Mariagabriella Pugliese1, Antonia Amatiello, Marco Correra, Vincenzo Stoia, Vincenzo Cerciello, Vincenzo Roca, Filomena Loffredo, Francesco Fiore, Giuseppe La Verde.
Abstract
BACKGROUND: Following recent epidemiological studies, which showed tissue reactions from ionizing radiation at significantly lower doses, the 2013/59 EURATOM Directive of 5th December 2013 lowered the limit on the equivalent dose to the eye lens from 150 mSv to 20 mSv per year. Therefore, as a precautionary measure, it is considered appropriate to perform a timely dose monitoring by using specific dosimeters.Entities:
Mesh:
Year: 2018 PMID: 30556537 PMCID: PMC7682182 DOI: 10.23749/mdl.v110i6.7286
Source DB: PubMed Journal: Med Lav ISSN: 0025-7818 Impact factor: 1.275
Performance of a TLD dosimeter for eye lens
| Measured variables | Hp(3) |
| Detectors | LiF(Mg, Cu, P), GR-200A |
| Filtration: | 1 mm ABS (about 20 mg/cm2) for the determination of Hp(3) values |
| Response interval in energy (n): | Photons: 20 keV- 3 MeV; |
| Dependence of the response in terms of Hp(3) in the measured energy range: | <±25% |
| Dependence of the response in terms of Hp(3) as a function of the angle of incidence (max 60°): | <±15% |
| Minimal detectable dose (with 95% confidence in routine procedures): | Not more than 20 µSv; |
| Minimum certifiable dose (with 95% confidence in routine procedures): | Not more than 50 µSv; |
| More information | Symmetrical dosimeter. |
Summary sheet of diagnostic exams of I and II Operators with related times and machine parameters
| Distance from the patient | Surgical interventions | Total running time | C-arm current | C-arm voltage |
| I Operator | ||||
| 0.5 m | Nephrostomy | 34 min | 3 mA | 90 kV |
| 0.5 m | Hepatic biliary drainage | 39 min | 3 mA | (86-90) kV |
| 0.5 m | Biliary stent replacement | 15 min | 3 mA | (86-90) kV |
| 0.5 m | Selective arteriography with coaxial catheter | 86 min | 3 mA | 86 kV |
| II Operator | ||||
| 0.5 m/1 m | Nephrostomy | 24 min | 3 mA | 90 kV |
| 1 m | Hepatic biliary drainage | 15 min | 3 mA | (86-90) kV |
| 0.5 m/1 m | Biliary stent replacement | 66 min | 3 mA | 86 kV |
| 0.5 m | Selective arteriography with coaxial catheter | 15 min | 3 mA | 86 kV |
Number of workers classified in A or B category stratified by Interventional Radiology Unit (IRU)
| IRU | Category A | Category B |
| NA01 | 11 | 0 |
| NA02 | 6 | 2 |
| NA03 | 4 | 2 |
| NA04 | 3 | 2 |
| NA05 | 7 | 0 |
| NA06 | 5 | 0 |
| NA07 | 5 | 0 |
| CE01 | 5 | 0 |
| Total | 46 (88%) | 6 (12%) |
Figure 1Number of answers (Yes, No) given by the operators to the question on the knowledge of the new dose limit to the lens of the eye by Interventional Radiology unit
Figure 2Number of operators for which the eye lens dosimetry is performed (twice a year), and the number of operators for which it is not carried out (never) by Interventional Radiology unit
Use of protection equipment in the examined structures
| Structures | Leaded coats | Collars | Glasses, masks, screens | Gloves |
| NA01 | 11 | 11 | 7 | 0 |
| NA02 | 7 | 5 | 2 | 0 |
| NA03 | 6 | 5 | 4 | 0 |
| NA04 | 5 | 5 | 0 | 0 |
| NA05 | 7 | 7 | 2 | 0 |
| NA06 | 5 | 4 | 3 | 0 |
| NA07 | 5 | 5 | 3 | 0 |
| CE01 | 5 | 5 | 2 | 0 |
| Total | 51 | 47 | 23 | 0 |