Formulation of objective indices to quantify machine failure risk analysis for interruptions in radiotherapy.

Daisuke Kawahara1, Hisashi Nakano2, Akito Saito1, Yusuke Ochi3, Yasushi Nagata1,4.   

Abstract

OBJECTIVES: To evaluate the effect of interruption in radiotherapy due to machine failure in patients and medical institutions using machine failure risk analysis (MFRA).
MATERIAL AND METHODS: The risk of machine failure during treatment is assigned to three scores (biological effect, B; occurrence, O; and cost of labor and repair parts, C) for each type of machine failure. The biological patient risk (BPR) and the economic institution risk (EIR) are calculated as the product of B and O ( B × O ) and C and O ( C × O ), respectively. The MFRA is performed in two linear accelerators (linacs). RESULT: The multileaf collimator (MLC) fault has the highest BPR and second highest EIR. In particular, TrueBeam has a higher BPR and EIR for MLC failures. The total EIR in TrueBeam was significantly higher than that in Clinac iX. The minor interlock had the second highest BPR, whereas a smaller EIR. Meanwhile, the EIR for the LaserGuard fault was the highest, and that for the monitor chamber fault was the second highest. These machine failures occurred in TrueBeam. The BPR and EIR should be evaluated for each linac. Further, the sensitivity of the BPR, it decreased with higher T 1 / 2 and α/β values. No relative difference is observed in the BPR for each machine failure when T 1 / 2 and α/β were varied.
CONCLUSION: The risk faced by patients and institutions in machine failure may be reduced using MFRA. ADVANCES IN KNOWLEDGE: For clinical radiotherapy, interruption can occur from unscheduled downtime with machine failures. Interruption causes sublethal damage repair. The current study evaluated the effect of interruption in radiotherapy owing to machine failure on patients and medical institutions using a new method, that is, machine failure risk analysis.
© 2020 The Authors. Journal of Applied Clinical Medical Physics published by Wiley Periodicals, LLC on behalf of American Association of Physicists in Medicine.

Entities:  

Keywords:  biological dose calculation; downtime; machine failure

Mesh:

Year:  2020        PMID: 33326695      PMCID: PMC7856522          DOI: 10.1002/acm2.13126

Source DB:  PubMed          Journal:  J Appl Clin Med Phys        ISSN: 1526-9914            Impact factor:   2.102


  18 in total

1.  Radiation response of mammalian cells grown in culture. 1. Repair of X-ray damage in surviving Chinese hamster cells.

Authors:  M M ELKIND; H SUTTON
Journal:  Radiat Res       Date:  1960-10       Impact factor: 2.841

2.  Recovery from sublethal damage during intermittent exposures in cultured tumor cells: implications for dose modification in radiosurgery and IMRT.

Authors:  Yuta Shibamoto; Masato Ito; Chikao Sugie; Hiroyuki Ogino; Masaki Hara
Journal:  Int J Radiat Oncol Biol Phys       Date:  2004-08-01       Impact factor: 7.038

3.  The role of radiotherapy in cancer treatment: estimating optimal utilization from a review of evidence-based clinical guidelines.

Authors:  Geoff Delaney; Susannah Jacob; Carolyn Featherstone; Michael Barton
Journal:  Cancer       Date:  2005-09-15       Impact factor: 6.860

4.  The report of Task Group 100 of the AAPM: Application of risk analysis methods to radiation therapy quality management.

Authors:  M Saiful Huq; Benedick A Fraass; Peter B Dunscombe; John P Gibbons; Geoffrey S Ibbott; Arno J Mundt; Sasa Mutic; Jatinder R Palta; Frank Rath; Bruce R Thomadsen; Jeffrey F Williamson; Ellen D Yorke
Journal:  Med Phys       Date:  2016-07       Impact factor: 4.071

5.  Characteristics of a serially transplanted mouse mammary tumor and its tissue-culture-adapted derivative.

Authors:  S C Rockwell; R F Kallman; L F Fajardo
Journal:  J Natl Cancer Inst       Date:  1972-09       Impact factor: 13.506

6.  Cost evaluation to optimise radiation therapy implementation in different income settings: A time-driven activity-based analysis.

Authors:  Jacob Van Dyk; Eduardo Zubizarreta; Yolande Lievens
Journal:  Radiother Oncol       Date:  2017-09-22       Impact factor: 6.280

7.  Comparative Analysis of Radiotherapy Linear Accelerator Downtime and Failure Modes in the UK, Nigeria and Botswana.

Authors:  L M Wroe; T A Ige; O C Asogwa; S C Aruah; S Grover; R Makufa; M Fitz-Gibbon; S L Sheehy
Journal:  Clin Oncol (R Coll Radiol)       Date:  2019-11-19       Impact factor: 4.126

8.  Evaluation of safety in a radiation oncology setting using failure mode and effects analysis.

Authors:  Eric C Ford; Ray Gaudette; Lee Myers; Bruce Vanderver; Lilly Engineer; Richard Zellars; Danny Y Song; John Wong; Theodore L Deweese
Journal:  Int J Radiat Oncol Biol Phys       Date:  2009-05-04       Impact factor: 7.038

9.  The linear-quadratic model is an appropriate methodology for determining isoeffective doses at large doses per fraction.

Authors:  David J Brenner
Journal:  Semin Radiat Oncol       Date:  2008-10       Impact factor: 5.934

10.  Conditions for the equivalence of continuous to pulsed low dose rate brachytherapy.

Authors:  D J Brenner; E J Hall
Journal:  Int J Radiat Oncol Biol Phys       Date:  1991-01       Impact factor: 7.038

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.