| Literature DB >> 30366222 |
Rohit Mahar1, Syeda Batool1, Farid Badar1, Yang Xia2.
Abstract
Both spin-echo (SE) and ultra-short echo (UTE) based MRI sequences were used on a 7 T µMRI system to quantify T2, T1ρ and T1 relaxation times from articular cartilage to the cartilage-bone interface on canine humeral specimens at 19.5 µm pixel resolution. A series of five relaxation-weighted images were acquired to calculate one relaxation map (T2, T1ρ or T1), from which the depth-dependent profiles were examined between the SE method and the UTE method, over the entire non-calcified cartilage and within the cartilage-bone interface. SE-based methods enabled the quantification of relaxation profiles over the noncalcified cartilage, from 0 µm (articular surface) to approximately 460 µm in depth (near the end of radial zone). Most of the cartilage-bone interface was imaged by the UTE-based methods, to a tissue depth of about 810 µm. Pixel-by-pixel calculation of the relaxation times between the independent SE and UTE methods correlated well with each other. A better understanding of the tissue properties reliably over the cartilage-bone interface region by a non-invasive MRI approach could contribute to the clinical diagnostics of trauma-induced osteoarthritis.Entities:
Keywords: Articular cartilage; Microscopic resolution; Quantitative MRI; T1; T1ρ; T2; UTE; Zone of calcified cartilage
Mesh:
Year: 2018 PMID: 30366222 PMCID: PMC6289866 DOI: 10.1016/j.jmr.2018.10.008
Source DB: PubMed Journal: J Magn Reson ISSN: 1090-7807 Impact factor: 2.229