| Literature DB >> 30511680 |
Xue-Ying Cao1, Ya-Ni He2, Jian-Hui Zhou1, Shi-Ren Sun3, Li-Ning Miao4, Wen Chen5, Jing-Ai Fang6, Ming Wang7, Nian-Song Wang8, Hong-Li Lin9, Jian Liu10, Zhao-Hui Ni11, Wen-Hu Liu12, Yu Na13, Jiu-Yang Zhao14, Zhi-Yong Guo15, Hong-Guang Zheng16, Wei Shi17, Geng-Ru Jiang18, Guang-Yan Cai1, Xiang-Mei Chen1.
Abstract
BACKGROUND: Automated peritoneal dialysis (APD) can cater to individual needs, provide treatment while asleep, take into account the adequacy of dialysis, and improve the quality of life. Currently, independent research and development of APD machines made in China are more conducive to patients. A randomized, multicenter, crossover study was conducted by comparing an APD machine made in China with an imported machine. The safety, effectiveness, and manipulability of the two machines were compared.Entities:
Keywords: Automated Peritoneal Dialysis; Effectiveness; Peritoneal Dialysis; Peritoneal Dialysis Machine; Safety
Mesh:
Year: 2018 PMID: 30511680 PMCID: PMC6278200 DOI: 10.4103/0366-6999.246079
Source DB: PubMed Journal: Chin Med J (Engl) ISSN: 0366-6999 Impact factor: 2.628
Figure 1Schematic flow of this study.
Baseline characteristics of the patients undergoing maintenance PD in the entry group (n = 260)
| Characteristics | Values |
|---|---|
| Age (years) | 49.3 ± 15.3 |
| Gender, male, | 158 (60.8) |
| Height (cm) | 166.5 ± 8.1 |
| Weight (kg) | 65.7 ± 13.2 |
| History of cardiovascular disease, | |
| Normal | 239 (91.9) |
| Abnormal | 21 (8.1) |
| Systolic pressure (mmHg) | 145.5 ± 20.2 |
| Diastolic pressure (mmHg) | 87.4 ± 12.7 |
| CAPD during the screening period (6 L/d), | |
| No | 15 (5.8) |
| Yes | 245 (94.2) |
Data were shown as mean ± SD, or n (%). SD: Standard deviation; CAPD: Continuous ambulatory peritoneal dialysis; PD: Peritoneal dialysis.
Removal of serum creatinine, urea, and correction of electrolyte and acid-base balance disorders before and after treatment (n = 260)
| Variables | FM-APD | HC-APD | ||
|---|---|---|---|---|
| BUN (mmol/L) | ||||
| Before | 18.55 ± 6.06 | 18.41 ± 5.91 | 0.26 | 0.7952 |
| After | 17.26 ± 5.62 | 16.92 ± 5.36 | 0.71 | 0.4768 |
| Serum creatinine (µmol/L) | ||||
| Before | 882.16 ± 317.14 | 869.13 ± 314.77 | 0.47 | 0.6390 |
| After | 837.78 ± 304.06 | 827.84 ± 302.69 | 0.37 | 0.7101 |
| Na (mmol/L) | ||||
| Before | 140.55 ± 3.62 | 140.59 ± 3.48 | −0.01 | 0.9930 |
| After | 140.55 ± 3.42 | 139.76 ± 9.22 | 1.34 | 0.1796 |
| K (mmol/L) | ||||
| Before | 3.91 ± 0.61 | 3.88 ± 0.59 | −0.05 | 0.9598 |
| After | 3.91 ± 0.61 | 3.91 ± 0.61 | −0.47 | 0.6359 |
| Ca (mmol/L) | ||||
| Before | 2.22 ± 0.23 | 2.25 ± 0.24 | −0.99 | 0.3233 |
| After | 2.57 ± 0.28 | 2.26 ± 0.21 | −0.73 | 0.4653 |
| P (mmol/L) | ||||
| Before | 2.43 ± 0.43 | 1.45 ± 0.44 | 1.01 | 0.3119 |
| After | 1.55 ± 0.47 | 1.46 ± 0.45 | −0.22 | 0.8242 |
| CO2 (mmol/L) | ||||
| Before | 26.33 ± 3.33 | 27.20 ± 3.97 | −0.64 | 0.5227 |
| After | 26.52 ± 3.40 | 30.78 ± 4.21 | −1.24 | 0.2163 |
FM-APD: Domestic FM peritoneal dialysis machine; HC-APD: Homechoice peritoneal dialysis machine; BUN: Blood urea nitrogen.
Comparison of occurrence and frequency of different machine adverse events (n = 260)
| Items | FM-APD | HC-APD | Total | |||
|---|---|---|---|---|---|---|
| Cases | Incidence (%) | Cases | Incidence (%) | Cases | Incidence (%) | |
| Finding intestinal tract | 1 | 0.38 | 0 | 0 | 1 | 0.19 |
| Abdominal pain | 0 | 0 | 1 | 0.38 | 1 | 0.19 |
| Urinary retention | 1 | 0.38 | 0 | 0 | 1 | 0.19 |
| Backache | 0 | 0 | 1 | 0.38 | 1 | 0.19 |
FM-APD: Domestic FM peritoneal dialysis machine; HC-APD: Homechoice peritoneal dialysis machine.
Comparison of the fuselage characteristics of two kinds of PD machines (n = 260)
| Items | FM-APD | HC-APD | ||
|---|---|---|---|---|
| Fuselage space, | ||||
| Space-saving | 108 (41.9) | 218 (83.9) | 102.8308 | <0.0001 |
| Requires part of the space | 148 (57.4) | 38 (14.6) | ||
| Requires a lot of space | 2 (0.8) | 4 (1.5) | ||
| Any noise in the fuselage, | ||||
| No noise | 144 (55.8) | 156 (6.0) | 3.139 | 0.2082 |
| Mild noise | 114 (44.2) | 102 (39.2) | ||
| Obvious noise | 0 (0) | 2 (0.8) | ||
| Convenience, | ||||
| Inconvenient | 3 (1.2) | 139 (53.5) | 178.0038 | <0.0001 |
| Convenient | 255 (98.8) | 121 (46.5) | ||
| Continued dialysate liquid storage space, | ||||
| No | 2 (0.8) | 168 (64.6) | 239.3773 | <0.0001 |
| Yes | 256 (99.2) | 92 (35.4) | ||
| Storage function after power failure, | ||||
| No | 30 (11.6) | 75 (28.9) | 23.7553 | <0.0001 |
| Yes | 228 (88.4) | 185 (71.2) | ||
| Convenient and clean fuselage, | ||||
| No | 0 (0) | 4 (1.5) | 4.0001 | 0.0455 |
| Yes | 258 (100) | 256 (98.5) |
FM-APD: Domestic FM peritoneal dialysis machine; HC-APD: Homechoice peritoneal dialysis machine; PD: Peritoneal dialysis.
Comparison of the operating characteristics of two kinds of PD machines (n = 260)
| Items | FM-APD | HC-APD | ||
|---|---|---|---|---|
| Set the operating prompt clearly, | ||||
| Very clear | 249 (96.5) | 195 (75.0) | 48.9989 | <0.0001 |
| Sometimes not clear | 9 (3.5) | 64 (24.6) | ||
| Vague | 0 (0) | 1 (0.4) | ||
| Man-machine dialog clearly at the alarm, | ||||
| Clear Chinese prompt | 245 (95.0) | 203 (78.1) | 31.7274 | <0.0001 |
| Code prompt | 13 (5.0) | 56 (21.5) | ||
| No prompt | 0 (0) | 1 (0.4) | ||
| Set comprehensively, | ||||
| No | 0 (0) | 25 (9.6) | 26.8151 | <0.0001 |
| Yes | 258 (100) | 235 (90.4) | ||
| Adjusted alarm sound, | ||||
| No | 3 (1.2) | 35 (13.5) | 28.8151 | <0.0001 |
| Yes | 255 (98.8) | 225 (86.5) | ||
| Changed during the process, | ||||
| No | 34 (13.2) | 129 (49.6) | 79.7249 | <0.0001 |
| Yes | 224 (86.8) | 131 (50.4) | ||
| Identified multiple information sources, | ||||
| No | 2 (0.8) | 129 (49.6) | 163.4915 | <0.0001 |
| Yes | 256 (99.2) | 131 (50.4) |
FM-APD: Domestic FM peritoneal dialysis machine; HC-APD: Homechoice peritoneal dialysis machine; PD: Peritoneal dialysis.