Colman Moore1, Yuting Bai1, Ali Hariri1, Joan B Sanchez2, Ching-Yu Lin1, Sreenivas Koka3,4, Parish Sedghizadeh2, Casey Chen2, Jesse V Jokerst1,5,6. 1. Department of NanoEngineering, University of California, San Diego, 9500 Gilman Drive. La Jolla, CA, 92092, USA. 2. Herman Ostrow School of Dentistry, University of Southern California, 925 West 34th Street, Los Angeles, CA, 90089, USA. 3. School of Dentistry, University of California, Los Angeles, 714 Tiverton Ave, Los Angeles, CA, 90024, USA. 4. Koka Dental Clinic, 8031 Linda Vista Rd, San Diego, CA, 92111, USA. 5. Materials Science Program, University of California, San Diego, 9500 Gilman Drive. La Jolla, CA, 92092, USA. 6. Department of Radiology, University of California, San Diego, 9500 Gilman Drive. La Jolla, CA, 92092, USA.
Abstract
The gold-standard periodontal probe is an aging tool that can detect periodontitis and monitor gingival health but is highly error-prone, does not fully characterize the periodontal pocket, and causes pain. Photoacoustic imaging is a noninvasive technique that can address these limitations. Here, a range of ultrasound frequencies between 16-40 MHz were used to image the periodontium and a contrast medium based on cuttlefish ink was used to label the pockets. A 40 MHz ultrasound frequency could spatially resolve the periodontal anatomy, including tooth, gum, gingival margin, and gingival thickness of tooth numbers 7-10 and 22-27. The photoacoustic-ultrasound measurements were more precise (0.01 mm) than those taken with physical probes by a dental hygienist. Furthermore, the full geometry of the pockets could be visualized with relative standard deviations of 10% (n = 5). This study shows the potential for non-invasive monitoring of periodontal health with photoacoustic-ultrasound imaging in the dental clinic.
The gold-standard periodontal probe is an aging tool that can detect periodontitis and monitor gingival health but is highly error-prone, does not fully characterize the periodontal pocket, and causes pain. Photoacoustic imaging is a noninvasive technique that can address these limitations. Here, a range of ultrasound frequencies between 16-40 MHz were used to image the periodontium and a contrast medium based on cuttlefish ink was used to label the pockets. A 40 MHz ultrasound frequency could spatially resolve the periodontal anatomy, including tooth, gum, gingival margin, and gingival thickness of tooth numbers 7-10 and 22-27. The photoacoustic-ultrasound measurements were more precise (0.01 mm) than those taken with physical probes by a dental hygienist. Furthermore, the full geometry of the pockets could be visualized with relative standard deviations of 10% (n = 5). This study shows the potential for non-invasive monitoring of periodontal health with photoacoustic-ultrasound imaging in the dental clinic.
Periodontitis affects nearly 50% of Americans and exerts both local and systemic effects on the body [1,2]. These range from mild discomfort to debilitating pain, tooth loss, and excessive activation of the immune system [[3], [4], [5]]. In fact, studies have identified the chronic inflammation from periodontitis as a risk factor for conditions such as cardiovascular disease [[6], [7], [8]], cancer [9], and dementia [10,11]. Thus, it is critical to diagnose periodontal disease early while the symptoms are mild and reversible [12].The current metrics for monitoring periodontal health include attachment level, probing depth, bone loss, mobility, recession, and degree of inflammation [13]. The gold standard for quantifying periodontitis progression is the measurement of clinical attachment loss with a periodontal probe. This tool provides a numerical metric that reflects the degree of apical epithelial attachment measured from the gingival margin and is critical for disease staging [14].However, the periodontal probe is a relatively unsophisticated tool that suffers from poor reproducibility [14,15]. These errors are attributed to the significant variation in probing force between operators, which can vary across multiple orders of magnitude [16]. Physical probing can penetrate inflamed tissue leading to patient discomfort, bleeding, and inaccurate measurements [5,17]. The probe can also only measure depth at the point of insertion—with no information on the full width or contour of the pocket. Furthermore, the benefit of traditional probing around dental implants is abrogated due to the implant threads that impede probe penetration along the implant surface [18,19]. This limits clinical assessment of these tissues potentially leading to peri-implantitis [20,21].We recently described an approach that uses photoacoustic imaging to potentially address these issues [22]. Photoacoustic imaging is a hybrid imaging modality that combines visible and near infrared excitation with acoustic detection [[23], [24], [25], [26]]. It extends the utility of ultrasound by enabling contrast based on optical absorption [27,28]. Traditional ultrasound operates under the principle of “sound in, sound out”, but photoacoustic imaging shifts this concept to “light in, sound out” [29]. Here, a near infrared laser excites a light-absorbing target. The target then undergoes spatially confined heating followed by thermoelastic expansion. This generates wideband acoustic waves that can be detected with ultrasound transducers for image generation. The clinical applications of photoacoustic imaging are growing; some recent examples include endogenous imaging of inflammatory arthritis, noninvasive assessment of Crohn’s Disease, and diagnosis of vascular malformations [[30], [31], [32], [33], [34], [35]]. Though photoacoustic-ultrasound (PA-US) signal can be limited by tissue penetration, it has two significant advantages over radiography, the most common dental imaging modality [36]: (1) it can image soft tissue; and (2) it does not use ionizing radiation. Additionally, the surfaces of hard tissues such as bone [37] and teeth [38] can be imaged.Our previous report demonstrated the first application of photoacoustics for periodontal health and used a biocompatible oral contrast agent based on food-grade cuttlefish ink [22]. That work utilized swine models and imaged nearly 40 teeth to validate the method. While bias values relative to gold standard were low (<0.2 mm), unresolved questions remained about its clinical utility. Here, we show that photoacoustic-ultrasound can image both the full depths and geometries of pockets in healthy human adults for non-invasive monitoring of gingival health after local irrigation of the pocket with contrast media.
Materials and methods
Reagents and equipment
Cuttlefish ink was purchased from Conservas de Cambados and contained ink, water, salt, and sodium carboxymethyl cellulose. Phosphate-buffered saline tablets were purchased from Sigma-Aldrich. Ultrasound gel was obtained from Next Medical Products. PA-US images were taken with a laser-integrated high frequency ultrasound system from Visualsonics (Vevo LAZR). A medical head immobilizer was purchased from DealMed. Disposable dental cheek retractors were obtained from Url Dental.Contrast agent solutions were prepared individually from stock solutions of cuttlefish ink for each imaging experiment. Stock cuttlefish ink solutions (50% w/v in 0.1 M PBS) were aliquoted and refrigerated. To prepare the contrast agent, an aliquot was diluted and mixed with corn starch to a final solution containing 5% ink w/v and 2% corn starch. It was briefly heated to boiling to achieve homogeneity. The spherical melanin nanoparticles within the contrast agent were previously characterized with transmission electron microscopy, which indicated a mean particle size of 125 nm from 500 nanoparticles; and dynamic light scattering, which showed a hydrodynamic radius of 266 nm with a polydispersity index of 0.116 [22]. It is a food-grade material and was approved for human use by the Institutional Review Board (IRB). Human subject exclusion criteria included a shellfish allergy or a kosher diet.
Volunteer recruitment
This case study enrolled a healthy 22-year old adult female with good oral hygiene. All work with human subjects was approved by the UCSD IRB and conducted according the ethical standards set forth by the IRB and the Helsinki Declaration of 1975. The participant gave written informed consent and teeth 7–10, 22–27 were imaged.
Periodontal probe measurements
Pockets were measured with the Williams and Marquis probes [39] by a licensed periodontist. The measurements were performed at the distobuccal, mesiobuccal, and buccal sites according to clinical convention [40]. For distal and mesial measurements, the probe was inserted at a 10° angle at the interproximal space between adjacent teeth. The buccal measurements were collected at the deepest point observed after walking the probe across the width of the pocket. Per standard clinical practice, measurements were rounded up to the nearest integer. For the Williams probe, all measurements <2 mm were recorded as 2 mm. For the Marquis probe, readings within the 3 mm intervals were estimated to be either in the lower 1.5 mm or upper 1.5 mm range and then rounded up to the nearest integer.
Pocket labeling
The subgingival pockets were labelled with ∼8 μL of contrast agent per tooth. A micropipette with a sterile 2–20 μL tip was placed in contact with the gingival sulcus and used to irrigate the region with contrast agent. Following imaging, the contrast agent was removed from the pockets by rinsing the mouth with water or gentle tooth brushing (< 10 s).
Imaging procedure
Photoacoustic imaging was performed by pulsing light through two optical fiber bundles integrated with both sides of a rectangular, linear array transducer. The laser excitation (Q-switched Nd:YAG) used 5 ns pulses at 20 Hz (6 Hz frame rate). The ultrasound resolution was controlled by changing between three transducers: LZ-201 (center frequency = 16 MHz), LZ-250 (center frequency = 21 MHz), and LZ-550 (center frequency = 40 MHz). Typical gains were 15 dB for photoacoustic signal and 10 dB for ultrasound. The subject and operator both wore near-infrared protective laser goggles during experiments. The subject was seated in front of the imaging system and rested their chin on a flat surface in front of the transducer. The subject wore dental cheek retractors and a medical head immobilizer to reduce movements that contribute to motion-based imaging artifacts. A layer of ultrasound gel was applied to the transducer, which was adjusted to 1 cm from the tooth. The 680 nm laser was initialized, and the stepper motor was scanned 17 mm (0.076 mm step size) to obtain a 3D PA-US image via maximum intensity projection [41].
Image analysis and statistics
Following raw data acquisition, sagittal cross-sections were examined in ImageJ to determine the penetration of contrast agent and quantify pocket depths. First, images were converted to 8-bit images and exported as two files: one that displayed only the photoacoustic signal and another that displayed only the ultrasound. For these images, a line profile was manually drawn parallel to the gingival margin through both images (Fig. S2). From this, the pixel intensity could be plotted with respect to the position along the line for both the photoacoustic and ultrasound images. Then, a minimum pixel intensity of 4% was used as a threshold, above which signal was considered significant. If both signals were above the threshold at a given position, then that point was considered part of the pocket depth. Taken together, all of these points formed a length and its magnitude was recorded as the pocket depth.To avoid bias during quantitative comparison to physical probing, sagittal planes were chosen the same way each time: for distobuccal and mesiobuccal sites, we chose the first 8 sagittal planes (0.6 mm-wide sections) with a measurable pocket depth on each lateral side of the tooth. These sections mimicked the diameters of the physical probes. For the buccal sites, we selected 0.6 mm-wide sections from the images at the deepest part of the pocket. This dimension mimicked the diameter of the physical probes and the typical procedure of recording the lowest number obtained by walking the probe across the pocket width. For replicate measurements, images were collected on nonconsecutive days across two weeks.The buccal contours of the pocket geometry were mapped by averaging five separate imaging events. The width of the pocket consisted of dozens of sagittal planes. For each plane, two measurements were taken: the distance from the crown of the tooth to the gingival margin, and the distance from the gingival margin to the edge of photoacoustic signal (the subgingival terminus). Together, these measurements determined the top and bottom of the pocket for that plane. Once these measurements were taken for all planes, they were used to reconstruct a full map of the pocket to overlap on top of the ultrasound image.
Results
High-resolution ultrasound imaging of teeth and soft tissues
Photoacoustic and ultrasound images were collected by adapting a commercial PA-US system for human operation (Fig. 1). Ultrasound imaging was used to generate frontal and sagittal cross-sections of 10 teeth (tooth numbers 7–10, 22–27) and their associated soft tissues from a healthy volunteer (Fig. 2). They were labelled with the numbering system used by the American Dental Association [42]. These anterior teeth were imaged because the size of the transducer limited access to the posterior region of the oral cavity.
Fig. 1
Overview of the imaging setup, periodontal anatomy, and workflow. (A) The PA-US transducer was connected to a stepper motor for axial scanning and a sliding positioner for lateral control—these components were attached to a larger frame for general positioning. The head immobilizer (cross-section shown) rested on a flat surface. (B) The simplified diagram shows a tooth, gingiva, gingival margin, subgingival terminus, and pocket depth for an arbitrary cross-section. (C) Anatomical terms for periodontal reference: Mesial refers to a location on a tooth that is closer to the midline while distal refers to one that is further. Lingual refers to the side of a tooth facing the tongue while buccal refers to the side facing the cheek. For example, mesiobuccal and distobuccal refer to the areas of the buccal face of a tooth that are closer to and farther from the midline of the mouth, respectively. (D) Experimental workflow. Briefly, a contrast agent containing melanin nanoparticles (TEM inset) derived from cuttlefish ink was used to irrigate the pocket of a target tooth followed by photoacoustic imaging. The pocket could be rinsed with water and then images could be analyzed to measure pocket depths.
Fig. 2
US images of ten teeth acquired from one subject at 40 MHz and comparison to images at lower frequencies. (A) Frontal and sagittal images of the bottom front incisors (24 and 25) at increasing US frequencies. Frontal images are inset within the magnified sagittal images of tooth 25 at the gingival margin. Scale bars are 1 mm and apply to all frequencies for the sagittal images. (B) Frontal and sagittal cross-sections of teeth 7–10 and 22–27. The frontal and sagittal views are composites of planes from ten separate scans. The gingiva, gingival margin, and tooth are resolved in each sagittal image. The 3-mm scale bar is for sagittal images.
Overview of the imaging setup, periodontal anatomy, and workflow. (A) The PA-US transducer was connected to a stepper motor for axial scanning and a sliding positioner for lateral control—these components were attached to a larger frame for general positioning. The head immobilizer (cross-section shown) rested on a flat surface. (B) The simplified diagram shows a tooth, gingiva, gingival margin, subgingival terminus, and pocket depth for an arbitrary cross-section. (C) Anatomical terms for periodontal reference: Mesial refers to a location on a tooth that is closer to the midline while distal refers to one that is further. Lingual refers to the side of a tooth facing the tongue while buccal refers to the side facing the cheek. For example, mesiobuccal and distobuccal refer to the areas of the buccal face of a tooth that are closer to and farther from the midline of the mouth, respectively. (D) Experimental workflow. Briefly, a contrast agent containing melanin nanoparticles (TEM inset) derived from cuttlefish ink was used to irrigate the pocket of a target tooth followed by photoacoustic imaging. The pocket could be rinsed with water and then images could be analyzed to measure pocket depths.US images of ten teeth acquired from one subject at 40 MHz and comparison to images at lower frequencies. (A) Frontal and sagittal images of the bottom front incisors (24 and 25) at increasing US frequencies. Frontal images are inset within the magnified sagittal images of tooth 25 at the gingival margin. Scale bars are 1 mm and apply to all frequencies for the sagittal images. (B) Frontal and sagittal cross-sections of teeth 7–10 and 22–27. The frontal and sagittal views are composites of planes from ten separate scans. The gingiva, gingival margin, and tooth are resolved in each sagittal image. The 3-mm scale bar is for sagittal images.Increasing the ultrasound transducer frequency (Fig. 2A) significantly improved our ability to identify the gingival margin. The spatial resolution of the ultrasound mode was approximately 100 μm at 40 MHz and 300 μm for the photoacoustic mode [22,43]. At 15 and 21 MHz, the gingival-tooth interface and subgingival tooth could not be identified. These features were clearly resolved at 40 MHz, and we used this frequency for all subsequent images. Frontal views captured surface features of the teeth and tissue, while sagittal views revealed the gingival margin, gingival thickness, and surface topology (Fig. 2B).
Periodontal labeling and photoacoustic imaging
The contrast agent possessed broad photoacoustic absorption from 680 to 970 nm due to the presence of melanin nanoparticles (Fig. S1) [22,44,45]. Following administration of contrast agent, we collected PA-US images and overlaid the 680-nm signal on ultrasound to measure the periodontal pocket depths. Frontal and sagittal views of the lower central incisors in Fig. 3A–D show the signal from the mid-buccal pocket after applying the contrast agent and imaging. The mid-buccal pocket depth of tooth number 25 was 1.31 mm (Fig. 3D); signal from the supragingival surface of the tooth did not contribute to the measurement. The gingival thickness could also be quantified. When measured at the mid-buccal plane, 2 mm below the gingival margin, the gingival thickness was 1.01 ± 0.05 mm for tooth 25 and 0.97 ± 0.04 mm for tooth 24 (n = 5). These values matched the average thickness reported for a 16–24 year old age group [46].
Fig. 3
Photoacoustic-ultrasound images of the bottom front incisors (24, 25) before and after administration of contrast agent with demonstrated reproducibility. (A) Frontal cross-section of teeth 24 and 25 with photoacoustic signal overlaid on ultrasound prior to administration of contrast agent. (B) Sagittal cross-section of tooth 25, indicated by the dashed yellow line in Panel A, with tooth, gingival margin, and gingival thickness labelled. (C) Frontal cross-section of teeth 24 and 25 with photoacoustic signal overlaid on ultrasound after administration of contrast agent. (D) The sagittal cross-section of tooth 25 (dashed yellow line in Panel C) shows the soft-hard tissue interface and pocket revealed by the photoacoustic contrast agent below the gingival margin. (E, F) Reproducibility of measurements for representative sagittal planes at mesiobuccal, central buccal, and distobuccal probing locations across five replicates in tooth 25; labeling and imaging procedures were performed independently from beginning to end. Each PA-US image in F corresponds to the mesiobuccal probing depth measurement (squares) for its given replicate number on the x-axis in E. Light yellow lines show the pocket measurements. White: ultrasound, red: photoacoustic.
Photoacoustic-ultrasound images of the bottom front incisors (24, 25) before and after administration of contrast agent with demonstrated reproducibility. (A) Frontal cross-section of teeth 24 and 25 with photoacoustic signal overlaid on ultrasound prior to administration of contrast agent. (B) Sagittal cross-section of tooth 25, indicated by the dashed yellow line in Panel A, with tooth, gingival margin, and gingival thickness labelled. (C) Frontal cross-section of teeth 24 and 25 with photoacoustic signal overlaid on ultrasound after administration of contrast agent. (D) The sagittal cross-section of tooth 25 (dashed yellow line in Panel C) shows the soft-hard tissue interface and pocket revealed by the photoacoustic contrast agent below the gingival margin. (E, F) Reproducibility of measurements for representative sagittal planes at mesiobuccal, central buccal, and distobuccal probing locations across five replicates in tooth 25; labeling and imaging procedures were performed independently from beginning to end. Each PA-US image in F corresponds to the mesiobuccal probing depth measurement (squares) for its given replicate number on the x-axis in E. Light yellow lines show the pocket measurements. White: ultrasound, red: photoacoustic.To demonstrate reproducibility, we conducted five independent replicates by performing the entire labelling, imaging, cleaning, and processing procedure from start to finish on different days (Fig. S2). In Fig. 3E–F, the pocket depths are plotted across these replicates for buccal, mesiobuccal, and distobuccal planes. The means and standard deviations for these locations were 1.24 ± 0.04 mm, 0.74 ± 0.12 mm, and 0.68 ± 0.06 mm respectively. These corresponded to relative standard deviations of 3%, 16%, and 9%.The pocket depth could be determined for a single plane, a region of planes, or the full buccal width of the pocket. The distance between each plane was 0.076 mm, corresponding to roughly 50–120 measurements per tooth. Fig. 4 shows the steps to construct a visualization of the full pocket overlaid on a frontal US image of the tooth. The measurements for these steps were performed five times (n = 5) on a lower central incisor across its measurable width (Fig. 4A–B). The total mean for the full pocket width was 1.02 ± 0.21 mm. The mean standard deviation of these measurements, i.e. the average of the standard deviations for every measurement plane across five replicates, was 0.104 mm, resulting in a relative standard deviation of 10.2% (Fig. 4E). Furthermore, the range of the measurements was 0.57 mm, reflecting the variation in pocket depth according to probing location. The distobuccal, buccal, and mesiobuccal regions of planes (0.6 mm wide each) had measurements of 0.84 ± 0.17 mm, 1.15 ± 0.05, and 0.79 ± 0.05. Fig. 4C shows a sagittal plane from Fig. 4B split into photoacoustic and ultrasound signal only with a line profile drawn parallel to the gingival margin. Fig. 4D quantifies the intensity of each signal along the line profile, which indicates a pocket depth of 1.01 mm for the plane. The final mapping of the pocket geometry is shown overlaid on the ultrasound image in Fig. 4F.
Fig. 4
Procedure for mapping the full width and contours of the periodontal pocket. (A) US frontal image of tooth 25 prior to administration of contrast agent. (B) PA-US frontal image of the same tooth following contrast agent with the subgingival region of interest highlighted by a black box, prior to image processing. (C) A sagittal view of the tooth at the yellow dashed line from (A) and (B) used for image processing. The top of panel C is an overlay and the lower panels are ultrasound and photoacoustic layers. A line (solid turquoise) was manually drawn parallel to the gingival margin for both of the layers using ImageJ. (D) The 8-bit pixel intensities of the ultrasound and photoacoustic signal along the line were plotted with respect to the length of the line. A minimum intensity threshold (dashed line) of 10 was chosen to minimize the influence of background signal, and the pocket depth was taken as the total length for which the photoacoustic and ultrasound signals were both higher than the threshold, i.e., from 2.71 mm to 3.72 mm. (E) Pocket measurements for every sagittal plane across the full buccal width of the tooth– a total of 40 planes were measured. The x-axis represents each sagittal slice across the width of the tooth. The shaded region represents the geometry of the pocket and the white region represents the normal area of attachment between tooth and epithelia. Error bars show the standard error of the mean for five replicates (n = 5). (F) The final mapping of the pocket generated by our photoacoustic method. It was generated by taking the measurements for each sagittal slice from (E) and overlaying them on the ultrasound image. This step removes the non-specific photoacoustic signal from (B).
Procedure for mapping the full width and contours of the periodontal pocket. (A) US frontal image of tooth 25 prior to administration of contrast agent. (B) PA-US frontal image of the same tooth following contrast agent with the subgingival region of interest highlighted by a black box, prior to image processing. (C) A sagittal view of the tooth at the yellow dashed line from (A) and (B) used for image processing. The top of panel C is an overlay and the lower panels are ultrasound and photoacoustic layers. A line (solid turquoise) was manually drawn parallel to the gingival margin for both of the layers using ImageJ. (D) The 8-bit pixel intensities of the ultrasound and photoacoustic signal along the line were plotted with respect to the length of the line. A minimum intensity threshold (dashed line) of 10 was chosen to minimize the influence of background signal, and the pocket depth was taken as the total length for which the photoacoustic and ultrasound signals were both higher than the threshold, i.e., from 2.71 mm to 3.72 mm. (E) Pocket measurements for every sagittal plane across the full buccal width of the tooth– a total of 40 planes were measured. The x-axis represents each sagittal slice across the width of the tooth. The shaded region represents the geometry of the pocket and the white region represents the normal area of attachment between tooth and epithelia. Error bars show the standard error of the mean for five replicates (n = 5). (F) The final mapping of the pocket generated by our photoacoustic method. It was generated by taking the measurements for each sagittal slice from (E) and overlaying them on the ultrasound image. This step removes the non-specific photoacoustic signal from (B).We evaluated the persistence of contrast agent in the pockets following oral rinses with water as well as brushing (Fig. 5). The reduction of pocket measurements was averaged from six representative planes across teeth 24 and 25 following 15 s rinsing, 30 s rinsing, and brushing; these steps resulted in 12.5%, 25.2%, and 100% reductions respectively when averaged across five replicates.
Fig. 5
Removal of contrast agent following oral rinsing with water and teeth brushing. (A) Representative PA-US images show the reduction in photoacoustic signal/pocket measurement following oral rinsing with water for 15 s, 30 s, and brushing. Yellow lines indicate the pocket measurements. (B) Fraction of original measurement following washing steps for five representative sagittal planes taken from teeth 24 and 25. The dotted line indicates the original measurement and error bars show SEM (n = 5). Asterisks denote significant difference from original measurement (unpaired t-test), ** (P ≤ 0.01), *** (P ≤ 0.001).
Removal of contrast agent following oral rinsing with water and teeth brushing. (A) Representative PA-US images show the reduction in photoacoustic signal/pocket measurement following oral rinsing with water for 15 s, 30 s, and brushing. Yellow lines indicate the pocket measurements. (B) Fraction of original measurement following washing steps for five representative sagittal planes taken from teeth 24 and 25. The dotted line indicates the original measurement and error bars show SEM (n = 5). Asterisks denote significant difference from original measurement (unpaired t-test), ** (P ≤ 0.01), *** (P ≤ 0.001).Finally, we compared our pocket measurements with the gold standard (Williams and Marquis) probes. The lowest possible measurement from the probes, as determined by a periodontist, was 2 mm (any lower values were rounded up); this value was recorded for the buccal pockets of teeth 24 and 25 with both probes. We measured 1.34 mm and 1.15 mm (n = 5) for teeth numbers 24 and 25 with PA-US, respectively, which agreed with the gold standard measurements while providing more precision.
Discussion
We report the first application of PA imaging for monitoring dental and periodontal health in humans by using a food-grade contrast agent. This technique is ideal for imaging periodontal pockets, gingival soft tissue and thickness, and the surfaces of teeth. These features could be distinguished both by visual inspection— ultrasound signal from tooth is brighter than tissue— and more quantitatively with line profiles of photoacoustic and ultrasound pixel intensities (Fig. 4D, Fig. S2). We determined that a 40 MHz ultrasound frequency provided sufficient resolution for discerning these features without sacrificing the requisite penetration depth. The US-only mode is particularly suited for applications that require knowledge of how a tooth is situated within the gingiva—one potential example would be diagnosis and monitoring of patients with delayed tooth eruption [47].The flow of contrast agent into the gingival sulcus allowed the imaging and measurement of pocket depths, eliminating the need for probing. In the case of patients with peri-implantitis, this technique has potential to allow pocket depth measurements when physical probing is risky or impossible due to threading along the implant surface and tissue sensitivity. It was common for the agent to coat the majority of the tooth surface even when it was locally administered to the gingival margin (Fig. 3A); however, this did not adversely affect measurements. The beginning of the pocket could always be identified by overlaying the ultrasound image to reveal the gingival margin and gingival thickness (Fig. 3B). The imaging procedure took roughly 5 min per tooth but could be greatly improved with a mouthpiece transducer capable of scanning all teeth at once.The PA imaging technique is highly reproducible (Figs. 3E, 4 E) with relative standard deviations of 10%. This precision offers a major improvement over the Williams probe, which can only record integer values for pocket depths and has inter-operator error of up to 40% [48]. The periodontal pocket is a dynamic structure due to the constant battle between bacterial invasion and the immune response [49], and integers are a poor reflection of this process. In a typical oral exam, six measurements are taken per tooth (distolingual, lingual, mesiolingual, distobuccal, buccal, and mesiobuccal) to approximate the full profile of the pocket [14]. We show that photoacoustic imaging can be used to reveal the entirety of the buccal pocket with sub-millimeter (0.01 mm) precision (Fig. 4). The overlay in Fig. 4F was created manually but it represents a convenient visualization tool for clinicians; in the future, the conversion from raw data to final image with highlighted pocket would be performed by automated algorithms [50]. Imaging the whole pocket over the patients’ lifetimes has previously been impossible and could supersede periodontal charting— this practice could reveal the development of abnormalities that would otherwise be overlooked.We observed that contrast agent could be easily removed from the gingival sulcus after imaging (Fig. 5). Oral rinsing with water reduced the contrast agent both on the surface of the teeth and within the sulcus; a few seconds of normal brushing completely removed the agent. We saw that our PA-US measurements (#24 = 1.34 mm, #25 = 1.17 mm) were in good agreement with those acquired by the Williams and Marquis probes, which recorded 2 mm for both lower central incisors. The probes were not capable of distinguishing any values lower than 2 mm; in addition, they could only record integers and relied on subjective estimation. PA-US imaging lowers ambiguity and improves the precision of measurements.We recognize some limitations in this study. Due to the size of the PA—US transducer, only anterior teeth could be imaged. In principle, all 32 teeth and the periodontium can be imaged with this technique but the geometry of our PA—US transducer physically limited access. Fortunately, several groups have reported custom transducer geometries, and future work includes development of a mouthpiece-shaped device to overcome this limitation [[51], [52], [53]]. A mouthpiece transducer would also reduce imaging artifacts caused by motion from the subject during scanning because the subject could bite down on the device. Partial penetration of contrast agent into the sulcus is also a possibility but the extent of penetration is consistent across irrigation and scanning events (Fig. 3C). The subject of the study had good oral hygiene and did not have deep (≥ 4 mm) pockets indicative of periodontal disease. However, we previously showed that deep pockets could be imaged in swine, a good model of the human gingiva [22,54]. Finally, because the periodontal pocket is a dynamic structure, imaging and probing across different days may have contributed to small variations in measurements.
Conclusion
This work demonstrates the first use of photoacoustic imaging for monitoring periodontal health in humans by quantitating pocket depths. The measurements acquired with this technique agreed with those acquired by a trained periodontist, while providing significantly more spatial information and precision. The gold-standard periodontal probe is an established but limited tool. This technique presents an opportunity for dental clinicians to move away from manual probing and toward photoacoustic image-guided diagnostics. Ultimately, we envision the development of a compact, PA-US mouthpiece as a platform for fast, non-invasive collection of critical periodontal metrics.
Conflicts of interest
The authors declare that there are no conflicts of interest regarding the authorship and publication of this article.
Authors: Ferdinand Knieling; Clemens Neufert; Arndt Hartmann; Jing Claussen; Alexander Urich; Cornelia Egger; Marcel Vetter; Sarah Fischer; Lukas Pfeifer; Alexander Hagel; Christian Kielisch; Rüdiger S Görtz; Dane Wildner; Matthias Engel; Jens Röther; Wolfgang Uter; Jürgen Siebler; Raja Atreya; Wolfgang Rascher; Deike Strobel; Markus F Neurath; Maximilian J Waldner Journal: N Engl J Med Date: 2017-03-30 Impact factor: 91.245
Authors: Carmel M Moran; Stephen D Pye; William Ellis; Anna Janeczko; Keith D Morris; Alan S McNeilly; Hamish M Fraser Journal: Ultrasound Med Biol Date: 2011-01-21 Impact factor: 2.998
Authors: Colman Moore; Yong Cheng; Natalia Tjokro; Brendan Zhang; Matthew Kerr; Mohammed Hayati; Kai Chiao Joe Chang; Nisarg Shah; Casey Chen; Jesse V Jokerst Journal: Angew Chem Int Ed Engl Date: 2022-06-13 Impact factor: 16.823
Authors: Colman Moore; Raina M Borum; Yash Mantri; Ming Xu; Pavla Fajtová; Anthony J O'Donoghue; Jesse V Jokerst Journal: ACS Sens Date: 2021-05-26 Impact factor: 9.618