| Literature DB >> 30560229 |
Christopher Smith1, Daniel Kaitis1, Jordan Winegar1, Sean Edelstein1, Matthew Council1, George Kontadakis2, Rocio Bentivegna1, Mohamed Abou Shousha3.
Abstract
PURPOSE: This study compared the effectiveness of endothelial/Descemet's membrane complex thickness obtained using high-definition anterior segment optical coherence tomography with endothelial cell density obtained using confocal microscopy as diagnostic tools in predicting corneal transplant rejection.Entities:
Keywords: Descemet’s membrane; corneal graft rejection; endothelial cell density; optical coherence tomography
Year: 2018 PMID: 30560229 PMCID: PMC6293363 DOI: 10.1177/2515841418814187
Source DB: PubMed Journal: Ther Adv Ophthalmol ISSN: 2515-8414
Characteristics of corneal grafts and mean values for ECD, En/DMT, CCT, polymegathism, and DRI.
| Clear group | Rejection group | ||
|---|---|---|---|
| Number of participants | 18 | 13 | |
| Number of eyes | 24 | 13 | |
| Age (years) | 67 ± 19 (27–92) | 53 ± 19 (31–80) | |
| Gender | |||
| Men | 13 | 5 | |
| Women | 5 | 8 | |
| Postoperative time (years) | 7 ± 7 (1–22) | 6.4 ± 11 (1–36) | |
| ECD (cells/mm2) | 1411 ± 714 (264–2069) | 870 ± 524 (212–1068) | |
| En/DMT (µm) | 15.0 ± 2.9 (12.0–20.0) | 43.0 ± 45.7 (16.0–151.6) | |
| CCT (µm) | 526 ± 67 (606–315) | 775 ± 243 (471–1117) | |
| Polymegathism (%) | 37.5 ± 12 (66.1–22.3) | 29.5 ± 7.5 (16.3–41.2) | |
| DRI | 0.95 ± 0.19 (0.65–1.25) | 1.66 ± 1.37 (0.76–4.77) | |
Values presented as means ± standard deviation (range).
CCT, central corneal thickness; DRI, En/DMT rejection index, defined as En/DMT divided by CCT multiplied by 33; ECD, endothelial cell density; En/DMT, Descemet’s membrane complex thickness.
Figure 1.The four images (a–d) represent a comparison of anterior segment high-definition optical coherence tomography images between clear (a and b) and rejection corneal grafts (c and d).Presets (b and d) show magnified images of the posterior part of the corresponding cornea (a and c), respectively. Descemet’s membrane complex is represented by the area posterior to the hyper-reflective line. This is delineated by arrows seen in b and d.
Median values for ECD, En/DMT, CCT, polymegathism, and DRI.
| Clear group | Rejection group | |
|---|---|---|
| ECD (cells/mm2) | 597 | 437 |
| En/DMT (µm) | 16.0 | 23.9 |
| CCT (µm) | 539 | 814 |
| DRI | 0.95 | 1.13 |
Values presented as median values.
CCT, central corneal thickness; DRI, En/DMT rejection index, defined as En/DMT divided by CCT multiplied by 33; ECD, endothelial cell density; En/DMT, Descemet’s membrane complex thickness.
Nonparametric correlation between ECD, En/DMT, CCT, and polymegathism (R value).
| En/DMT | CCT | Polymegathism | DRI | |
|---|---|---|---|---|
| ECD | −0.293 ( | −0.294 ( | −0.182 ( | −0.128 ( |
| En/DMT | x | 0.585 ( | −0.007 ( | 0.705 ( |
| CCT | x | x | 0.023 ( | 0.003 ( |
| Polymegathism | x | X | x | 0.012 ( |
Values presented as correlation coefficient (R value). x represents that the value is not applicable.
CCT, central corneal thickness. DRI, En/DMT rejection index; ECD, endothelial cell density. En/DMT, Descemet’s membrane complex thickness.
ROC curve data which represent the diagnostic performance of En/DMT, CCT, ECD, and polymegathism.
| Clear | ||
|---|---|---|
| En/DMT | <0.0001 | |
| AUC | 0.909 | |
| Sensitivity | 86% | |
| Specificity | 81% | |
| Cutoff | >16.0 µm | |
| CCT | 0.0029 | |
| AUC | 0.794 | |
| Sensitivity | 73% | |
| Specificity | 71% | |
| Cutoff | >563 µm | |
| ECD | 0.053 | |
| AUC | 0.744 | |
| Sensitivity | 71% | |
| Specificity | 71% | |
| Cutoff | ⩽897 cells/mm2 | |
| Polymegathism | 0.1 | |
| AUC | 0.292 | |
| Sensitivity | 71% | |
| Specificity | 70% | |
| Cutoff | ⩽30% |
AUC, area under the curve; CCT, central corneal thickness; ECD, endothelial cell density; En/DMT, Descemet’s membrane complex thickness; ROC, receiver operating characteristic.
Figure 2.This graph represents the combined receiver operating characteristic curves of ECD, En/DMT, CCT, and polymegathism. Increased area under the curve represents a better sensitivity and specificity for the diagnostic test. En/DMT had the highest sensitivity and specificity compared with the other parameters. CCT, central corneal thickness; ECD, endothelial cell density; En/DMT, Descemet’s membrane complex thickness.