; E von Elm
Journal:  Cochrane Database Syst Rev       Date:  2007-04-18

4.  Comparison of registered and published primary outcomes in randomized controlled trials.

Authors:  Sylvain Mathieu; Isabelle Boutron; David Moher; Douglas G Altman; Philippe Ravaud
Journal:  JAMA       Date:  2009-09-02       Impact factor: 56.272

5.  Vocal fold immobility after thyroidectomy with intraoperative recurrent laryngeal nerve monitoring.

Authors:  Irene de Pedro Netto; Jose Guilherme Vartanian; Jose Guilherme Vartarian; Pablo Rodrigo Rocha Ferraz; Priscila Salgado; Juliana Bueno Meirelles de Azevedo; Ronaldo Nunes Toledo; José Ricardo Gurgel Testa; Elisabete Carrara-de-Angelis; Luiz Paulo Kowalski
Journal:  Sao Paulo Med J       Date:  2007-05-03       Impact factor: 1.044

6.  Comparing the morbidity of external laryngeal nerve injury in thyroid surgery with and without identifying the nerve using intraoperative neuromonitoring.

Authors:  A O Khaled; M Irfan; A Baharudin; H Shahid
Journal:  Med J Malaysia       Date:  2012-06

7.  Recurrent laryngeal nerve electrophysiologic monitoring in thyroid surgery: the standard of care?

Authors:  Robert L Witt
Journal:  J Voice       Date:  2005-09       Impact factor: 2.009

Review 8.  "The final countdown": Is intraoperative, intermittent neuromonitoring really useful in preventing permanent nerve palsy? Evidence from a meta-analysis.

Authors:  Celestino Pio Lombardi; Giulia Carnassale; Gianfranco Damiani; Anna Acampora; Marco Raffaelli; Carmela De Crea; Rocco Bellantone
Journal:  Surgery       Date:  2016-08-24       Impact factor: 3.982

9.  The importance of surgeon experience for clinical and economic outcomes from thyroidectomy.

Authors:  J A Sosa; H M Bowman; J M Tielsch; N R Powe; T A Gordon; R Udelsman
Journal:  Ann Surg       Date:  1998-09       Impact factor: 12.969

10.  Protective Effects of Intraoperative Nerve Monitoring (IONM) for Recurrent Laryngeal Nerve Injury in Thyroidectomy: Meta-analysis.

Authors:  Binglong Bai; Wuzhen Chen
Journal:  Sci Rep       Date:  2018-05-17       Impact factor: 4.379

View more
  16 in total

Review 1.  Morbidity following thyroid surgery: acceptable rates and how to manage complicated patients.

Authors:  M N Minuto; S Reina; E Monti; G L Ansaldo; E Varaldo
Journal:  J Endocrinol Invest       Date:  2019-05-23       Impact factor: 4.256

Review 2.  Seeing Is Not Believing: Intraoperative Nerve Monitoring (IONM) in the Thyroid Surgery.

Authors:  Anuja Deshmukh; Anand Ebin Thomas; Harsh Dhar; Parthiban Velayutham; Gouri Pantvaidya; Prathamesh Pai; Devendra Chaukar
Journal:  Indian J Surg Oncol       Date:  2021-05-17

3.  The Effect of Energy Devices, Nerve Monitors, and Drains on Thyroidectomy Outcomes: An American College of Surgeons National Surgical Quality Improvement Project Database Analysis.

Authors:  John D Vossler; Kameko M Karasaki; Reid C Mahoney; Stacey L Woodruff; Kenric M Murayama
Journal:  Hawaii J Health Soc Welf       Date:  2021-11

Review 4.  Improving Voice Outcomes after Thyroid Surgery and Ultrasound-Guided Ablation Procedures.

Authors:  Pia Pace-Asciak; Jon O Russell; Ralph P Tufano
Journal:  Front Surg       Date:  2022-05-04

5.  Intraoperative neuromonitoring versus visual nerve identification for prevention of recurrent laryngeal nerve injury in adults undergoing thyroid surgery.

Authors:  Roberto Cirocchi; Alberto Arezzo; Vito D'Andrea; Iosief Abraha; Georgi I Popivanov; Nicola Avenia; Chiara Gerardi; Brandon Michael Henry; Justus Randolph; Marcin Barczyñski
Journal:  Cochrane Database Syst Rev       Date:  2019-01-19

6.  Intraoperative Neuromonitoring for Thyroid Surgery: The Proven Benefits and Limitations.

Authors:  Sung Yong Choi; Young-Ik Son
Journal:  Clin Exp Otorhinolaryngol       Date:  2019-10-02       Impact factor: 3.372

7.  The utility of intraoperative nerve monitoring in secondary and tertiary Hyperparathyroid surgery.

Authors:  Samuel Dudley; Nosratollah Nezakatgoo; Tyler M Rist; Theodore Klug; Courtney B Shires
Journal:  Clin Case Rep       Date:  2021-01-27

8.  The Impact of Coexistent Hashimoto's Thyroiditis on Central Compartment Lymph Node Metastasis in Papillary Thyroid Carcinoma.

Authors:  Yang Liu; Hongjun Lv; Shaoqiang Zhang; Bingyin Shi; Yushi Sun
Journal:  Front Endocrinol (Lausanne)       Date:  2021-11-16       Impact factor: 5.555

9.  Feasibility of Attachable Ring Stimulator for Intraoperative Neuromonitoring during Thyroid Surgery.

Authors:  Jongjin Kim; Hyeon Jong Moon; Young Jun Chai; Jung-Man Lee; Ki-Tae Hwang; Che-Wei Wu; Gianlorenzo Dionigi; Hoon Yub Kim; Kyung Sik Park; Sang Wan Kim; Ka Hee Yi
Journal:  Int J Endocrinol       Date:  2020-01-30       Impact factor: 3.257

10.  Role of intraoperative neuromonitoring of recurrent laryngeal nerve in thyroid and parathyroid surgery.

Authors:  Yuwei Ling; Jing Zhao; Ye Zhao; Kaifu Li; Yajun Wang; Hua Kang
Journal:  J Int Med Res       Date:  2020-09       Impact factor: 1.671

View more