Munyaradzi Tambo1, Mary Mwinga2, Davis R Mumbengegwi1.
Abstract
Malaria cases sometimes go undetected using RDTs due to their inaccurate use, poor storage conditions and failure to detect low parasitaemia (<50parasites/μL). This could result in continuous transmission of malaria and sustenance of parasite reservoirs. Molecular diagnostic tools are more sensitive and specific than RDTs in the detection of plasmodium parasites. However, the Polymerase Chain Reaction (PCR) is not routinely used because equipment and reagents are expensive and requires highly skilled personnel. Loop-mediated isothermal amplification (LAMP) is a relatively new molecular diagnostic tool for malaria with all the advantages of PCR (sensitive and specific) without the mentioned disadvantages. However, it has not been evaluated extensively as a point of care diagnostic in the field. One hundred and fifteen used RDTs were collected from health facilities in Northern Namibia in a blind study and PCR and LAMP were used to determine the presence of Plasmodium DNA. The sensitivities and PPV were 40.91% and 90% respectively for RDTs, 72.73% and 100% respectively for PCR with LAMP as the golden standard. In low malaria transmission settings, LAMP can be also be considered for use as a surveillance tool to detect all sources of malaria and determine proportion of low parasitaemia infections in order to eliminate them.Entities:
Mesh:
Substances:
Year: 2018 PMID: 30540744 PMCID: PMC6291133 DOI: 10.1371/journal.pone.0206848
Source DB: PubMed Journal: PLoS One ISSN: 1932-6203 Impact factor: 3.240
Fig 1A gel showing PCR amplicons in the form of bands, indicating malaria infection.
Lane 1- positive control, lanes 2 & 11- negative control, lanes 3, 4, 5, 6, 7, & 8 are positive samples, lanes 9 & 10 are negative samples.
RDT vs LAMP diagnostic efficacy.
| 10.44% | |
| 0.9298 to 19.9331 | |
| 5.211 | |
| 1 | |
| P = 0.0224 |
PCR vs RDT diagnostic efficacy.
| 5.22% | |
| -3.6167 to 14.1067 | |
| 1.556 | |
| 1 | |
| P = 0.2122 |
Diagnostic evaluation testof RDTs and PCR with LAMP as the golden standard.
| Statistic | Sensitivity % (95% CI) | Specificity % (95% CI) | PPV % (95% CI) | NPV % (95% CI) |
|---|---|---|---|---|
| RDT | 40.91 (20.71–63.65) | 98.92 (94.15–99.97) | 90 (54.59–98.54) | 87.62 (83.32–90.95) |
| PCR | 72.73 (49.78–89.27) | 100 (96.11–100) | 100 | 93.94 (88.68–96.84) |
| LAMP | Reference | Reference | Reference | Reference |
Diagnostic perfomance of PCR vs RDTs with LAMP as the golden standard.
| Diagnostic tools | Number of true positives | Number of true negatives | Number of false positives | Number of false negatives |
|---|---|---|---|---|
| CareStart Rapid Diagnostic Test (RDT) | 9 | 93 | 1 | 12 |
| Polymerase Chain Reaction (PCR) | 16 | 93 | 0 | 6 |
| Loop-mediated isothermal AMPlification (LAMP) | Ref | Ref | Ref | Ref |