Scalar localisation of peri-modiolar electrodes and speech perception outcomes.

C Shaul1, A S Dragovic1, A K Stringer1, S J O'Leary1, R J Briggs1.   

Abstract

OBJECTIVE: To identify the intracochlear electrode position in cochlear implant recipients and determine the correlation to speech perception for two peri-modiolar electrode arrays.
METHODS: Post-operative cone-beam computed tomography images of 92 adult recipients of the 'CI512' electrode and 18 adult recipients of the 'CI532' electrode were analysed. Phonemes scores were recorded pre-implantation, and at 3 and 12 months post-implantation.
RESULTS: All CI532 electrodes were wholly within scala tympani. Of the 79 CI512 electrodes intended to be in scala tympani, 58 (73 per cent) were in scala tympani, 14 (17 per cent) were translocated and 7 (9 per cent) were wholly in scala vestibuli. Thirteen CI512 electrodes were deliberately inserted into scala vestibuli. Speech perception scores for post-lingual recipients were higher in the scala tympani group (69.1 per cent) compared with the scala vestibuli (54.2 per cent) and translocation (50 per cent) groups (p < 0.05). Electrode location outside of scala tympani independently resulted in a 10.5 per cent decrease in phoneme scores.
CONCLUSION: Cone-beam computed tomography was valuable for demonstrating electrode position. The rate of scala tympani insertion was higher in CI532 than in CI512 electrodes. Scala vestibuli insertion and translocation were associated with poorer speech perception outcomes.

Entities:  

Keywords:  X-Ray; Cochlear Implantation; Computed Tomography; Scala Tympani; Scala Vestibuli

Mesh:

Year:  2018        PMID: 30370884     DOI: 10.1017/S0022215118001871

Source DB:  PubMed          Journal:  J Laryngol Otol        ISSN: 0022-2151            Impact factor:   1.469


  8 in total

1.  Effect of Scala Tympani Height on Insertion Depth of Straight Cochlear Implant Electrodes.

Authors:  William G Morrel; Jourdan T Holder; Benoit M Dawant; Jack H Noble; Robert F Labadie
Journal:  Otolaryngol Head Neck Surg       Date:  2020-02-25       Impact factor: 3.497

Review 2.  Review on cochlear implant electrode array tip fold-over and scalar deviation.

Authors:  Anandhan Dhanasingh; Claude Jolly
Journal:  J Otol       Date:  2019-01-09

3.  Evaluating cochlear insertion trauma and hearing preservation after cochlear implantation (CIPRES): a study protocol for a randomized single-blind controlled trial.

Authors:  Saad Jwair; Ralf A Boerboom; Huib Versnel; Robert J Stokroos; Hans G X M Thomeer
Journal:  Trials       Date:  2021-12-09       Impact factor: 2.279

4.  Suitable Electrode Choice for Robotic-Assisted Cochlear Implant Surgery: A Systematic Literature Review of Manual Electrode Insertion Adverse Events.

Authors:  Paul Van de Heyning; Peter Roland; Luis Lassaletta; Sumit Agrawal; Marcus Atlas; Wolf-Dieter Baumgartner; Kevin Brown; Marco Caversaccio; Stefan Dazert; Wolfgang Gstoettner; Rudolf Hagen; Abdulrahman Hagr; Greg Eigner Jablonski; Mohan Kameswaran; Vladislav Kuzovkov; Martin Leinung; Yongxin Li; Andreas Loth; Astrid Magele; Robert Mlynski; Joachim Mueller; Lorne Parnes; Andreas Radeloff; Chris Raine; Gunesh Rajan; Joachim Schmutzhard; Henryk Skarzynski; Piotr H Skarzynski; Georg Sprinzl; Hinrich Staecker; Timo Stöver; Dayse Tavora-Viera; Vedat Topsakal; Shin-Ichi Usami; Vincent Van Rompaey; Nora M Weiss; Wilhelm Wimmer; Mario Zernotti; Javier Gavilan
Journal:  Front Surg       Date:  2022-03-24

5.  High-speed flat-detector computed tomography for temporal bone imaging and postoperative control of cochlear implants.

Authors:  Felix Eisenhut; Lava Taha; Michael Manhart; Vivian Thimsen; Konstantinos Mantsopoulos; Heinrich Iro; Joachim Hornung; Arnd Dörfler; Stefan Lang
Journal:  Neuroradiology       Date:  2022-04-12       Impact factor: 2.995

6.  Vertigo Associated With Cochlear Implant Surgery: Correlation With Vertigo Diagnostic Result, Electrode Carrier, and Insertion Angle.

Authors:  Charlotte Weinmann; Uwe Baumann; Martin Leinung; Timo Stöver; Silke Helbig
Journal:  Front Neurol       Date:  2021-06-11       Impact factor: 4.003

7.  Outcomes for a clinically representative cohort of hearing-impaired adults using the Nucleus® CI532 cochlear implant.

Authors:  Matthias Hey; Nicole Neben; Timo Stöver; Uwe Baumann; Alexander Mewes; Tim Liebscher; Mark Schüssler; Antje Aschendorff; Thomas Wesarg; Andreas Büchner; Paula Greenham; Ulrich Hoppe
Journal:  Eur Arch Otorhinolaryngol       Date:  2020-03-05       Impact factor: 2.503

8.  Intracochlear electrode array position and cochlear implant outcomes using the nucleus slim modiolar electrode and the extended round window approach: a follow-up study.

Authors:  Tim M Klabbers; Floris Heutink; Wendy J Huinck; Willem-Jan van der Woude; Berit M Verbist; Emmanuel A M Mylanus
Journal:  Eur Arch Otorhinolaryngol       Date:  2022-01-18       Impact factor: 3.236

  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.