Eva Braig1,2, Jessica Böhm3, Martin Dierolf3, Christoph Jud3, Benedikt Günther3,4, Korbinian Mechlem3,5, Sebastian Allner3, Thorsten Sellerer3, Klaus Achterhold3, Bernhard Gleich3, Peter Noël5, Daniela Pfeiffer5, Ernst Rummeny5, Julia Herzen3, Franz Pfeiffer3,5.
Abstract
Dual-energy CT has opened up a new level of quantitative X-ray imaging for many diagnostic applications. The energy dependence of the X-ray attenuation is the key to quantitative material decomposition of the volume under investigation. This material decomposition allows the calculation of virtual native images in contrast enhanced angiography, virtual monoenergetic images for beam-hardening artifact reduction and quantitative material maps, among others. These visualizations have been proven beneficial for various diagnostic questions. Here, we demonstrate a new method of 'virtual dual-energy CT' employing grating-based phase-contrast for quantitative material decomposition. Analogue to the measurement at two different energies, the applied phase-contrast measurement approach yields dual information in form of a phase-shift and an attenuation image. Based on these two image channels, all known dual-energy applications can be demonstrated with our technique. While still in a preclinical state, the method features the important advantages of direct access to the electron density via the phase image, simultaneous availability of the conventional attenuation image at the full energy spectrum and therefore inherently registered image channels. The transfer of this signal extraction approach to phase-contrast data multiplies the diagnostic information gained within a single CT acquisition. The method is demonstrated with a phantom consisting of exemplary solid and fluid materials as well as a chicken heart with an iodine filled tube simulating a vessel. For this first demonstration all measurements have been conducted at a compact laser-undulator synchrotron X-ray source with a tunable X-ray energy and a narrow spectral bandwidth, to validate the quantitativeness of the processing approach.Entities:
Year: 2018 PMID: 30401876 PMCID: PMC6219573 DOI: 10.1038/s41598-018-34809-6
Source DB: PubMed Journal: Sci Rep ISSN: 2045-2322 Impact factor: 4.379
Figure 1Measurements and results of virtual dual-energy processing for a material phantom. (A) Photography of the sample consisting of 6 different solid or liquid materials. From the reconstructed attenuation (B) and phase-contrast (C) data an effective interaction energy of was assigned via the literature value for the linear attenuation coefficient and the refractive index decrement of PMMA. The effective atomic number map (D) shows the distribution of Zeff ≈ 6.25 for Nylon to Zeff ≈ 7.99 for iodine. The quantitative iodine map (E) shows positive values of [I] ≈ 4.6 mg/ml only for the iodine solution. The virtual non-contrast image (F) is the conventional attenuation image where the identified iodine containing pixels are replaced with the attenuation value of water. The virtual monoenergetic image at (G) looks very similar to the conventional attenuation image. For the higher energies (H,I), the contribution of the electron density increases and the virtual monoenergetic image at looks very similar to the electron density image (which is simply proportional to the refractive index decrement image).
Comparison of measured linear attenuation coefficients μ, and refractive index decrements δ with literature values for the material phantom at an effective energy of Eeff = 24.6 keV for the conventional attenuation image and Eeff = 23.8 keV for the phase-contrast image.
| NaI (5.9) mg/ml | 0.597 ± 0.006 | 0.609 | 4.08 ± 0.07 | 4.07 |
| Blood | 0.568 ± 0.006 | 0.574 | 4.31 ± 0.07 | 4.27 |
| Ethanol | 0.323 ± 0.006 | 0.325 | 3.23 ± 0.06 | 3.27 |
| PMMA | 0.470 ± 0.006 | 0.470 | 4.70 ± 0.06 | 4.70 |
| POM | 0.628 ± 0.006 | 0.628 | 5.55 ± 0.06 | 5.55 |
| Nylon | 0.423 ± 0.006 | 0.419 | 4.61 ± 0.06 | 4.58 |
| Water | 0.523 ± 0.006 | 0.523 | 4.07 ± 0.06 | 4.07 |
The subscripts m and l indicate measured and literature values. The literature values are calculated from tabulated values from the NIST database[25], from[26] for blood and from[20] for the NaI solution. The given uncertainty includes the standard deviation of the image region and the systematic error.
Quantitative results for the effective atomic numbers Zeff and the electron density ρe for the material phantom in comparison to different literature values.
| NaI (5.9 mg/ml) | 7.97 ± 0.06 | — | — | 3.34 ± 0.07 | — |
| Blood | 7.60 ± 0.06 | — | 7.74 | 3.54 ± 0.08 | — |
| Ethanol | 6.52 ± 0.07 | 6.35 | — | 2.69 ± 0.07 | 2.68 |
| PMMA | 6.58 ± 0.03 | 6.47 | 6.56 | 3.86 ± 0.06 | 3.86 |
| POM | 7.05 ± 0.05 | 6.95 | 7.03 | 4.56 ± 0.06 | 4.56 |
| Nylon | 6.24 ± 0.06 | 6.12 | 6.21 | 3.79 ± 0.06 | 3.76 |
| Water | 7.51 ± 0.05 | 7.42 | 7.51 | 3.34 ± 0.06 | 3.34 |
The subscript m indicates measured values. For the effective atomic number two different literature sources are given, denoted with l1 for values from Qi et al.[16] and l2 for values from the XmuDat library[25,26]. The given uncertainty includes the standard deviation of the image region and the systematic error.
Figure 2Measurements and results of virtual dual-energy processing for a biological soft tissue sample (chicken heart). (A) Photography of a fresh chicken heart next to the measurement container. The conventional attenuation image (B) shows very limited contrast only between the fatty tissue and the experimentally simulated iodine filled vessel. The phase-contrast image (C) reveals good contrast for the different anatomical structures like muscle, fat and blood vessels (most likely the aorta and two vessels of the low pressure system) but no contrast between contrast agent and the surrounding water. The effective atomic number map (D) reflects the situation of the conventional attenuation image with very low variations between the different structures besides fat and contrast agent. The quantitative iodine map (E) shows positive iodine concentrations of [I] ≈ 43 mg/ml only for the region of the contrast agent filled tube. The virtual non-contrast image (F) is the conventional attenuation image with the iodine containing pixels replaced with the attenuation value of water. At , the virtual monoenergetic image (G) looks very similar to the conventional attenuation image. For the low atomic number soft tissue materials the Compton effect dominates the image formation already at (H) and a difference to the virtual monoenergetic image at (I) and the phase-contrast image is only visible for the iodine filled tube.